P-100 CORRELATION OF TWO TRANSIENT ELASTOGRAPHY EQUIPMENT FOR ESTIMATION OF LIVER STEATOSIS AND LIVER FIBROSIS
Visits
7
Raul Alfonso Gomez Mendoza1, Eva Juarez Hernandez2, Montserrat Berrospe Alfaro2, Regina Lopez De Cardenas Rojo2, Martha Helena Ramos Ostos3, Graciela Castro Narro1, Ivan Lopez Mendez1
1 Department of Gastroenterology and Hepatology, Medica Sur, Mexico City, Mexico
2 Translational Research Unit, Medica Sur, Mexico City, Mexico
3 CIDyT, Medica Sur, Mexico City, Mexico
This item has received
Article information
Special issue
This article is part of special issue:
Vol. 29. Issue S3

Abstracts of the 2023 Annual Meeting of the ALEH

More info
Conflict of interest

No

Introduction and Objectives

Transient elastography (TE) is an alternative diagnostic tool for estimating liver steatosis and fibrosis. Two ET devices, iLivTouch®/UAP and Fibroscan®/CAP, have shown similar diagnostic accuracy; however, no studies have evaluated the correlation of liver steatosis and fibrosis measurements between both devices in the Mexican population. Aim: To evaluate the correlation of measurement values of liver steatosis and liver fibrosis using iLivTouch®/UAP and Fibroscan®/CAP.

Patients / Materials and Methods

This prospective study included adult patients who attended a check-up unit. The evaluation of liver steatosis and liver fibrosis was performed using TE with two devices (iLivTouch® FT100 and Fibroscan® 502 Touch), meeting the reliability parameters. The correlation of measurements was evaluated with the Pearson correlation coefficient; meanwhile, a Student's t-test was performed to determine differences in dB/m and kPa means.

Results and Discussion

A total of 69 patients were included; 57% (n=40) were men, with a mean age and body mass index of 45±10 years and 22.4±4.5 kg/m2, respectively. The prevalence of diabetes mellitus was 7.2% (n=5). The mean dB/m and kPa were 266±42 dB/m and 5.8±1.2 kPa with iLivTouch® and 243±56 dB/m and 4.0±0.8 kPa with Fibroscan®. The prevalence of liver steatosis was 29% (n=20) by iLivTouch® and 20% (n=14) by Fibroscan® (p=0.32), while fibrosis was 49.3% (n=34) and 2.9%. (n=2), respectively, (p=0.0001). The mean difference for dB/m was -22.6, p≤0.0001, while for kPa, the difference was 1.72, p≤0.0001. According to the correlation analysis, this was r=0.73 (strong), p≤0.0001 for the estimation of liver steatosis, and r=0.35 (moderate), p=0.003 for the estimation of liver fibrosis.

Conclusions

TE devices iLivTouch® and Fibroscan® show a strong correlation for estimating liver steatosis and a moderate correlation for estimating liver fibrosis.