Correlación de los datos isocinéticos y la resonancia magnética en las lesiones de cofia rotadora
Correlation of the isokinetics data and magnetic resonance in the lesions of the rotator cuff
Visits
2864
Mª.L. Catasús Clavéa,
Corresponding author
mcatasus@csub.scs.es

Correspondencia: C/ Feixa Llarga, s/n 08907 Hospitalet de Llobregat. Barcelona
, U. Díaz Aristizabalb, C. García Almazánb, A. Ribera Alcazarb, A. Guillén Soláb, J. Santín Paredesc
a Médico adjunto. L’Hospitalet de Llobregat. Barcelona
b Médico residente. L’Hospitalet de Llobregat. Barcelona
c Fisioterapeuta. Servicio de Rehabilitación y Medicina Física. Ciutat Sanitaria i Universitaria de Bellvitge. L’Hospitalet de Llobregat. Barcelona
This item has received
Article information
Resumen
Objetivos

Analizar los resultados isocinéticos en lesiones de cofia rotadora del hombro y la búsqueda de patrones isocinéticos que tengan correlación con la lesión anatómica confirmada por RM.

Material y método

Recogimos una muestra de 25 pacientes (15 hombres y 10 mujeres) de edades comprendidas entre los 42 y los 70 años, todos ellos con criterios clínicos de lesión de cofia rotadora, confirmada posteriormente con un estudio por imagen (RM) con secuencias en SE T2 y FFE T2 en los tres planos del espacio a nivel del hombro afecto. La etiología era degenerativa en todos los casos.

Este grupo lo hemos comparado con un grupo control de 25 sujetos.

Hemos utilizado un dinamómetro isocinético CYBEX NORM II. El test se realizó en sedestación, con el tronco fijado y el brazo en el plano de la escápula (anteflexión de 30° y abdución de 45°), siempre en régimen concéntrico, valorándose la fuerza (Peak torque) y el conjunto del trabajo total (Set Total Work) de rotación externa y rotación interna, y a velocidades angulares de 90° y 180°.

Análisis estadístico con programa SPSS, Test de Mann-Whitney y test de Wilcoxon.

Resultados

Grupo control-grupo afecto: Diferencia estadisticamente significativa (p < 0,05) para peak torque y set total Work.

Grupo control-grupo afecto por sexos p < 0,05 para peak toeque y set total work.

Conclusiones

El test isocinético es una prueba útil para valorar la pérdida de fuerza en lesiones de cofia rotadora.

No hemos encontrado un patrón característico para las distintas lesiones de cofia rotadora.

Hemos apreciado una disminución global de fuerza en el grupo antagonista.

Palabras clave:
Resonancia magnética
Isocinesia
Cofia rotadora
Summary
Objectives

Analyze the isokinetic results in lesions of the rotator cuff of the shoulder and the search for isokinetic patterns that have a correlation with the anatomic lesion confirmed by the MRI.

Material and methods

We gathered a sample of 25 patients (15 men and 10 women) whose ages ranged from 42 to 70 years, all of them with clinical criteria of rotator cuff lesion, confirmed after with an imaging study (MRI) with sequences in SE T2 and FFE T2 on the three planes of the space in the affected shoulder. Etiology is degenerative in all the cases.

We have compared this group with a control group of 25 subjects.

We have used a CYBEX NORM II isokinetic dynamometer. The test was performed in a sitting position, with the trunk fixed and the arms on the plane of the scapula (anteflexion of 30° and abduction of 45°), always in concentric regime (Peak torque) and Set Total Work of external rotation and internal rotation, and angular rates of 90° and 180° were assessed.

Statistic analysis with SPSS program, Mann-Whitney Test and Wilcoxon Test.

Results

Control group-affected group: statistically significant difference (p < 0.05) of peak torque and set total Work. Control group affected group by gender p < 0.05 for peak torque and set total work.

Conclusions

The isokinetics test is a useful test to assess force loss in rotator cuff lesions. We have not found a characteristic pattern for the different rotator cuff lesions. We have observed a global decrease in force in the antagonist group.

Key words:
Magnetic resonance
Isokinesis
Rotator cuff

Article

These are the options to access the full texts of the publication Rehabilitación
Subscriber
Subscriber

If you already have your login data, please click here .

If you have forgotten your password you can you can recover it by clicking here and selecting the option “I have forgotten my password”
Subscribe
Subscribe to

Rehabilitación

Purchase
Purchase article

Purchasing article the PDF version will be downloaded

Price 19.34 €

Purchase now
Contact
Phone for subscriptions and reporting of errors
From Monday to Friday from 9 a.m. to 6 p.m. (GMT + 1) except for the months of July and August which will be from 9 a.m. to 3 p.m.
Calls from Spain
932 415 960
Calls from outside Spain
+34 932 415 960
E-mail
Article options
es en pt

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos