covid
Buscar en
Revista Brasileira de Cardiologia Invasiva (English Edition)
Toda la web
Inicio Revista Brasileira de Cardiologia Invasiva (English Edition) Immediate Results and Clinical Follow-up of Patients Undergoing Transcatheter Ao...
Journal Information
Vol. 20. Issue 3.
Pages 260-266 (June 2012)
Share
Share
Download PDF
More article options
Visits
1639
Vol. 20. Issue 3.
Pages 260-266 (June 2012)
Open Access
Immediate Results and Clinical Follow-up of Patients Undergoing Transcatheter Aortic-Valve Implantation
Visits
1639
Mohamad Said Ghandour1,
Corresponding author
ghandour@terra.com.br

Mohamad Said Ghandour. Rua Maestro Cardim, 769 – 1o subsolo – Bloco 1 – sala 73 – Paraíso – São Paulo, SP, Brasil – CEP 01323-001
, Júlio Flávio Meirelles Marchini2, Alfredo Moreira da Rocha Neto3, Eduardo Erudilho4, Nádia Mendonça Carnieto5, Bruno Stefani Lelis Silva6, Bruno Veras Bezerra7, Daniel Oliveira Neto Barbosa8, Ingrid Olah do Nascimento9, Henry Abensur10, Januário Manoel de Souza11, Marco Antônio Praça de Oliveira12, Sérgio Almeida de Oliveira13, Salvador André Bavaresco Cristovão14, Maria Fernanda Zuliani Mauro15, José Armando Mangione16
1 Trainee at Ariê Interventional Cardiology Team of the Hospital Beneficência Portuguesa de São Paulo. São Paulo, SP, Brazil
2 Trainee at Ariê Interventional Cardiology Team of the Hospital Beneficência Portuguesa de São Paulo. São Paulo, SP, Brazil
3 Trainee at Ariê Interventional Cardiology Team of Hospital Beneficência Portuguesa de São Paulo. São Paulo, SP, Brazil
4 Assistant Physician of Ariê Interventional Cardiology Team of the Hospital Beneficência Portuguesa de São Paulo. São Paulo, SP, Brazil
5 Assistant Physician of Ariê Interventional Cardiology Team of the Hospital Beneficência Portuguesa de São Paulo. São Paulo, SP, Brazil
6 Trainee at Ariê Interventional Cardiology Team of the Hospital Beneficência Portuguesa de São Paulo. São Paulo, SP, Brazil
7 Trainee at Ariê Interventional Cardiology Team of the Hospital Beneficência Portuguesa de São Paulo. São Paulo, SP, Brazil
8 Trainee at Ariê Interventional Cardiology Team of the Hospital Beneficência Portuguesa de São Paulo. São Paulo, SP, Brazil
9 Research Coordinator of Ariê Interventional Cardiology Team of the Hospital Beneficência Portuguesa de São Paulo. São Paulo, SP, Brazil
10 Head of Echocardiography Team of Hospital Beneficência Portuguesa de São Paulo. São Paulo, SP, Brazil.
11 Head of Cardiovascular Surgery Team of the Hospital Beneficência Portuguesa de São Paulo. São Paulo, SP, Brazil
12 Assistant Physician of Cardiovascular Surgery Team of the Hospital Beneficência Portuguesa de São Paulo. São Paulo, SP, Brazil.
13 Head of Cardiovascular Surgery Team of the Hospital Beneficência Portuguesa de São Paulo. São Paulo, SP, Brazil.
14 Assistant Physician of Ariê Interventional Cardiology Team of the Hospital Beneficência Portuguesa de São Paulo. São Paulo, SP, Brazil.
15 Assistant Physician of Ariê Interventional Cardiology Team of the Hospital Beneficência Portuguesa de São Paulo. São Paulo, SP, Brazil.
16 Head of the Interventional Cardiology Team of the Hospital Beneficência Portuguesa de São Paulo. São Paulo, SP, Brazil.
Ver más
This item has received

Under a Creative Commons license
Article information
Abstract
Full Text
Bibliography
Download PDF
Statistics
Figures (3)
Show moreShow less
Tables (4)
TABLE 1. Baseline clinical characteristics of patients undergoing transcatheter aortic valve implantation (TAVI)
TABLE 2. Pre-transcatheter aortic valve implantation echocardiographic characteristics
TABLE 3. Characteristics related to the procedure and in-hospital clinical events
TABLE 4. ECG abnormalities after transcatheter aortic valve implantation
Show moreShow less
ABSTRACT
Background

One third of the elderly population with symptomatic calcified aortic stenosis cannot undergo surgery due to their high operative risk. The transcatheter aortic-valve implantation (TAVI) has emerged as an alternative therapy for this group of patients.

Methods

All patients submitted to TAVI from November 2008 to April 2012 were included in our study. We report the baseline clinical characteristics, procedural data, hospital outcomes and clinical follow-up of this population. Definitions were based on the Valve Academic Research Consortium criteria.

Results

TAVI was performed in 23 patients, with 79 ± 6.7 years of age, and 56% were female. The EuroSCORE was 20.4 ± 11.1%. The CoreValve® prosthesis was used in 19 patients (82.6%) and the Edwards SAPIEN™ valve was used in the remaining ones. Procedure success rate was 96%. The mean follow-up was 22 ± 12.8 months, with 6 deaths (26.1%) in this period, 3 of which were observed in the first 30 days (13%) and other 2 (21.7%) by the end of the first year. One patient had a transient ischemic attack during hospitalization (4.3%), but there were no episodes of stroke or myocardial infarction in the periprocedural period or in the follow-up. The composite safety endpoint at 30 days was observed in 5 patients (21.7%) and the composite efficacy endpoint at 12 months was 78.3%.

Conclusions

The results of this study demonstrate that TAVI is an attractive procedure for the treatment of patients with calcified aortic stenosis and high operative risk.

RESUMO

Resultados Imediatos e Seguimento Clínico dos Pacientes Submetidosa Implante Valvar Aórtico Transcateter

Introdução

Um terço da população idosa portadora de estenose aórtica calcificada sintomática não apresenta condições cirúrgicas em decorrência do elevado risco operatório. O implante valvar aórtico transcateter (IVAT) surgiu como uma alternativa terapêutica para esses pacientes.

Métodos

Incluímos, no período de novembro de 2008 a abril de 2012, todos os pacientes submetidos a IVAT em nosso serviço. Relatamos as características clínicas basais, os dados dos procedimentos, os resultados hospitalares e o seguimento clínico dessa população. As definições utilizadas foram baseadas nos critérios do Valve Academic Research Consortium.

Resultados

O IVAT foi realizado em 23 pacientes, com 79+6,7 anos de idade, 56% do sexo feminino. O EuroSCORE foi de 20,4+11,1%. A prótese CoreValve® foi utilizada em 19 pacientes (82,6%) e a Edwards SAPIEN™ nos demais. A taxa de sucesso do procedimento foi de 96%. O tempo médio de seguimento clínico foi de 22+12,8 meses, observando-se 6 óbitos (26,1%) nesse período, 3 dos quais ocorreram nos primeiros 30 dias (13%) e outros 2 (21,7%), até o final do primeiro ano. Um paciente apresentou ataque isquêmico transitório na fase hospitalar (4,3%), mas não ocorreram episódios de acidente vascular encefálico ou de infarto do miocárdio no período periprocedimento ou no acompanhamento tardio. O desfecho combinado de segurança aos 30 dias ocorreu em 5 pacientes (21,7%) e o desfecho combinado de eficácia aos 12 meses foi de 78,3%.

Conclusões

Os resultados obtidos neste estudo demonstram o IVAT como procedimento atrativo para o tratamento de pacientes portadores de estenose aórtica calcificada de alto risco cirúrgico.

DESCRIPTORS:
Aortic valve stenosis
Heart valve prosthesis
Heart valve prosthesis implantation
DESCRITORES:
Estenose da valva aórtica
Próteses valvulares cardíacas
Implante de prótese de valva cardíaca
Full Text

Calcified aortic stenosis is the most commonly diagnosed valve disease in the United States and Europe,1 and its severe form affects 2% of the population older than 65 years of age.2 Considering that this is the rate of prevalent severe aortic stenosis in the Brazilian population, 380,000 cases are expected to be diagnosed by 2020 and approximately 1,000,000 cases by 2050.3

Initially considered to be natural valve tissue degeneration, current data supports the hypothesis that calcified aortic stenosis is caused by an inflammatory process, similar to atherosclerosis. The accumulation of oxidised lipoproteins is observed in the intima, as well as in endothelial activation and the recruitment of T lymphocytes and monocytes.4 Several mechanisms have been proposed for valve tissue calcification, and the transdifferentiation of myofibroblasts into osteoblasts is one of them.5

In adult patients with aortic stenosis, the obstruction to blood flow increases gradually over the years. During this period, there is a gradual adaptation of the left ventricle to pressure overload through the development of concentric myocardial hypertrophy. This results in diastolic dysfunction, reduced coronary flow reserve, myocardial ischaemia, and eventually left ventricular systolic dysfunction.6

Typically, patients with aortic stenosis remain asymptomatic for a long period of time; however, after the onset of symptoms, the prognosis is poor, with an interval to death of two years for those with heart failure, three years for syncope, and five years for angina pectoris.7

The standard treatment for symptomatic patients is surgical replacement of the aortic valve.8 However, approximately one-third of the elderly population with this disease cannot benefit from treatment due to the presence of comorbidities or other conditions that are unfavourable to surgery.9,10 These patients have adverse outcomes, with a one to two year survival rate of 50%.11,12

Transcatheter aortic valve implantation (TAVI), introduced by Cribier et al.,13 has emerged as an attractive alternative treatment for these patients, and its use has rapidly increased. The initial results were favourable,14,15 and the use of TAVI in patients with contraindications to surgery was shown to be superior to medical therapy.12,16 As such, this technique has been adapted by several departments throughout the world, beginning with this department in 2008. This study aims to present the immediate and long-term outcomes in this group of patients, using the criteria proposed by the Valve Academic Research Consortium (VARC).17

METHODSPatients

In November of 2008, a multidisciplinary team consisting of specialists in anesthesia, cardiology, echocardiography, haemodynamics, interventional radiology, and cardiac surgery at the Hospital Beneficência Portuguesa de São Paulo started the TAVI program. All patients at this hospital who underwent TAVI between November of 2008 and April of 2012 were included in this registry. This study was approved by the institution ethics and research committee under protocol number 669-11/CAAE: 0104.1.360.000-11.

Selection criteria

Individuals were selected based on clinical criteria and the morphological parameters of the aortic valve/ aorta and the access route to determine the technical feasibility of percutaneous valve replacement.18 For that purpose, transthoracic echocardiography with colour Doppler, multi-detector CT angiotomography, and angiography of the ascending aorta, iliac-femoral territory, and coronary arteries were performed.

Patients in this registry included those with an aortic valve area < 1cm2 or with an indexed aortic valve area ≤ 0.6cm2/m2, symptomatic patients with an aortic valve annulus diameter ≥ 20mm and ≤ 27mm, an ascending aortic diameter ≤ 45mm, and a logistic EuroSCORE ≥ 15% or those at high surgical risk as established by a multidisciplinary assessment, and considering the following comorbidities: hepatic cirrhosis, pulmonary hypertension > 60mmHg, extensively calcified aorta (porcelain aorta), recurrent pulmonary embolism, right ventricular failure, or refusal of the patient to undergo surgery.

Device and procedure description

One of two commercially available prostheses was used: the third generation CoreValve® prosthesis (Medtronic Inc. – Minneapolis, USA) or the Edwards SAPIEN™ XT prosthesis (Edwards Lifesciences Inc. – Irvine, California, USA).

The CoreValve® prosthesis (Figure 1) consists of three porcine pericardium leaflets mounted on a Nitinol self-expandable stent. The prosthesis diameters used were 26mm, for a valve annulus of 20mm to 23mm, and 29mm, for a valve annulus > 23mm to 27mm. The Edwards SAPIEN™ XT prosthesis (Figure 2) consists of three bovine pericardial leaflets mounted on a cobalt-chromium balloon-expandable stent. This device is available in Brazil with diameters of 23mm, for an annulus of 21mm to 23mm, and of 26mm, for a valve annulus > 23 to 24.5mm. Both systems use 18F diameter sheaths.

Figure 1.

– Aortography showing the CoreValve® prosthesis in the aortic position without regurgitation.

(0.11MB).
Figure 2.

– Aortography showing the Edwards SAPIEN™ XT prosthesis in the aortic position without regurgitation.

(0.11MB).

The procedures were performed under general anesthesia or sedation, depending on the patient’s clinical condition and the prosthesis chosen, because it is necessary to use transoesophageal echocardiography to guide prosthetic implantations with the Edwards SAPIENTM XT. All patients received 200mg of acetylsalicylic acid and 300mg of clopidogrel orally 24 hours before the procedure and were maintained for six months postprocedure with 100mg of acetylsalicylic acid and 75mg of clopidogrel daily. At the beginning of each procedure, unfractionated heparin was administered intravenously (100 IU/kg) with adequate corrections to achieve an activated clotting time between 300 and 350 seconds.

The implantation techniques for these two prostheses have been previously described in detail.14,19 Briefly, the retrograde access was used, preferably by puncture of the femoral artery. Following the selective catheterisation of the left ventricle, aortic valvuloplasty was performed with a balloon when using the Edwards SAPIEN™ XT prosthesis or, when necessary, when using the CoreValve® prosthesis. During balloon pre-dilation, a temporary pacemaker was used to raise the heart rate to between 180 beats to 220 beats per minute, thus preventing balloon displacement at the time of inflation. The stent was then implanted. At the end of the procedure, a control ascending aortography was performed to assess the presence of perivalvular regurgitation. Hemostasis was performed with the vascular occlusion device Perclose ProGlide (Abbott, Inc. – Abbott Park, USA). Patients were referred to the intensive care unit, where they remained for at least 48 hours. They also underwent preventive maintenance with a transvenous pacemaker during that period, as there was a possibility complete atrioventricular block.

Clinical follow-up was performed through medical consultation or by contacting patients by telephone.

Study definitions and endpoints

Procedural success was defined as the implantation of a properly functioning single prosthesis in the correct position.

The primary study endpoint was defined as the number of all-cause mortality at 30 days and at one year. The combined safety endpoint at 30 days was defined as the number of occurrences of death from any cause, major stroke, periprocedural myocardial infarctions, life-threatening bleeding, need for dialyses, or severe vascular complications within 30 days of the procedure. The combined efficacy endpoint at one year was defined as the number of survivals, free of death by any cause, treatment failure for aortic stenosis, or prosthesis dysfunction.

Other assessed outcomes were cardiovascular mortality at 30 days and one year, the incidence of stroke and acute myocardial infarction, the occurrence of major vascular complications, the need for implantation of a permanent pacemaker, and new left bundle of His branch block.

RESULTS

From November of 2008 to April of 2012, 23 patients were included in the registry, and their outcomes were followed for a mean period of 22 ± 12.8 months, ranging from three months to 44 months. The mean age was 79 ± 6.7 years, and 56% of the patients were female. All patients were white; 87% of individuals had hypertension, 65% had dyslipidemia, and 30% had diabetes mellitus.

The patients had an average aortic valve area of 0.53 ± 0.14cm2, 35% had reported episodes of syncope, 30% had angina, and 96% had dyspnoea on exertion, with 9% in functional class II, 74% in functional class III, and 17% in functional class IV, as categorised according to the criteria set forth by the New York Heart Association (NYHA). The mean logistic EuroSCORE was 20.4 ± 11.1%. The main baseline characteristics are presented in Table 1.

TABLE 1.

Baseline clinical characteristics of patients undergoing transcatheter aortic valve implantation (TAVI)

Characteristics  n = 23 
Female gender, n (%)  13 (56) 
Age, years  79 ± 6.7 
Functional class NYHA, n (%)
2 (9) 
III  16 (74) 
IV  4 (17) 
Syncope, n (%)  8 (35) 
Angina, n (%)  7 (30) 
Systemic arterial hypertension, n (%)  20 (87) 
Dyslipidaemia, n (%)  14 (61) 
Diabetes mellitus, n (%)  8 (35) 
Coronary artery disease*, n (%)  14 (70) 
Previous acute myocardial infarction, n (%)  4 (17) 
Previous coronary artery bypass grafting, n (%)  4 (17) 
Previous percutaneous coronary intervention, n (%)  10 (43) 
Creatinine clearance, mL/min  48 ± 17.12 
Peripheral vascular disease, n (%)  5 (22) 
Pulmonary hypertension, n (%)  5 (22) 
Chronic obstructive pulmonary   
disorder, n (%)  3 (13) 
Mean logistic EuroSCORE, %  20.4 ± 11.1 

n = number of patients; NYHA = New York Heart Association.

*

Stenosis > 50% on coronary angiography.

Systolic pulmonary artery pressure > 60mmHg.

Chronic use of bronchodilators or steroids for pulmonary disease or Spirometry with forced expiratory volume in one second < 60% predicted or PaO2 < 60mmHg in ambient air and PaCO2 & > 50mmHg in ambient air.

The annulus diameter of the aorta was 22.4 ± 2.2mm and the ascending aorta was 35.3 ± 3.9mm. The left ventricle ejection fraction was 66 ± 11%, with three patients showing left ventricular systolic dysfunction. The maximum peak gradient observed between the left ventricle and the aorta was 92.1 ± 23.2mmHg and the mean was 54.1 ± 14.5mmHg. Echocardiographic data are shown in Table 2.

TABLE 2.

Pre-transcatheter aortic valve implantation echocardiographic characteristics

Pre-implantation echocardiographic variables  n = 23 
Left ventricular ejection fraction, %  66 ± 11 
Aortic valve area, cm2  0.53 ± 0.14 
Peak systolic gradient,mmHg  92.1 ± 23.2 
Mean systolic gradient,mmHg  54.1 ± 14.5 
Aortic valve annulus,mm  22.4 ± 2.2 
Ascending aorta,mm  35.3 ± 3.9 
Aortic regurgitation, n (%)
Absent/mild  21 (91.3) 
Moderate  2 (8.7) 
Severe 

n = number of patients.

Procedural data

Sedation was used in 12 procedures (52%), while the remainder of the procedures were performed under general anesthesia. A right femoral artery puncture was used in 15 (65%), while the other patients underwent surgical dissection. The CoreValve® prosthesis was implanted in 19 cases and the Edwards SAPIEN™ XT, in four. A pre-dilatation balloon catheter was used in seven (37%) patients receiving the CoreValve® prosthesis. Postdilatation was performed in seven (30%) procedures. The main procedural data are shown in Table 3.

TABLE 3.

Characteristics related to the procedure and in-hospital clinical events

Variables  n = 23 
Access route, n (%)
Femoral puncture  15 (65) 
Femoral dissection  8 (35) 
Anaesthesia, n (%)
Sedation  12 (52) 
General anaesthesia  11 (48) 
Prosthesis used
CoreValve®, n (%)
26mm  8 (35) 
29mm  11 (48) 
Edwards SAPIEN™ XT
26mm  3 (13) 
29mm  1 (4) 
Pre-dilation CoreValve®, n (%)  7 (37) 
Post-dilation CoreValve®, n (%)/Edwards SAPIEN™ XT, n (%)  7 (30) 
Post-implantation aortic regurgitation degree, n (%)
0 or I  20 (87) 
II  2 (8.7) 
III  1 (4.3) 
Clinical events, n (%)
Cardiovascular death  3 (13) 
Stroke/TIA  1 (4.3) 
AMI 
Definitive pacemaker  4 (17.4) 
Vascular complication  2 (8.6) 
Arterial thrombosis  1 (4.3) 
Haematoma with blood transfusion  1 (4.3) 
Combined 30-day endpoint of safety  5 (21.7) 

TIA= transient ischaemic attack; AMI = acute myocardial infarction.

Technical success of the implant was achieved in 95.7% of cases. In one patient, the prosthesis failed to implant in the correct position, and progressed to moderate heart failure. This called for a repositioning of the prosthesis, which was successfully performed using a loop snare. The maximum gradient between the left ventricle and the aorta after valve implantation was 17.9 ± 7.3mmHg. The left ventricular ejection fraction achieved was 71% ± 6.6% after the procedure, with a percentage increase of 7.3% from before the procedure. Mild or absent aortic regurgitation was observed in 17 (74%) patients.

Clinical outcomes

The mean follow-up time for patients was 22 ± 12.8 months. Six deaths, regardless of cause, were observed (26.1%) during this period, three of which occurred in the first 30 days (13%) and two of which (21.7%) occurred by the end of the first year. There were three cardiovascular deaths, all in the first month after the procedure, three non-cardiovascular deaths at four months due to pneumonia, and another two at 12 months and 33 months after the procedure due to gastrointestinal complications.

The combined safety endpoint at 30 days was 21.7% (5/23). There were three cardiovascular deaths, one arterial thrombosis and one bleeding event requiring transfusion. One patient (4.3%) had a transient ischaemic attack in the hospital, but there were no episodes of stroke or myocardial infarction in both the periprocedural period or during the clinical follow-up period.

Among the study patients, four had pacemakers implanted before the procedure, four required definitive pacemaker implants in the hospital, and one patient had the device implanted three months after TAVI. Left bundle branch blockage before the procedure was observed in five patients, while nine patients (50%) presented new left bundle branch blockages. Electrocardiographic alterations after TAVI are described in Table 4.

TABLE 4.

ECG abnormalities after transcatheter aortic valve implantation

Electrocardiographic abnormalities after implantation
Complete left bundle blockage, n (%)  9/18 (50) 
AVB First degree, n (%)  3/19 (15.8)* 
AVB Second degree Mobitz type II, n (%)  1/19 (5.3)* 
Total AVB, n (%)  4/19 (21.1)* 
*

All cases with CoreValve® implant. AVB = atrioventricular block.

The functional class of patients during evolution is shown in Figure 3.

Figure 3.

– Functional class evolution of patients submitted to transcatheter aortic valve implantation. n = number of patients.

(0.12MB).

The combined efficacy endpoint at 12 months was 78.3% (18/23), with the occurrence of two additional non-cardiovascular deaths between 30 days and one year. There were no admissions due to aortic stenosis treatment failure during this period and no malfunctions were detected in the implanted prosthesis.

DISCUSSION

TAVI enables the approach of patients with surgical contraindications, such as high-risk patients or those with exclusionary features, such as a porcelain aorta. In the present sample, patients with high-risk profiles, those who were elderly (mean age of 79 years), and those with multiple risk factors for atherosclerosis and a EuroSCORE of 20.4 were selected.

TAVI was successful in 22 of the 23 patients. The individual defined as a failure presented moderate aortic regurgitation at the end of the procedure, but was resolved with prosthesis repositioning two days after the implantation.

A 13% overall mortality at 30 days is similar to previously published reports: Gilard et al.,20 9.7%; Lemos et al.,21 8.4%; Gurvitch et al.,22 9.4%; and Stähli et al.,23 11.5%. The 21.7% safety outcome obtained in this series is comparable to that of Gurvitch et al.22 (18.4%) and Stähli et al.23 (20.8%). This outcome evaluates events related to the technical complications of the procedure within 30 days. The 78.3% combined efficacy endpoint at one year was similar to the endpoint reported by Stähli et al.23 (70.2%). This outcome of effectiveness evaluates therapeutic failure-free survival, with the proper performance of the prosthesis.

TAVI provided marked clinical improvement, as shown in Figure 3. At baseline, 74% of patients were in functional class III and 17% were in functional class IV. Immediate after the operation and at their 30-day assessment, 94.7% and 91.3% of patients, respectively, were in functional class I or II, and remained at satisfactory levels in late follow-up.

Procedural completion through the femoral puncture technique and hemostasis using the Perclose ProGlide system was feasible, with only one vascular complication in the present study. One patient required surgical treatment for femoral artery thrombosis.

The main adverse effect observed with this technique was the need to implant a pacemaker, which occurred in four patients (21.1%), all of whom had received the CoreValve™ prosthesis. In comparison, this complication can occur in up to 8% of those undergoing the conventional surgical procedure.24,25 Some series have shown rates of up to 30.4% for patients needing pacemakers after TAVI.25

Study limitations

The study was limited by its observational nature, its single-centre design, and its small sample size.

CONCLUSIONS

These results demonstrate that TAVI is an attractive alternative for the treatment of severe calcified aortic stenosis in high-risk surgical patients. A thorough evaluation is crucial for procedural success and to reduce complications. However, the long-term effectiveness needs to be demonstrated, and technological developments may further improve outcomes and expand the indications for patients at lower risk.

CONFLICT OF INTEREST

The authors declare no conflicts of interest.

REFERENCES
[1.]
B. Iung, G. Baron, E.G. Butchart, F. Delahaye, C. Gohlke-Barwolf, O.W. Levang, et al.
A prospective survey of patients with valvular heart disease in Europe: the Euro Heart Survey on Valvular Heart Disease.
Eur Heart J, 24 (2003), pp. 1231-1243
[2.]
C.M. Otto, B.K. Lind, D.W. Kitzman, B.J. Gersh, D.S. Siscovick.
Association of aortic-valve sclerosis with cardiovascular mortality and morbidity in the elderly.
N Engl J Med, 341 (1999), pp. 142-147
[3.]
J.C. Oliveira, F.R.P. Albuquerque, I.B. Lins, Instituto Brasileiro de Geografia e Estatística (IBGE).
Projeção da população do Brasil por sexo e idade para o período 1980-2050 – revisão 2004. Rio de Janeiro.
[4.]
R.V. Freeman, C.M. Otto.
Spectrum of calcific aortic valve disease: pathogenesis, disease progression, and treatment strategies.
Circulation, 111 (2005), pp. 3316-3326
[5.]
N.M. Rajamannan, M. Subramaniam, D. Rickard, S.R. Stock, J. Donovan, M. Springett, et al.
Human aortic valve calcification is associated with an osteoblast phenotype.
Circulation, 107 (2003), pp. 2181-2184
[6.]
B.A. Carabello, W.J. Paulus.
Aortic stenosis.
[7.]
B.A. Carabello, L.H. Green, W. Grossman, L.H. Cohn, J.K. Koster, J.J. Collins Jr..
Hemodynamic determinants of prognosis of aortic valve replacement in critical aortic stenosis and advanced congestive heart failure.
Circulation, 62 (1980), pp. 42-48
[8.]
F. Tarasoutchi, M.W. Montera, M. Grinberg, M.R. Barbosa, D.J. Piñeiro, C.R.M. Sanches, et al.
Diretriz Brasileira de Valvopatias – SBC 2011. I Diretriz Interamericana de Valvopatias – SIAC 2011.
Arq Bras Cardiol, 97 (2011), pp. 1-67
[9.]
B. Iung, A. Cachier, G. Baron, D. Messika-Zeitoun, F. Delahaye, P. Tornos, et al.
Decision-making in elderly patients with severe aortic stenosis: why are so many denied surgery?.
Eur Heart J, 26 (2005), pp. 2714-2720
[10.]
B.J. Bouma, R.B. van Den Brink, J.H. van Der Meulen, H.A. Verheul, E.C. Cheriex, H.P. Hamer, et al.
To operate or not on elderly patients with aortic stenosis: the decision and its consequences.
Heart, 82 (1999), pp. 143-148
[11.]
P. Varadarajan, N. Kapoor, R.C. Bansal, R.G. Pai.
Clinical profile and natural history of 453 nonsurgically managed patients with severe aortic stenosis.
Ann Thorac Surg, 82 (2006), pp. 2111-2115
[12.]
M.B. Leon, C.R. Smith, M. Mack, D.C. Miller, J.W. Moses, L.G. Svensson, et al.
Transcatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery.
N Engl J Med, 363 (2010), pp. 1597-1607
[13.]
A. Cribier, H. Eltchaninoff, A. Bash, N. Borenstein, C. Tron, F. Bauer, et al.
Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis: first human case description.
Circulation, 106 (2002), pp. 3006-3008
[14.]
E. Grube, G. Schuler, L. Buellesfeld, U. Gerckens, A. Linke, P. Wenaweser, et al.
Percutaneous aortic valve replacement for severe aortic stenosis in high-risk patients using the second-and current third-generation self-expanding Corevalve prosthesis: device success and 30-day clinical outcome.
J Am Coll Cardiol, 50 (2007), pp. 69-76
[15.]
R. Sarmento-Leite, A.S. Quadros, P.R.L. Prates, L.G. Zanatta, P.A. Salgado Filho, T. Grando, et al.
Implante valvular aórtico percutâneo: experiência inicial do sul do Brasil.
Rev Bras Cardiol Invasiva, 16 (2008), pp. 398-405
[16.]
C.R. Smith, M.B. Leon, M.J. Mack, D.C. Miller, J.W. Moses, L.G. Svensson, et al.
Transcatheter versus surgical aortic-valve replacement in high-risk patients.
N Engl J Med, 364 (2011), pp. 2187-2198
[17.]
M.B. Leon, N. Piazza, E. Nikolsky, E.H. Blackstone, D.E. Cutlip, A.P. Kappetein, et al.
Standardized endpoint definitions for Transcatheter Aortic Valve Implantation clinical trials: a consensus report from the Valve Academic Research Consortium.
J Am Coll Cardiol, 57 (2011), pp. 253-269
[18.]
N. Piazza, P. de Jaegere, C. Schultz, A.E. Becker, P.W. Serruys, R.H. Anderson.
Anatomy of the aortic valvar complex and its implications for transcatheter implantation of the aortic valve.
Circ Cardiovasc Interv, 1 (2008), pp. 74-81
[19.]
J.G. Webb, M. Chandavimol, C.R. Thompson, D.R. Ricci, R.G. Carere, B.I. Munt, et al.
Percutaneous aortic valve implantation retrograde from the femoral artery.
Circulation, 113 (2006), pp. 842-850
[20.]
M. Gilard, H. Eltchaninoff, B. Iung, P. Donzeau-Gouge, K. Chevreul, J. Fajadet, et al.
Registry of transcatheter aorticvalve implantation in high-risk patients.
N Engl J Med, 366 (2012), pp. 1705-1715
[21.]
P.A. Lemos, F. Saia, A. Marzocchi, J. Mariani Jr., A. Esteves Filho, L.J. Kajita, et al.
Implante transcateter de prótese valvular aórtica: perfil clínico e evolução de uma série consecutiva de 75 casos do registro conjunto Universidade de Bolonha/ Hospital Sírio-Libanês.
Rev Bras Cardiol Invasiva, 19 (2011), pp. 24-27
[22.]
R. Gurvitch, S. Toggweiler, A.B. Willson, N. Wijesinghe, A. Cheung, D.A. Wood, et al.
Outcomes and complications of transcatheter aortic valve replacement using a balloon expandable valve according to the Valve Academic Research Consortium (VARC) guidelines.
EuroIntervention, 7 (2011), pp. 41-48
[23.]
B.E. Stähli, R. Bunzli, J. Grunenfelder, I. Buhler, C. Felix, D. Bettex, et al.
Transcatheter Aortic Valve Implantation (TAVI) outcome according to standardized endpoint definitions by the Valve Academic Research Consortium (VARC).
J Invasive Cardiol, 23 (2011), pp. 307-312
[24.]
Z. El-Khally, B. Thibault, C. Staniloae, P. Theroux, M. Dubuc, D. Roy, et al.
Prognostic significance of newly acquired bundle branch block after aortic valve replacement.
Am J Cardiol, 94 (2004), pp. 1008-1011
[25.]
P. Kolh, L. Lahaye, P. Gerard, R. Limet.
Aortic valve replacement in the octogenarians: perioperative outcome and clinical follow-up.
Eur J Cardiothorac Surg, 16 (1999), pp. 68-73
Copyright © 2012. Elsevier Editora Ltda. and Sociedade Brasileira de Hemodinâmica e Cardiologia Intervencionista
Article options