Implantación del Esfínter Urinario Artificial “FlowSecureTM” en uretra bulbar: Descripción de la técnica quirúrgica paso a paso
“flowsecure™” artificial urinary sphincter: a new adjustable artificial urinary sphincter concept with conditional occlusion for stress urinary incontinence.
Visitas
1971
F. García Montes**,1
Autor para correspondencia
fgmontes@wanadoo.es

Correspondencia Autor: Dr. F. García Montes Hospital Universitario Son Dureta Andrea Doria 55 - 07014 Palma de Mallorca (Islas Baleares) Teléfono: 971 175 000
, A. Vicens Vicens**, M. Ozonas Moragues**, P. Pizá Reus**, A. Mora Salvá**, A.R. Mundy*, M.D. Craggs*
* Institute of Urology and Nephrology. University College of London.
** Hospital Universitario Son Dureta. Palma de Mallorca.
Este artículo ha recibido
Información del artículo
Resumen
Introducción

En Octubre de 2006 se implantó por primera vez en nuestra Institución el nuevo esfínter urinario artificial FlowSecure TM. El prototipo fue concebido y diseñado por los Profesores Craggs MD y Mundy AR y los resultados clínicos preliminares de 9 pacientes fueron publicados a mediados del 2006. Nuestro objetivo es el de difundir la descripción detallada de la técnica quirúrgica para la implantación de esta nueva prótesis.

Materiales y métodos

Se requiere una incisión perineal para la exposición de la uretra bulbar y una abdominal para la creación de un espacio paravesical y de un bolsillo escrotal. Se coloca un trocar con un estilete desde la incisión abdominal a la perineal para pasar el manguito desinflado, que se ajusta a la uretra y se asegura con tres puntos de Prolene. Tras volver a llenar el manguito se extrae líquido hasta que se forma una muesca en el reservorio de asistencia al estrés, indicando que el sistema está a presión atmosférica 0. Se coloca entonces la bomba en el escroto y los reservorios en el espacio paravesical.

Resultados

El esfínter fue colocado en un paciente con incontinencia de esfuerzo tras una RTU de próstata. La técnica es simple, el paciente se fue de alta al cuarto día y no fue necesario presurizar el sistema porque el paciente recuperó la continencia desde la retirada de la sonda.

Discusión

La implantación en uretra bulbar es muy sencilla. Los resultados en este paciente concuerdan con los de Knigth et al., que describen tiempo quirúrgico de 30 a 40 minutos, media de estancia hospitalaria de 4 días y presurización innecesaria en tres de sus nueve pacientes. Parece que la técnica y los resultados son reproducibles.

Palabras clave:
Esfínter urinario artificial
Oclusión condicional
Abstract
introduction

We have implanted the FlowSecure artificial sphincter for the first time in October 2006. The prototype was originally conceived and designed by Professor Craggs M. D. and Professor Mundy A.R. Preliminary clinical results were reported in nine patients early this year. Our objective is to spread technique for surgical implantation.

Methods

Combined perineal and abdominal incisions are required for exposure of bulbar urethra, creation of a cavity in the para-vesical space and dissection of a pocket under de scrotal wall. A trocar with a stylet is routed from the abdominal incision to the perineal incision to pass the deflated cuff to the perineal site. The cuff is placed around the urethra and secured with Prolene sutures. After refilling the cuff, fluid is removed from the system until the stress relief balloon becomes just indented (atmospheric pressure 0). The pump is placed in the scrotum and the balloons in the paravesical space.

Results

We have implanted our first FlowSecure artificial sphincter in a patient with severe stress incontinence following a T.U.R.P. The surgical technique is simple and associated with little handling. He was discharged from hospital 4 days after the procedure and it was decided that pressurisation was unnecessary.

Discussion

Surgical implantation of the new FlowSecure artificial urinary sphincter is an easy procedure in males with stress urinary incontinence. Knigth et al. reported 30 to 40 minutes operating time, 4 days mean hospital stay and unnecessary pressurisation procedure in 3 out of their 9 patients. It seems that their results are reproducible.

Key words:
Artificial Urinary Sphincter
Conditional occlusion

Artículo

Opciones para acceder a los textos completos de la publicación Actas Urológicas Españolas
Suscriptor
Suscriptor de la revista

Si ya tiene sus datos de acceso, clique aquí.

Si olvidó su clave de acceso puede recuperarla clicando aquí y seleccionando la opción "He olvidado mi contraseña".
Suscribirse
Suscribirse a:

Actas Urológicas Españolas

Comprar
Comprar acceso al artículo

Comprando el artículo el PDF del mismo podrá ser descargado

Precio 19,34 €

Comprar ahora
Contactar
Teléfono para suscripciones e incidencias
De lunes a viernes de 9h a 18h (GMT+1) excepto los meses de julio y agosto que será de 9 a 15h
Llamadas desde España
932 415 960
Llamadas desde fuera de España
+34 932 415 960
E-mail
Opciones de artículo
es en pt

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos