Efecto de la segmentación por tejidos en los mapas de atenuación sobre la cuantificación PET con especial hincapié en grandes arterias
The effect of tissue-segmented attenuation maps on PET quantification with a special focus on large arteries
Visitas
600
A. Mota-Cobiana, J.C. Alonso-Fartoa,b, L. Fernández-Frieraa,c, J. Sánchez-Gonzálezd, B. López-Melgara,c, L.J. Jiménez-Borregueroa,e, V. Fustera,f, J. Ruiz-Cabelloa,g, S. Españaa,
Autor para correspondencia
sespana@cnic.es

Autor para correspondencia.
a Centro Nacional de Investigaciones Cardiovasculares Carlos III (CNIC), Madrid, España
b Hospital General Universitario Gregorio Marañón, Madrid, España
c Hospital Universitario Montepríncipe, Madrid, España
d Philips Healthcare Iberia, España
e Hospital Universitario La Princesa, Madrid, España
f Mount Sinai School of Medicine, New York, NY, EE.UU.
g Ciber de Enfermedades Respiratorias (CIBERES), Madrid, España
Este artículo ha recibido
Disponible módulo formativo: Unidad 2. Saber más
Información del artículo
Resumen
Texto completo
Bibliografía
Descargar PDF
Estadísticas
Figuras (5)
Mostrar másMostrar menos
Tablas (2)
Tabla 1. Características de los pacientes
Tabla 2. Media y RMSEa de las desviaciones en la distribución de vóxels entre las imágenes PET con MRAC y CTAC obtenidos para pulmón, grasa, tejidos blandos y hueso
Mostrar másMostrar menos
Resumen
Objetivos

La precisión cuantitativa en PET requiere una adecuada aplicación de la corrección de atenuación, lo cual representa uno de los mayores retos que aún están por resolver en la técnica PET/RM. El propósito de este estudio es evaluar el efecto de utilizar mapas de atenuación basados en RM y el uso de antenas flexibles sobre la precisión cuantitativa en PET con especial hincapié en grandes arterias.

Materiales y métodos

Se utilizaron datos PET/TC de 8 pacientes oncológicos. Los datos PET fueron reconstruidos utilizando mapas de atenuación con diferente nivel de detalle emulando las distintas aproximaciones utilizadas actualmente por lo equipos PET/RM. Las imágenes PET obtenidas con mapas de atenuación basados en TC y RM fueron comparadas para evaluar las desviaciones cuantitativas obtenidas. El efecto producido por las antenas flexibles fue también estudiado.

Resultados

El uso de mapas de atenuación más simplificados da lugar a un incremento en las desviaciones cuantitativas en grasa, tejido blando y hueso. Las desviaciones en pulmón son altas debido a su heterogeneidad y a la variabilidad entre pacientes. La cuantificación en grandes arterias muestra pequeñas desviaciones excepto en el caso de utilizar antenas flexibles. La cuantificación mediante TBR proporciona menores desviaciones al cancelarse las desviaciones medidas en arterias y las venas utilizadas como referencia.

Conclusiones

Los mapas de atenuación simplificados que se utilizan en PET/RM producen un incremento significativo de variabilidad cuantitativa especialmente en pulmones y huesos. La cuantificación aplicada al estudio de la aterosclerosis produce pequeñas desviaciones, especialmente cuando se utiliza el TBR.

Palabras clave:
PET/RM
Corrección de atenuación
Aterosclerosis
Abstract
Objectives

Accuracy on quantitative PET image analysis relies on the correct application of attenuation correction which is one of the major challenges for PET/MRI that remains to be solved. The purpose of this study is to evaluate the effect of MRI-based attenuation maps and the use of flexible coils on the quantitative accuracy of PET images with a special focus on large arteries.

Materials and methods

PET/CT data from eight oncologic patients was used. PET data was reconstructed using attenuation maps with different level of detail emulating several approaches available on current PET/MRI scanners. PET images obtained with CT-based and MRI-based attenuation maps were compared to evaluate the quantitative biases obtained. The quantitative effect produced by flexible MRI receiver coils on the attenuation maps was also studied.

Results

The use of simpler attenuation maps produced increased biases between PET data reconstructed with CT-based and MRI-based attenuation maps for fat, non-fat soft-tissues and bone. Biases in lung were very high due to the large heterogeneity and inter-patient variability of the lung. The quantification on large arteries had small deviations except for the case when flexible coils were used. The TBR provided smaller biases in all cases as it cancelled out the similar deviations obtained for arteries and reference veins.

Conclusions

Simplified attenuation maps used on PET/MRI significantly increase the quantitative variability of PET images especially on lungs and bones. The quantification of PET images acquired with PET/MRI scanners applied to studies of atherosclerosis has small deviations, especially when the TBR is considered.

Keywords:
PET/MRI
Attenuation correction
Atherosclerosis

Artículo

Opciones para acceder a los textos completos de la publicación Revista Española de Medicina Nuclear e Imagen Molecular
Socio
Si es usted socio de la Sociedad Española de Medicina Nuclear e Imagen Molecular (SEMNIM) puede acceder al texto completo de los contenidos de la Revista Española de Medicina Nuclear e Imagen Molecular desde los enlaces a la revista publicados en la web de la SEMNIN (enlace a https://semnim.es/iniciar-sesion/), previo inicio de sesión como socio. Si tiene problemas de acceso puede contactar con la Secretaría Técnica de la SEMNIM en el correo electrónico secretaria.tecnica@semnim.es o en el teléfono: + 34 619 594 780
Suscriptor
Suscriptor de la revista

Si ya tiene sus datos de acceso, clique aquí.

Si olvidó su clave de acceso puede recuperarla clicando aquí y seleccionando la opción "He olvidado mi contraseña".
Suscribirse
Suscribirse a:

Revista Española de Medicina Nuclear e Imagen Molecular

Comprar
Comprar acceso al artículo

Comprando el artículo el PDF del mismo podrá ser descargado

Precio 19,34 €

Comprar ahora
Contactar
Teléfono para suscripciones e incidencias
De lunes a viernes de 9h a 18h (GMT+1) excepto los meses de julio y agosto que será de 9 a 15h
Llamadas desde España
932 415 960
Llamadas desde fuera de España
+34 932 415 960
E-mail
Opciones de artículo
es en pt

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos