Update in Radiology
A practical introduction to the hemodynamic analysis of the cardiovascular system with 4D Flow MRI
Introducción práctica al análisis hemodinámico del sistema cardiovascular mediante la técnica «4D Flow»
J.A. Pineda Zapataa,
Corresponding author
julian.pineda21@gmail.com

Corresponding author.
, J.A. Delgado de Bedouta, S. Rascovsky Ramíreza, C. Bustamantea, S. Mesab, V.D. Calvo Betancura
a Grupo de Investigación, Instituto de Alta Tecnologi¿a Médica (IATM), Medellín, Antioquia, Colombia
b Universidad CES, Medellín, Antioquia, Colombia
Read
2324
Times
was read the article
369
Total PDF
1955
Total HTML
Share statistics
 array:22 [
  "pii" => "S2173510714000603"
  "issn" => "21735107"
  "doi" => "10.1016/j.rxeng.2014.08.001"
  "estado" => "S300"
  "fechaPublicacion" => "2014-11-01"
  "aid" => "759"
  "copyright" => "SERAM"
  "copyrightAnyo" => "2013"
  "documento" => "article"
  "subdocumento" => "fla"
  "cita" => "Radiologia. 2014;56:485-95"
  "abierto" => array:3 [
    "ES" => false
    "ES2" => false
    "LATM" => false
  ]
  "gratuito" => false
  "lecturas" => array:2 [
    "total" => 2306
    "formatos" => array:2 [
      "HTML" => 1943
      "PDF" => 363
    ]
  ]
  "Traduccion" => array:1 [
    "es" => array:19 [
      "pii" => "S0033833814001301"
      "issn" => "00338338"
      "doi" => "10.1016/j.rx.2014.08.001"
      "estado" => "S300"
      "fechaPublicacion" => "2014-11-01"
      "aid" => "759"
      "copyright" => "SERAM"
      "documento" => "article"
      "crossmark" => 1
      "subdocumento" => "fla"
      "cita" => "Radiologia. 2014;56:485-95"
      "abierto" => array:3 [
        "ES" => false
        "ES2" => false
        "LATM" => false
      ]
      "gratuito" => false
      "lecturas" => array:2 [
        "total" => 2508
        "formatos" => array:3 [
          "EPUB" => 22
          "HTML" => 1716
          "PDF" => 770
        ]
      ]
      "es" => array:13 [
        "idiomaDefecto" => true
        "cabecera" => "<span class="elsevierStyleTextfn">ACTUALIZACI&#211;N</span>"
        "titulo" => "Introducci&#243;n pr&#225;ctica al an&#225;lisis hemodin&#225;mico del sistema cardiovascular mediante la t&#233;cnica &#171;4D Flow&#187;"
        "tienePdf" => "es"
        "tieneTextoCompleto" => "es"
        "tieneResumen" => array:2 [
          0 => "es"
          1 => "en"
        ]
        "paginas" => array:1 [
          0 => array:2 [
            "paginaInicial" => "485"
            "paginaFinal" => "495"
          ]
        ]
        "titulosAlternativos" => array:1 [
          "en" => array:1 [
            "titulo" => "A practical introduction to the hemodynamic analysis of the cardiovascular system with 4<span class="elsevierStyleHsp" style=""></span>D Flow MRI"
          ]
        ]
        "contieneResumen" => array:2 [
          "es" => true
          "en" => true
        ]
        "contieneTextoCompleto" => array:1 [
          "es" => true
        ]
        "contienePdf" => array:1 [
          "es" => true
        ]
        "resumenGrafico" => array:2 [
          "original" => 0
          "multimedia" => array:7 [
            "identificador" => "fig0025"
            "etiqueta" => "Figura 5"
            "tipo" => "MULTIMEDIAFIGURA"
            "mostrarFloat" => true
            "mostrarDisplay" => false
            "figura" => array:1 [
              0 => array:4 [
                "imagen" => "gr5.jpeg"
                "Alto" => 2666
                "Ancho" => 2917
                "Tamanyo" => 781380
              ]
            ]
            "descripcion" => array:1 [
              "es" => "<p id="spar0035" class="elsevierStyleSimplePara elsevierViewall">M&#233;todos de visi&#243;n 3<span class="elsevierStyleHsp" style=""></span>D en la aorta&#46; Im&#225;genes <span class="elsevierStyleItalic">4</span><span class="elsevierStyleHsp" style=""></span><span class="elsevierStyleItalic">D Flow</span>&#46; A&#41; Las l&#237;neas de flujo &#171;streamlines&#187; dan una perspectiva 3<span class="elsevierStyleHsp" style=""></span>D del movimiento de la sangre en las estructuras vasculares&#46; B&#41; Los trazadores de part&#237;culas &#171;pathlines&#187; dan informaci&#243;n de la trayectoria&#44; la velocidad&#44; la direcci&#243;n y la evoluci&#243;n de los patrones de flujo sangu&#237;neo en cada fase del ciclo card&#237;aco&#46; C<span class="elsevierStyleBold">&#41;</span> Los gr&#225;ficos de vectores muestran la magnitud de la velocidad y la direcci&#243;n para cada part&#237;cula dentro del fluido en un plano 2<span class="elsevierStyleHsp" style=""></span>D&#46; En la imagen se presentan la aorta y el tronco pulmonar&#46; D&#41; Representaci&#243;n del perfil de velocidad de la sangre en la aorta ascendente y descendente mediante dos mallas 3<span class="elsevierStyleHsp" style=""></span>D&#46;</p>"
            ]
          ]
        ]
        "autores" => array:1 [
          0 => array:2 [
            "autoresLista" => "J&#46;A&#46; Pineda Zapata, J&#46;A&#46; Delgado de Bedout, S&#46; Rascovsky Ram&#237;rez, C&#46; Bustamante, S&#46; Mesa, V&#46;D&#46; Calvo Betancur"
            "autores" => array:6 [
              0 => array:2 [
                "nombre" => "J&#46;A&#46;"
                "apellidos" => "Pineda Zapata"
              ]
              1 => array:2 [
                "nombre" => "J&#46;A&#46;"
                "apellidos" => "Delgado de Bedout"
              ]
              2 => array:2 [
                "nombre" => "S&#46;"
                "apellidos" => "Rascovsky Ram&#237;rez"
              ]
              3 => array:2 [
                "nombre" => "C&#46;"
                "apellidos" => "Bustamante"
              ]
              4 => array:2 [
                "nombre" => "S&#46;"
                "apellidos" => "Mesa"
              ]
              5 => array:2 [
                "nombre" => "V&#46;D&#46;"
                "apellidos" => "Calvo Betancur"
              ]
            ]
          ]
        ]
      ]
      "idiomaDefecto" => "es"
      "Traduccion" => array:1 [
        "en" => array:9 [
          "pii" => "S2173510714000603"
          "doi" => "10.1016/j.rxeng.2014.08.001"
          "estado" => "S300"
          "subdocumento" => ""
          "abierto" => array:3 [
            "ES" => false
            "ES2" => false
            "LATM" => false
          ]
          "gratuito" => false
          "lecturas" => array:1 [
            "total" => 0
          ]
          "idiomaDefecto" => "en"
          "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S2173510714000603?idApp=UINPBA00004N"
        ]
      ]
      "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S0033833814001301?idApp=UINPBA00004N"
      "url" => "/00338338/0000005600000006/v3_201706012056/S0033833814001301/v3_201706012056/es/main.assets"
    ]
  ]
  "itemSiguiente" => array:18 [
    "pii" => "S2173510714000615"
    "issn" => "21735107"
    "doi" => "10.1016/j.rxeng.2012.11.004"
    "estado" => "S300"
    "fechaPublicacion" => "2014-11-01"
    "aid" => "650"
    "copyright" => "SERAM"
    "documento" => "article"
    "subdocumento" => "fla"
    "cita" => "Radiologia. 2014;56:515-23"
    "abierto" => array:3 [
      "ES" => false
      "ES2" => false
      "LATM" => false
    ]
    "gratuito" => false
    "lecturas" => array:2 [
      "total" => 858
      "formatos" => array:2 [
        "HTML" => 648
        "PDF" => 210
      ]
    ]
    "en" => array:13 [
      "idiomaDefecto" => true
      "cabecera" => "<span class="elsevierStyleTextfn">Original article</span>"
      "titulo" => "Preoperative lymphoscintigraphy and tumor histologic grade are associated with surgical detection of the sentinel lymph node"
      "tienePdf" => "en"
      "tieneTextoCompleto" => "en"
      "tieneResumen" => array:2 [
        0 => "en"
        1 => "es"
      ]
      "paginas" => array:1 [
        0 => array:2 [
          "paginaInicial" => "515"
          "paginaFinal" => "523"
        ]
      ]
      "titulosAlternativos" => array:1 [
        "es" => array:1 [
          "titulo" => "La linfogammagraf&#237;a prequir&#250;rgica y el grado histol&#243;gico del tumor se asocian a la detecci&#243;n quir&#250;rgica del ganglio centinela"
        ]
      ]
      "contieneResumen" => array:2 [
        "en" => true
        "es" => true
      ]
      "contieneTextoCompleto" => array:1 [
        "en" => true
      ]
      "contienePdf" => array:1 [
        "en" => true
      ]
      "resumenGrafico" => array:2 [
        "original" => 0
        "multimedia" => array:7 [
          "identificador" => "fig0005"
          "etiqueta" => "Figure 1"
          "tipo" => "MULTIMEDIAFIGURA"
          "mostrarFloat" => true
          "mostrarDisplay" => false
          "figura" => array:1 [
            0 => array:4 [
              "imagen" => "gr1.jpeg"
              "Alto" => 2407
              "Ancho" => 1649
              "Tamanyo" => 461139
            ]
          ]
          "descripcion" => array:1 [
            "en" => "<p id="spar0045" class="elsevierStyleSimplePara elsevierViewall">&#40;A&#41; <span class="elsevierStyleSup">99m</span>Tc-albumin nanocolloid radioisotope transportation through lead syringe&#46; &#40;B&#41; Node with change in echogeneicity after intratumor injection of the radioisotope&#46; &#40;C&#41; Radioisotope marking of microcalcifications through an internal stereotactic metallic marker&#46; &#40;D&#41; Lymphogammagraphy &#40;anterior projection&#41; with masking and axillary sentinel node&#46; &#40;E&#41; Location of breast lesion and sentinel lymph node and design of surgical approach&#46; &#40;F&#41; Gamma-detecting portable probe model Europrobe<span class="elsevierStyleSup">&#174;</span> II&#46; &#40;G&#41; Axillary incision and tracking through gamma-detecting probe at the OR&#46; &#40;H&#41; Gamma-probe-guided ex vivo checking through of the activity of the extracted nodes&#46;</p>"
          ]
        ]
      ]
      "autores" => array:1 [
        0 => array:2 [
          "autoresLista" => "M&#46; Arias Ortega, M&#46;Y&#46; Torres Sousa, B&#46; Gonz&#225;lez Garc&#237;a, R&#46; Pardo Garc&#237;a, A&#46; Gonz&#225;lez L&#243;pez, M&#46; Delgado Portela"
          "autores" => array:6 [
            0 => array:2 [
              "nombre" => "M&#46;"
              "apellidos" => "Arias Ortega"
            ]
            1 => array:2 [
              "nombre" => "M&#46;Y&#46;"
              "apellidos" => "Torres Sousa"
            ]
            2 => array:2 [
              "nombre" => "B&#46;"
              "apellidos" => "Gonz&#225;lez Garc&#237;a"
            ]
            3 => array:2 [
              "nombre" => "R&#46;"
              "apellidos" => "Pardo Garc&#237;a"
            ]
            4 => array:2 [
              "nombre" => "A&#46;"
              "apellidos" => "Gonz&#225;lez L&#243;pez"
            ]
            5 => array:2 [
              "nombre" => "M&#46;"
              "apellidos" => "Delgado Portela"
            ]
          ]
        ]
      ]
    ]
    "idiomaDefecto" => "en"
    "Traduccion" => array:1 [
      "es" => array:9 [
        "pii" => "S003383381200286X"
        "doi" => "10.1016/j.rx.2012.11.008"
        "estado" => "S300"
        "subdocumento" => ""
        "abierto" => array:3 [
          "ES" => false
          "ES2" => false
          "LATM" => false
        ]
        "gratuito" => false
        "lecturas" => array:1 [
          "total" => 0
        ]
        "idiomaDefecto" => "es"
        "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S003383381200286X?idApp=UINPBA00004N"
      ]
    ]
    "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S2173510714000615?idApp=UINPBA00004N"
    "url" => "/21735107/0000005600000006/v1_201412120104/S2173510714000615/v1_201412120104/en/main.assets"
  ]
  "en" => array:21 [
    "idiomaDefecto" => true
    "cabecera" => "<span class="elsevierStyleTextfn">Update in Radiology</span>"
    "titulo" => "A practical introduction to the hemodynamic analysis of the cardiovascular system with 4D Flow MRI"
    "tieneTextoCompleto" => true
    "paginas" => array:1 [
      0 => array:2 [
        "paginaInicial" => "485"
        "paginaFinal" => "495"
      ]
    ]
    "autores" => array:1 [
      0 => array:4 [
        "autoresLista" => "J&#46;A&#46; Pineda Zapata, J&#46;A&#46; Delgado de Bedout, S&#46; Rascovsky Ram&#237;rez, C&#46; Bustamante, S&#46; Mesa, V&#46;D&#46; Calvo Betancur"
        "autores" => array:6 [
          0 => array:4 [
            "nombre" => "J&#46;A&#46;"
            "apellidos" => "Pineda Zapata"
            "email" => array:1 [
              0 => "julian&#46;pineda21&#64;gmail&#46;com"
            ]
            "referencia" => array:2 [
              0 => array:2 [
                "etiqueta" => "<span class="elsevierStyleSup">a</span>"
                "identificador" => "aff0005"
              ]
              1 => array:2 [
                "etiqueta" => "<span class="elsevierStyleSup">&#42;</span>"
                "identificador" => "cor0005"
              ]
            ]
          ]
          1 => array:3 [
            "nombre" => "J&#46;A&#46;"
            "apellidos" => "Delgado de Bedout"
            "referencia" => array:1 [
              0 => array:2 [
                "etiqueta" => "<span class="elsevierStyleSup">a</span>"
                "identificador" => "aff0005"
              ]
            ]
          ]
          2 => array:3 [
            "nombre" => "S&#46;"
            "apellidos" => "Rascovsky Ram&#237;rez"
            "referencia" => array:1 [
              0 => array:2 [
                "etiqueta" => "<span class="elsevierStyleSup">a</span>"
                "identificador" => "aff0005"
              ]
            ]
          ]
          3 => array:3 [
            "nombre" => "C&#46;"
            "apellidos" => "Bustamante"
            "referencia" => array:1 [
              0 => array:2 [
                "etiqueta" => "<span class="elsevierStyleSup">a</span>"
                "identificador" => "aff0005"
              ]
            ]
          ]
          4 => array:3 [
            "nombre" => "S&#46;"
            "apellidos" => "Mesa"
            "referencia" => array:1 [
              0 => array:2 [
                "etiqueta" => "<span class="elsevierStyleSup">b</span>"
                "identificador" => "aff0015"
              ]
            ]
          ]
          5 => array:3 [
            "nombre" => "V&#46;D&#46;"
            "apellidos" => "Calvo Betancur"
            "referencia" => array:1 [
              0 => array:2 [
                "etiqueta" => "<span class="elsevierStyleSup">a</span>"
                "identificador" => "aff0005"
              ]
            ]
          ]
        ]
        "afiliaciones" => array:2 [
          0 => array:3 [
            "entidad" => "Grupo de Investigaci&#243;n&#44; Instituto de Alta Tecnologi¿a M&#233;dica &#40;IATM&#41;&#44; Medell&#237;n&#44; Antioquia&#44; Colombia"
            "etiqueta" => "a"
            "identificador" => "aff0005"
          ]
          1 => array:3 [
            "entidad" => "Universidad CES&#44; Medell&#237;n&#44; Antioquia&#44; Colombia"
            "etiqueta" => "b"
            "identificador" => "aff0015"
          ]
        ]
        "correspondencia" => array:1 [
          0 => array:3 [
            "identificador" => "cor0005"
            "etiqueta" => "&#8270;"
            "correspondencia" => "Corresponding author&#46;"
          ]
        ]
      ]
    ]
    "titulosAlternativos" => array:1 [
      "es" => array:1 [
        "titulo" => "Introducci&#243;n pr&#225;ctica al an&#225;lisis hemodin&#225;mico del sistema cardiovascular mediante la t&#233;cnica &#171;4D Flow&#187;"
      ]
    ]
    "resumenGrafico" => array:2 [
      "original" => 0
      "multimedia" => array:7 [
        "identificador" => "fig0045"
        "etiqueta" => "Figure 9"
        "tipo" => "MULTIMEDIAFIGURA"
        "mostrarFloat" => true
        "mostrarDisplay" => false
        "figura" => array:1 [
          0 => array:4 [
            "imagen" => "gr9.jpeg"
            "Alto" => 1334
            "Ancho" => 3252
            "Tamanyo" => 343176
          ]
        ]
        "descripcion" => array:1 [
          "en" => "<p id="spar0055" class="elsevierStyleSimplePara elsevierViewall">4D Flow and 2D PCA&#46; &#40;A&#41; Quantification of the average velocity of flow in the ascending aorta through the 2D cine PCA modality&#46; To the left we can see flow direction for every vessel and to the right the chart of velocity values in the cardiac cycle&#46; &#40;B&#41; The same quantification was performed with 4D Flow&#46; Note the similarity of values of the average velocity of blood measured through 2D cine PCA and 4D Flow modalities &#40;B&#41;&#46; &#40;C&#41; Selection of the ascending aorta and the pulmonary artery in one single 4D Flow study&#46; The charts of average velocity of aorta &#40;up&#41; and pulmonary artery blood &#40;down&#41; are shown here as well&#46; In charts A and B <span class="elsevierStyleItalic">y</span>-axis is blood velocity &#40;cm&#47;s&#41; while <span class="elsevierStyleItalic">x</span>-axis is time &#40;s&#41;&#46; However&#44; in chart C <span class="elsevierStyleItalic">x</span>-axis represents the cardiac cycle divided into sixteen &#40;16&#41; different phases&#46;</p>"
        ]
      ]
    ]
    "textoCompleto" => "<span class="elsevierStyleSections"><span id="sec0005" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0025">Introduction</span><p id="par0005" class="elsevierStylePara elsevierViewall">During the last years we have seen important innovations in magnetic resonance images &#40;MRI&#41;&#44; especially in the field of cardiovascular imaging&#59; this means that today not only the anatomical study but also the functional study of heart and vessels is possible&#46; The very first maps of in vivo vascular velocities were reported back in the early 1980s<a class="elsevierStyleCrossRef" href="#bib0005"><span class="elsevierStyleSup">1</span></a> and since then angioresonance sequences like phase contrast angiography &#40;PCA&#41;&#8211;consisting of 2D retrospective acquisitions with the cardiac cycle &#40;2D cine PCA&#41; are available in most MRI equipments&#46; This sequence is essential for the functional evaluation of the cardiovascular system and it can be acquired during one apnea only&#46;<a class="elsevierStyleCrossRef" href="#bib0010"><span class="elsevierStyleSup">2</span></a></p><p id="par0010" class="elsevierStylePara elsevierViewall">The time-resolved three-dimensional flow-sensitive MRI with three-directional velocity encoding &#40;4D Flow&#41; modality consists of MRI acquired though a phase contrast angiography that acquires blood flow data in three &#40;3&#41; spatial directions &#40;3D&#41; during the whole cardiac cycle<a class="elsevierStyleCrossRef" href="#bib0015"><span class="elsevierStyleSup">3</span></a> for the qualitative and quantitative evaluation of vascular hemodynamics&#46;<a class="elsevierStyleCrossRef" href="#bib0020"><span class="elsevierStyleSup">4</span></a> In this case acquisition lasts more &#40;5&#8211;20<span class="elsevierStyleHsp" style=""></span>min&#41; and it is necessary to synchronize it with breathing&#46; With 4D Flow data we can both see and analyze quantitatively blood flow patterns in heart and large vessels through 3D cine representations during the whole cardiac cycle by means of pathlines&#44; streamlines&#44; vector graphics and 3D velocities&#46; It also allows us to perform hemodynamic measurements of mid&#44; peak and minimum velocities&#44; ejection volumes&#44; vessel wall shear forces and pressure gradients&#46;<a class="elsevierStyleCrossRef" href="#bib0025"><span class="elsevierStyleSup">5</span></a> Given it is a 3D modality among the multiples advantages of 4D Flow the possibility of quantifying the flow in any interesting views and vessels by postprocessing images is included when the patient is no longer in the resonance equipment without the need for any more sequences in other views as it happens with the 2D cine PCA&#46; Also 4D Flow is a modality that does not use any ionizing radiation or contrast media&#46;<a class="elsevierStyleCrossRefs" href="#bib0030"><span class="elsevierStyleSup">6&#44;7</span></a> Its clinical potential is enormous because it allows us to perform the anatomic and hemodynamic evaluation of the patient&#39;s vessels with cardiac or cardiovascular diseases like&#58; congenital heart diseases&#44; heart failure&#44; arteriovenous malformations&#44; aneurysms&#44; fistulas or stenoses in blood vessels&#44; among others&#46;</p><p id="par0015" class="elsevierStylePara elsevierViewall">In this review we will describe the 4D Flow modality in ten &#40;10&#41; volunteers through a 3T MRI and we will be discussing the advantages&#44; limitations and possible clinical alterations&#46;</p></span><span id="sec0010" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0030">Principles of the 2D cine PCA</span><p id="par0020" class="elsevierStylePara elsevierViewall">The 2D cine PCA is an MRI modality synchronized with the electrocardiogram &#40;ECG&#41; that delivers phase contrast images velocity-sensitive&#46; 2D cine PCA images are acquired in different stages of the cardiac cycle &#40;retrospective acquisition&#41; and they can be seen dynamically&#46; Their main goal is to evaluate hemodynamic parameters like flows&#44; volumes&#44; and velocities&#46;</p><p id="par0025" class="elsevierStylePara elsevierViewall">Flow is quantified by measuring spin-echo transverse magnetization in two &#40;2&#41; different time frames after applying one bipolar pulsed magnetic field gradient with a positive &#171;lobule&#187; followed by a negative one of the same magnitude but different direction&#46; The coding of blood velocity in any spatial directions is based on the phase difference between the moving spin transverse magentization vector and static spins&#46;<a class="elsevierStyleCrossRefs" href="#bib0040"><span class="elsevierStyleSup">8&#44;9</span></a> This is possible due to the fact that the static spins do not accumulate phase difference because when they experience the positive gradient they rotate to a certain angle and when they experience the negative gradient they rotate to the same angle but in a different direction&#46; This means that the net phase displacement is zero&#46; However&#44; moving spins accumulate phase difference due to the fact that the intensity experienced by the positive and negative gradients is not the same due to movement and change of position &#40;<a class="elsevierStyleCrossRef" href="#fig0005">Fig&#46; 1</a>&#41;&#46; The phase difference angle is directly proportional to both the velocity and movement of the spins&#46;<a class="elsevierStyleCrossRef" href="#bib0050"><span class="elsevierStyleSup">10</span></a></p><elsevierMultimedia ident="fig0005"></elsevierMultimedia><p id="par0030" class="elsevierStylePara elsevierViewall">It is important to highlight that the maximum velocity detected is determined by the Velocity Encoding parameter&#8211;expressed in cm&#47;s&#46; Blood flow velocities beyond this parameter are coded erroneously&#44; so aliasing<a class="elsevierStyleCrossRef" href="#bib0045"><span class="elsevierStyleSup">9</span></a> occurs&#46; With the data acquired magnitude and phase images are built &#40;<a class="elsevierStyleCrossRef" href="#fig0010">Fig&#46; 2</a>&#41; through the complex difference and the phase difference of the transverse magnetization vectors&#46;<a class="elsevierStyleCrossRef" href="#bib0055"><span class="elsevierStyleSup">11</span></a></p><elsevierMultimedia ident="fig0010"></elsevierMultimedia></span><span id="sec0015" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0035">4D Flow</span><p id="par0035" class="elsevierStylePara elsevierViewall">The 2D cine PCA modality can be extended to the 3D spatial acquisition in an effort to code the blood flow in the three &#40;3&#41; directions of Cartesian coordinate system in different phases of the cardiac cycle&#46; Because we need three &#40;3&#41; directions for flow coding we need four &#40;4&#41; sequences&#8211;three &#40;3&#41; that are sensitive to velocity and one &#40;1&#41; reference sequence the total acquisition time is longer than with 2D modality&#46; This modality is known as 4D Flow since it is named after four &#40;4&#41; dimensions&#8211;three &#40;3&#41; spatial dimensions plus a 4th temporal dimension&#46;</p><p id="par0040" class="elsevierStylePara elsevierViewall">4D Flow images allow us to quantitatively evaluate the regional and global dynamics of blood flow through hemodynamic measurements like the average&#44; maximum and minimum velocities&#44; the average beat flow&#44; regurgitant and progressive flows&#44; ejection volumes and mathematical approximations to pressure gradients and shear forces in the vessel walls&#46; It also allows us to analyze quantitatively the temporal evolution of the complex patterns of blood flow through streamlines&#44; pathlines&#44; 3D velocity charts&#44; and vectors&#59; representations that describe blood trajectory across the cardiovascular system&#46;<a class="elsevierStyleCrossRefs" href="#bib0060"><span class="elsevierStyleSup">12&#8211;14</span></a></p><p id="par0045" class="elsevierStylePara elsevierViewall">At the beginning this modality was known thanks to its capacity to give qualitative information through the 3D representation of cardiovascular flow&#46; However&#44; recently with 4D Flow some methods have been developed for the quantitative analysis of blood flow&#46; Also the new technological advancements in both hardware and software allow us to reduce the sequence time of acquisition and apply it in the clinical setting&#46;<a class="elsevierStyleCrossRef" href="#bib0065"><span class="elsevierStyleSup">13</span></a></p><span id="sec0020" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0040">Technique of acquisition</span><p id="par0050" class="elsevierStylePara elsevierViewall">For the gathering of time-based volumetric data combined with velocity coding data in the three &#40;3&#41; spatial directions we use <span class="elsevierStyleItalic">k</span>-space segmentation methods synchronized with the ECG signal&#8211;for a total description the reader can use the quote&#46;<a class="elsevierStyleCrossRef" href="#bib0075"><span class="elsevierStyleSup">15</span></a> The Cartesian segmentation method uses the <span class="elsevierStyleItalic">k</span>-space with its three &#40;3&#41; perpendicular axes &#40;<span class="elsevierStyleItalic">kx</span>&#44; <span class="elsevierStyleItalic">ky</span>&#44; <span class="elsevierStyleItalic">kz</span>&#41; and divides it into a 3D matrix &#40;Nx&#44; Ny&#44; Nz&#41;&#46; Then during each beat of the heart it acquires a set of single-cut <span class="elsevierStyleItalic">k</span>-space lines &#40;Nseg&#41; for all phases of the cardiac cycle&#46;<a class="elsevierStyleCrossRefs" href="#bib0020"><span class="elsevierStyleSup">4&#44;16</span></a> If Ny represents the lines all across the phase image &#40;<span class="elsevierStyleItalic">ky</span>-axis&#41; and Nz represents the cuts &#40;<span class="elsevierStyleItalic">kz</span>-axis&#41; the acquisition of the whole 3D volume will last for a certain number of beats equal to NyNz&#47;Nseg and a time equal to <span class="elsevierStyleItalic">T</span><span class="elsevierStyleInf">adq</span><span class="elsevierStyleHsp" style=""></span>&#61;<span class="elsevierStyleHsp" style=""></span>&#40;NyNz&#47;Nseg&#41;<span class="elsevierStyleItalic">T</span><span class="elsevierStyleInf">ECG</span>&#44; where <span class="elsevierStyleItalic">T</span><span class="elsevierStyleInf">adq</span> is the duration of all acquisition and <span class="elsevierStyleItalic">T</span><span class="elsevierStyleInf">ECG</span> is the duration of a single beat&#46; For every line of <span class="elsevierStyleItalic">k</span>-space four &#40;4&#41; different data need to be gathered&#58; one reference acquisition and three &#40;3&#41; velocity codifications resulting from the application of bipolar gradients in the <span class="elsevierStyleItalic">x</span>&#44; <span class="elsevierStyleItalic">y</span>&#44; <span class="elsevierStyleItalic">z</span> directions&#46;<a class="elsevierStyleCrossRefs" href="#bib0020"><span class="elsevierStyleSup">4&#44;16</span></a> After the gathering of data the 3D images are reconstructed&#8211;including one &#40;1&#41; anatomic or magnitude image and three &#40;3&#41; phase images representing blood velocity in the three &#40;3&#41; axes of the coordinate system &#40;<a class="elsevierStyleCrossRef" href="#fig0015">Fig&#46; 3</a>&#41;&#46;</p><elsevierMultimedia ident="fig0015"></elsevierMultimedia></span><span id="sec0025" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0045">Image acquisition</span><p id="par0055" class="elsevierStylePara elsevierViewall">We studied 10 volunteers &#40;9 healthy ones and one with a transplanted kidney&#41; using a Philips Ingenia 3T kit &#40;Philips Healthcare&#44; Eindhoven&#44; The Netherlands&#41; with 4D Flow in some of the following body areas&#58; chest&#44; abdomen&#44; neck and skull&#46; Acquisition parameters can be seen in <a class="elsevierStyleCrossRef" href="#tbl0005">Table 1</a>&#46; We proceeding with free breathing&#44; electrocardiographic synchronization and a chest&#44; head&#44; or neck cardiac coil based on the examination area&#46; Also in certain volunteers 2D cine PCA with parameters of temporal resolution similar to those of 4D Flow were acquired&#46;</p><elsevierMultimedia ident="tbl0005"></elsevierMultimedia></span><span id="sec0030" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0050">Image processing</span><p id="par0060" class="elsevierStylePara elsevierViewall">Obtained data were extracted in RAR-REC format and postprocessing was performed using the GTFlow software &#40;Gyrotools LLC&#44; Zurich&#44; Switzerland&#41; that allowed us to estimate each point&#44; blood velocity in the three &#40;3&#41; spatial directions and in all phases of cardiac cycle&#46; Postprocess is detailed in the three &#40;3&#41; simplified steps shown in <a class="elsevierStyleCrossRef" href="#fig0020">Fig&#46; 4</a>&#46;</p><elsevierMultimedia ident="fig0020"></elsevierMultimedia><p id="par0065" class="elsevierStylePara elsevierViewall">Additionally the following hemodynamic variables were estimated in different vessels&#58; ejected volume&#44; average velocity&#44; maximum and minimum velocity<a class="elsevierStyleCrossRef" href="#bib0085"><span class="elsevierStyleSup">17</span></a> by picking out 2D regions of interest from one &#40;1&#41; automatic edge detection algorithm&#46;</p></span></span><span id="sec0035" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0055">Methods of representation</span><p id="par0070" class="elsevierStylePara elsevierViewall">4D Flow data are commonly viewed as streamlines&#44; pathlines&#44; vector charts and 3D meshes that allow us to represent 3D blood flow patters in time&#46;<a class="elsevierStyleCrossRef" href="#bib0090"><span class="elsevierStyleSup">18</span></a> In <a class="elsevierStyleCrossRef" href="#fig0025">Fig&#46; 5</a> the different kinds of 3D representation that can be obtained through this modality can be seen&#46;<ul class="elsevierStyleList" id="lis0005"><li class="elsevierStyleListItem" id="lsti0005"><span class="elsevierStyleLabel">1&#46;</span><p id="par0075" class="elsevierStylePara elsevierViewall"><span class="elsevierStyleItalic">Streamlines</span>&#46; Streamlines are curves that connect elements &#40;particles&#41; of a fluid in space with the characteristic that the curve is tangential to the velocity vector of several particles in a given time&#46; This is why streamlines give us a general view of the flow pattern described by a fluid element in the cardiac cycle&#46;<a class="elsevierStyleCrossRef" href="#bib0095"><span class="elsevierStyleSup">19</span></a></p></li><li class="elsevierStyleListItem" id="lsti0010"><span class="elsevierStyleLabel">2&#46;</span><p id="par0080" class="elsevierStylePara elsevierViewall"><span class="elsevierStyleItalic">Pathlines</span>&#46; Pathlines indicate the trajectory followed by one particle of fluid&#46; This representation can be regarded as the pathway followed by one fluid element within the flow at a given time&#46; Pathlines are used to analyze the temporal evolution of blood flow patterns during the cardiac cycle&#46; They can be coded with colors representing flow velocity too&#46;<a class="elsevierStyleCrossRef" href="#bib0100"><span class="elsevierStyleSup">20</span></a></p></li><li class="elsevierStyleListItem" id="lsti0015"><span class="elsevierStyleLabel">3&#46;</span><p id="par0085" class="elsevierStylePara elsevierViewall"><span class="elsevierStyleItalic">Vector charts and meshes&#46;</span> For every particle vectors charts associate one vector with the magnitude and direction of flow velocity at that point so that for every point the evolution of the flow velocity pattern can be observed through time&#46; You can also draw regions of interest in any vessels in order to analyze the behavior of blood flow through any of the aforementioned methods&#46; You can also build 3D meshes to get profile information of blood flow velocity on one vessel or region if interest&#46;</p></li></ul></p><elsevierMultimedia ident="fig0025"></elsevierMultimedia></span><span id="sec0040" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0060">Artifacts and difficulties</span><p id="par0090" class="elsevierStylePara elsevierViewall">Phase contrast images are susceptible to errors induced by eddy currents and other heterogeneities of the magnetic field<a class="elsevierStyleCrossRef" href="#bib0105"><span class="elsevierStyleSup">21</span></a> that have a tendency to increase as the studied region moves away from the magnetic isocenter&#46; Aliasing can occur since the velocity encoding &#40;Venc&#41; sampling is lower than blood velocity&#46; It is important to correct data when these artifacts show up in order to guarantee the vision and quantification of flow&#46;<a class="elsevierStyleCrossRefs" href="#bib0020"><span class="elsevierStyleSup">4&#44;7&#44;9</span></a> To prevent these artifacts from happening due to aliasing the use of one 2D cine PCA sequence is recommended and based on the estimated peak blood velocity the picking of the most adequate value for Venc in the 4D Flow acquisition always bearing in mind that a high Venc eliminates artifacts due to aliasing but it also limits the low velocity-flow sensitivity&#46;<a class="elsevierStyleCrossRefs" href="#bib0110"><span class="elsevierStyleSup">22&#44;23</span></a> When it comes to artifacts caused by respiratory motion there are several compensation modalities by using sensors &#40;navigators&#41; capable of detecting the patient&#39;s respiration through the diaphragmatic motion while synchronizing such respiration with 4D Flow acquisition&#46;<a class="elsevierStyleCrossRefs" href="#bib0020"><span class="elsevierStyleSup">4&#44;7&#44;24&#44;25</span></a></p><p id="par0095" class="elsevierStylePara elsevierViewall">The main setback of 4D Flow is the duration of 4D Flow acquisition sequences&#8211;time especially increases when more spatial and temporal resolution is needed or simply when the acquired volume is increased&#46; To solve this problem there are speed-up methods including parallel images&#44; sampling strategies exploring the correlations in the space&#8211;time domain and radial acquisitions of <span class="elsevierStyleItalic">k</span>-space&#46;<a class="elsevierStyleCrossRefs" href="#bib0130"><span class="elsevierStyleSup">26&#8211;28</span></a> These modalities can produce overlapping artifacts or reduce the temporal resolution of images&#46; Some of these strategies are currently under investigations and validation&#46;</p><p id="par0100" class="elsevierStylePara elsevierViewall">The difficulty of clinically applying 4D Flow is due to both the time of sequence acquisition and its complexity&#46; However&#44; both scientific investigation and the interest of the industry on MRIs predict a bright future for this modality&#46;</p></span><span id="sec0045" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0065">Applications of the <span class="elsevierStyleItalic">4D Flow</span> modality</span><p id="par0105" class="elsevierStylePara elsevierViewall">The 4D Flow modality allows us to assess blood flow both quantitatively and qualitatively in any vessels of interest and without any means of contrast&#46; It allows us to assess the heart large vessels&#44; its ventricles and valves in systole and diastole in one acquisition only&#46; Also thanks to 3D acquisition it allows us to select within the acquired volume 2D views through any angles&#46; Similarly and based on the area of interest abdominal&#44; neck&#44; and skull vessels can be assessed among other vascular structures&#46;<a class="elsevierStyleCrossRefs" href="#bib0145"><span class="elsevierStyleSup">29&#8211;31</span></a></p><p id="par0110" class="elsevierStylePara elsevierViewall">To better illustrate the capacity of quantitative analysis this modality has the videos that can be seen in files <span class="elsevierStyleItalic">4D Flow</span> 1 and <span class="elsevierStyleItalic">4D Flow</span> 2&#46;<ul class="elsevierStyleList" id="lis0010"><li class="elsevierStyleListItem" id="lsti0020"><span class="elsevierStyleLabel">1&#46;</span><p id="par0115" class="elsevierStylePara elsevierViewall"><span class="elsevierStyleItalic">Aorta and heart</span> &#40;<a class="elsevierStyleCrossRef" href="#fig0030">Fig&#46; 6</a>&#41;&#46; One of the various potential clinical applications of 4D Flow is the study of flow in the ascending aorta&#46; Hope et al&#46;<a class="elsevierStyleCrossRef" href="#bib0155"><span class="elsevierStyleSup">31</span></a> evaluated one subgroup of patients with bivalve aortas and found within this group an eccentric abnormal blood flow associated with major shear forces in the vessel walls as well as a greater risk of aneurysms&#46; Consequently the risk of aneurysm in patients with bivalve aorta can be stratified by quantifying the shear forces obtained through 4D Flow&#44; which allows more objectivity both in the diagnosis and possible interventions in the ascending aorta of these patients&#46; Also possibly though 4D Flow we will be able to quantify both the abnormal flow and the hemodynamic load&#59; define the risk of vascular diseases like aneurysms or dissections before developing symptoms&#46;<a class="elsevierStyleCrossRef" href="#bib0155"><span class="elsevierStyleSup">31</span></a></p><elsevierMultimedia ident="fig0030"></elsevierMultimedia><p id="par0230" class="elsevierStylePara elsevierViewall">The 4D Flow modality will possibly be useful in numerous clinical situations associated with congenital thoracic cardiovascular anomalies like persistent arterial ductus&#44; conotruncal anomalies like the Fallot tetralogy&#44; partial or total anomalous pulmonary venous returns and in patients with left-to-right shunting&#46; Also with this modality we can identify patients at risk of pulmonary hypertension and further development of Eisenmenge syndrome&#46;<a class="elsevierStyleCrossRef" href="#bib0160"><span class="elsevierStyleSup">32</span></a> Similarly the MRI is the postoperative chosen modality since it allows us to assess both anatomic and functionally the cardiac cavities and the integrity and functionality of surgical corrections like interatrial baffles&#44; outflow septostomies and Fontan conduits&#44; among others&#46; This is why 4D Flow images have a greater clinical value because they give us information of both intra and extra cardiac blood flow patterns in these situations&#46;<a class="elsevierStyleCrossRefs" href="#bib0085"><span class="elsevierStyleSup">17&#44;33</span></a></p></li><li class="elsevierStyleListItem" id="lsti0025"><span class="elsevierStyleLabel">2&#46;</span><p id="par0120" class="elsevierStylePara elsevierViewall"><span class="elsevierStyleItalic">Abdomen</span>&#46; The Doppler ultrasound has been the less invasive modality for the study of blood flows and the quantification of hemodynamic parameters in the vena cava&#44; the portal system and vascular diseases &#40;<a class="elsevierStyleCrossRef" href="#fig0035">Fig&#46; 7</a>&#41;&#46; However&#44; it can be greatly affected by factors such as obesity&#44; intestinal air or edema&#46; This is why the MR-angiography is more and more important&#8211;because it does an excellent anatomical evaluation&#46; Markl et al&#46;<a class="elsevierStyleCrossRef" href="#bib0020"><span class="elsevierStyleSup">4</span></a> published their results showing the potentiality of the 4D Flow modality both for the quantification and characterization of blood flow of the different intra-abdominal vascular structures and in some pathological situations like portal hypertension&#46;<a class="elsevierStyleCrossRef" href="#bib0170"><span class="elsevierStyleSup">34</span></a></p><elsevierMultimedia ident="fig0035"></elsevierMultimedia><p id="par0235" class="elsevierStylePara elsevierViewall">In patients with renal transplant the MRIs can be very useful because they give us important information like perfusion defects&#44; tumors&#44; post-transplant proliferative conditions and collections&#44; among others&#46; Through the 4D Flow modality we will be able to evaluate the size&#44; orientation and relation to other abdominal organs and integrity of the intra-renal vascular structures&#46; It is also possible to evaluate the anastomoses of the renal artery to the iliac artery and of the renal vein to the iliac vein while watching and quantifying blood flow through these anastomoses &#40;<a class="elsevierStyleCrossRef" href="#fig0035">Fig&#46; 7</a>&#41;&#46;<a class="elsevierStyleCrossRef" href="#bib0020"><span class="elsevierStyleSup">4</span></a></p></li><li class="elsevierStyleListItem" id="lsti0030"><span class="elsevierStyleLabel">3&#46;</span><p id="par0125" class="elsevierStylePara elsevierViewall"><span class="elsevierStyleItalic">Carotid and cerebral arteries</span>&#46; The digital subtraction angiography is the chosen method to evaluate neck and intra-cranial arteries but it is also an invasive method requiring iodinated contrast media&#46; Thus techniques like CT-angiography &#40;with iodinated contrast too though&#41; and MR-angiography are more and more important for the study of carotid and cerebral arteries&#46; The carotid Doppler is a very valuable low cost-tool thanks to its great accuracy for the study of carotid bifurcations yet&#44; it is limited to small fields&#44; it depends on the operator and is sensitive to a poor acoustic window&#46; The 4D Flow modality has all the characteristics of the aforementioned methods while giving us anatomic and hemodynamic information &#40;<a class="elsevierStyleCrossRef" href="#fig0040">Fig&#46; 8</a>&#41;&#46; It allows us to assess the vascular structures of the neck and the distribution of the carotid walls shear forces&#44; find carotid stenoses and quantify blood velocity in the common carotid artery and the carotid bifurcation&#46;<a class="elsevierStyleCrossRef" href="#bib0175"><span class="elsevierStyleSup">35</span></a> It is also useful to evaluate and quantify both flow and velocity patterns in aneurysms and intra- and extra-cranial arteriovenous malformations&#46;<a class="elsevierStyleCrossRefs" href="#bib0180"><span class="elsevierStyleSup">36&#8211;38</span></a> Studies performed in 7<span class="elsevierStyleHsp" style=""></span>T ultra-high magnetic field resonance imaging devices have shown that smaller caliber-vessels can adequately be assessed &#40;e&#46;g&#46; the posterior cerebral artery&#41; and their pulsatility index be measured which can be applied in processes like trigeminal neuralgia&#46;<a class="elsevierStyleCrossRef" href="#bib0195"><span class="elsevierStyleSup">39</span></a></p><elsevierMultimedia ident="fig0040"></elsevierMultimedia></li></ul></p></span><span id="sec0050" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0070"><span class="elsevierStyleItalic">4D Flow</span> and 2D cine PCA</span><p id="par0130" class="elsevierStylePara elsevierViewall">Today 2D cine PCA is widely used for the study of blood flow in cardiovascular MRIs&#46; However&#44; its main setback is that is uses 2D cut views that need to be selected during the examination and require a highly skilled operator managing the resonance equipment&#46; In contrast in the 4D Flow modality the whole heart is 3D volume-studied&#44; so the planning of the sequence is greatly simplified without contribution from the operator &#40;<a class="elsevierStyleCrossRef" href="#fig0045">Fig&#46; 9</a>&#41;&#46; However&#44; the greater advantage of the 4D Flow modality lies in its capacity to perform multiplanar image analyses in the postprocessing since it allows the quantification of blood flow in any arbitrary views and angles without the need for acquiring new images&#46; Other than measuring flow and velocity it also allows us to estimate hemodynamic biomarkers such as shear stress in the blood vessel walls&#44; pressure gradients and pulse wave velocity without the use of any venous contrast media&#46;<a class="elsevierStyleCrossRefs" href="#bib0200"><span class="elsevierStyleSup">40&#44;41</span></a></p><elsevierMultimedia ident="fig0045"></elsevierMultimedia><p id="par0135" class="elsevierStylePara elsevierViewall">Some studies have confirmed that the 4D Flow measurements of flow and velocity strongly correlate to that of 2D cine PCA&#46;<a class="elsevierStyleCrossRefs" href="#bib0210"><span class="elsevierStyleSup">42&#44;43</span></a> On the other hand these modalities have a tendency to underestimate 20&#8211;25&#37; of the maximum velocity or average systolic peak observed through Doppler ultrasound&#46;<a class="elsevierStyleCrossRef" href="#bib0175"><span class="elsevierStyleSup">35</span></a> However&#44; we need to understand that in the MRIs velocity is always averaged since both multiple cardiac beats and a greater voxel size are needed to be able to obtain 3D images which in turn translates into a lower temporal and spatial resolution than that of Doppler ultrasound&#46; When the resolution of the 4D Flow sequence is very high only those voxels displaying the fastest velocity will show values more proximal to that of Doppler ultrasound&#46;</p></span><span id="sec0055" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0075">Conclusion</span><p id="par0140" class="elsevierStylePara elsevierViewall">In this review the 4D Flow modality is shown to be valid&#44; non-invasive and allowing us to perform quantitative and qualitative analyses of blood flow in large and medium-size vessels without the need for exogenous contrasts&#46; This modality has a great clinical potential since it gives us complete morphologic and anatomical information on the cardiovascular system by estimating hemodynamic biomarkers and visualizing blood flow patterns&#46;</p></span><span id="sec0060" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0080">Conflict of interests</span><p id="par0145" class="elsevierStylePara elsevierViewall">The authors reported no conflicts of interests&#46;</p></span><span id="sec0065" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0085">Ethical responsibilities</span><span id="sec0070" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0090">Protection of people and animals</span><p id="par0150" class="elsevierStylePara elsevierViewall">Authors confirm that no experiments have been performed on human beings or animals&#46;</p></span><span id="sec0075" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0095">Data confidentiality</span><p id="par0155" class="elsevierStylePara elsevierViewall">Authors confirm that the protocols of their centers have been followed on matters concerning the publishing of data from patients&#46; They also confirm that all patients included in this study have been given enough information and handed over their written informed consent for their participation in this study&#46;</p></span><span id="sec0080" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0100">Right to privacy and informed consent</span><p id="par0160" class="elsevierStylePara elsevierViewall">Authors confirm that in this report there are no personal data of patients&#46;</p></span></span><span id="sec0085" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0105">Authors&#47;collaborators</span><p id="par0165" class="elsevierStylePara elsevierViewall"><ul class="elsevierStyleList" id="lis0015"><li class="elsevierStyleListItem" id="lsti0035"><span class="elsevierStyleLabel">1&#46;</span><p id="par0170" class="elsevierStylePara elsevierViewall">Manager of the integrity of the study&#58; JAPZ&#44; JADB&#44; SRR&#46;</p></li><li class="elsevierStyleListItem" id="lsti0040"><span class="elsevierStyleLabel">2&#46;</span><p id="par0175" class="elsevierStylePara elsevierViewall">Original idea of the study&#58; JAPZ&#44; JADB&#44; SRR&#46;</p></li><li class="elsevierStyleListItem" id="lsti0045"><span class="elsevierStyleLabel">3&#46;</span><p id="par0180" class="elsevierStylePara elsevierViewall">Study design&#58; JAPZ&#44; VDCB&#44; CBA&#44; SMV&#46;</p></li><li class="elsevierStyleListItem" id="lsti0050"><span class="elsevierStyleLabel">4&#46;</span><p id="par0185" class="elsevierStylePara elsevierViewall">Data mining&#58; JAPZ&#44; CBA&#46;</p></li><li class="elsevierStyleListItem" id="lsti0055"><span class="elsevierStyleLabel">5&#46;</span><p id="par0190" class="elsevierStylePara elsevierViewall">Data analysis and interpretation&#58; JAPZ&#44; VDCB&#44; CBA&#44; SMV&#46;</p></li><li class="elsevierStyleListItem" id="lsti0060"><span class="elsevierStyleLabel">6&#46;</span><p id="par0195" class="elsevierStylePara elsevierViewall">Statistical analysis&#58; VDCB&#46;</p></li><li class="elsevierStyleListItem" id="lsti0065"><span class="elsevierStyleLabel">7&#46;</span><p id="par0200" class="elsevierStylePara elsevierViewall">Reference search&#58; CBA&#44; SMV&#46;</p></li><li class="elsevierStyleListItem" id="lsti0070"><span class="elsevierStyleLabel">8&#46;</span><p id="par0205" class="elsevierStylePara elsevierViewall">Writing&#58; JAPZ&#44; VDCB&#44; SMV&#44; CBA&#44; JADB&#44; SRR&#46;</p></li><li class="elsevierStyleListItem" id="lsti0075"><span class="elsevierStyleLabel">9&#46;</span><p id="par0210" class="elsevierStylePara elsevierViewall">Manuscript critical review with intellectually relevant contributions&#58; JADB&#44; SRR&#46;</p></li><li class="elsevierStyleListItem" id="lsti0080"><span class="elsevierStyleLabel">10&#46;</span><p id="par0215" class="elsevierStylePara elsevierViewall">Final version approval&#58; JAPZ&#44; VDCB&#44; SMV&#44; CBA&#44; JADB&#44; SRR&#46;</p></li></ul></p></span></span>"
    "textoCompletoSecciones" => array:1 [
      "secciones" => array:16 [
        0 => array:2 [
          "identificador" => "xres389418"
          "titulo" => "Abstract"
        ]
        1 => array:2 [
          "identificador" => "xpalclavsec367707"
          "titulo" => "Keywords"
        ]
        2 => array:2 [
          "identificador" => "xres389417"
          "titulo" => "Resumen"
        ]
        3 => array:2 [
          "identificador" => "xpalclavsec367706"
          "titulo" => "Palabras clave"
        ]
        4 => array:2 [
          "identificador" => "sec0005"
          "titulo" => "Introduction"
        ]
        5 => array:2 [
          "identificador" => "sec0010"
          "titulo" => "Principles of the 2D cine PCA"
        ]
        6 => array:3 [
          "identificador" => "sec0015"
          "titulo" => "4D Flow"
          "secciones" => array:3 [
            0 => array:2 [
              "identificador" => "sec0020"
              "titulo" => "Technique of acquisition"
            ]
            1 => array:2 [
              "identificador" => "sec0025"
              "titulo" => "Image acquisition"
            ]
            2 => array:2 [
              "identificador" => "sec0030"
              "titulo" => "Image processing"
            ]
          ]
        ]
        7 => array:2 [
          "identificador" => "sec0035"
          "titulo" => "Methods of representation"
        ]
        8 => array:2 [
          "identificador" => "sec0040"
          "titulo" => "Artifacts and difficulties"
        ]
        9 => array:2 [
          "identificador" => "sec0045"
          "titulo" => "Applications of the 4D Flow modality"
        ]
        10 => array:2 [
          "identificador" => "sec0050"
          "titulo" => "4D Flow and 2D cine PCA"
        ]
        11 => array:2 [
          "identificador" => "sec0055"
          "titulo" => "Conclusion"
        ]
        12 => array:2 [
          "identificador" => "sec0060"
          "titulo" => "Conflict of interests"
        ]
        13 => array:3 [
          "identificador" => "sec0065"
          "titulo" => "Ethical responsibilities"
          "secciones" => array:3 [
            0 => array:2 [
              "identificador" => "sec0070"
              "titulo" => "Protection of people and animals"
            ]
            1 => array:2 [
              "identificador" => "sec0075"
              "titulo" => "Data confidentiality"
            ]
            2 => array:2 [
              "identificador" => "sec0080"
              "titulo" => "Right to privacy and informed consent"
            ]
          ]
        ]
        14 => array:2 [
          "identificador" => "sec0085"
          "titulo" => "Authors&#47;collaborators"
        ]
        15 => array:1 [
          "titulo" => "References"
        ]
      ]
    ]
    "pdfFichero" => "main.pdf"
    "tienePdf" => true
    "fechaRecibido" => "2013-11-15"
    "fechaAceptado" => "2014-08-14"
    "PalabrasClave" => array:2 [
      "en" => array:1 [
        0 => array:4 [
          "clase" => "keyword"
          "titulo" => "Keywords"
          "identificador" => "xpalclavsec367707"
          "palabras" => array:5 [
            0 => "<span class="elsevierStyleItalic">4D Flow</span>"
            1 => "Cardiovascular magnetic resonance imaging"
            2 => "Phase contrast angiography"
            3 => "Hemodynamics"
            4 => "Blood flow"
          ]
        ]
      ]
      "es" => array:1 [
        0 => array:4 [
          "clase" => "keyword"
          "titulo" => "Palabras clave"
          "identificador" => "xpalclavsec367706"
          "palabras" => array:5 [
            0 => "<span class="elsevierStyleItalic">4D Flow</span>"
            1 => "Resonancia magn&#233;tica cardiovascular"
            2 => "Angiograf&#237;a con contraste de fase"
            3 => "Hemodin&#225;mica"
            4 => "Flujo sangu&#237;neo"
          ]
        ]
      ]
    ]
    "tieneResumen" => true
    "resumen" => array:2 [
      "en" => array:2 [
        "titulo" => "Abstract"
        "resumen" => "<p id="spar0005" class="elsevierStyleSimplePara elsevierViewall">The <span class="elsevierStyleItalic">4D Flow</span> MRI technique provides a three-dimensional representation of blood flow over time&#44; making it possible to evaluate the hemodynamics of the cardiovascular system both qualitatively and quantitatively&#46; In this article&#44; we describe the application of the <span class="elsevierStyleItalic">4D Flow</span> technique in a 3<span class="elsevierStyleHsp" style=""></span>T scanner&#59; in addition to the technical parameters&#44; we discuss the advantages and limitations of the technique and its possible clinical applications&#46; We used <span class="elsevierStyleItalic">4D Flow</span> MRI to study different body areas &#40;chest&#44; abdomen&#44; neck&#44; and head&#41; in 10 volunteers&#46; We obtained 3D representations of the patterns of flow and quantitative hemodynamic measurements&#46; The technique makes it possible to evaluate the pattern of blood flow in large and midsize vessels without the need for exogenous contrast agents&#46;</p>"
      ]
      "es" => array:2 [
        "titulo" => "Resumen"
        "resumen" => "<p id="spar0010" class="elsevierStyleSimplePara elsevierViewall">La t&#233;cnica de resonancia magn&#233;tica <span class="elsevierStyleItalic">4D Flow</span> permite evaluar cualitativa y cuantitativamente la hemodin&#225;mica del sistema cardiovascular representando en tres dimensiones los patrones de flujo sangu&#237;neo en el tiempo y cuantificando variables hemodin&#225;micas&#46; En este trabajo describimos la t&#233;cnica <span class="elsevierStyleItalic">4D Flow</span> en un equipo de resonancia de 3<span class="elsevierStyleHsp" style=""></span>T y adicionalmente se exponen&#44; adem&#225;s de los par&#225;metros t&#233;cnicos&#44; las ventajas&#44; las limitaciones y las posibles aplicaciones cl&#237;nicas&#46; Para esto estudiamos a diez voluntarios con la t&#233;cnica <span class="elsevierStyleItalic">4D Flow</span> en diferentes &#225;reas corporales &#40;t&#243;rax&#44; abdomen&#44; cuello y cr&#225;neo&#41; con la que obtuvimos representaciones tridimensionales de los patrones del flujo y medidas cuantitativas hemodin&#225;micas&#46; La t&#233;cnica permite evaluar los patrones de flujo sangu&#237;neo en vasos grandes y medianos sin la necesidad de contrastes ex&#243;genos&#46;</p>"
      ]
    ]
    "NotaPie" => array:1 [
      0 => array:2 [
        "etiqueta" => "&#9734;"
        "nota" => "<p class="elsevierStyleNotepara" id="npar0005">Please cite this article as&#58; Pineda Zapata JA&#44; Delgado de Bedout JA&#44; Rascovsky Ram&#237;rez S&#44; Bustamante C&#44; Mesa S&#44; Calvo Betancur VD&#46; Introducci&#243;n pr&#225;ctica al an&#225;lisis hemodin&#225;mico del sistema cardiovascular mediante la t&#233;cnica &#171;4D Flow&#187;&#46; Radiolog&#237;a&#46; 2014&#59;56&#58;485&#8211;495&#46;</p>"
      ]
    ]
    "apendice" => array:1 [
      0 => array:1 [
        "seccion" => array:1 [
          0 => array:4 [
            "apendice" => "<p id="par0225" class="elsevierStylePara elsevierViewall">The following are the supplementary data to this article&#58;<elsevierMultimedia ident="upi0005"></elsevierMultimedia><elsevierMultimedia ident="upi0010"></elsevierMultimedia></p>"
            "etiqueta" => "Appendix A"
            "titulo" => "Supplementary data"
            "identificador" => "sec0095"
          ]
        ]
      ]
    ]
    "multimedia" => array:12 [
      0 => array:7 [
        "identificador" => "fig0005"
        "etiqueta" => "Figure 1"
        "tipo" => "MULTIMEDIAFIGURA"
        "mostrarFloat" => true
        "mostrarDisplay" => false
        "figura" => array:1 [
          0 => array:4 [
            "imagen" => "gr1.jpeg"
            "Alto" => 2009
            "Ancho" => 2161
            "Tamanyo" => 178907
          ]
        ]
        "descripcion" => array:1 [
          "en" => "<p id="spar0015" class="elsevierStyleSimplePara elsevierViewall">Velocity coding principle in 2D cine PCA images&#46; In time 1 a positive gradient is applied resulting in a phase displacement that is the same both for the static &#40;circle without asterisk&#41; and mobile spin &#40;circle with asterisk&#41;&#46; In time 2 a gradient of the same magnitude but different opposed to the 1st one is applied&#46; It is now that the mobile spin has been displaced from its original position while perceiving a different gradient magnitude from the static spin&#46; After the application of the bipolar gradient the result is that static spin phase between time 1 and time 2 is zero&#59; however&#44; the mobile spin rotates different angles in two &#40;2&#41; times and consequently accumulates a phase proportional to the velocity and distance gone by&#46;</p>"
        ]
      ]
      1 => array:7 [
        "identificador" => "fig0010"
        "etiqueta" => "Figure 2"
        "tipo" => "MULTIMEDIAFIGURA"
        "mostrarFloat" => true
        "mostrarDisplay" => false
        "figura" => array:1 [
          0 => array:4 [
            "imagen" => "gr2.jpeg"
            "Alto" => 581
            "Ancho" => 1300
            "Tamanyo" => 46639
          ]
        ]
        "descripcion" => array:1 [
          "en" => "<p id="spar0020" class="elsevierStyleSimplePara elsevierViewall">2D cine PCA axial images at the level of the pulmonary artery bifurcation&#46; &#40;A&#41; Magnitude image&#46; Blood vessel anatomic information&#46; &#40;B&#41; Phase image&#46; Blood velocity is coded through the intensity of pixels&#8211;bright color meaning flow coming out of the image and dark color meaning flow coming in&#46;</p>"
        ]
      ]
      2 => array:7 [
        "identificador" => "fig0015"
        "etiqueta" => "Figure 3"
        "tipo" => "MULTIMEDIAFIGURA"
        "mostrarFloat" => true
        "mostrarDisplay" => false
        "figura" => array:1 [
          0 => array:4 [
            "imagen" => "gr3.jpeg"
            "Alto" => 2588
            "Ancho" => 2919
            "Tamanyo" => 336472
          ]
        ]
        "descripcion" => array:1 [
          "en" => "<p id="spar0025" class="elsevierStyleSimplePara elsevierViewall">Image acquisition scheme through the 4D Flow MRI modality&#46; For each phase &#40;&#916;<span class="elsevierStyleItalic">t</span>&#41; of the cardiac cycle &#40;synchronized with the ECG signal&#41; four &#40;4&#41; different sets of 3D data are collected&#8211;one &#40;1&#41; reference set of data and three &#40;3&#41; sets of data to code flow&#46; After data acquisition the phase image reconstruction is performed with the existing phase difference between the reference data and the phase codification data for every direction&#46; The magnitude image is reconstructed through a complex difference&#46; This figure from left to right shows both the magnitude image and velocity images in the <span class="elsevierStyleItalic">x</span>-&#44; <span class="elsevierStyleItalic">y</span>-&#44; <span class="elsevierStyleItalic">z</span>-axes&#44; respectively&#46;</p>"
        ]
      ]
      3 => array:7 [
        "identificador" => "fig0020"
        "etiqueta" => "Figure 4"
        "tipo" => "MULTIMEDIAFIGURA"
        "mostrarFloat" => true
        "mostrarDisplay" => false
        "figura" => array:1 [
          0 => array:4 [
            "imagen" => "gr4.jpeg"
            "Alto" => 597
            "Ancho" => 1751
            "Tamanyo" => 267943
          ]
        ]
        "descripcion" => array:1 [
          "en" => "<p id="spar0030" class="elsevierStyleSimplePara elsevierViewall">4D Flow image postprocessing&#46; Firstly the vascular structures in the magnitude image were segmented by choosing an appropriate intensity threshold and secondly a 3D isosurface was created based on the maximum velocity data included in the segmentation&#46; Finally the vessel isosurface was used to generate a certain number of seeds &#40;particles&#41; inside the isosurface that were used to estimate the streamlines&#44; the pathlines and the 3D vector charts and meshes&#46;</p>"
        ]
      ]
      4 => array:7 [
        "identificador" => "fig0025"
        "etiqueta" => "Figure 5"
        "tipo" => "MULTIMEDIAFIGURA"
        "mostrarFloat" => true
        "mostrarDisplay" => false
        "figura" => array:1 [
          0 => array:4 [
            "imagen" => "gr5.jpeg"
            "Alto" => 2661
            "Ancho" => 2919
            "Tamanyo" => 789058
          ]
        ]
        "descripcion" => array:1 [
          "en" => "<p id="spar0035" class="elsevierStyleSimplePara elsevierViewall">3D view methods of the aorta&#46; 4D Flow images&#46; &#40;A&#41; The streamlines give a 3D perspective of the movement of blood at the vascular structures&#46; &#40;B&#41; The pathlines give us information on the trajectory&#44; velocity&#44; direction&#44; and evolution of blood flow patters in every phase of cardiac cycle&#46; &#40;C&#41; Vector charts show the magnitude of velocity and direction for every particle inside the flow in a 2D view&#46; At the image both the aorta and pulmonary trunk are shown&#46; &#40;D&#41; Representation of the profile of blood velocity in the ascending and descending aorta through two &#40;2&#41; 3D meshes&#46;</p>"
        ]
      ]
      5 => array:7 [
        "identificador" => "fig0030"
        "etiqueta" => "Figure 6"
        "tipo" => "MULTIMEDIAFIGURA"
        "mostrarFloat" => true
        "mostrarDisplay" => false
        "figura" => array:1 [
          0 => array:4 [
            "imagen" => "gr6.jpeg"
            "Alto" => 1024
            "Ancho" => 3252
            "Tamanyo" => 366419
          ]
        ]
        "descripcion" => array:1 [
          "en" => "<p id="spar0040" class="elsevierStyleSimplePara elsevierViewall">4D Flow images in large vessels and heart&#46; &#40;A&#41; Segmentation of aorta and representation of blood flow through pathlines&#46; &#40;B&#41; View of blood trajectory in the aortic arch and the pulmonary artery in one healthy volunteer&#46; &#40;C&#41; Heart ventricular filling in diastole&#46;</p>"
        ]
      ]
      6 => array:7 [
        "identificador" => "fig0035"
        "etiqueta" => "Figure 7"
        "tipo" => "MULTIMEDIAFIGURA"
        "mostrarFloat" => true
        "mostrarDisplay" => false
        "figura" => array:1 [
          0 => array:4 [
            "imagen" => "gr7.jpeg"
            "Alto" => 2263
            "Ancho" => 3252
            "Tamanyo" => 606413
          ]
        ]
        "descripcion" => array:1 [
          "en" => "<p id="spar0045" class="elsevierStyleSimplePara elsevierViewall">4D Flow images in abdominal vessels and in one patient with a renal transplant&#46; &#40;A&#41; Healthy volunteer&#46; Abdominal aorta&#44; main efferent branches and main intra-abdominal venous structures&#46; &#40;B&#41; Detail from the origin of the celiac trunk&#44; the superior mesenteric artery and the portal vein&#46; &#40;C&#41; Saggital image with anatomic reference of the abdominal aorta and some of its main branches&#46; &#40;D&#41; Patients with a renal transplant&#46; Coronal view where 3D streamlines of the vascular structures irrigating the transplanted kidney can be seen&#46; &#40;E&#41; Detail of the arterial &#40;caudal vessel&#41; and venous flow &#40;celiac vessel&#41; and blood velocity coded based on a scale of colors&#46; &#40;F&#41; Pathlines of former case&#46;</p>"
        ]
      ]
      7 => array:7 [
        "identificador" => "fig0040"
        "etiqueta" => "Figure 8"
        "tipo" => "MULTIMEDIAFIGURA"
        "mostrarFloat" => true
        "mostrarDisplay" => false
        "figura" => array:1 [
          0 => array:4 [
            "imagen" => "gr8.jpeg"
            "Alto" => 1271
            "Ancho" => 3252
            "Tamanyo" => 334403
          ]
        ]
        "descripcion" => array:1 [
          "en" => "<p id="spar0050" class="elsevierStyleSimplePara elsevierViewall">Blood flow patterns at the main cerebral arteries and neck&#46; &#40;A&#41; 4D Flow of the carotid and vertebrobasilar system of one healthy volunteer performed in 3T&#46; Color patterns are coded according to velocity&#46; &#40;B&#41; View through pathlines at the bifurcation of the right carotid and vertebral arteries&#46; &#40;C&#41; Velocity patterns at the right carotid vertebral arteries showing greater peak velocity in the carotid artery&#46; Note the parabolic profiles so characteristic of laminar flow in both arteries&#46;</p>"
        ]
      ]
      8 => array:7 [
        "identificador" => "fig0045"
        "etiqueta" => "Figure 9"
        "tipo" => "MULTIMEDIAFIGURA"
        "mostrarFloat" => true
        "mostrarDisplay" => false
        "figura" => array:1 [
          0 => array:4 [
            "imagen" => "gr9.jpeg"
            "Alto" => 1334
            "Ancho" => 3252
            "Tamanyo" => 343176
          ]
        ]
        "descripcion" => array:1 [
          "en" => "<p id="spar0055" class="elsevierStyleSimplePara elsevierViewall">4D Flow and 2D PCA&#46; &#40;A&#41; Quantification of the average velocity of flow in the ascending aorta through the 2D cine PCA modality&#46; To the left we can see flow direction for every vessel and to the right the chart of velocity values in the cardiac cycle&#46; &#40;B&#41; The same quantification was performed with 4D Flow&#46; Note the similarity of values of the average velocity of blood measured through 2D cine PCA and 4D Flow modalities &#40;B&#41;&#46; &#40;C&#41; Selection of the ascending aorta and the pulmonary artery in one single 4D Flow study&#46; The charts of average velocity of aorta &#40;up&#41; and pulmonary artery blood &#40;down&#41; are shown here as well&#46; In charts A and B <span class="elsevierStyleItalic">y</span>-axis is blood velocity &#40;cm&#47;s&#41; while <span class="elsevierStyleItalic">x</span>-axis is time &#40;s&#41;&#46; However&#44; in chart C <span class="elsevierStyleItalic">x</span>-axis represents the cardiac cycle divided into sixteen &#40;16&#41; different phases&#46;</p>"
        ]
      ]
      9 => array:7 [
        "identificador" => "tbl0005"
        "etiqueta" => "Table 1"
        "tipo" => "MULTIMEDIATABLA"
        "mostrarFloat" => true
        "mostrarDisplay" => false
        "tabla" => array:2 [
          "leyenda" => "<p id="spar0065" class="elsevierStyleSimplePara elsevierViewall">ET&#44; echo time&#59; RT&#44; repetition time&#59; Venc&#44; velocity encoding&#46;</p>"
          "tablatextoimagen" => array:1 [
            0 => array:2 [
              "tabla" => array:1 [
                0 => """
                  <table border="0" frame="\n
                  \t\t\t\t\tvoid\n
                  \t\t\t\t" class=""><thead title="thead"><tr title="table-row"><td class="td" title="\n
                  \t\t\t\t\ttable-head\n
                  \t\t\t\t  " align="" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t" style="border-bottom: 2px solid black">&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-head\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t" style="border-bottom: 2px solid black">Thorax&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-head\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t" style="border-bottom: 2px solid black">Abdomen&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-head\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t" style="border-bottom: 2px solid black">Neck and skull&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr></thead><tbody title="tbody"><tr title="table-row"><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">Field of view &#40;mm&#41;&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">350<span class="elsevierStyleHsp" style=""></span>&#215;<span class="elsevierStyleHsp" style=""></span>350<span class="elsevierStyleHsp" style=""></span>&#215;<span class="elsevierStyleHsp" style=""></span>120&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">350<span class="elsevierStyleHsp" style=""></span>&#215;<span class="elsevierStyleHsp" style=""></span>350<span class="elsevierStyleHsp" style=""></span>&#215;<span class="elsevierStyleHsp" style=""></span>120&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">270<span class="elsevierStyleHsp" style=""></span>&#215;<span class="elsevierStyleHsp" style=""></span>270<span class="elsevierStyleHsp" style=""></span>&#215;<span class="elsevierStyleHsp" style=""></span>70&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">Voxel size &#40;mm&#41;&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">2&#46;5<span class="elsevierStyleHsp" style=""></span>&#215;<span class="elsevierStyleHsp" style=""></span>2&#46;5<span class="elsevierStyleHsp" style=""></span>&#215;<span class="elsevierStyleHsp" style=""></span>5&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">2&#46;5<span class="elsevierStyleHsp" style=""></span>&#215;<span class="elsevierStyleHsp" style=""></span>2&#46;5<span class="elsevierStyleHsp" style=""></span>&#215;<span class="elsevierStyleHsp" style=""></span>5&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">1&#46;5<span class="elsevierStyleHsp" style=""></span>&#215;<span class="elsevierStyleHsp" style=""></span>1&#46;5<span class="elsevierStyleHsp" style=""></span>&#215;<span class="elsevierStyleHsp" style=""></span>1&#46;5&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">Images&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">48&#8211;56&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">40&#8211;50&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">40&#8211;60&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">RT &#40;ms&#41;&#47;ET &#40;ms&#41;&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">4&#47;2&#46;1&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">4&#47;2&#46;1&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">4&#47;2&#46;1&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">Venc &#40;cm&#47;s&#41;&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">200&#8211;150&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">100&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">110&#8211;160&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">SENSE &#40;factor&#41;&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">2&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">2&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">2&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">Angle of inclination &#40;&#176;&#41;&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">6&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">6&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">6&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr><tr title="table-row"><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">Phases&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">16&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">20&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td><td class="td" title="\n
                  \t\t\t\t\ttable-entry\n
                  \t\t\t\t  " align="left" valign="\n
                  \t\t\t\t\ttop\n
                  \t\t\t\t">16&nbsp;\t\t\t\t\t\t\n
                  \t\t\t\t</td></tr></tbody></table>
                  """
              ]
              "imagenFichero" => array:1 [
                0 => "xTab599941.png"
              ]
            ]
          ]
        ]
        "descripcion" => array:1 [
          "en" => "<p id="spar0060" class="elsevierStyleSimplePara elsevierViewall">Acquisition parameters for the 4D Flow modality based on the area of study&#46;</p>"
        ]
      ]
      10 => array:5 [
        "identificador" => "upi0005"
        "tipo" => "MULTIMEDIAECOMPONENTE"
        "mostrarFloat" => false
        "mostrarDisplay" => true
        "Ecomponente" => array:3 [
          "fichero" => "mmc1.mp4"
          "ficheroTamanyo" => 5403240
          "Video" => array:2 [
            "flv" => array:5 [
              "fichero" => "mmc1.flv"
              "poster" => "mmc1.jpg"
              "tiempo" => 0
              "alto" => 0
              "ancho" => 0
            ]
            "mp4" => array:2 [
              "fichero" => "mmc1.m4v"
              "poster" => "mmc1.jpg"
            ]
          ]
        ]
      ]
      11 => array:5 [
        "identificador" => "upi0010"
        "tipo" => "MULTIMEDIAECOMPONENTE"
        "mostrarFloat" => false
        "mostrarDisplay" => true
        "Ecomponente" => array:3 [
          "fichero" => "mmc2.mp4"
          "ficheroTamanyo" => 2783731
          "Video" => array:2 [
            "flv" => array:5 [
              "fichero" => "mmc2.flv"
              "poster" => "mmc2.jpg"
              "tiempo" => 0
              "alto" => 0
              "ancho" => 0
            ]
            "mp4" => array:2 [
              "fichero" => "mmc2.m4v"
              "poster" => "mmc2.jpg"
            ]
          ]
        ]
      ]
    ]
    "bibliografia" => array:2 [
      "titulo" => "References"
      "seccion" => array:1 [
        0 => array:2 [
          "identificador" => "bibs0005"
          "bibliografiaReferencia" => array:43 [
            0 => array:3 [
              "identificador" => "bib0005"
              "etiqueta" => "1"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "NMR blood flow imaging using multiecho&#44; phase contrast sequences"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:1 [
                            0 => "M&#46; O&#8217;Donnell"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1118/1.595736"
                      "Revista" => array:6 [
                        "tituloSerie" => "Med Phys"
                        "fecha" => "1985"
                        "volumen" => "12"
                        "paginaInicial" => "59"
                        "paginaFinal" => "64"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/3974526"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            1 => array:3 [
              "identificador" => "bib0010"
              "etiqueta" => "2"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Effect of protocol choice on phase contrast cardiac magnetic resonance flow measurement in the ascending aorta&#58; breath-hold and non-breath-hold"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "M&#46;A&#46; Bolen"
                            1 => "R&#46;M&#46; Setser"
                            2 => "R&#46;S&#46; Gabriel"
                            3 => "R&#46;D&#46; Renapurkar"
                            4 => "Y&#46; Tandon"
                            5 => "M&#46;L&#46; Lieber"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1007/s10554-012-0047-z"
                      "Revista" => array:6 [
                        "tituloSerie" => "Int J Cardiovasc Imaging"
                        "fecha" => "2013"
                        "volumen" => "29"
                        "paginaInicial" => "113"
                        "paginaFinal" => "120"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/22527258"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            2 => array:3 [
              "identificador" => "bib0015"
              "etiqueta" => "3"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Non-contrast enhanced MR angiography&#58; physical principles"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:2 [
                            0 => "A&#46;J&#46; Wheaton"
                            1 => "M&#46; Miyazaki"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1002/jmri.23641"
                      "Revista" => array:6 [
                        "tituloSerie" => "J Magn Reson Imaging"
                        "fecha" => "2012"
                        "volumen" => "36"
                        "paginaInicial" => "286"
                        "paginaFinal" => "304"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/22807222"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            3 => array:3 [
              "identificador" => "bib0020"
              "etiqueta" => "4"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "4D flow MRI"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:5 [
                            0 => "M&#46; Markl"
                            1 => "A&#46; Frydrychowicz"
                            2 => "S&#46; Kozerke"
                            3 => "M&#46; Hope"
                            4 => "O&#46; Wieben"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1002/jmri.23632"
                      "Revista" => array:6 [
                        "tituloSerie" => "J Magn Reson Imaging"
                        "fecha" => "2012"
                        "volumen" => "36"
                        "paginaInicial" => "1015"
                        "paginaFinal" => "1036"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/23090914"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            4 => array:3 [
              "identificador" => "bib0025"
              "etiqueta" => "5"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Comparison of hemodynamics of intracranial aneurysms between MR fluid dynamics using 3D cine phase-contrast MRI and MR-based computational fluid dynamics"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "H&#46; Isoda"
                            1 => "Y&#46; Ohkura"
                            2 => "T&#46; Kosugi"
                            3 => "M&#46; Hirano"
                            4 => "M&#46;T&#46; Alley"
                            5 => "R&#46; Bammer"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1007/s00234-009-0634-4"
                      "Revista" => array:6 [
                        "tituloSerie" => "Neuroradiology"
                        "fecha" => "2010"
                        "volumen" => "52"
                        "paginaInicial" => "913"
                        "paginaFinal" => "920"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/19967532"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            5 => array:3 [
              "identificador" => "bib0030"
              "etiqueta" => "6"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "4D phase-contrast flow cardiovascular magnetic resonance&#58; comprehensive quantification and visualization of flow dynamics in atrial septal defect and partial anomalous pulmonary venous return"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [
                            0 => "I&#46; Valverde"
                            1 => "J&#46; Simpson"
                            2 => "T&#46; Schaeffter"
                            3 => "P&#46; Beerbaum"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1007/s00246-010-9782-x"
                      "Revista" => array:7 [
                        "tituloSerie" => "Pediatr Cardiol"
                        "fecha" => "2010"
                        "volumen" => "31"
                        "paginaInicial" => "1244"
                        "paginaFinal" => "1248"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/20848278"
                            "web" => "Medline"
                          ]
                        ]
                        "itemHostRev" => array:3 [
                          "pii" => "S0169500211005150"
                          "estado" => "S300"
                          "issn" => "01695002"
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            6 => array:3 [
              "identificador" => "bib0035"
              "etiqueta" => "7"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Comprehensive 4D velocity mapping of the heart and great vessels by cardiovascular magnetic resonance"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [
                            0 => "M&#46; Markl"
                            1 => "P&#46;J&#46; Kilner"
                            2 => "T&#46; Ebbers"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1186/1532-429X-13-7"
                      "Revista" => array:5 [
                        "tituloSerie" => "J Cardiovasc Magn Reson"
                        "fecha" => "2011"
                        "volumen" => "13"
                        "paginaInicial" => "7"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/21235751"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            7 => array:3 [
              "identificador" => "bib0040"
              "etiqueta" => "8"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Cardiovascular applications of phase-contrast MRI"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [
                            0 => "M&#46;B&#46; Srichai"
                            1 => "R&#46;P&#46; Lim"
                            2 => "S&#46; Wong"
                            3 => "V&#46;S&#46; Lee"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.2214/AJR.07.3744"
                      "Revista" => array:6 [
                        "tituloSerie" => "AJR Am J Roentgenol"
                        "fecha" => "2009"
                        "volumen" => "192"
                        "paginaInicial" => "662"
                        "paginaFinal" => "675"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/19234262"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            8 => array:3 [
              "identificador" => "bib0045"
              "etiqueta" => "9"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Cardiovascular flow measurement with phase-contrast MR imaging&#58; basic facts and implementation"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [
                            0 => "J&#46; Lotz"
                            1 => "C&#46; Meier"
                            2 => "A&#46; Leppert"
                            3 => "M&#46; Galanski"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1148/radiographics.22.3.g02ma11651"
                      "Revista" => array:6 [
                        "tituloSerie" => "Radiographics"
                        "fecha" => "2002"
                        "volumen" => "22"
                        "paginaInicial" => "651"
                        "paginaFinal" => "671"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/12006694"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            9 => array:3 [
              "identificador" => "bib0050"
              "etiqueta" => "10"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Blood flow velocity assessment"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:1 [
                            0 => "D&#46; Firmin"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "LibroEditado" => array:5 [
                        "titulo" => "Cardiovascular magnetic resonance"
                        "paginaInicial" => "91"
                        "paginaFinal" => "99"
                        "edicion" => "2nd ed&#46;"
                        "serieFecha" => "2010"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            10 => array:3 [
              "identificador" => "bib0055"
              "etiqueta" => "11"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Angiographic pulse sequences"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:1 [
                            0 => "M&#46;A&#46; Bernstein"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "LibroEditado" => array:4 [
                        "titulo" => "Handbook of MRI pulse sequences"
                        "paginaInicial" => "648"
                        "paginaFinal" => "701"
                        "serieFecha" => "2004"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            11 => array:3 [
              "identificador" => "bib0060"
              "etiqueta" => "12"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Time-resolved three-dimensional phase-contrast MRI"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "M&#46; Markl"
                            1 => "F&#46;P&#46; Chan"
                            2 => "M&#46;T&#46; Alley"
                            3 => "K&#46;L&#46; Wedding"
                            4 => "M&#46;T&#46; Draney"
                            5 => "C&#46;J&#46; Elkins"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1002/jmri.10272"
                      "Revista" => array:6 [
                        "tituloSerie" => "J Magn Reson Imaging"
                        "fecha" => "2003"
                        "volumen" => "17"
                        "paginaInicial" => "499"
                        "paginaFinal" => "506"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/12655592"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            12 => array:3 [
              "identificador" => "bib0065"
              "etiqueta" => "13"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Flow imaging&#58; cardiac applications of 3D cine phase-contrast MRI"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:1 [
                            0 => "T&#46; Ebbers"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Curr Cardiovasc Imaging Rep"
                        "fecha" => "2011"
                        "volumen" => "4"
                        "paginaInicial" => "127"
                        "paginaFinal" => "133"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            13 => array:3 [
              "identificador" => "bib0070"
              "etiqueta" => "14"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Four-dimensional flow MRI using spiral acquisition"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [
                            0 => "A&#46; Sigfridsson"
                            1 => "S&#46; Petersson"
                            2 => "C&#46;J&#46; Carlh&#228;ll"
                            3 => "T&#46; Ebbers"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1002/mrm.23297"
                      "Revista" => array:6 [
                        "tituloSerie" => "Magn Reson Med"
                        "fecha" => "2012"
                        "volumen" => "68"
                        "paginaInicial" => "1065"
                        "paginaFinal" => "1073"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/22161650"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            14 => array:3 [
              "identificador" => "bib0075"
              "etiqueta" => "15"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Surcando el espacio-k para mejorar la imagen por resonancia magn&#233;tica"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:5 [
                            0 => "D&#46; Moratal-P&#233;rez"
                            1 => "L&#46; Mart&#237;-Bonmat&#237;"
                            2 => "M&#46;E&#46; Brummer"
                            3 => "J&#46; Millet-Roig"
                            4 => "F&#46; Castells"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Radiolog&#237;a"
                        "fecha" => "2004"
                        "volumen" => "46"
                        "paginaInicial" => "133"
                        "paginaFinal" => "150"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            15 => array:3 [
              "identificador" => "bib0080"
              "etiqueta" => "16"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Semiautomated method for noise reduction and background phase error correction in MR phase velocity data"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:6 [
                            0 => "P&#46;G&#46; Walker"
                            1 => "G&#46;B&#46; Cranney"
                            2 => "M&#46;B&#46; Scheidegger"
                            3 => "G&#46; Waseleski"
                            4 => "G&#46;M&#46; Pohost"
                            5 => "A&#46;P&#46; Yoganathan"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "J Magn Reson Imaging"
                        "fecha" => "1993"
                        "volumen" => "3"
                        "paginaInicial" => "521"
                        "paginaFinal" => "530"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/8324312"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            16 => array:3 [
              "identificador" => "bib0085"
              "etiqueta" => "17"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Assessment of normal flow patterns in the pulmonary circulation by using 4D magnetic resonance velocity mapping"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "P&#46; B&#228;chler"
                            1 => "N&#46; Pinochet"
                            2 => "J&#46; Sotelo"
                            3 => "G&#46; Crelier"
                            4 => "P&#46; Irarrazaval"
                            5 => "C&#46; Tejos"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1016/j.mri.2012.06.036"
                      "Revista" => array:6 [
                        "tituloSerie" => "Magn Reson Imaging"
                        "fecha" => "2013"
                        "volumen" => "31"
                        "paginaInicial" => "178"
                        "paginaFinal" => "188"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/22898700"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            17 => array:3 [
              "identificador" => "bib0090"
              "etiqueta" => "18"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Kinematics of fluido&#46; Introduction to fluid mechanics and fluid machines"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:2 [
                            0 => "S&#46;K&#46; Som"
                            1 => "G&#46; Biswas"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Libro" => array:6 [
                        "edicion" => "2nd ed&#46;"
                        "fecha" => "2007"
                        "paginaInicial" => "78"
                        "paginaFinal" => "107"
                        "editorial" => "Tata McGraw-Hill"
                        "editorialLocalizacion" => "New Delhi"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            18 => array:3 [
              "identificador" => "bib0095"
              "etiqueta" => "19"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Visualizing three-dimensional flow with simulated streamlines and three-dimensional phase-contrast MR imaging"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [
                            0 => "S&#46; Napel"
                            1 => "D&#46;H&#46; Lee"
                            2 => "R&#46; Frayne"
                            3 => "B&#46;K&#46; Rutt"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "J Magn Reson Imaging"
                        "fecha" => "1992"
                        "volumen" => "2"
                        "paginaInicial" => "143"
                        "paginaFinal" => "153"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/1562765"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            19 => array:3 [
              "identificador" => "bib0100"
              "etiqueta" => "20"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Particle trace visualization of intracardiac flow using time-resolved 3D phase contrast MRI"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "L&#46; Wigstr&#246;m"
                            1 => "T&#46; Ebbers"
                            2 => "A&#46; Fyrenius"
                            3 => "M&#46; Karlsson"
                            4 => "J&#46; Engvall"
                            5 => "B&#46; Wranne"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Magn Reson Med"
                        "fecha" => "1999"
                        "volumen" => "41"
                        "paginaInicial" => "793"
                        "paginaFinal" => "799"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/10332856"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            20 => array:3 [
              "identificador" => "bib0105"
              "etiqueta" => "21"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Influence of eddy current&#44; Maxwell and gradient field corrections on 3D flow visualization of 3D CINE PC-MRI data"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:6 [
                            0 => "R&#46; Lorenz"
                            1 => "J&#46; Bock"
                            2 => "J&#46; Snyder"
                            3 => "J&#46;G&#46; Korvink"
                            4 => "B&#46;A&#46; Jung"
                            5 => "M&#46; Markl"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1002/mrm.24885"
                      "Revista" => array:6 [
                        "tituloSerie" => "Magn Reson Med"
                        "fecha" => "2014"
                        "volumen" => "72"
                        "paginaInicial" => "33"
                        "paginaFinal" => "40"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/24006013"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            21 => array:3 [
              "identificador" => "bib0110"
              "etiqueta" => "22"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Accelerated time-resolved three-dimensional MR velocity mapping of blood flow patterns in the aorta using SENSE and k-t BLAST"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [
                            0 => "A&#46; Stadlbauer"
                            1 => "W&#46; van der Riet"
                            2 => "G&#46; Crelier"
                            3 => "E&#46; Salomonowitz"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1016/j.ejrad.2009.06.009"
                      "Revista" => array:6 [
                        "tituloSerie" => "Eur J Radiol"
                        "fecha" => "2010"
                        "volumen" => "75"
                        "paginaInicial" => "e15"
                        "paginaFinal" => "e21"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/19581063"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            22 => array:3 [
              "identificador" => "bib0115"
              "etiqueta" => "23"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "4D FLOW&#58; una nueva herramienta de diagn&#243;stico para cardiopat&#237;as cong&#233;nitas"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "N&#46; Pinochet"
                            1 => "P&#46; B&#227;chler"
                            2 => "C&#46; Tejos"
                            3 => "G&#46; Crelier"
                            4 => "R&#46; Parra"
                            5 => "J&#46;M&#46; Allendes"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Rev Chil Radiol"
                        "fecha" => "2011"
                        "volumen" => "17"
                        "paginaInicial" => "134"
                        "paginaFinal" => "140"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            23 => array:3 [
              "identificador" => "bib0120"
              "etiqueta" => "24"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Four-dimensional &#40;4D&#41; flow of the whole heart and great vessels using real-time respiratory self-gating"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:6 [
                            0 => "S&#46; Uribe"
                            1 => "P&#46; Beerbaum"
                            2 => "T&#46;S&#46; S&#248;rensen"
                            3 => "A&#46; Rasmusson"
                            4 => "R&#46; Razavi"
                            5 => "T&#46; Schaeffter"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1002/mrm.22090"
                      "Revista" => array:6 [
                        "tituloSerie" => "Magn Reson Med"
                        "fecha" => "2009"
                        "volumen" => "62"
                        "paginaInicial" => "984"
                        "paginaFinal" => "992"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/19672940"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            24 => array:3 [
              "identificador" => "bib0125"
              "etiqueta" => "25"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Generalized reconstruction of phase contrast MRI&#58; analysis and correction of the effect of gradient field distortions"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "M&#46; Markl"
                            1 => "R&#46; Bammer"
                            2 => "M&#46;T&#46; Alley"
                            3 => "C&#46;J&#46; Elkins"
                            4 => "M&#46;T&#46; Draney"
                            5 => "A&#46; Barnett"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1002/mrm.10582"
                      "Revista" => array:7 [
                        "tituloSerie" => "Magn Reson Med"
                        "fecha" => "2003"
                        "volumen" => "50"
                        "paginaInicial" => "791"
                        "paginaFinal" => "801"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/14523966"
                            "web" => "Medline"
                          ]
                        ]
                        "itemHostRev" => array:3 [
                          "pii" => "S0169500205003181"
                          "estado" => "S300"
                          "issn" => "01695002"
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            25 => array:3 [
              "identificador" => "bib0130"
              "etiqueta" => "26"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Accuracy and reproducibility in phase contrast imaging using SENSE"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [
                            0 => "P&#46; Thunberg"
                            1 => "M&#46; Karlsson"
                            2 => "L&#46; Wigstr&#246;m"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1002/mrm.10634"
                      "Revista" => array:6 [
                        "tituloSerie" => "Magn Reson Med"
                        "fecha" => "2003"
                        "volumen" => "50"
                        "paginaInicial" => "1061"
                        "paginaFinal" => "1068"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/14587017"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            26 => array:3 [
              "identificador" => "bib0135"
              "etiqueta" => "27"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Accelerating cine phase-contrast flow measurements using k-t BLAST and k-t SENSE"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:6 [
                            0 => "C&#46; Baltes"
                            1 => "S&#46; Kozerke"
                            2 => "M&#46;S&#46; Hansen"
                            3 => "K&#46;P&#46; Pruessmann"
                            4 => "J&#46; Tsao"
                            5 => "P&#46; Boesiger"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1002/mrm.20730"
                      "Revista" => array:7 [
                        "tituloSerie" => "Magn Reson Med"
                        "fecha" => "2005"
                        "volumen" => "54"
                        "paginaInicial" => "1430"
                        "paginaFinal" => "1438"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/16276492"
                            "web" => "Medline"
                          ]
                        ]
                        "itemHostRev" => array:3 [
                          "pii" => "S155608641533166X"
                          "estado" => "S300"
                          "issn" => "15560864"
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            27 => array:3 [
              "identificador" => "bib0140"
              "etiqueta" => "28"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "k-t PCA&#58; temporally constrained k-t BLAST reconstruction using principal component analysis"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:5 [
                            0 => "H&#46; Pedersen"
                            1 => "S&#46; Kozerke"
                            2 => "S&#46; Ringgaard"
                            3 => "K&#46; Nehrke"
                            4 => "W&#46;Y&#46; Kim"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1002/mrm.22052"
                      "Revista" => array:6 [
                        "tituloSerie" => "Magn Reson Med"
                        "fecha" => "2009"
                        "volumen" => "62"
                        "paginaInicial" => "706"
                        "paginaFinal" => "716"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/19585603"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            28 => array:3 [
              "identificador" => "bib0145"
              "etiqueta" => "29"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "4D flow magnetic resonance imaging in bicuspid aortic valve disease demonstrates altered distribution of aortic blood flow helicity"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "R&#46; Lorenz"
                            1 => "J&#46; Bock"
                            2 => "A&#46;J&#46; Barker"
                            3 => "F&#46; von Knobelsdorff-Brenkenhoff"
                            4 => "W&#46; Wallis"
                            5 => "J&#46;G&#46; Korvink"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1002/mrm.24802"
                      "Revista" => array:6 [
                        "tituloSerie" => "Magn Reson Med"
                        "fecha" => "2014"
                        "volumen" => "71"
                        "paginaInicial" => "1542"
                        "paginaFinal" => "1553"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/23716466"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            29 => array:3 [
              "identificador" => "bib0150"
              "etiqueta" => "30"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Aortic hemodynamics in patients with and without repair of aortic coarctation&#58; in vivo analysis by 4D flow-sensitive magnetic resonance imaging"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "A&#46; Frydrychowicz"
                            1 => "M&#46; Markl"
                            2 => "D&#46; Hirtler"
                            3 => "A&#46; Harloff"
                            4 => "C&#46; Schlensak"
                            5 => "J&#46; Geiger"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1097/RLI.0b013e3182034fc2"
                      "Revista" => array:6 [
                        "tituloSerie" => "Invest Radiol"
                        "fecha" => "2011"
                        "volumen" => "46"
                        "paginaInicial" => "317"
                        "paginaFinal" => "325"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/21285892"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            30 => array:3 [
              "identificador" => "bib0155"
              "etiqueta" => "31"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "4D flow CMR in assessment of valve-related ascending aortic disease"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "M&#46;D&#46; Hope"
                            1 => "T&#46;A&#46; Hope"
                            2 => "S&#46;E&#46; Crook"
                            3 => "K&#46;G&#46; Ordovas"
                            4 => "T&#46;H&#46; Urbania"
                            5 => "M&#46;T&#46; Alley"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1016/j.jcmg.2011.05.004"
                      "Revista" => array:6 [
                        "tituloSerie" => "JACC Cardiovasc Imaging"
                        "fecha" => "2011"
                        "volumen" => "4"
                        "paginaInicial" => "781"
                        "paginaFinal" => "787"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/21757170"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            31 => array:3 [
              "identificador" => "bib0160"
              "etiqueta" => "32"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Evaluation of elevated mean pulmonary arterial pressure based on magnetic resonance 4D velocity mapping&#58; comparison of visualization techniques"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "U&#46; Reiter"
                            1 => "G&#46; Reiter"
                            2 => "G&#46; Kovacs"
                            3 => "A&#46;F&#46; Stalder"
                            4 => "M&#46;A&#46; Gulsun"
                            5 => "A&#46; Greiser"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1371/journal.pone.0082212"
                      "Revista" => array:5 [
                        "tituloSerie" => "PLOS ONE"
                        "fecha" => "2013"
                        "volumen" => "8"
                        "paginaInicial" => "e82212"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/24349224"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            32 => array:3 [
              "identificador" => "bib0165"
              "etiqueta" => "33"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Postoperative pulmonary and aortic 3D haemodynamics in patients after repair of transposition of the great arteries"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "J&#46; Geiger"
                            1 => "D&#46; Hirtler"
                            2 => "J&#46; B&#252;rk"
                            3 => "B&#46; Stiller"
                            4 => "R&#46; Arnold"
                            5 => "B&#46; Jung"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1007/s00330-013-2998-4"
                      "Revista" => array:6 [
                        "tituloSerie" => "Eur Radiol"
                        "fecha" => "2014"
                        "volumen" => "24"
                        "paginaInicial" => "200"
                        "paginaFinal" => "208"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/23995974"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            33 => array:3 [
              "identificador" => "bib0170"
              "etiqueta" => "34"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "In vivo validation of 4D flow MRI for assessing the hemodynamics of portal hypertension"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "A&#46; Rold&#225;n-Alzate"
                            1 => "A&#46; Frydrychowicz"
                            2 => "E&#46; Niespodzany"
                            3 => "B&#46;R&#46; Landgraf"
                            4 => "K&#46;M&#46; Johnson"
                            5 => "O&#46; Wieben"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1002/jmri.23906"
                      "Revista" => array:6 [
                        "tituloSerie" => "J Magn Reson Imaging"
                        "fecha" => "2013"
                        "volumen" => "37"
                        "paginaInicial" => "1100"
                        "paginaFinal" => "1108"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/23148034"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            34 => array:3 [
              "identificador" => "bib0175"
              "etiqueta" => "35"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Comparison of blood flow velocity quantification by 4D flow MR imaging with ultrasound at the carotid bifurcation"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:6 [
                            0 => "A&#46; Harloff"
                            1 => "T&#46; Zech"
                            2 => "F&#46; Wegent"
                            3 => "C&#46; Strecker"
                            4 => "C&#46; Weiller"
                            5 => "M&#46; Markl"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.3174/ajnr.A3419"
                      "Revista" => array:7 [
                        "tituloSerie" => "AJNR Am J Neuroradiol"
                        "fecha" => "2013"
                        "volumen" => "34"
                        "paginaInicial" => "1407"
                        "paginaFinal" => "1413"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/23413247"
                            "web" => "Medline"
                          ]
                        ]
                        "itemHostRev" => array:3 [
                          "pii" => "S016950020800411X"
                          "estado" => "S300"
                          "issn" => "01695002"
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            35 => array:3 [
              "identificador" => "bib0180"
              "etiqueta" => "36"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Intracranial 4D Flow MRI&#58; Toward individualized assessment of arteriovenous malformation hemodynamics and treatment-induced changes"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "S&#46;A&#46; Ansari"
                            1 => "S&#46; Schnell"
                            2 => "T&#46; Carroll"
                            3 => "P&#46; Vakil"
                            4 => "M&#46;C&#46; Hurley"
                            5 => "C&#46; Wu"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.3174/ajnr.A3537"
                      "Revista" => array:6 [
                        "tituloSerie" => "AJNR Am J Neuroradiol"
                        "fecha" => "2013"
                        "volumen" => "34"
                        "paginaInicial" => "1922"
                        "paginaFinal" => "1928"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/23639564"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            36 => array:3 [
              "identificador" => "bib0185"
              "etiqueta" => "37"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Complete intracranial arterial and venous blood flow evaluation with 4D flow MR imaging"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "M&#46;D&#46; Hope"
                            1 => "D&#46;D&#46; Purcell"
                            2 => "T&#46;A&#46; Hope"
                            3 => "C&#46; von Morze"
                            4 => "D&#46;B&#46; Vigneron"
                            5 => "M&#46;T&#46; Alley"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.3174/ajnr.A1138"
                      "Revista" => array:7 [
                        "tituloSerie" => "AJNR Am J Neuroradiol"
                        "fecha" => "2009"
                        "volumen" => "30"
                        "paginaInicial" => "362"
                        "paginaFinal" => "366"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/18653687"
                            "web" => "Medline"
                          ]
                        ]
                        "itemHostRev" => array:3 [
                          "pii" => "S0169500209005960"
                          "estado" => "S300"
                          "issn" => "01695002"
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            37 => array:3 [
              "identificador" => "bib0190"
              "etiqueta" => "38"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Four-dimensional phase contrast magnetic resonance angiography&#58; potential clinical applications"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [
                            0 => "A&#46; Frydrychowicz"
                            1 => "C&#46;J&#46; Fran&#231;ois"
                            2 => "P&#46;A&#46; Turski"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1016/j.ejrad.2011.01.094"
                      "Revista" => array:6 [
                        "tituloSerie" => "Eur J Radiol"
                        "fecha" => "2011"
                        "volumen" => "80"
                        "paginaInicial" => "24"
                        "paginaFinal" => "35"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/21333479"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            38 => array:3 [
              "identificador" => "bib0195"
              "etiqueta" => "39"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "4D flow measurements in the superior cerebellar artery at 7 Tesla&#58; feasibility and potential for applications in patients with trigeminal neuralgia"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:6 [
                            0 => "S&#46; Schmitter"
                            1 => "B&#46;D&#46; Jagadeesan"
                            2 => "A&#46;W&#46; Grande"
                            3 => "J&#46; Sein"
                            4 => "K&#46; Ugurbil"
                            5 => "P&#46; Van de Moortele"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "J Cardiovasc Magn Reson"
                        "fecha" => "2013"
                        "volumen" => "15"
                        "numero" => "Suppl&#46; 1"
                        "paginaInicial" => "W21"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            39 => array:3 [
              "identificador" => "bib0200"
              "etiqueta" => "40"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "In vivo assessment of wallshear stress in the atherosclerotic aorta using flow-sensitive 4D MRI"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "A&#46; Harloff"
                            1 => "A&#46; Nussbaumer"
                            2 => "S&#46; Bauer"
                            3 => "A&#46;F&#46; Stalder"
                            4 => "A&#46; Frydrychowicz"
                            5 => "C&#46; Weiller"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1002/mrm.22383"
                      "Revista" => array:7 [
                        "tituloSerie" => "Magn Reson Med"
                        "fecha" => "2010"
                        "volumen" => "63"
                        "paginaInicial" => "1529"
                        "paginaFinal" => "1536"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/20512856"
                            "web" => "Medline"
                          ]
                        ]
                        "itemHostRev" => array:3 [
                          "pii" => "S1556086415319110"
                          "estado" => "S300"
                          "issn" => "15560864"
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            40 => array:3 [
              "identificador" => "bib0205"
              "etiqueta" => "41"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Aortic pulse wave velocity measurements with under sampled 4D flow-sensitive MRI&#58; comparison with 2D and algorithm determination"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => true
                          "autores" => array:6 [
                            0 => "A&#46;L&#46; Wentland"
                            1 => "O&#46; Wieben"
                            2 => "C&#46;J&#46; Fran&#231;ois"
                            3 => "C&#46; Boncyk"
                            4 => "A&#46; Munoz del Rio"
                            5 => "K&#46;M&#46; Johnson"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1002/jmri.23877"
                      "Revista" => array:6 [
                        "tituloSerie" => "J Magn Reson Imaging"
                        "fecha" => "2013"
                        "volumen" => "37"
                        "paginaInicial" => "853"
                        "paginaFinal" => "859"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/23124585"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            41 => array:3 [
              "identificador" => "bib0210"
              "etiqueta" => "42"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Improved cardiovascular flow quantification with time-resolved volumetric phase-contrast MRI"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:6 [
                            0 => "A&#46; Hsiao"
                            1 => "M&#46;T&#46; Alley"
                            2 => "P&#46; Massaband"
                            3 => "R&#46;J&#46; Herfkens"
                            4 => "F&#46;P&#46; Chan"
                            5 => "S&#46;S&#46; Vasanawala"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1007/s00247-010-1932-z"
                      "Revista" => array:6 [
                        "tituloSerie" => "Pediatr Radiol"
                        "fecha" => "2011"
                        "volumen" => "41"
                        "paginaInicial" => "711"
                        "paginaFinal" => "720"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/21221566"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            42 => array:3 [
              "identificador" => "bib0215"
              "etiqueta" => "43"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Three dimensional three component whole heart cardiovascular magnetic resonance velocity mapping&#58; comparison of flow measurements from 3D and 2D acquisitions"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:5 [
                            0 => "L&#46; Brix"
                            1 => "S&#46; Ringgaard"
                            2 => "A&#46; Rasmusson"
                            3 => "T&#46;S&#46; S&#248;rensen"
                            4 => "W&#46;Y&#46; Kim"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1186/1532-429X-11-3"
                      "Revista" => array:5 [
                        "tituloSerie" => "J Cardiovasc Magn Reson"
                        "fecha" => "2009"
                        "volumen" => "11"
                        "paginaInicial" => "3"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/19232119"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
          ]
        ]
      ]
    ]
  ]
  "idiomaDefecto" => "en"
  "url" => "/21735107/0000005600000006/v1_201412120104/S2173510714000603/v1_201412120104/en/main.assets"
  "Apartado" => array:4 [
    "identificador" => "8097"
    "tipo" => "SECCION"
    "en" => array:2 [
      "titulo" => "Updates in radiology"
      "idiomaDefecto" => true
    ]
    "idiomaDefecto" => "en"
  ]
  "PDF" => "https://static.elsevier.es/multimedia/21735107/0000005600000006/v1_201412120104/S2173510714000603/v1_201412120104/en/main.pdf?idApp=UINPBA00004N&text.app=https://www.elsevier.es/"
  "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S2173510714000603?idApp=UINPBA00004N"
]
Article information
ISSN: 21735107
Original language: English
The statistics are updated each day
Year/Month Html Pdf Total
2024 May 2 2 4
2024 January 1 0 1
2023 October 2 0 2
2023 August 1 0 1
2023 March 4 2 6
2023 January 1 0 1
2022 December 1 2 3
2018 May 3 1 4
2018 April 39 15 54
2018 March 36 2 38
2018 February 24 2 26
2018 January 22 12 34
2017 December 27 4 31
2017 November 31 2 33
2017 October 34 7 41
2017 September 22 4 26
2017 August 56 4 60
2017 July 42 4 46
2017 June 48 6 54
2017 May 76 10 86
2017 April 67 12 79
2017 March 52 29 81
2017 February 103 2 105
2017 January 48 3 51
2016 December 54 3 57
2016 November 83 9 92
2016 October 87 12 99
2016 September 118 14 132
2016 August 40 7 47
2016 July 59 3 62
2016 June 76 18 94
2016 May 93 27 120
2016 April 59 32 91
2016 March 87 18 105
2016 February 65 22 87
2016 January 48 12 60
2015 December 59 16 75
2015 November 48 7 55
2015 October 53 9 62
2015 September 41 7 48
2015 August 42 5 47
2015 July 41 11 52
2015 June 27 5 32
2015 May 31 5 36
2015 February 1 1 2
2015 January 0 1 1
2014 December 1 0 1
Show all

Follow this link to access the full text of the article

es en pt

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos