Grado de concordancia y diferencias entre valores maculares en niños mediante dos tipos de tomografía de coherencia óptica espectral
Agreement and differences between macular values in children using two types of spectral optical coherence tomography
Visits
52
J.M. Ruiz Caro Larrea
Corresponding author
drivertad@hotmail.com

Autor para correspondencia.
, L. Cabrejas Martínez, M.A. Alonso Peralta, I. Mahíllo Fernández, I. Jiménez-Alfaro Morote
Servicio de Oftalmología, Hospital Universitario Fundación Jiménez Díaz, Madrid, España
This item has received
Article information
Abstract
Full Text
Bibliography
Download PDF
Statistics
Figures (3)
Show moreShow less
Tables (4)
Tabla 1. Valores de normalidad OCT Cirrus y OCT Spectralis
Tabla 2. Valores de grosor de capas intrarretinianas con OCT Spectralis
Tabla 3. Correlación entre la longitud axial y los valores maculares
Tabla 4. Grado de concordancia absoluto entre los valores de OCT Cirrus y Spectralis
Show moreShow less
Resumen
Objetivo

Determinar el grado de concordancia y las diferencias entre los parámetros maculares en función de sexo, edad, lateralidad, etnia y tipo de tomografía de coherencia óptica (OCT) en pacientes pediátricos.

Materiales y métodos

Estudio prospectivo, transversal y comparativo. Se hizo uso de la OCT Spectralis y la OCT Cirrus para obtener los valores maculares de ambos ojos en 100 pacientes entre 5 y 15años sin patología ocular. Debido a la semejanza entre los valores de ambos ojos, en la muestra final solo consideramos los ojos derechos (OD) (n=100).

Resultados

La media del grosor macular central (GMC) fue de 244,2μm con Cirrus y 261,7μm con Spectralis. El grado de concordancia para el GMC fue moderado (coeficiente de correlación intraclase: 0,66) y de bajo a muy bajo en el resto de cuadrantes. Los valores de GMC de ambos sistemas fueron significativamente mayores en varones que en mujeres (p<0,05). La capa de fibras nerviosas de la retina macular (CFNRm), la capa de células ganglionares (CCG), la capa plexiforme interna (CPI) y la capa nuclear interna (CNI) tuvieron valores significativamente mayores en varones que en mujeres (p<0,01). No hubo diferencias estadísticamente significativas en el GMC según edad y lateralidad; sin embargo, fue significativamente mayor en niños de etnia europea que en niños de etnia latinoamericana; Cirrus (p=0,042) y Spectralis (p=0,033).

Conclusiones

Los parámetros maculares obtenidos mediante OCT Spectralis y OCT Cirrus no deberían ser intercambiables en pacientes pediátricos.

Palabras clave:
Grado de concordancia
Pacientes pediátricos
Tomografía de coherencia óptica
Mácula
Abstract
Objective

To determine the degree of agreement and differences between macular parameters according to sex, age, laterality, ethnicity and type of optical coherence tomography (OCT) in paediatric patients.

Materials and methods

Prospective, transversal, and comparative study. OCT Spectralis and OCT Cirrus were used to obtain the macular values of both eyes in 100 patients between 5 and 15years without ocular pathology. Due to the similarity between the values of both eyes, in the final sample we only considered the right eyes (OD) (n=100).

Results

The average central macular thickness (CMT) was 244.2μm with Cirrus and 261.7μm with Spectralis. The degree of agreement for the GMC was moderate (intraclass correlation coefficient: 0.66) and from low to very low in the other quadrants. The GMC values of both systems were significantly higher in men than in women (P<.05). The macular retinal nerve fibre layer (mRNC), ganglion cell layer (GCC), the inner plexiform layer (IPLC), and the inner nuclear layer (INFL) had significantly higher values in men than in women (P<.01). There were no statistically significant differences in the GMC by age and laterality, however it was significantly higher in ethnic European children than in ethnic Latin American children; Cirrus (P=.042) and Spectralis (P=.033).

Conclusions

Macular parameters obtained by OCT Spectralis and OCT Cirrus should not be interchangeable in pediatric patients.

Keywords:
Degree of agreement
Pediatric patients
Optical coherence tomography
Macula

Article

These are the options to access the full texts of the publication Archivos de la Sociedad Española de Oftalmología
Subscriber
Subscriber

If you already have your login data, please click here .

If you have forgotten your password you can you can recover it by clicking here and selecting the option “I have forgotten my password”
Subscribe
Subscribe to

Archivos de la Sociedad Española de Oftalmología

Purchase
Purchase article

Purchasing article the PDF version will be downloaded

Price 19.34 €

Purchase now
Contact
Phone for subscriptions and reporting of errors
From Monday to Friday from 9 a.m. to 6 p.m. (GMT + 1) except for the months of July and August which will be from 9 a.m. to 3 p.m.
Calls from Spain
932 415 960
Calls from outside Spain
+34 932 415 960
E-mail
Article options
es en pt

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos