metricas
covid
Buscar en
Acta Otorrinolaringológica Española
Toda la web
Inicio Acta Otorrinolaringológica Española Parámetros de validación de otoemisiones con el echocheck®
Journal Information
Vol. 53. Issue 3.
Pages 151-155 (March 2002)
Share
Share
Download PDF
More article options
Vol. 53. Issue 3.
Pages 151-155 (March 2002)
Parámetros de validación de otoemisiones con el echocheck®
Eoaes (evoked oto acoustic emissions) validation parameters with echocheck®
Visits
735
P. Torrico Román
,
Corresponding author
med003858@saludalia.com

Correspondencia: Servicio de ORL. Hospital Don Benito-Villanueva. Avda. Vegas Altas, s/n. 06400 Don Benito (Badajoz)
, J. López-Ríos Velasco*, M.C. De Cáceres Morillo*, M. Serrano Berrocal**, G. Trinidad Ramos**
* Subunidad de sorderas. Unidad de minusvalías (junta de extremadura-servicio extremeño de salud). Hospital D. Benito-Villanueva. Badajoz
** Subunidad de sorderas. Unidad de minusvalías (junta de extremadura-servicio extremeño de salud). Hospital infanta cristina. badajoz
This item has received
Article information
Resumen
Objetivos

Aunque existen trabajos que demuestran la utilidad del Echocheck® en el despistaje de la hipoacusia en neonatos, ninguno lo compara con los sistemas ILO® ni tampoco han sido cotejados los resultados que el aparato muestra mediante un código de colores con los parámetros internos que genera. Se pretende demostrar que ambas técnicas son equivalentes y válidas.

Métodos

Se estudiaron mediante otoemisiones acústicas 494 oídos de recién nacidos vivos usando el sistema Echocheck. Se clasificaron en Normales, No Válidos, Falta e Inválidos según un código de colores. Posteriormente se cotejaron con los resultados internos registrados por el aparato. Por otra parte se detectaron las otoemisiones en 135 oídos con ambos sistemas y se compararon los resultados obtenidos.

Resultados

La correlación entre los resultados externos determinados mediante el código de colores y los registros internos fue absoluta según los parámetros previamente determinados. En la comparación mediante los dos sistemas, los resultados coincidieron excepto en dos pruebas en las que faltaron las otoemisiones con el Echocheck estando presentes con el ILO.

Conclusiones

Queda demostrada la validez de la prueba con este nuevo sistema, así como la equiparación de sus resultados con los obtenidos por los sistemas ILO®, recomendándose su uso para la detección precoz universal de todos los recién nacidos vivos por la sencillez de su manejo.

Palabras clave:
Otoemisiones acústicas
Echosensor
Echocheck
Detección precoz de hipoacusia infantil
Abstract
Goals

There are several papers that support the use of Echocheck® system to exclude neonatal hypoacusis, but none of them compare it with the ILO® systems, neither compare the internal parameters obtained versus those from a colour coded system. We are trying to demonstrate that both techniques are equivalent and valid.

Methods

Acoustic otoemissions were studied in four hundred and ninety four (494) newborn ears using Echocheck. They were classified as Normal, Not valid, No EOAEs and Invalid using a colour coded system. After this they were compared with the internal results registered by the system. On the other hand, otoemissions were obtained using both methods (Echocheck and ILO-88) in 135 ears and the results compared.

Results

The correlation between the results obtained through the colour coded system and those obtained from the internal register was absolute for those parameters previously determined. For the comparison between both systems, results did coincide in all cases except for two, in which there were only results with ILO and not with Echocheck.

Conclusions

The validity of the test using this new system as well as the comparison of the results with those obtained from ILO systems has been demonstrated, therefore its use has been recommended for the early detection of hypoacusis in newborn due to the fact that it is simple to perform.

Key words:
Acoustic otoemissions
Echosensor
Echocheck
Early detection of hypoacusis in children

Article

These are the options to access the full texts of the publication Acta Otorrinolaringológica Española
Subscriber
Subscriber

If you already have your login data, please click here .

If you have forgotten your password you can you can recover it by clicking here and selecting the option “I have forgotten my password”
Subscribe
Subscribe to

Acta Otorrinolaringológica Española

Purchase
Purchase article

Purchasing article the PDF version will be downloaded

Price 19.34 €

Purchase now
Contact
Phone for subscriptions and reporting of errors
From Monday to Friday from 9 a.m. to 6 p.m. (GMT + 1) except for the months of July and August which will be from 9 a.m. to 3 p.m.
Calls from Spain
932 415 960
Calls from outside Spain
+34 932 415 960
E-mail
Article options
es en pt

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos