metricas
covid
Buscar en
Revista Española de Medicina Nuclear e Imagen Molecular
Toda la web
Inicio Revista Española de Medicina Nuclear e Imagen Molecular Determinación del factor de calibración para el samario-153 en el activímetro...
Información de la revista
Vol. 26. Núm. 6.
Páginas 354-358 (noviembre 2007)
Compartir
Compartir
Descargar PDF
English PDF
Más opciones de artículo
Vol. 26. Núm. 6.
Páginas 354-358 (noviembre 2007)
Determinación del factor de calibración para el samario-153 en el activímetro de dosis CRC®-35R
Determination of dose calibrator setting of the capintec CRC®-35R for samarium-153
Visitas
9245
MC. Plancha Mansaneta, J. Félix Fontestadb, EV. Caballero Calabuiga, MD. Reyes Ojedaa, P. Sopena Novalesa, EM. del Valle del Vala, P. Abreu Sáncheza, MC. Cano Terola, R. Sopena Monfortea
a Servicio de Medicina Nuclear. Hospital Universitario Dr. Peset. Valencia.
b Servicio de Radiofísica y Protección Radiológica. Hospital Universitario Dr. Peset. Valencia.
Este artículo ha recibido
Información del artículo
Resumen
Texto completo
Bibliografía
Descargar PDF
Estadísticas
Tablas (4)
Tabla 1. PRINCIPALES EMISIONES DEL 153SM
Fig. 1. --A) Geometría del vial y B) de la jeringa.
Fig. 2. --Factor de corrección por volumen en vial.
Tabla 2. MEDIDAS DE ACTIVIDAD (MBq) EN VIAL Y JERINGA. RELACIÓN ENTRE LA ACTIVIDAD EN VIAL RESPECTO DE JERINGA
Mostrar másMostrar menos
Objetivo. Determinar el factor de calibración del activímetro para el radionucleido samario-153 (153Sm) y estudiar la influencia de la geometría en la medida de la actividad, de forma que se pueda determinar con exactitud la actividad administrada al paciente. Material y métodos. El factor de calibración del activímetro para el 153Sm se determinó a partir del valor medio de la respuesta del detector en un canal de factor de calibración conocido, utilizando fuentes de 153Sm de uso clínico. El factor de corrección según geometría se calculó a partir del valor de actividad real indicada por el fabricante y las medidas de actividad realizadas en vial y jeringa. Resultados. El factor de calibración obtenido es de 239 ± 4. El factor de corrección que tiene en cuenta la geometría es de 0,87 ± 0,07. La actividad medida en jeringa es un 18 % superior a la medida en vial. Conclusiones. El método empleado permite medir correctamente la actividad de 153Sm, tanto en vial como en jeringa para dosificar con exactitud la actividad de Samario [153Sm] lexidronam administrada al paciente.
Palabras clave:
activímetro, samario-153, factor de calibración
Objetive. The aim of this work is to determine the calibration factor for Sm-153 and evaluate the influence of the geometry in the activity measurements in order to administer accurately the activity to the patient. Methods. The calibration factor for Sm-153 was determinated using the Sm-153 sources commonly used in clinical practice and the calibrator response in a known calibration setting. The geometry correction factor for the vial and the plastic syringe was calculated using the real activity indicated by the manufacturer and the activity measured in the vial and in the plastic syringe. Results. The calibration factor obtained is 239 ± 4 and the correction factor which takes in consideration the geometry is 0,87 ± 0,07. The activity measured in the syringe is eighteen percent higher to activity measured in the vial. Conclusion. This method allows to measure accurately the activity of Sm-153 in vial and syringe applying this geometry correction factor to determine accurately the activity administered to the patient.
Keywords:
dose calibrator, sammarium-153, calibration factor

Artículo

Opciones para acceder a los textos completos de la publicación Revista Española de Medicina Nuclear e Imagen Molecular
Socio
Si es usted socio de la Sociedad Española de Medicina Nuclear e Imagen Molecular (SEMNIM) puede acceder al texto completo de los contenidos de la Revista Española de Medicina Nuclear e Imagen Molecular desde los enlaces a la revista publicados en la web de la SEMNIN (enlace a https://semnim.es/iniciar-sesion/), previo inicio de sesión como socio. Si tiene problemas de acceso puede contactar con la Secretaría Técnica de la SEMNIM en el correo electrónico secretaria.tecnica@semnim.es o en el teléfono: + 34 619 594 780
Suscriptor
Suscriptor de la revista

Si ya tiene sus datos de acceso, clique aquí.

Si olvidó su clave de acceso puede recuperarla clicando aquí y seleccionando la opción "He olvidado mi contraseña".
Suscribirse
Suscribirse a:

Revista Española de Medicina Nuclear e Imagen Molecular

Comprar
Comprar acceso al artículo

Comprando el artículo el PDF del mismo podrá ser descargado

Precio 19,34 €

Comprar ahora
Contactar
Teléfono para suscripciones e incidencias
De lunes a viernes de 9h a 18h (GMT+1) excepto los meses de julio y agosto que será de 9 a 15h
Llamadas desde España
932 415 960
Llamadas desde fuera de España
+34 932 415 960
E-mail
Opciones de artículo
es en pt

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos