metricas
covid
Buscar en
Revista Española de Cirugía Ortopédica y Traumatología (English Edition)
Toda la web
Inicio Revista Española de Cirugía Ortopédica y Traumatología (English Edition) Pedicular dynamic stabilization system. Functional outcomes and implant-related ...
Journal Information
Vol. 58. Issue 2.
Pages 85-91 (March - April 2014)
Visits
2874
Vol. 58. Issue 2.
Pages 85-91 (March - April 2014)
Original Article
Full text access
Pedicular dynamic stabilization system. Functional outcomes and implant-related complications for the treatment of degenerative lumbar disc disease with a minimum follow-up of 4 years
Sistema pedicular de estabilización dinámica. Resultados funcionales y complicaciones del implante en pacientes con enfermedad degenerativa discal lumbar, tras un seguimiento mínimo de 4 años
Visits
2874
M. Segura-Trepichioa,
Corresponding author
manusegura5@gmail.com

Corresponding author.
, D. Ferrández-Sempereb, F. López-Pratsa, J. Segura-Ibáñezb, L. Maciá-Solerc
a Departamento de Cirugía Ortopédica y Traumatología, Hospital General Universitario de Elche, Elche, Alicante, Spain
b Departamento de Neurocirugía, Hospital General Universitario de Elche, Elche, Alicante, Spain
c Departamento de Ciencias de la Salud, Universidad Jaume I de Castellón, Castellón de la Plana, Spain
This item has received
Article information
Abstract
Full Text
Bibliography
Download PDF
Statistics
Figures (5)
Show moreShow less
Tables (2)
Table 1. Baseline characteristics of the patients in the sample.
Table 2. Clinical results and complications of the implant after 4 years follow-up.
Show moreShow less
Abstract
Introduction

The Dynesys® system is a non-fusion pedicular dynamic stabilization system. The aim of our study is to evaluate the clinical outcomes in patients with degenerative disc disease and/or stenosis, and to measure the prevalence of screw loosening and breakage after 4 years of follow up.

Material and methods

All patients who underwent surgery with Dynesys® system in 2008 were reviewed. The surgery was performed in cases of low back pain of more than 6 months duration and a positive MRI for degenerative disc disease and/or stenosis.

Results

A total of 22 patients (11 females, 11 males) with a mean age of 44.40±11 years were included, 20 patients (91%) underwent Dynesys® without any associated decompression maneuver. The evaluation of back and leg pain (0–10mm) showed a mean decrease of 2.4±2.06mm (P=.0001). The preoperative value of the Oswestry disability index was 52.36±16.56% (severe functional limitation). After surgery, this value was 34.27±17.87% (moderate functional limitation) (P=.001) with a decrease of 18.09±16.03% (P=.001). A total of 4 (18%) patients showed signs of loosening screws. One patient (4.5%) had a screw breakage.

Conclusions

Surgery with Dynesys® shows favorable long term clinical results, however the range of improvement in our series is lower than those reported in other studies. Comparative studies between Dynesys® and decompression need to be performed in order to isolate the benefit of the dynamic stabilization system. Implant-related complications are not uncommon.

Keywords:
Pedicular dynamic stabilization system
Degenerative disc disease
Lumbar stenosis
Functional outcomes
Screw loosening
Resumen
Introducción

El sistema de estabilización Dynesys® es un sistema pedicular de estabilización dinámica sin fusión. El objetivo de nuestro estudio es evaluar los resultados clínicos en pacientes con enfermedad degenerativa discal y/o estenosis, así como medir la prevalencia de aflojamiento de tornillos tras 4 años de seguimiento.

Material y métodos

Se trata de un estudio de serie de casos retrospectivo donde fueron incluidos todos los pacientes intervenidos desde enero a diciembre de 2008 con Dynesys®. Se indicó la cirugía si presentaban dolor lumbar de más de 6 meses de evolución y una RM positiva para enfermedad degenerativa discal y/o estenosis.

Resultados

Veintidós pacientes (11 mujeres y 11 varones) con una edad media de 44,40±11 años fueron evaluados. Veinte pacientes (91%) recibieron el implante Dynesys® sin ninguna maniobra de descompresión asociada. La evaluación del dolor de espalda y piernas (0–10mm) registró una disminución media de 2,4±2,06mm (p=0,0001). El valor preoperatorio del índice de discapacidad de Oswestry fue de 52,36±16,56% (limitación funcional severa). Tras la cirugía este valor fue de 34,27±17,87% (limitación funcional moderada) con una disminución de 18,09±16,03% (p=0,001). Cuatro pacientes (18%) mostraron signos de aflojamiento de tornillos. Un paciente (4,5%) presentó rotura de tornillo.

Conclusiones

La cirugía con Dynesys® muestra resultados clínicos favorables, sin embargo el rango de mejoría en nuestra serie es menor a los comunicados por otros autores. Estudios comparativos entre Dynesys® y descompresión deberían realizarse para poder aislar el beneficio de la estabilización dinámica del obtenido por la descompresión. Las complicaciones relacionadas con el implante no son infrecuentes.

Palabras clave:
Sistema de estabilización dinámica
Enfermedad degenerativa discal
Estenosis de canal lumbar
Resultados funcionales
Aflojamiento de tornillos
Full Text
Introduction

Degenerative spine diseases have a great impact on functional capacity and pain. Among them is degenerative disc disease, associated or not with spinal canal stenosis. Traditionally, the accepted surgical treatment for these diseases has been decompression associated to fusion.1 Such treatment has been associated to certain complications, such as loosening of screws, breakage of stems, hypermobility of the adjacent segment and nonunion of the instrumented segment, which in many cases lead to revision surgery.2 Several dynamic stabilization systems, like the Dynesys® system (Zimmer, Inc., Warsaw, IN, USA), have been developed in order to avoid these undesirable effects. The Dynesys® dynamic stabilization system was presented by Dr. Gilles Dubois and was first used in France in 1994.3,4 It was introduced into clinical practice in Europe in the year 2000 and was approved in the USA in 2009 to provide spinal stabilization for patients with degenerative spinal diseases causing lumbar and radicular pain.5,6 The system replaces rigid fusion stems by pedicular screws manufactured with Ti-Al-Nb and joined together by polyethylene terephthalate (Sulene-PET®) strings running through the center of a polycarbonate-urethane (Sulene-PCU®) cylindrical spacer, thus allowing a certain degree of flexibility of the instrumented segments7 (Fig. 1). There is scarce literature assessing the functional outcomes and complications of long-term implantation in patients with degenerative lumbar disc disease. Only 1 study has been published in our country.8 The aim of our study is to evaluate the functional and lumbar pain results in these patients, and to evaluate the complications related to the material by measuring the prevalence of breakage and loosening of screws after a follow-up period of 4 years.

Figure 1.

Dynesys® dynamic stabilization system.

(0.23MB).
MethodsType of study

This was a retrospective case series study based on registry data conducted at Hospital General Universitario of Elche with patients undergoing interventions with the Dynesys® technique between January and December 2008 and with postoperative follow-up during the 4 years after surgery.

Inclusion criteria

Patients with lumbar and/or radicular pain resistant to conservative treatment with non-opioid analgesics (NSAIDs, paracetamol, metamizole), codeine, tramadol and/or buprenorphine, oxycodone and/or physical therapy for at least 6 months.

Positive magnetic resonance imaging (MRI) scan for disc degeneration in stages 2, 3 and 4 of the Pfirrmann classification9 (Figs. 2 and 3).

Figures 2 and 3.

Patient suffering degenerative disc disease, Pfirrmann 3 and 4, in segments L2–L3–L4, stabilized with the Dynesys® system. The image shows the MRI obtained before (above) and after (below), with reabsorption of a small L2–L3 disc protusion (white arrows).

(0.18MB).

Changes in disc morphology consisting in disc prolapse, disc hernia and/or lumbar canal stenosis.

Exclusion criteria

  • Patients with degenerative disc changes of Pfirrmann grade 5.

  • Disc extrusion or disc sequestration.

  • Spondylolisthesis.

  • Lumbar scoliosis greater than 10 degrees.

  • Presence of malignant tumors.

  • Body mass index over 30kg/m2.

  • History of alcohol or drug abuse.

All patients gave their informed consent to be assessed and the study was approved by the Bioethics Committee of our hospital.

Surgical procedure

All surgical procedures were performed by the same surgeon (D.F-S) through a Wiltse paramedian approach.10 The titanium screws were anchored in the pedicle and vertebral body preserving the facet joints, under intraoperative radiographic control. After the pedicular screws were placed, the interpedicular distance was measured with the dynamometric compass included in the instrumentation, and distraction was performed up to mark number 1 of the 3 preset marks in the instrument. Next, the Sulene-PCU® cylinder was cut to the resulting length and introduced using the Sulene-PET® string (Fig. 4. In those cases requiring associated nerve root decompression, this was performed through an incision in the midline and unilateral or bilateral laminotomy, preserving the spinous apophysis and supra- and interspinous ligament.

Figure 4.

Dynamometric compass used to measure the length of the interpedicular spacer cylinder (the circle indicates the 3 markings or distraction strength levels. The greater the strength, the greater the distraction and, therefore, the greater the length of the spacer cylinder).

(0.09MB).

Sitting and walking were allowed on the day after surgery, without a lumbar orthosis or any limitation of activities of daily living.

Clinical and radiographic variables were collected from the medical records of patients and a personal interview conducted during the last outpatient visit.

Clinical assessment

Functional outcomes were measured using the Oswestry disability index (ODI) in its Spanish version, whilst back and leg pain was assessed by means of a visual analog scale (VAS).11 The ODI is represented on a scale of 0–100%, where higher values describe greater functional limitations. Between 0% and 20%: minimal functional limitations; 20–40%: moderate; 40–60%: severe; 60–80%: disability, and above 80%: maximum functional limitation.

Radiographic assessment

Lumbar radiographs in anteroposterior and lateral projections were obtained using the Risolution® (MediAlfa Corp., Miami, FL, USA) digitized system. Each image was magnified and contrast was adjusted to obtain the best possible visualization. An orthopedic surgeon (M.S-T), reviewed the presence of breakage and loosening of the screws, considering as positive the presence, in both projections, of a radiolucent circumference around the screw of 1mm or more, surrounded by a radiopaque bone line with greater density; “Double halo sign”12 (Fig. 5).

Figure 5.

Right L4 screws showing a “double halo sign” (radiolucent line surrounded by another radiopaque line of bone with greater density).

(0.09MB).
Statistical analysis

We used the mean and standard deviation for quantitative variables with normal distribution and the frequency for categorical variables. We performed the Wilcoxon test for the analysis of repeated functional outcomes over time and the Mann–Whitney test was used to detect differences between different groups. The level of statistical significance was set at P<.05. Data were assessed using the software package SPSS® version 15.0 (SPSS Inc., Chicago, IL, USA).

Results

The total sample consisted of 23 patients, of whom 22 were included in the study (1 patient could not be examined due to a change of residence). The gender distribution was 11 females and 11 males, and the mean age was 44.40±11 years. The mean follow-up period was 53.45 months (range: 48–60 months). A total of 20 patients (91%) were implanted with the Dynesys® system through a Wiltse paramedian approach without any associated decompression maneuvers, whilst 2 patients (9%) were implanted after a unilateral laminectomy and discectomy due to a herniated disc. The mean body mass index (BMI) was 26.74±2.57kg/cm2. A monosegmental stabilization was performed in 8 patients (36.4%), whilst 2 segments were instrumented in 11 (36.4%) cases, and 3 segments in 3 cases (13.6%) (Table 1).

Table 1.

Baseline characteristics of the patients in the sample.

Age (years)a  44.40±11 
Genderb
Male  11/22 (50%) 
Female  11/22 (50%) 
BMI (kg/m2)  26.74±2.57 
VAS (010mm)  8.04±1.28 
ODI (0%–100%)  52.36±16.56 
Instrumented lumbar segments  1 level 8 (36.4%) 
  2 levels 11 (50%) 
  3 levels 3 (13.6%) 
Follow-up (months)  57.13±8.06 
Intervention
Dynesys® without decompression  20 (91%) 
Dynesys® with decompression c  2 (9%) 

BMI: body mass index; ODI: Oswestry disability index; VAS: analog visual scale.

a

Data expressed as mean±standard deviation.

b

Data expressed as number of subjects (percentage).

c

Decompression was carried out as a partial laminectomy following placement of the Dynesys® device.

Clinical monitoring

The assessment of back and leg pain by a VAS (0–10mm) improved from a preoperative value of 8.04±1.28 to 5.63±1.48, with a mean decrease of 2.4±2.06mm (P=.0001) (Table 2).

Table 2.

Clinical results and complications of the implant after 4 years follow-up.

Clinical results  Preoperative  Postoperative  Mean decrease 
VAS (0–10mm)a  8.04±1.28  5.63±1.48  2.4±2.06 P<.001* 
ODI (0–100%)a  52.36±16.56  34.27±17.87  18.09±16.03 P<.001* 
Screw looseningb    4/22 (18%) patients   
Screw breakageb    1/22 (4.5%) patients   

ODI: Oswestry disability index; VAS: analog visual scale.

Significant difference using the Wilcoxon test, P<.05.

a

Data expressed as mean±standard deviation.

b

Data expressed as number of subjects (percentage).

The preoperative value of the ODI on the absolute scale from 0% to 100% was 52.36±16.56% (severe functional limitation). After surgery, this value was 34.27±17.87% (moderate functional limitation), with a mean decrease of 18.09±16.03% (P=.001). The results were slightly better in the 2 patients in whom laminectomy was associated, with a decrease in the ODI of 30% in one case and 20% in the other.

When comparing the clinical outcomes by gender, males presented better results on the ODI scale, without this difference being statistically significant (P=.075). No correlation between functional outcome and BMI or age was found.

A total of 19 patients (86.4%) reported that their symptoms improved after surgery, while 3 patients (13.6%) reported a worse outcome after surgery. Of these 3 patients, 2 worsened their ODI scores by 10 points, while the third patient worsened by 30 points. The 3 patients were female and did not differ in age and BMI compared to patients who reported improvement.

Radiographic monitoring

A total of 4 (18%) patients presented signs of loosening of screws. The improvement in ODI among these patients was lower than in patients without these signs, specifically 9.75 points compared to 19.94 points (on the absolute scale of 0–100%). This difference was not statistically significant (P=.10).

One patient (4.5%) suffered breakage of 1 screw. The patient did not suffer a worsening of clinical status relative to the preoperative condition (Figs. 6 and 7).

Figures 6 and 7.

Tear of the right L5 screw of the Dynesys® dynamic stabilization system.

(0.19MB).
Discussion

This retrospective observational study presents the experience of the authors with the Dynesys® system. According to the study published by Ostelo et al. (establishing the relevant minimum difference as 10 points in the ODI and 1.5 points in the VAS scale) patients included in our study achieved a clinically relevant reduction of pain and disability after surgery.13 When we compare our results with those reported by other published series of cases, we can see that the margin of improvement in our patients was not as extensive as those reported by other studies. In this regard, Lee et al.14 published the largest decrease in the ODI recorded to date, with a variation of 57.43%, from 79.58% to 22.17% (P<.001), in 19 patients with spondylolisthesis and lumbar stenosis. Sapkas et al.15 also obtained better results than those in our study, with a variation in the ODI of 35% in 114 patients with degenerative disc disease and spinal canal stenosis. Recently, Fay et al.16 have reported a decrease of 4.1 points on the VAS scale (P<.001) and 23.3% in the ODI (P<.001) in patients with similar characteristics to the previous set. These differences between the aforementioned studies and our own, assuming that the patients were comparable, could be due to the fact that the former associated decompression in 100% of patients, whereas in our study only 2 (9%) patients underwent decompression along with the Dynesys® device. It seems clear that, in cases in which a decompressive maneuver is associated, this could act as a confounding factor, making it impossible to isolate the effect of Dynesys® per se. In this sense, our study, in which 91% of patients did not undergo decompression, provides an approximate reference to the functional improvement that may be obtained by using the Dynesys® device alone.

The ODI results were slightly better in the 2 cases (9%) in which decompression was associated. This is consistent with the results published by Grob et al. and Bothmann et al.,17,18 who found better results when dynamic stabilization was combined with decompression of the spinal canal or nerve root. All this evidence suggests that, as is the case with fusion, the main maneuver is decompression, with dynamic stabilization providing scarce additional benefit in those cases requiring decompression of the spinal canal or foramen due to a stenosing lesion.19

Dynesys® versus fusion

The Dyensys® system has not demonstrated superior functional results compared to fusion. In this regard, Yu20 found no significant differences in the ODI when comparing a group of patients treated with the Dynesys® system versus another group treated with posterior interbody fusion (P=.254). Moreover, during the clinical trial conducted to obtain approval of the Dynesys® system by the FDA in the USA in 2009 the Dynesys® system was compared against rigid posterolateral fusion and the difference in the functional results was not statistically significant at 2 years follow-up (P=.34). However, it should be mentioned that some authors argue that, with all results being equal, the Dynesys® system provides other benefits, as it does not require an extraction of iliac crest graft and it also shortens the surgical time.21

In the Spanish context, 1 study has been published which reported positive results regarding the improvement in disability at 2 years follow-up. However, this study did not specify whether these differences were statistically significant nor the percentage of patients undergoing decompression together with the Dynesys® system.10 At an international level, only 4 studies have published results with the Dynesys® system with more than 4 years follow-up.15,21–23 In this sense, the long-term results are still uncertain, with the most extensive records to date being those reported by Hoppe,22 with 7.1 years follow-up.

In the analysis of screw loosening we obtained similar results to those published by other authors with this technique. In this sense, Wu et al.24 found 19.8% of cases with radiographic signs of loosening among 126 patients treated with this technique. Some authors argue that the percentage of screw loosening with the dynamic system may be lower than with rigid systems because the flexible stems would allow some degree of mobility, reducing the load on the pedicular screws.23 However, the studies by Yu et al.20,25 compared these 2 systems and obtained a similar percentage of screw loosening with the Dynesys® system compared to the rigid fusion system; 14.3% versus 20% of patients, respectively (P=.728).

Limitations

There are several stages within degenerative disc disease.9 We present our experience in cases suffering mild disc degeneration, Pfirrmann stages 2, 3 and 4, and with healthy joints facets, where most cases (91%) were treated with the Dynesys® device alone. In our clinical practice we would opt for fusion surgery and/or decompression in those cases presenting a more advanced stage of disc degeneration and/or facet arthrosis. This means that we must be cautious when comparing our results to other populations with different stages of disc disease and undergoing interventions which combine dynamic stabilization after decompression.

Conclusion

Surgery with the Dynesys® device presents favorable long-term clinical outcomes in patients with degenerative lumbar disc disease. However, the range of improvement in our series was lower than those reported by other studies. This could be due to the difference in the percentage of decompressions associated with Dynesys® in the results reported by other authors. Comparative studies between the Dynesys® device and decompression should be performed in order to isolate the benefit offered by dynamic stabilization from that offered by decompression. Complications associated with loosening of the material are not uncommon.

Level of evidence

Level of evidence IV.

Ethical responsibilitiesProtection of people and animals

The authors declare that this investigation adhered to the ethical guidelines of the Committee on Responsible Human Experimentation, as well as the World Medical Association and the Declaration of Helsinki.

Confidentiality of data

The authors declare that they have followed the protocols of their workplace on the publication of patient data and that all patients included in the study received sufficient information and gave their written informed consent to participate in the study.

Right to privacy and informed consent

The authors declare having obtained written informed consent from patients and/or subjects referred to in the work. This document is held by the corresponding author.

Conflict of interests

The authors declare that they do not have any financial or personal relationship with any institution related to the content of this publication which could entail a conflict of interests.

References
[1]
R. Chou, J. Baisden, E.J. Carragee, D.K. Resnick, W.O. Shaffer, J.D. Loeser.
Surgery for low back pain: a review of the evidence for an American Pain Society Clinical Practice Guideline.
Spine (Phila Pa 1976), 34 (2009), pp. 1094-1109
[2]
J. Beastall, E. Karadimas, M. Siddiqui, M. Nicol, J. Hughes, F. Smith, et al.
The Dynesys lumbar spinal stabilization system: a preliminary report on positional magnetic resonance imaging findings.
Spine (Phila Pa 1976), 32 (2007), pp. 685-690
[3]
T.M. Stoll, G. Dubois, O. Schwarzenbach.
The dynamic neutralization system for the spine: a multi-center study of a novel non fusion system.
Eur Spine J, 11 (2002), pp. S170-S178
[4]
B. Dubois, B. de Germay, N.S. Schaerer, P. Fennema.
Dynamic neutralization: a new concept for restabilization of the spine.
Lumbar segmental instability, pp. 233-240
[5]
C.C. Würgler-Hauri, A. Kalbarczyk, M. Wiesli, H. Landolt, J. Fandino.
Dynamic neutralization of the lumbar spine after microsurgical decompression in acquired lumbar spinal stenosis and segmental instability.
Spine (Phila Pa 1976), 33 (2008), pp. E66-E72
[6]
Food and Drug Administration (FDA). FDA Executive Summary for Zimmer Spine's Dynesys Spinal System, Orthopedic and Rehabilitation Devices Panel [accessed 10 Mar 2010]. Available in: http://www.fda.gov/downloads/advisorycommittees/committeesmeetingmaterials/medicaldevices/medicaldevicesadvisorycommittee/orthopaedicandrehabilitationdevicespanel/ucm188734.pdf.
[7]
C.A. Niosi, Q.A. Zhu, D.C. Wilson, O. Keynan, D.R. Wilson, T.R. Oxland.
Biomechanical characterization of the three-dimensional kinematic behaviour of the Dynesys dynamic stabilization system: an in vitro study.
Eur Spine J, 15 (2006), pp. 913-922
[8]
M. Bordes-Monmeneu, V. Bordes-García, F. Rodrigo-Baeza, D. Baeza.
System of dynamic neutralization in the lumbar spine: experience on 94 cases.
Neurocirugia (Astur), 16 (2005), pp. 499-506
[9]
C.W. Pfirrmann, A. Metzdorf, M. Zanetti, J. Hodler, N. Boos.
Magnetic resonance classification of lumbar intervertebral disc degeneration.
Spine (Phila Pa 1976), 26 (2001), pp. 1873-1878
[10]
T. Nydegger, G. Dubois.
Dynamic neutralization system for the spine: Biomechanical aspects.
Proceedings of the International 19th Course for Percutaneous Endoscopic Spinal Surgery and Complementary Techniques,
[11]
S. Alcántara-Bumbiedro, M.T. Flórez-García, C. Echávarri-Pérez, Y.F. García-Pérez.
Escala de incapacidad por dolor lumbar de Oswestry.
Rehabilitacion (Madr), 40 (2006), pp. 150-158
[12]
F. Dakhil-Jerew, H. Jadeja, A. Cohen, J.A. Shepperd.
Inter-observer reliability of detecting Dynesys pedicle screw using plain X-rays: a study on 50 post-operative patients.
Eur Spine J, 18 (2009), pp. 1486-1493
[13]
R.W. Ostelo, R.A. Deyo, P. Stratford, G. Waddell, P. Croft, M. von Korff, et al.
Interpreting change scores for pain and functional status in low back pain: Towards international consensus regarding minimal important change.
Spine (Phila Pa 1976), 33 (2008), pp. 90-94
[14]
S.E. Lee, S.B. Park, T.A. Jahng, C.K. Chung, H.J. Kim.
Clinical experience of the dynamic stabilization system for the degenerative spine disease.
J Korean Neurosurg Soc, 43 (2008), pp. 221-226
[15]
G. Sapkas, A.F. Mavrogenis, K.A. Starantzis, K. Soultanis, Z.T. Kokkalis, P.J. Papagelopoulos.
Outcome of a dynamic neutralization system for the spine.
Orthopedics, 35 (2012), pp. e1497-e1502
[16]
L.Y. Fay, J.C. Wu, T.Y. Tsai, T.H. Tu, C.L. Wu, W.C. Huang, et al.
Dynamic stabilization for degenerative spondylolisthesis: evaluation of radiographic and clinical outcomes.
Clin Neurol Neurosurg, 115 (2013), pp. 535-541
[17]
D. Grob, A. Benini, A. Junge, A.F. Mannion.
Clinical experience with the Dynesys semirigid fixation system for the lumbar spine: Surgical and patient-oriented outcome in 50 cases after an average of 2 years.
Spine (Phila Pa 1976), 30 (2005), pp. 324-331
[18]
M. Bothmann, E. Kast, G.J. Boldt.
Dynesys fixation for lumbar spine degeneration.
Neurosurg Rev, 31 (2008), pp. 189-196
[19]
A. Hallett, J.S. Huntley, J.N. Gibson.
Foraminal stenosis and single-level degenerative disc disease: a randomized controlled trial comparing decompression with decompression and instrumented fusion.
Spine (Phila Pa 1976), 32 (2007), pp. 1375-1380
[20]
S.W. Yu, S.C. Yang, C.H. Ma, C.H. Wu, C.Y. Yen, Y.K. Tu.
Comparison of Dynesys posterior stabilization and posterior lumbar interbody fusion for spinal stenosis L4L5.
Acta Orthop Belg, 78 (2010), pp. 230-239
[21]
M. Di Silvestre, F. Lolli, G. Bakaloudis, P. Parisini.
Dynamic stabilization for degenerative lumbar scoliosis in elderly patients.
Spine (Phila Pa 1976), 35 (2010), pp. 227-234
[22]
S. Hoppe, O. Schwarzenbach, E. Aghayev, H. Bonel, U. Berlemann.
Long-term outcome after monosegmental L4/5 stabilization for degenerative sponylolisthesis with the dynesys device.
J Spinal Disord Tech, (2012),
[23]
S. Schaeren, I. Broger, B. Jeanneret.
Minimum four-year follow-up of spinal stenosis with degenerative spondylolisthesis treated with decompression and dynamic stabilization.
Spine (Phila Pa 1976), 33 (2008), pp. E636-E642
[24]
J.C. Wu, W.C. Huang, H.W. Tsai, C.C. Ko, C.L. Wu, T.H. Tu, et al.
Pedicle screw loosening in dynamic stabilization: incidence, risk, and outcome in 126 patients.
Neurosurg. Focus, 31 (2011), pp. E9
[25]
S.W. Yu, C.Y. Yen, C.H. Wu, F.C. Kao, Y.H. Kao, Y.K. Tu.
Radiographic and clinical results of posterior dynamic stabilization for the treatment of multisegment degenerative disc disease with a minimum follow-up of 3 years.
Arch Orthop Trauma Surg, 132 (2012), pp. 583-589

Please cite this article as: Segura-Trepichio M, Ferrández-Sempere D, López-Prats F, Segura-Ibáñez J, Maciá-Soler L. Sistema pedicular de estabilización dinámica. Resultados funcionales y complicaciones del implante en pacientes con enfermedad degenerativa discal lumbar, tras un seguimiento mínimo de 4 años. Rev Esp Cir Ortop Traumatol. 2014;58:85–91.

Copyright © 2013. SECOT
Article options
es en pt

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos