metricas
covid
Buscar en
Enfermedades Infecciosas y Microbiología Clínica
Toda la web
Inicio Enfermedades Infecciosas y Microbiología Clínica Análisis de resultados del Programa de Control de Calidad Externo SEIMC de carg...
Journal Information
Vol. 29. Issue S3.
Programa Externo de Control de Calidad SEIMC. Año 2009
Pages 8-13 (March 2011)
Share
Share
Download PDF
More article options
Vol. 29. Issue S3.
Programa Externo de Control de Calidad SEIMC. Año 2009
Pages 8-13 (March 2011)
Full text access
Análisis de resultados del Programa de Control de Calidad Externo SEIMC de carga viral del VIH-1 y del VHC. Año 2009
Analysis of the results of the HIV-1 and HCV viral load of the SEIMC External Quality Control Program. Year 2009
Visits
2760
Nieves Orta Miraa,b,
Corresponding author
niormi@gmail.com

Autor para correspondencia.
, María del Remedio Guna Serranoa,c, José-Carlos Latorre Martínezd, María Rosario Oviesa, Marta Povedaa, Enrique Ruiz de Gopeguia,e, José L. Péreza,e, Concepción Gimeno Cardonaa,c,f
a Programa de Control de Calidad Externo SEIMC, España
b Unidad de Microbiología, Hospital Francesc de Borja, Gandía, Valencia, España
c Servicio de Microbiología, Consorcio Hospital General Universitario, Centro de Diagnóstico Biomédico, Valencia, España
d Servicio de Microbiología, Hospital Clínico Universitario, Valencia, España
e Servicio de Microbiología, Hospital Son Espases, Palma de Mallorca, España
f Facultad de Medicina, Universidad de Valencia, Valencia, España
This item has received
Article information
Resumen

Las determinaciones de la carga viral de los virus de la inmunodeficiencia humana tipo 1 (VIH-1) y de la hepatitis C (VHC) son marcadores fundamentales para el seguimiento y control de los pacientes infectados por estos virus. Los laboratorios de microbiología deben disponer de herramientas que garanticen la fiabilidad de sus resultados, entre ellas se encuentran los programas de intercomparación externos. En el presente número se muestra el análisis de resultados del Programa de Control de Calidad SEIMC de carga viral de ambos virus y del genotipado del VHC, realizados durante el año 2009.

En el control del VIH-1 se remitieron 5 estándares, de los que 1 (plasma humano seronegativo) no contenía el virus, y los otros 4 consistían en plasma de 3 pacientes virémicos distintos en un intervalo de concentraciones entre 2–5 log10 copias/ml; 2 de ellos eran idénticos, con el fin de analizar la repetibilidad. Una parte significativa de los laboratorios obtuvo resultados fuera de los límites aceptables (media ± 0,2 log10 copias/ml), dependiendo del estándar y del método empleado, en promedio el 21,5%. La repetibilidad fue muy buena, y más del 95% de los laboratorios obtuvieron resultados aceptables (Δ < 0,5 log10 copias/ml). Se detectaron errores postanalíticos de transcripción de los resultados en el control de carga viral VIH-1. En el control de VHC se remitieron 2 estándares con diferente contenido del virus. La mayor parte de los participantes (79,7%) obtuvo ambos resultados dentro de los límites de la media ± 1,96 DE log10 UI/ml.

Los resultados obtenidos ponen de manifiesto la utilidad de los controles externos para asegurar la calidad de los resultados analíticos, incluyendo la fase postanalítica. Debido a la variabilidad interlaboratorio, es aconsejable utilizar un mismo método y el mismo laboratorio en el seguimiento de los pacientes.

Palabras clave:
VHC
VIH-1
Carga viral
Control de calidad externo
Intercomparación
Abstract

Human immunodeficiency virus type 1 (HIV-1) and hepatitis C virus (HCV) viral load determinations are among the most important markers in the follow-up of patients infected with these viruses. External quality control tools are crucial to ensure the accuracy of the results obtained by microbiology laboratories. This article summarizes the results obtained from the SEIMC's External Quality Control Program for HIV-1 and HCV viral loads in 2009.

In the HIV-1 program, a total of five standards were sent. One standard consisted of seronegative human plasma, while the remaining four contained plasma from three different viremic patients, in the range of 2–5 log10 copies/mL; two of these standards were identical, aiming to determine repeatability. A significant proportion of the laboratories (21.5% on average) obtained values outside the accepted range (mean ± 0.2 log10 copies/mL), depending on the standard and on the method used for quantification. Repeatability was very good, with up to 95 % of laboratories reporting results within the accepted limits (Δ<0.5 log10 copies/mL). Post-analytical errors due to mistranscription of the results were detected for HIV-1. The HCV program consisted of two standards with different viral load contents. Most of the participants (79.7%) obtained results within the accepted range (mean ± 1.96 SD log10 UI/mL).

Data from this analysis reinforce the utility of proficiency programs to ensure the quality of the results obtained by a particular laboratory, as well as the importance of the post-analytical phase in overall quality. Due to marked interlaboratory variability, use of the same method and the same laboratory for patient follow-up is advisable.

Keywords:
HCV
HIV-1
Viral load
External quality control
Proficiency
Full text is only aviable in PDF
Bibliografía
[1.]
P.R. Bevington, D.K. Robinson.
Data reduction and error analysis for the physical sciences.
3rd ed., McGraw-Hill, (2003),
[2.]
Niesters HGM. QCMD 2005 human immunodeficiency virus type 1 (HIVRNA05) proficiency programme. Final report. Glasgow: Quality Control for Molecular Diagnostics, 2005; Disponible en: www.qcmd.org
[3.]
N. Orta, M.R. Guna, J.C. Latorre, J.L. Pérez, C. Gimeno.
Análisis de resultados del Programa de Control de Calidad Externo de Carga Viral del VIH-1 y del VHC, año 2006.
Enferm Infecc Microbiol Clin, 25 (2007), pp. 8-13
[4.]
N. Orta Mira, M.R. Guna Serrano, J.C. Latorre Martínez, J.L. Pérez, C. Gimeno Cardona.
Análisis de resultados del Programa Externo de Control de Calidad SEIMC de carga viral del VIH-1 y del VHC. Año 2007.
Enferm Infecc Microbiol Clin, 26 (2008), pp. 8-13
[5.]
N. Orta Mira, M.R. Guna Serrano, J.C. Latorre Martínez, J.L. Pérez, C. Gimeno Cardona.
Análisis de resultados del Programa Externo de Control de Calidad SEIMC de carga viral del VIH-1 y del VHC. Año 2008.
Enferm Infecc Microbiol Clin, 28 (2010), pp. 7-11
[6.]
S.J. Best, A.P. Gust, E.J.M. Johnson, C.H. McGavin, E.M. Dax.
Quality of human virus viral load testing in Australia.
J Clin Microbiol, 38 (2000), pp. 4015-4020
[7.]
D.J. Brambilla, S. Granger, C. Jennings, J.W. Bremer.
Multisite comparison of reproducibility and recovery from the standard and ultrasensitive Roche Amplicor HIV Monitor assays.
J Clin Microbiol, 29 (2001), pp. 1221-1223
[8.]
G. Muyldermans, L. Debaisieux, K. Fransen, D. Marissens, K. Miller, D. Vaira, et al.
Blinded, multicenter quality control study for the quantification of human immunodeficiency virus type 1 RNA in plasma by the Belgian AIDS reference laboratories.
Clin Microbiol Infect, 6 (2000), pp. 213-217
[9.]
Niesters HGM. QCMD 2005 hepatitis C virus (HCVRNA05) proficiency programme. Final report. Glasgow: Quality Control for Molecular Diagnostics, 2005. Disponible en: www.qcmd.org
[10.]
G. Gentili, K. Cristiano, G. Pisani, G.M. Bisso, M. Miceli, M. Wirz, et al.
Collaborative study for the calibration of a new Italian HCV RNA reference preparation against the international standard.
Ann Ist Super Sanità, 39 (2003), pp. 183-187
[11.]
L. Fanning, E. Kenny-Walsh, J. Levis, K.R. Choudhury, B. Cannon, M. Sheehan, et al.
Natural fluctuations of hepatitis C virus load in a homogeneous patient population: a prospective study.
Hepatology, 35 (2000), pp. 225-229
[12.]
E. Martínez-Bauer, J. Crespo, M. Romero-Gómez, R. Moreno-Otero, R. Sola, N. Tesei, et al.
Development and validation of two models for early prediction of response to therapy in genotype 1 chronic hepatitis C.
Hepatology, 43 (2006), pp. 72-80
Copyright © 2011. Elsevier España S.L.. Todos los derechos reservados
Download PDF
Article options
es en pt

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos