covid
Buscar en
Revista Portuguesa de Estomatologia, Medicina Dentária e Cirurgia Maxilofacial
Toda la web
Inicio Revista Portuguesa de Estomatologia, Medicina Dentária e Cirurgia Maxilofacial Torsional Failure Characteristics of a NiTi file based on a Case Report
Información de la revista
Vol. 51. Núm. 2.
Páginas 85-89 (abril - junio 2010)
Compartir
Compartir
Descargar PDF
Más opciones de artículo
Vol. 51. Núm. 2.
Páginas 85-89 (abril - junio 2010)
Caso Clínico
Open Access
Torsional Failure Characteristics of a NiTi file based on a Case Report
Visitas
4122
Jorge N.R. Martins
,
Autor para correspondencia
jnr_martins@yahoo.com.br

Correspondência para.
, Joseph DiBernardo**
* Doctor of Dental Surgery, University of Lisbon, Portugal; Private practice limited to Endondontics, Instituto de Implantologia, Lisboa
** Doctor of Dental Surgery, Stony Brook University, USA; Private practice limited to Endondontics, Smithtown, NY, USA
Este artículo ha recibido

Under a Creative Commons license
Información del artículo
Abstract

The use of nickel-titanium rotary files has been accepted worldwide. One of the major concerns about its use is the chance of file separation. Two types of failure have already been identified: cyclic fatigue and torsional failure. A scanning electron microscope analysis was performed on a recent case of file separation. The scan was used to show the main characteristics of torsional failure and to compare these characteristics with a known example of cyclic fatigue type failure. The clinical reasons that may lead to such failure are also discussed.

Key-words:
Nickel-titanium rotary file
Torsional failure
File separation
Scanning electron microscopy
Resumo

A instrumentação mecanizada com limas de níquel-titânio é usada mundialmente, sendo a separação e fractura destes instrumentos considerada uma grande preocupação. São documentados dois tipos de fracturas: por fadiga cíclica ou por fractura por torção. Uma análise de microscópio electrónico de varredura foi realizada num recente caso de fractura de uma lima. A análise é usada para documentar as características da fractura por torção e compará-la com a fadiga cíclica. As razoes clínicas que podem levar a este tipo de falha são também debatidas.

Palavras-Chave:
Limas mecanizadas níquel-titânio
Fractura por torção
Fractura de lima
Microscópio electrónico de varrimento
El Texto completo está disponible en PDF
References
[1-]
G.N. Glickman.
Nickel titanium rotary instruments for root canal preparation: are we there yet?.
Alpha Omegan, 90 (1997), pp. 29-38
[2-]
A.O. Peters.
Current challenges and concepts in the preparation of root canal systems: a review.
J Endodon, 30 (2004), pp. 559-567
[3-]
M.K. Iqbal, M.R. Kohli, J.S. Kim.
A retrospective clinical study of incidence of root canal instruments separation in an endodontics graduate program: a PennEndo Database study.
J Endodon, 32 (2006), pp. 1048-1052
[4-]
G.N. Tzanetakis, E.G. Kontakiotis, D.V. Maurikou, M.P. Marzelou.
Prevalence and management of instrument fracture in the postgraduate endodontic program at the Dental School of Athens: a five-year retrospective clinical study.
J Endodon, 34 (2008), pp. 675-678
[5-]
S. Wolcott, J. Wolcott, D. Ishley, W. Kennedy, S. Johnson, S. Minnich, J. Meyers.
Separation incidence of protaper rotary instruments: a large cohort clinical evaluation.
J Endodon, 32 (2006), pp. 1139-1141
[6-]
F.H. Klevant, C.O. Eggink.
The effect of the canal preparation on periapical disease.
Int Endod J, 16 (1983), pp. 68-75
[7-]
J.C. Donnelly.
Resolution of a periapical radiolucency without root canal filling.
J Endodon, 16 (1990), pp. 394-395
[8-]
O.A. Peters, C.I. Peters.
Cleaning and shaping the root canal system.
Pathways of the, pulp., 9th, edi., pp. 290-357
[9-]
B. Sattapan, G.J. Nervo, J.E. Palamara, H.H. Messer.
Defects in rotary nickel-titanium files after clinical use.
J Endodon, 26 (2000), pp. 161-165
[10-]
S.A. Thompson.
An overview of nickel-titanium alloys used in dentistry.
Int Endodon J, 33 (2000), pp. 297-310
[11-]
K.J. Anusavice.
Dental casting alloys.
Phillips’ Science of dental materials, 10th edi., pp. 491-524
[12-]
G. Kuhn, L. Jordan.
Fatigue and mechanical properties of nickel-titanium endodontic instruments.
J Endodon, 28 (2002), pp. 716-720
[13-]
E. Berutti, G. Cantatore.
Rotary instruments in nickel-titanium.
1st edi., pp. 518-547
[14-]
P. Parashos, H.H. Messer.
Rotary NiTi instrument fracture and its consequences.
J Endodon, 32 (2006), pp. 1031-1043
[15-]
J.Y. Blum, A. Cohen, P. Machtou, J.P. Micallef.
Analysis of forces developed during mechanical preparation of extracted teeth using ProFile NiTi rotary instruments.
Int Endodon J, 32 (1999), pp. 24-30
[16-]
L.F. Morgan, S. Montgomery.
An evaluation of the crown-down pressureles technique.
J Endodon, 10 (1984), pp. 491-499
[17-]
S.A. Thompson, P.H. Dummer.
Shaping ability of Quantec Series 2000 rotary nickel-titanium instruments in simulated root canals. Part 1.
Int Endodon J, 31 (1997), pp. 259-262
[18-]
X. Wei, J. Ling, J. Jiang, X. Huang, L. Liu.
Modes of failure of ProTaper nickel-titanium rotary instruments after clinical use.
J Endodon, 33 (2007), pp. 276-279
[19-]
E. Berutti, A.R. Negro, M. Lendini, D. Pasqualini.
Influence of manual preflaring and torque on the failure rate of ProTaper rotary instruments.
J Endodon, 30 (2004), pp. 228-230
[20-]
K.S. Herold, B.R. Johnson, C.S. Wenckus.
A scanning electron microscopy evaluation of microfractures, deformation and separation in EndoSequence and Profile nickel-titanium rotary files using an extracted molar tooth model.
J Endodon, 33 (2007), pp. 712-714

(Martins JNR, DiBernardo J. Torsional Failure Characteristics of a NiTi file based on a Case Report. Rev Port Estomatol Med Dent Cir Maxilofac 2010;51:85–89)

Copyright © 2010. Sociedade Portuguesa de Estomatologia e Medicina Dentária
Descargar PDF
Opciones de artículo
es en pt

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos