Utilidad de la técnica ROLL en la exéresis de lesiones no palpables de mama
Use of the ROLL technique for lumpectomy in non-palpable breast lesions
Visitas
1007
D. Carreraa,
Autor para correspondencia
david.dcarrera@gmail.com

Autor para correspondencia.
, L. Martínb, M. de la Florc, F. Guspíd, J. Picase, V. Izquierdoa, S. Martínezf, C. Jordàg, R. Siuranaa, M. Martínez-Casalsh, J.M. Jaéni, A. Pujola, A. Benítezj
a Medicina Nuclear, H.U. Joan XXIII, Tarragona, España
b Radiología, H.U. Joan XXIII, Tarragona, España
c Ginecología, H.U. Joan XXIII, Tarragona, España
d Cirugía, Hospital Verge de la Cinta, Tortosa, España
e Cirugía, Hospital Sant Pau i Santa Tecla, Tarragona, España
f Anatomía Patológica, H.U. Joan XXIII, Tarragona, España
g Cirugía, Hospital Sant Joan de Reus, Reus, España
h Cirugía, Hospital Pius Valls, Valls, España
i Anatomía Patológica, Hospital Verge de la Cinta, Tortosa, España
j Medicina Nuclear, H.U. Bellvitge, Hospitalet de LLobtregat, España
Ver más
Este artículo ha recibido
Información del artículo
Resumen
Texto completo
Bibliografía
Descargar PDF
Estadísticas
Figuras (1)
Tablas (4)
Tabla 1. Características de las pacientes en el momento diagnóstico
Tabla 2. Variables independientes valoradas en estudio multivariante
Tabla 3. Resultados poblacionales
Tabla 4. Comparación de resultados entre técnica ROLL y arpón
Mostrar másMostrar menos
Resumen
Objetivo

Evaluar la eficacia de la localización radioguiada de lesiones no palpables de mama (LNPM) respecto al arpón quirúrgico.

Método

Se han estudiado prospectivamente 161 mujeres con LNPM, 80 marcadas con arpón (grupo 1) y 81 con inyección intratumoral de 99mTc nanocoloide (grupo 2). Las lesiones se localizaron por ecografía o estereotaxia. Las tumorectomías se realizaron, en el grupo 1 siguiendo la dirección del arpón y en el grupo 2 con la ayuda de una sonda gammadetectora. Posteriormente se comprobaron los márgenes quirúrgicos, determinando la necesidad de ampliación si el margen era menor a 5mm en el estudio intraoperatorio y menor a 2mm en el estudio diferido. Se recogieron datos de porcentaje de detección quirúrgica, afectación de márgenes quirúrgicos, número de ampliaciones, presencia de lesión residual en la ampliación, número de reintervenciones, volumen de la tumorectomía y volumen total extraído, ratio volumen/tumor y complicaciones.

Resultados

No hubo diferencias significativas entre ambos grupos en porcentaje de detección, afectación de márgenes, número de ampliaciones, presencia de lesión residual en la ampliación, reintervenciones, volumen de la tumorectomía, volumen total extraído, ratio volumen/tumor y complicaciones. El análisis multivariante mostró que los factores condicionantes del volumen extraído son la técnica de marcaje radiológico y el cirujano.

Conclusiones

La técnica de localización radioguiada de lesiones ocultas permite la detección y exeresis de las LNPM con la misma eficacia que el arpón y añade la posibilidad de detección simultánea del ganglio centinela. Los condicionantes del volumen extraído son la técnica de marcaje radiológico y el cirujano.

Palabras clave:
Cáncer de mama
ROLL
Arpón
Abstract
Objective

To evaluate the efficiency of radioguided occult lesion localising in non-palpable breast lesions (NPBL) compared to the surgical wire technique.

Method

A prospective study was conducted on 161 women with NPBL, of whom 80 marked with the wire (group 1), whereas 81 women were marked with an intratumour injection of 99mTc-nanocoloid (group 2). The NPBL were located by ultrasound or stereotactic guidance. The lumpectomies were performed following the wire direction in group 1, and with the aid of a gamma-probe in group 2. Surgical margins were then checked, determining the need of extension if the margin was less than 5mm in the intra-surgical study, and less than 2mm in the deferred study. Data were collected on the mean number detected by surgery, surgical margins, number of extensions, presence of residual tumour in the extension, second surgeries, lumpectomy volume, as well as total resected volume, volume/tumour ratio, and complications.

Results

No significant differences were observed between the two groups in the mean number detected, surgical margins, number of extensions, presence of residual tumour in the extension, second surgeries, lumpectomy volume, total resected volume, volume/tumour ratio or complications. The multivariate analysis showed the determining factors of the resected volume were the radiological guidance technique, as well as the surgeon.

Conclusions

The radioguided occult lesion localising technique helps in the detection and resection of NPBL with the same efficiency as the surgical wire, and adds the possibility of sentinel node detection in the same surgery. The determining factors of the resected volume were the radiological guidance technique and the surgeon.

Keywords:
Breast cancer
ROLL
Surgical wire

Artículo

Opciones para acceder a los textos completos de la publicación Revista Española de Medicina Nuclear e Imagen Molecular
Socio
Si es usted socio de la Sociedad Española de Medicina Nuclear e Imagen Molecular (SEMNIM) puede acceder al texto completo de los contenidos de la Revista Española de Medicina Nuclear e Imagen Molecular desde los enlaces a la revista publicados en la web de la SEMNIN (enlace a https://semnim.es/iniciar-sesion/), previo inicio de sesión como socio. Si tiene problemas de acceso puede contactar con la Secretaría Técnica de la SEMNIM en el correo electrónico secretaria.tecnica@semnim.es o en el teléfono: + 34 619 594 780
Suscriptor
Suscriptor de la revista

Si ya tiene sus datos de acceso, clique aquí.

Si olvidó su clave de acceso puede recuperarla clicando aquí y seleccionando la opción "He olvidado mi contraseña".
Suscribirse
Suscribirse a:

Revista Española de Medicina Nuclear e Imagen Molecular

Comprar
Comprar acceso al artículo

Comprando el artículo el PDF del mismo podrá ser descargado

Precio 19,34 €

Comprar ahora
Contactar
Teléfono para suscripciones e incidencias
De lunes a viernes de 9h a 18h (GMT+1) excepto los meses de julio y agosto que será de 9 a 15h
Llamadas desde España
932 415 960
Llamadas desde fuera de España
+34 932 415 960
E-mail
Opciones de artículo
es en pt

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos