covid
Buscar en
Endocrinología y Nutrición (English Edition)
Toda la web
Inicio Endocrinología y Nutrición (English Edition) Influence of the immunoassay used in measurement of serum vitamin D levels
Información de la revista
Vol. 61. Núm. 3.
Páginas 123-129 (marzo 2014)
Visitas
6228
Vol. 61. Núm. 3.
Páginas 123-129 (marzo 2014)
Original article
Acceso a texto completo
Influence of the immunoassay used in measurement of serum vitamin D levels
Influencia del inmunoensayo empleado en la determinación de vitamina D sérica
Visitas
6228
Silvia Górriz Pintado
Autor para correspondencia
silgopin@hotmail.com

Corresponding author.
, Pedro Luis Estela Burriel
Área de Diagnóstico Biológico, Hospital Universitario de la Ribera, Alzira, Valencia, Spain
Este artículo ha recibido
Información del artículo
Resumen
Texto completo
Bibliografía
Descargar PDF
Estadísticas
Figuras (2)
Tablas (2)
Table 1. Demographic characteristics and statistical parameters of the patient groups studied by 25(OH)D assay.
Table 2. Passing-Bablok and correlation concordance analyses between the Elecsys Vitamin D Total (Roche) and ADVIA Centaur Vitamin D Total (Siemens) assays in all patient groups.
Mostrar másMostrar menos
Abstract
Introduction

Serum 25-hydroxyvitamin D [25(OH)D] levels are the best indicator of vitamin D levels in the body. Precision, reproducibility, and lack of standardization are the main problems in such measurements. The aim of this study was to compare the 25(OH)D levels measured using Elecsys Vitamin D Total (Roche) and ADVIA Centaur Vitamin D Total (Siemens).

Materials and methods

25(OH)D levels were tested in 166 patients using both methods. Patients were subsequently divided into two groups: a “supplemented group” consisting of patients receiving vitamin D supplements, and an “untreated group” consisting of the rest of patients.

Results

25(OH)D mean levels measured by the Roche and Siemens methods in the overall group were 33.6±16.0 and 19.8±12.4ng/mL, respectively. 54.2% of patients were receiving vitamin D supplements. In this group, mean 25(OH)D levels measured by the Roche and Siemens methods were 40.6±14.5 and 25.4±13.1ng/mL, respectively. In the untreated group, the respective values were 24.9±13.2 and 12.8±6.6ng/mL. Prevalence of vitamin D deficiency (serum 25(OH)D levels less than 20ng/mL) was higher in samples analyzed using the Siemens method (60.2%) as compared to those tested using the Roche method (23.5%).

Conclusion

The assays evaluated are not comparable to each other. Laboratory technicians should inform clinicians of the features of the method used for measuring 25(OH)D because this will have a direct impact on interpretation of the results and medical decisions.

Keywords:
Vitamin D
Roche
Siemens
Methods comparison
Resumen
Introducción

La determinación en suero de 25-hidroxivitamina D [25(OH)D] es el mejor indicador del nivel de vitamina D en el organismo. El principal problema en su determinación es la precisión, reproducibilidad y la falta de estandarización de los métodos disponibles. Nuestro objetivo es comparar los resultados de 25(OH)D medidos por Elecsys Vitamin D Total (Roche) y ADVIA Centaur Vitamin D Total (Siemens)

Material y métodos

Se analizaron las concentraciones de 25(OH)D en 166 pacientes por ambos métodos. Tras el estudio del total de los pacientes, se consideraron 2 grupos: el grupo suplementado formado por los pacientes que recibían suplementos de vitamina D y el grupo sin tratamiento formado por el resto de pacientes.

Resultados

El valor medio de 25(OH)D en el grupo general medido por Roche y por Siemens fue 33,6±16,0 y 19,8±12,4ng/ml, respectivamente. El 54,2% de los pacientes recibían suplementos de vitamina D. En este grupo el valor medio de 25(OH)D medido por Roche y por Siemens fue 40,6±14,5 y 25,4±13,1ng/ml; y en el grupo de pacientes sin tratamiento fue 24,9±13,2 y 12,8±6,6ng/ml, respectivamente. La prevalencia de la insuficiencia de vitamina D (concentraciones séricas de 25(OH)D inferiores a 20ng/ml) fue más alta para las muestras medidas por Siemens (60,2%) que para las medidas por Roche (23,5%).

Conclusión

Los análisis evaluados no son comparables entre sí. Los especialistas del laboratorio deben informar a los clínicos de las características del método empleado en la determinación de 25(OH)D ya que tendrá consecuencias directas en la interpretación y decisión médica.

Palabras clave:
Vitamina D
Roche
Siemens
Comparación de métodos
Texto completo
Introduction

Interest in vitamin D has increased in recent years with advances in our understanding of its effects on both bone and other tissues, which may have significant physiological and pathological implications in different immune diseases, cancer, and cardiovascular disease.1–4

There are two forms of vitamin D, vitamin D3 or cholecalciferol, which is physiological in humans (the most abundant in circulation), and vitamin D2 or ergocalciferol, obtained from diet after eating food of vegetable origin. Both forms are available in pharmacological supplements or fortified foods.

Calcidiol or 25-hydroxyvitamin D [25(OH)D] is the main precursor of calcitriol, a biologically active metabolite of the vitamin D hormonal system. Its half-life and high serum concentration have made it the best marker of vitamin D levels.

The measurement of 25(OH)D poses some challenges, such as the high lipophilicity of the molecule, transport protein binding, the existence of two similar structural forms of different origin (endogenous or exogenous) with multiple metabolites,5 and the precision and reproducibility of the available methods.6

In 2010, Tai et al.7 developed a liquid chromatography–tandem mass spectrometry (LC–MS/MS) method which was later recognized as the reference method for 25(OH)D measurement by the Joint Committee for Traceability in Laboratory Medicine. Its complexity and the need for staff highly specialized in its use make it a procedure of little value in daily practice.

The recent availability of automated immunoassays makes it possible to cope with the increased demand for vitamin D measurement in health care.

The study objective was to compare the 25(OH)D levels measured using two of the currently marketed immunoassays and to check their impact on patient classification in terms of whether patients are regarded as having sufficient, deficient, or insufficient vitamin D levels.

Patients and methodsStudy setting

An observational, descriptive, cross-sectional study conducted at the biological diagnosis area of Hospital Universitario La Ribera (Alzira). This hospital covers a population of 260,000 inhabitants and is located in the Valencian Community (5,111,706 inhabitants).

Patients

Serum samples from 166 patients (57.2% females and 42.8% males; mean age, 64.7±15.5 years) attending Hospital Universitario La Ribera or its primary care center during May and June 2012 for routine blood tests including serum vitamin D measurement were tested. The study sample consisted of patients with chronic renal disease, parathyroid disease, osteoporosis, fractures, endocrine or autoimmune diseases with renal involvement, ricketts, or osteomalacia.

Ninety-two patients (52.4%) were receiving vitamin D supplements.

Samples were stored at −20°C until analysis. They were tested in batches, using a newly frozen aliquot for each test.

Method comparison

25(OH)D levels measured using the Elecsys Vitamin D Total assay in a Cobas e411 (Roche) and the ADVIA Centaur Vitamin D Total assay in an Advia Centaur XP (Siemens) were compared. Both methods were performed in accordance with the manufacturer's instructions.

Elecsys Vitamin D Total (Roche, Mannheim, Germany)

Elecsys Vitamin D Total (Roche) is a competitive protein binding immunoassay with electrochemiluminescence (ECLIA) detection. The latest version of this assay, marketed after modification by the manufacturer, was used.

Dithiothreitol and sodium hydroxide were used to free 25(OH)D from its binding plasma protein. Samples were then incubated with vitamin D binding protein labelled with ruthenium. Biotinylated 25(OH)D was subsequently added to occupy the vacant sites in the binding protein. When an electrical current is applied to the reading cell, the chemiluminescent reaction occurs.

The functional sensitivity of the assay is 4.0ng/mL (10.0nmol/L). Assay specificity, reflected by the percentage of cross reactivity with other metabolites, is 98% for 25(OH)D3, 81% for 25(OH)D2, and 93% for the C3 epimer of 25(OH)D3.

ADVIA Centaur Vitamin D Total (Siemens, Frimley Camberley, United Kingdom)

ADVIA Centaur Vitamin D Total (Siemens) is a one-step competitive immunoassay. This assay uses a releasing agent in saline buffer with sodium azide, an anti-fluorescein mouse monoclonal antibody covalently bound to paramagnetic particles, a mouse anti-25(OH)D monoclonal antibody labeled with acridinium ester, and a vitamin D analogue labeled with fluorescein.

The functional sensitivity of the assay is 3,3ng/mL (8.3nmol/L). Assay specificity, reflected by the percentage of cross reactivity with other metabolites, is 97.4% for 25(OH)D3, 106.2% for 25(OH)D2, and 1% for the C3 epimer of 25(OH)D3.

Statistical analysis

The results of the 166 samples were analyzed using a Bland–Altman plot, Passing-Bablok regression, concordance correlation coefficient (CCC), and kappa (k). Data distribution was analyzed using box plots. 25(OH)D values were expressed in ng/mL (1ng/dL is equivalent to 2.5nmol/L). Medcalc version 12.3.0 software was used for statistical analysis. Values of p<0.005 were considered statistically significant.

After testing all samples, two groups were established: a supplemented group, consisting of the 92 samples of patients receiving supplemental Vitamin D3, and an untreated group consisting of samples from the 74 patients not receiving vitamin D3 supplements. No patient was being treated with vitamin D2 supplements. Patient categorization was done after a literature review which suggested that some assays measuring 25(OH)D may not be able to recognize its exogenous fraction, i.e. 25(OH)D administered with pharmacological supplements or in diet.8

Bland–Altman analysis

Bland–Altman analysis9,10 is a graphic method for comparing two measurement procedures. In this method, differences (or proportions) between the two procedures are represented against the means of both. The limits of agreement are defined as the mean difference ±1.96 times the standard deviation of the differences. If these limits are not clinically relevant, either procedure may be used.

Passing-Bablok regression

Passing-Bablok regression11 is a linear regression procedure that calculates a regression equation including the 95% confidence interval (CI) of the intercept (A) and slope (B). This procedure requires that the variables are continuous and that a linear relationship exists between the two methods.

A is a measure of constant systematic differences between the two methods. The 95% CI of A may be used to test the hypothesis that A=0. If the hypothesis is rejected, there are constant systematic differences between them.

B is a measure of proportional systematic differences between the two methods. The 95% CI of B may be used to test the hypothesis that B=1. If the hypothesis is rejected, there are proportional systematic differences between them.

The residual standard deviation (RSD) is a measure of random differences between the two methods. Ninety-five percent of random differences are expected to be within the range of ±1.96 times RSD. If this range is wide, theymay not be comparable.

The linearity hypothesis is verified using the cumulative sum linearity test (CUSUM test). A value of p<0.05 suggests a significant deviation from linearity.

Concordance correlation coefficient

The concordance correlation coefficient (CCC)12,13 assesses the degree to which pairs of measurements fall on the line that crosses the origin at a 45° angle. It is the result of the product of Pearson's correlation coefficient (a measure of precision) and the bias correction factor (a measure of accuracy). Results higher than 0.99 suggest excellent concordance; those between 0.99 and 0.95, substantial concordance; values ranging from 0.90 to 0.94, moderate concordance; and those <0.9, poor concordance.14

Assessment of concordance of 25(OH)D concentrations

The position statement on vitamin D requirements and optimum concentrations prepared by the Spanish Society of Bone and Mineral Metabolism Research (SEIOMM) and other related societies was used as a guide.15 Vitamin D sufficiency was defined as serum 25(OH)D concentrations higher than 30ng/ml (>75nmol/L), vitamin D insufficiency as serum 25(OH)D concentrations ranging from 20 to 30ng/mL (50–75nmol/L), and vitamin D deficiency as serum 25(OH)D concentrations less than 20ng/mL (<50nmol/L).

Inter-assay agreement in assessment of vitamin D status was evaluated using k,16,17 with values of k less than 0.4 indicating poor agreement; values ranging from 0.4 to 0.75, fair to good agreement; and values higher than 0.75, excellent agreement.

ResultsAssay comparison

Table 1 summarizes the 25(OH)D values measured by Elecsys Vitamin D Total (Roche) and ADVIA Centaur Vitamin D Total (Siemens) and group characteristics. Individual data distribution in the overall group illustrates the significant differences between the assays for these samples (Fig. 1).

Table 1.

Demographic characteristics and statistical parameters of the patient groups studied by 25(OH)D assay.

  Overall group  Untreated group  Treated group 
Demographic characteristics
Total  166  74  92 
Males  71 (42.8%.  33 (44.6%)  38 (41.3%) 
Females  95 (57.2%)  41 (55.4%)  54 (58.7%) 
Age  64.7 (15.5)a  65.6 (16.2)a  64.1 (15.0)a 
  Roche  Siemens  Roche  Siemens  Roche  Siemens 
Statistical summary
Mean (ng/mL)  33.6  19.8  24.9  12.8  40.6  25.4 
95% CI  31.2–36.0  17.9–21.7  21.8.27.9  11.3–14.3  37.6–43.6  22.7–28.1 
Variance  254.5  153.4  174.4  43.0  210.2  172.3 
Standard deviation  16.0  12.4  13.2  6.6  14.5  13.1 
Median (ng/mL)  32.2  16.2  22.5  11.1  39.9  22.7 
95% CI  29.6–35.1  14.0–18.8  18.1–26.8  9.8–12.2  35.8–45.1  19.9–26.4 
Minimum (ng/mL)  5.0  4.5  5.0  4.5  9.9  7.3 
Maximum (ng/mL)  68.5  63.3  68.1  38.7  68.5  63.3 
Normal distribution  0.0005  <0.0001  0.0108  <0.0001  0.1088  0.0051 

CI: confidence interval.

a

Values with figures in brackets are mean and standard deviation (SD).

Figure 1.

Box plot showing the distribution of 25(OH)D concentrations (ng/mL) measured by the Elecsys Vitamin D Total (Roche) and the ADVIA Centaur Vitamin D Total (Siemens) assays.

(0.08MB).

Fig. 2 shows, using Bland–Altman analysis, agreement between the results obtained with the Elecsys Vitamin D Total (Roche) and ADVIA Centaur Vitamin D Total (Siemens) assays. The mean value of bias was higher for the treated group (15.2ng/mL [38.0nmol/L]) as compared to the untreated group (12.1ng/mL [30.3nmol/L]). Graphically, greater agreement was seen between the two assays when patients receiving vitamin D supplements were excluded. The scattering of the results obtained (positive trend in both groups) indicated that the degree of agreement decreased the higher the 25(OH)D concentration measured.

Figure 2.

Method comparison using Bland–Altman plots. Plots show bias in ng/mL (1ng/mL is equivalent to 2.5nmol/L). (A) All patients; (B) supplemented group; and (C) group without supplement.

(0.4MB).

Table 2 shows the results of both Passing-Bablok regression and CCC analysis.

Table 2.

Passing-Bablok and correlation concordance analyses between the Elecsys Vitamin D Total (Roche) and ADVIA Centaur Vitamin D Total (Siemens) assays in all patient groups.

Group  Passing-Bablok analysisCorrelation concordance analysis
  A  95% CI  B  95% CI  RSD  ±1.96 RSD  Cusum test  CCC  95% CI  r  Cb 
Overall  −1.41  −3.51 to −0.19  0.64  0.58–0.71  5.58  −10.94 to 10.94  p<0.01  0.56  0.49–0.62  0.85  0.66 
Treated  −8.75  −14.35 to −4.31  0.85  0.72–0.98  5.87  −11.51 to 11.51  p=0.13  0.51  0.41–0.60  0.82  0.62 
Untreated  2.59  1.52–3.75  0.40  0.35–0.45  3.62  −7.09 to 7.09  p=0.88  0.40  0.30–0.49  0.84  0.47 

A: intercept; B: slope; Cb: bias correction factor; CCC: concordance correlation coefficient; CI: confidence interval; r: Pearson's correlation coefficient; RSD: residual standard deviation.

Agreement in evaluation of vitamin D status

A comparison was made of the proportion of samples classified into the different vitamin D states based on serum 25(OH)D concentrations measured by both procedures.

If we consider the three possible vitamin D states proposed by the SEIOMM, similar k values were found both for the overall group and for the treated and untreated groups (k=0.2; k=0.1; k=0.1, respectively). If 20ng/mL (50nmol/L) was only used as the cut-off point, lower values being considered clearly pathological, similar k values were also found both overall and in the treated and untreated groups (k=0.3; k=0.2; k=0.2, respectively).

Siemens proposes different cut-off values for its ADVIA Centaur Vitamin D Total assay (10631296_ES Rev.C, 2011-01). This manufacturer defines vitamin D deficiency as serum 25(OH)D concentrations less than 10ng/mL (<25nmol/L). When this cut-off point was used to classify the results of this assay, similar k values were obtained both in the overall group and in the treated and untreated groups (k=0.8; k=0.7; k=0.8, respectively).

Discussion

In this study, lower 25(OH)D concentrations (mean value=19.8ng/mL [49.5nmol/L]) were obtained with the ADVIA Centaur Vitamin D Total assay (Siemens) as compared to the Elecsys Vitamin D Total assay (Roche) (mean value=33.6ng/mL [84.0nmol/L]).

Significant differences were found in all the statistical tests performed. The Bland–Altman analysis showed greater bias or systematic error in the group of patients treated with supplements (Fig. 2). Although acceptable values of Pearson's correlation coefficient were found (Table 2), the results of CCC lower than 0.9 found in all groups suggest poor agreement between these two methods. As regards Passing-Bablok regression, proportional and constant errors were simultaneously found in the study groups (Table 2).

These differences between the methods could be due to the assay design itself. If the specificity reported by the manufacturers is analyzed in more detail, special mention should be made of cross-reactivity to the C3 epimer of 25(OH)D3. Cross-reactivity is 93% and 1% for the Roche and Siemens assays, respectively. It is now known that this epimer is present in serum samples from both children and adults. The concentrations found range from 0.1ng/mL to 47.0ng/mL (0.25–117.5nmol/L),18,19 and some authors therefore state that their measurement in adult patients is of special interest when 25(OH)D levels are low.20 In addition, both methods use different pre-treatment steps to release 25(OH)D from vitamin D binding protein, which results in a lower level of 25(OH)D suitable for measurement. The design of the antibodies used in the Siemens assay may also influence the differences found. The monoclonal nature of these antibodies could result in the non-recognition of some derivative of exogenous vitamin D that does not express the epitope against which they were designed.

Our study has some limitations. One limitation was the use of frozen serum samples to compare the methods. Samples were frozen at −20°C and thawed at room temperature only once. However, prior studies on samples subject to multiple freezing–thawing cycles21 show that 25(OH)D concentrations should not have been affected.

An additional limitation was that no measurement was taken using the reference method. This limitation is common to most published studies, because the availability of LC–MS/MS in the clinical setting is restricted due to both the complexity of the method and to the increased demand for this measurement.

In prior studies, both methods showed a good correlation when compared to LC–MS/MS, but the results found with ADVIA Centaur Vitamin D Total (Siemens) showed a negative systematic error (−4.3%), while the results of Elecsys Vitamin D Total (Roche) showed a positive systematic error (17.4%).22

A final limitation was that not all assays available in Spain for 25(OH)D measurement were studied due to the lack of adequate means.

As regards patient classification, no agreement was found between the methods when values proposed by the SEIOMM were used (k<0.4). There were significant differences in the prevalence of vitamin D insufficiency between the tested assays. In all patient groups, a much higher prevalence was found with the Siemens assay (60.2%, 40.2%, and 85.1% in the overall, treated and untreated groups, respectively) as compared to the Roche assay (23.5, 7.6, and 43.2%, respectively). This means that some patients may receive a pharmacological treatment they do not need if the Siemens assay is used, or receive no treatment if the Roche assay is used. Agreement in patient classification improves (k=0.7–0.8) when the reference values for the ADVIA Centaur Vitamin D Total assay proposed by the manufacturer are used, although these do not agree with those proposed by scientific bodies.

It has been widely reported that the prevalence of vitamin D deficiency represents a public health problem affecting more than 50% of the general population.3 The prevalence of vitamin D deficiency in Spain is 28% in young people23 and approximately 65% in healthy adults.24 This generalized hypovitaminosis state may partly be influenced by the testing method used to measure 25(OH)D concentrations.

Because of these discrepancies, the standardization of immunoassays for measuring vitamin D with adequate reference material is an imperative need which has previously been suggested by other groups.25 With the launching of standard reference material (SRM) 972 by the US National Institute of Standards and Technology (NIST) and the implementation in 2010 of the Vitamin D Standardization Program (VDSP), the harmonization of all vitamin D assays is expected.26,27

In conclusion, the assays evaluated in this study are not comparable, despite their acceptable correlation with the reference method shown in prior publications. Patients should therefore be monitored with the same method throughout their treatment.

If the method used to assess vitamin D is changed, the laboratory should inform clinicians on the characteristics of the new method, because differences between the vitamin D levels measured by the different assays will have direct consequences on both medical interpretation and decision making.

Conflicts of interest

The authors state that they have no conflicts of interest.

References
[1]
M.T. Cantorna.
Vitamin D and autoimmunity: is vitamin D status an environmental factor affecting autoimmune disease prevalence?.
Proc Soc Exp Biol Med, 223 (2000), pp. 230-233
[2]
E. Giovannucci, Y. Liu, E.B. Rimm, B.W. Hollis, C.S. Fuchs, M.J. Stampfer, et al.
Prospective study of predictors of vitamin D status and cancer incidence and mortality in men.
J Natl Cancer Inst, 98 (2006), pp. 451-459
[3]
M.F. Holick.
Vitamin D: importance in the prevention of cancers, type 1 diabetes, heart disease, and osteoporosis.
Am J Clin Nutr, 79 (2004), pp. 362-371
[4]
T.J. Wang, M.J. Pencina, S.L. Booth, P.F. Jacques, E. Ingelsson, K. Lanier, et al.
Vitamin D deficiency and risk of cardiovascular disease.
Circulation, 117 (2008), pp. 503-511
[5]
B.W. Hollis.
Measuring 25-hydroxyvitamin D in a clinical environment: challenges and needs.
Am J Clin Nutr, 88 (2008), pp. 507S-510S
[6]
N. Binkley, D. Krueger, D. Gemar, M.K. Drezner.
Correlation among 25-hydroxy-vitamin D assays.
J Clin Endocrinol Metab, 93 (2008), pp. 1804-1808
[7]
S.S. Tai, M. Bedner, K.W. Phinney.
Development of a candidate reference measurement procedure for the determination of 25-hydroxyvitamin D3 and 25-hydroxyvitamin D2 in human serum using isotope-dilution liquid chromatography–tandem mass spectrometry.
Anal Chem, 82 (2010), pp. 1942-1948
[8]
J.E. Dowd, D. Stafford.
The vitamin D cure.
1st ed., John Wiley & Sons, (2008),
[9]
J.M. Bland, D.G. Altman.
Statistical methods for assessing agreement between two methods of clinical measurement.
Lancet, 1 (1986), pp. 307-310
[10]
J.M. Bland, D.G. Altman.
Measuring agreement in method comparison studies.
Stat Methods Med Res, 8 (1999), pp. 135-160
[11]
H.B. Passing.
A new biometrical procedure for testing the equality of measurements from two different analytical methods. Application of linear regression procedures for method comparison studies in clinical chemistry. Part I.
J Clin Chem Clin Biochem, 21 (1983), pp. 709-720
[12]
L.I. Lin.
A concordance correlation coefficient to evaluate reproducibility.
Biometrics, 45 (1989), pp. 255-268
[13]
J.J. Liao, J.W. Lewis.
A note on concordance correlation coefficient.
PDA J Pharm Sci Technol, 54 (2000), pp. 23-26
[14]
McBride GB. A proposal for strength-of-agreement criteria for Lin's Concordance Correlation Coefficient. NIWA Client Report: HAM 2005-062.
[15]
M.J. Gómez de Tejada, M. Sosa, J. del Pino, E. Jodar, J.M. Quesada, M.J. Cancelo, et al.
Sociedad Española de Investigación Ósea y del Metabolismo Mineral (SEIOMM) y Sociedades afines, Documento de posición sobre las necesidades y niveles óptimos de vitamina D.
Rev Osteoporos Metab Miner, 3 (2011), pp. 53-64
[16]
J. Cohen.
A coefficient of agreement for nominal scales.
Educ Psychol Meas, 20 (1960), pp. 37-46
[17]
J.L. Fleiss, B. Levin, M.C. Paik.
Statistical methods for rates and proportions.
3rd ed., John Wiley & Sons, (2003),
[18]
G. Lensmeyer, M. Poquette, D. Wiebe, N. Binkley.
The C-3 epimer of 25-hydroxyvitamin D(3) is present in adult serum.
J Clin Endocrinol Metab, 97 (2012), pp. 163-168
[19]
F.G. Strathmann, K. Sadilkova, T.J. Laha, S.E. leSourd, J.A. Bornhorst, A.N. Hoofnagle, et al.
3-epi-25 hydroxyvitamin D concentrations are not correlated with age in a cohort of infants and adults.
Clin Chim Acta, 413 (2012), pp. 203-206
[20]
D. Bailey, K. Veljkovic, M. Yazdanpanah, K. Adeli.
Analytical measurement and clinical relevance of vitamin D(3) C3-epimer.
Clin Biochem, (2012), pp. 2-12
[21]
D.M. Antoniucci, D.M. Black, D.E. Sellmeyer.
Serum 25-hydroxyvitamin D is unaffected by multiple freeze–thaw cycles.
Clin Chem, 51 (2005), pp. 258-261
[22]
H.W. Moon, J.H. Cho, M. Hur, J. Song, G.Y. Oh, C.M. Park, et al.
Comparison of four current 25-hydroxyvitamin D assays.
Clin Biochem, 45 (2012), pp. 326-330
[23]
M. Calatayud, E. Jódar, R. Sánchez, S. Guadalix, F. Hawkins.
Prevalencia de concentraciones deficientes e insuficientes de vitamina D en una población joven y sana.
Endocrinol Nutr, 56 (2009), pp. 164-169
[24]
J.M. Mata-Granados, M.D. Luque de Castro, J.M. Quesada Gómez.
Inappropriate serum levels of retinol, alpha-tocopherol, 25-hydroxyvitamin D3 and 24,25-dihydroxyvitamin D3 levels in healthy Spanish adults: simultaneous assessment by HPLC.
Clin Biochem, 41 (2008), pp. 676-680
[25]
N. Binkley, D. Krueger, S. Cowgill, L. Plum, E. Hansen, F. DeLuca.
Assay variation confounds the diagnosis of hypovitaminosis D: a call for standardization.
J Clin Endocrinol Metab, 89 (2004), pp. 3152-3157
[26]
C.J. Farrell, S. Martin, B. McWhinney, I. Straub, P. Williams, M. Herrmann.
State-of-the-art vitamin D assays: a comparison of automated immunoassays with liquid chromatography–tandem mass spectrometry methods.
Clin Chem, 58 (2012), pp. 531-542
[27]
C.T. Sempos, H.W. Vesper, K.W. Phinney, L.M. Thienpont, P.M. Coates.
Vitamin D Standardization Program (VDSP). Vitamin D status as an international issue: national surveys and the problem of standardization.
Scand J Clin Lab Invest Suppl, 243 (2012), pp. 32-40

Please cite this article as: Górriz Pintado S, Estela Burriel PL. Influencia del inmunoensayo empleado en la determinación de vitamina D sérica. Endocrinol Nutr. 2014;61:123–129.

Copyright © 2013. SEEN
Opciones de artículo
es en pt

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos