Esfínter Urinario Artificial “FlowSecureTM”: Un nuevo concepto de esfínter artificial regulable y con oclusión condicional para la incontinencia urinaria de esfuerzo
“FlowsecureTM” artificial urinary sphincter: a new adjustable artificial urinary sphincter concept with conditional occlusion for stress urinary incontinence
Visitas
1683
F. García Montes1,**, S.L. Knight***, T. Greenwell*, A.R. Mundy*, M.D. Craggs*
* Institue of Urology and Nephrology.University College of London
** Hospital Universitario Son Dureta. Palma de Mallorca
Este artículo ha recibido
Información del artículo
Resumen
Introducción

Difundir el concepto de un nuevo esfínter urinario artificial con oclusión condicional para el tratamiento de la incontinencia urinaria de esfuerzo. El prototipo fue concebido y diseñado por los profesores Craggs M. y Mundy A.R. en el Instituto de Urología y Nefrología de Londres.

Materiales y Métodos

El esfínter FlowSecure es un prototipo de una sola pieza de silicona llena de suero salino que consta de: 1) Un reservorio regulador de presión, 2) un reservorio de asistencia al estrés, 3) una bomba de control con un puerto autosellable y 4) un manguito. El reservorio regulador determina la presión de la prótesis y se puede ajustar desde 0 a 80 cm H2O mediante la inyección o extracción de líquido a través del puerto autosellable (oclusión basal). El reservorio de asistencia al estrés transmite de forma reversible los aumentos de la presión intra-abdominal durante los periodos de esfuerzo (oclusión condicional).

Resultados

La técnica quirúrgica de implantación es simple, reduciendo la posibilidad de infección y errores de ensamblaje. La asociación en serie de un reservorio regulador de presión con un reservorio de asistencia al estrés permite establecer presiones de oclusión basal bajas y el aumento condicional de la misma durante los periodos de estrés, reduciendo el riesgo de atrofia y erosión. Tras la implantación, la presión es ajustable dependiendo de las necesidades clínicas del paciente.

Discusión

El esfínter FlowSecure está diseñado para corregir la incontinencia ajustando la presión de oclusión según las necesidades individuales de cada paciente, así como para reducir las complicaciones asociadas al exceso de presión de oclusión y los fallos mecánicos.

Palabras clave:
Esfínter urinario artificial
Oclusión condicional
Abstract
Introduction

To spread de concept of a new artificial urinary sphincter with conditional occlusion for stress incontinence. The new prototype was conceived and designed in The Institute of Urology and Nephrology of London by Professor Craggs M. and Professor Mundy A.R.

Methods

The FlowSecure sphincter consists of an adjustable pressure-regulating balloon, a stress relief reservoir, a control pump and valve assembly unit with self-sealing port and a urethral cuff. The pressure regulating balloon determinates de operating pressure of the device; the pressure is adjustable in the range 0-80 cm H2O and can be altered by injection or removal of normal saline through the self sealing port. The stress relief balloon transmits transient intrabdominal pressure to the cuff during periods of stress. An adjustable circular urethral cuff minimises creasing and possible stress fractures.

Results

The device is implanted as a one-piece assembly which is pre-filled with sterile saline. The surgical technique is simple and associated with little handling, reducing risk of infection and potential assembly errors. The adjustable pressure regulating balloon in association with the stress relief reservoir enables the cuff occluding pressure to be set at a low range, therefore reducing the risk for atrophy and erosion.

Discussion

The new FlowSecure urinary artificial sphincter with conditional occlusion is designed to provide good continence rates adjusting regulating pressures when needed and conceived to reduce the risk of potential complications associated with excessive occluding pressures and mechanical failures.

Key words:
Artificial urinary sphincter
Conditional occlusion

Artículo

Opciones para acceder a los textos completos de la publicación Actas Urológicas Españolas
Suscriptor
Suscriptor de la revista

Si ya tiene sus datos de acceso, clique aquí.

Si olvidó su clave de acceso puede recuperarla clicando aquí y seleccionando la opción "He olvidado mi contraseña".
Suscribirse
Suscribirse a:

Actas Urológicas Españolas

Comprar
Comprar acceso al artículo

Comprando el artículo el PDF del mismo podrá ser descargado

Precio 19,34 €

Comprar ahora
Contactar
Teléfono para suscripciones e incidencias
De lunes a viernes de 9h a 18h (GMT+1) excepto los meses de julio y agosto que será de 9 a 15h
Llamadas desde España
932 415 960
Llamadas desde fuera de España
+34 932 415 960
E-mail
Opciones de artículo
es en pt

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos