metricas
covid
Buscar en
Enfermedades Infecciosas y Microbiología Clínica (English Edition)
Toda la web
Inicio Enfermedades Infecciosas y Microbiología Clínica (English Edition) Direct bacterial identification from positive blood cultures using matrix-assist...
Información de la revista
Vol. 36. Núm. 8.
Páginas 484-492 (octubre 2018)
Compartir
Compartir
Descargar PDF
Más opciones de artículo
Visitas
1851
Vol. 36. Núm. 8.
Páginas 484-492 (octubre 2018)
Original article
Acceso a texto completo
Direct bacterial identification from positive blood cultures using matrix-assisted laser desorption/ionization time-of-flight (MALDI-TOF) mass spectrometry: A systematic review and meta-analysis
Identificación bacteriana directa a partir de hemocultivos positivos usando espectrometría de masas MALDI-TOF (matrix-assisted laser desorption/ionization time-of-flight): una revisión sistemática y metaanálisis
Visitas
1851
Jesús Ruiz-Aragóna, Mónica Ballestero-Télleza,
Autor para correspondencia
m.ballestero.t@gmail.com

Corresponding author.
, Belén Gutiérrez-Gutiérreza, Marina de Cuetoa,b, Jesús Rodríguez-Bañoa,c, Álvaro Pascuala,b
a Unidad Clínica de Intercentros de Enfermedades Infecciosas, Microbiología y Medicina Preventiva, Instituto de Biomedicina de Sevilla (IBiS), Hospitales Universitarios Virgen Macarena and Virgen del Rocío/CSIC/Universidad de Sevilla, Seville, Spain
b Departamento de Microbiología, Universidad de Sevilla, Seville, Spain
c Departamento de Medicina, Universidad de Sevilla, Seville, Spain
Este artículo ha recibido
Información del artículo
Resumen
Texto completo
Bibliografía
Descargar PDF
Estadísticas
Figuras (3)
Mostrar másMostrar menos
Tablas (3)
Table 1. Main characteristics of 32 studies included in the systematic review and meta-analysis.
Table 2. Meta-analysis of the accuracy of MALDI-TOF for included Gram-negative and Gram positive pathogens.
Table 3. Differences between in-house methods and the Sepsityper method among Gram-positive bacteria.
Mostrar másMostrar menos
Material adicional (1)
Abstract
Introduction

The rapid identification of bacteraemia-causing pathogens could assist clinicians in the timely prescription of targeted therapy, thereby reducing the morbidity and mortality of this infection. In recent years, numerous techniques that rapidly and directly identify positive blood cultures have been marketed, with matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) being one of the most commonly used.

Methods

The aim of this systematic review and meta-analysis was to evaluate the accuracy of MALDI-TOF (Bruker®) for the direct identification of positive blood culture bottles.

Results

A meta-analysis was performed to summarize the results of the 32 studies evaluated. The overall quality of the studies was moderate. For Gram-positive bacteria, overall rates of correct identification of the species ranged from 0.17 to 0.98, with a cumulative rate (random-effects model) of 0.72 (95% CI: 0.64–0.80). For Gram-negative bacteria, correct identification rates ranged from 0.66 to 1.00, with a cumulative effect of 0.92 (95% CI: 0.88–0.95). For Enterobacteriaceae, the rate was 0.96 (95% CI: 0.94–0.97).

Conclusion

MALDI-TOF mass spectrometry shows high accuracy for the correct identification of Gram-negative bacteria, particularly Enterobacteriaceae, directly from positive blood culture bottles, and moderate accuracy for the identification of Gram-positive bacteria (low for some species).

Keywords:
MALDI-TOF
Blood culture
Bacteraemia
Rapid identification
Resumen
Introducción

La identificación rápida de los patógenos causantes de bacteriemia orienta a los clínicos a prescribir con mayor celeridad un tratamiento dirigido y reducir así la morbimortalidad de dicha infección. Durante los últimos años, han aparecido en el mercado numerosas técnicas con la intención de cubrir esta necesidad, que logran una identificación rápida y directa a partir de los frascos de hemocultivos positivos. La espectrometría de masas matrix-assisted laser desorption/ionization with time-of-flight mass spectrometry (MALDI-TOF MS) es una de las tecnologías más utilizadas en este campo.

Métodos

El objetivo de ese estudio es realizar una revisión sistemática y metaanálisis que evalúe la precisión de MALDI-TOF MS (Bruker) para la identificación directa a partir de frascos de hemocultivos positivos.

Resultados

El metaanálisis fue realizado para sintetizar los resultados de los 32 estudios evaluados. La calidad total de los estudios fue moderada. Para las bacterias grampositivas, el ratio total de identificaciones correctas a nivel de especie fue del 0,17 al 0,98 con un ratio acumulativo (modelo de efectos aleatorios) de 0,72 (IC 95%: 0,64-0,80). Para las bacterias gramnegativas, el rango de identificaciones correctas fue del 0,66 al 1,00 con un efecto acumulativo de 0,92 (IC 95%: 0,88-0,95), llegando a un 0,96 (IC 95%: 0,94-0,97) en Enterobacteriaceae.

Conclusiones

La espectrometría de masas MALDI-TOF muestra una alta precisión para la correcta identificación de bacterias gramnegativas realizada directamente a partir de los frascos de hemocultivos positivos, siendo mayor en el grupo de las enterobacterias. Para grampositivas, la precisión es moderada, llegando a ser baja para alguna especie.

Palabras clave:
MALDI-TOF
Hemocultivo
Bacteriemia
Identificación rápida
Texto completo
Introduction

Bloodstream infection (BSI) is one of the leading causes of death around the world, with an estimated incidence of up to 19 million people worldwide every year.1 Rapid identification of the causative organism is essential to guide clinicians in the selection of the most appropriate targeted treatment for patients with BSI, and it is associated with improved patient outcomes.2 Many remarkable improvements have been made in the attempt to reduce the time required to identify the pathogen in positive blood cultures, including the direct inoculation of fluids from positive blood cultures into automated systems, fluorescent in situ hybridization (FISH), and PCR. All these methods, however, are expensive and require several hours of work.3

In recent years, matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) has emerged as the new standard in bacterial identification and is now being adopted in clinical laboratories worldwide.

MALDI-TOF technology uses a laser to softly ionize the many structural elements (primarily ribosomal proteins) of bacteria and yeasts and then separate the molecules according to their mass-to-charge ratio (m/z).4,5 Four commercial systems are currently in use worldwide: the MALDI-TOF BioTyper (Bruker Daltonics, Bremen, Germany); Saramis (AnagnosTec, Potsdam, Germany); Andromas (Andromas, Paris, France), and Vitek MS (bioMérieux, Marcy l’Étoile, France), although Bruker's BioTyper system is the most widely used in clinical practice and research.

Different studies have reported successful identification of bacteria directly from positive blood cultures using MALDI-TOF MS. Nevertheless, the results of these studies have varied, depending on the distribution of microbial isolates, the pre-treatment/extraction method applied, and where the log score cut-off values were defined.6 One of these methods has become commercially available as the Sepsityper Kit (Bruker Daltonics) to standardize the preparation of blood culture prior to spectrometric analysis. The method involves lysis of blood cells, followed by centrifugation and washing steps. The final result is a pellet of bacteria, which is further processed by standard methods for identification using MALDI-TOF.7,8

One disadvantage of the Sepsityper kit is the additional cost, so that many laboratories have developed their own in-house methods of bacterial extraction from positive blood culture bottles.9–11 These methods vary in their approach to removing human cellular components and enriching the microbes from blood culture fluids. Saponin12 and ammonium chloride13 to lyse the blood cells have been described. Separation of microorganisms from blood cells can be performed with differential centrifugation and gel separator tubes.14,15 Simple stepwise sedimentation of blood cells and microorganisms has also been described.5,16,17

The present systematic review and meta-analysis aimed to evaluate the accuracy of MALDI-TOF for the identification of bacteria directly from positive blood culture bottles. We analyzed the different pre-treatment methods and other variables that may have influenced the correct identification of bacteria to species level from positive blood culture bottles.

Methods

We designed a systematic review and meta-analysis of the scientific literature. Our study was performed according to the recommendations of the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines for systematic reviews.18

Information sources and searches

We systematically searched the Medline (PubMed), Embase, Cochrane Database of Systematic Reviews (CDSR), Center for Reviews and Dissemination (CRD), Network of Agencies for Health Technology Assessment database, and the clinical trial results database, “ClinicalTrials.gov”, to identify studies between January 1, 2011 and October 1, 2015 that assessed the accuracy of MALDI-TOF MS for direct identification of bacteria from blood cultures. To localize other undetected published articles, a manual search was performed for references in relevant studies and specialized medical journals. No language restrictions were applied. To achieve maximum sensitivity, the search was performed combining the following keywords and free terms: “bacteraemia”, “sepsis”, “bloodstream infection”, “humans”, “matrix-assisted laser desorption ionization time-of-flight mass spectrometry”, “MALDI-TOF mass spectrometry”, “clinical trial”, “prospective studies”, “accuracy”, “sensitivity and specificity”, “comparative study”, “evaluation studies”, and “diagnosis”.

Eligibility criteria and study selection

Two investigators screened the titles and abstracts of the references localized. The full text of potentially eligible studies was read and evaluated for definitive inclusion. Included were studies of MALDI-TOF MS (Bruker) performed to identify bacteria from blood cultures of patients, compared with routine bacterial identification (automated or manual methods: phenotypic, microbiological, molecular diagnosis). Only studies using the Bruker® system were included because it is the most frequently used system and allows for greater standardization in evaluation. Exclusion criteria were: studies that did not investigate blood cultures; studies identifying bacteria by mass spectrometry methods different from MALDI-TOF MS; studies without a comparator method, studies applying MALDI-TOF to subcultures of positive blood cultures; studies for the identification of mycobacteria, yeasts or parasites. We also excluded non-original articles, non-human studies, in vitro simulation studies, case-series studies, editorials and letters.

Data extraction and assessment of risk bias

Two independent researchers reviewed all references in order to identify articles that required full-text appraisal, with the final decision reached through consensus. All retrieved articles were evaluated for inclusion in the systematic review and meta-analysis. Study characteristics (design, year of publication, study country, period of study), population (size, age, sample type), intervention (previous processing, bacterial database version, score, procedures), comparator (gold standard, different comparators), and outcomes (proportion of identification, accuracy) were recorded.

We used the Quality Assessment of Diagnostic Accuracy Studies (QUADAS) checklist to assess the quality of the studies and potential risk of bias.19 Two researchers independently performed the quality assessment.

Statistical analysis

The main effect size index was the correct identification rate (CIR); specifically, the identification rate obtained by use of MALDI-TOF was compared with results obtained from the reference method. Some researchers have found that using a lower threshold (cut-off 2.0) than the one specified by the manufacturer provided the best identification compared to the reference method.

Statistical significance was defined as a p value of ≤0.001. Meta-analyses were performed with R statistical software (version 3.0.2; R Foundation for Statistical Computing, Vienna, Austria) and the package “meta” (version 3.1-2). The Freeman-Tukey transformation of inverse hyperbolic sine function was used to calculate the CIR. Both fixed-effect and random-effects (DerSimonian and Laird method) meta-analyses were performed and heterogeneity was evaluated on the basis of I2, the heterogeneity measure of Cochran's Q test. Heterogeneity was evaluated quantitatively using I2 in which I2 values of 25%, 50%, and 75% indicate low, moderate, and high heterogeneity, respectively. When significant heterogeneity was observed among the included studies, the random-effects model was considered; otherwise, the fixed-effect model was used. In order to check whether publication bias might have influenced the validity of the results, funnel plots and Egger's test were applied.

ResultsStudy selection

Our systematic review identified a total of 280 references. After duplicates (16) were ruled out and titles and abstracts (264) screened, 74 potentially relevant studies were retrieved for inclusion criteria and data extraction. Forty-two studies were excluded for various reasons: data not shown, no intervention, use of a technique different from MALDI-TOF (Bruker), no proper comparator, and no clinical samples. A total of 32 studies fulfilled all inclusion criteria and were finally selected for performance of the systematic review and meta-analysis. Fig. 1 shows a flowchart of the process.

Fig. 1.

Flowchart of the published studies evaluated for inclusion in the meta-analysis.

(0.22MB).
Study characteristics

The main characteristics of the eligible studies are listed in Table 1. Thirty-two articles were finally included in this meta-analysis: 19 from Europe, 9 performed in North America (7 in USA and 2 in Canada), 2 carried out in Australia, and 2 in Asia. The population consisted of adults in four studies and children in 3 studies; in 19 studies, the population was not specified.

Table 1.

Main characteristics of 32 studies included in the systematic review and meta-analysis.

Author  Year  Period  Country  Patient  Samples  Previous treatment  Bruker Database  Comparator  Pathogen 
Angeletti20  2015  Jan 2012–Dec 2013  Italy  Adults  294  Inh  3.3.1  Vitek-2  G+, G−, yeast, anaerobic 
Riederer8  2015  Na  USA  Adults and children  96  Sep and inh  3.1  MALDI-TOF and Vitek-2  G− 
Egli21  2015  Jul 2013–Sep 2013  Switzerland  Immunodepressed  62  Sep  3.1  Vitek-2  G+, G− 
Chebotar22  2015  Na  Russia  Children sepsis  139  Inh  3.0  Vitek-2  G+, G−, yeast 
Martinez23  2014  Aug 2012–Dec 2012  USA  Na  174  Sep  Na  MALDI-TOF and Vitek-2  G+, G−, yeast 
Schieffer24  2014  Na  USA  Na  411  Sep  3.1.2  Phoenix and Vitek-2  G+, G−, yeast, anaerobic 
Haigh25  2014  Na  UK  Adults and children  297  Sep  3.1  Culture  G+, G−, yeast 
Rodriguez-Sanchez26  2014  Aug 2011–Jan 2013  Spain  Na  1085  inh  3.0  Microscan, API  G+, G−, yeast, anaerobic 
Mestas27  2014  Na  USA  Children  159  inh  3.0  Vitek-2  G+, G− 
Tadros28  2013  Feb 2011–Mar 2011  Canada  Children  80  Sep  2.0; 3.1.0.4  Phoenix  G+ y G− 
Gray29  2013  Aug 2012–Jan 2013  Australia  Adults and children  318  inh  3.1  Phoenix, API  G− 
Martiny30  2013  Sep 2011–Mar 2012  Belgium  Adults and children  277  inh  3.1.2.0  Culture  G+, G−, yeast 
Jamal31  2013  Jan 2012–May 2012  Kuwait  Adults  160  Sep  3.0  Vitek-2  G+, G−, yeast 
Leli32  2013  Oct 2011–Sep 2012  Italy  Adults  109  inh  Na  Phoenix, API  G+, G−, yeast 
Chen33  2013  Mar 2012–Jul 2012  China  Na  202  Sep and inh for Bruker  3.0  Vitek-2  G+, G− 
Nonneman34  2013  Jan 2011–Mar 2011  Denmark  Na  256  Sep and inh  2.0  Culture  G+, G−, yeast 
March–Rosselló35  2013  Na  Spain  Na  100  inh  3.0  Culture, Vitek-2  G+, G− 
Clerc36  2013  Jan2010–Dec2010  Switzerland  Adults  202  inh  2.0  MALDI-TOF  G- 
Hoyos-Mallecot37  2012  Feb2012–April 2012  Spain  Na  100  inh  3.0  Vitek-2, MALDI-TOF  G+, G−, yeast 
Lagacé-Wiens38  2012  Na  Canada  Na  61  Sep  3.0  Vitek-2, API  G+, G−, yeast 
Meex6  2012  Na  Belgium  Na  113  Sep and inh  3.1.2  Subculture and MALDI-TOF  G+, G−anaerobic 
Saffert39  2012  Na  USA  Na  180  Sep and two inh  3.0  Phoenix, sequencing  G+, G−, yeast 
Vlek40  2012  Feb2010–Apr il2010  Holland  Na  89  inh  Na  Phoenix  G+, G− 
Martiny41  2012  5 weeks  USA  Na  66  Sep and inh  3.1.1  Vitek, MALDI-TOF, API  G+, G− 
Buchan42  2012  Jan 2011–June 2011  USA  Na  164  Sep  Na  VItek-2, rapid-ID  G+, G− 
Loonen43  2012  Na  Holland  Na  Na  Sep, molYsis and inh  3.1.1.0  Vitek-2, MALDI-TOF, PCR  G+, G− 
Klein44  2012  May2010–Dec 2010  Germany  Na  Na  Sep and inh  Na  PCR, culture  G+, G− 
Juiz16  2012  4 weeks  Spain  Na  85  Sep and inh  Na  Phoenix, sequencing  G+, G− 
Fuglsang-Damgaard45  2011  April 2010–June 2010  Denmark  Adults and children  583  inh  2.0  Culture, Vitek-2  G+, G− 
Schubert46  2011  3 months  Germany  Na  500  Sep and inh  3.1.1.0  API, Phoenix, MALDI-TOF, sequencing  G+, G−, yeast 
Kok47  2011  March 2011–April 2011  Australia  Na  507  Sep  2.0  Phoenix  G+, G− 
Ferreira12  2011  Na  Spain  Na  318  inh  2.0  Wider, PCR  G+, G−, yeast 

Sep, Sepsityper; inh, in house; Na, not available; G+, Gram-positive; G-, Gram-negative.

The global number of analyzed samples was 7187. In two studies,43,44 the number of samples was not provided. All studies used clinical isolates. All included studies reported their pre-treatment method: 13 used an in-house method,12,20,22,26,27,29,30,32,35–37,40,45 7 used Sepsityper,21,24,25,28,31,38,42 and the rest used a combination of the two.6,8,16,33,34,39,41,43,44,46

Twenty-five of the 32 studies included used a commercial identification system as the comparator method and in the other 7 studies,6,25,30,34,35,44,45 the comparator was identification by MALDI-TOF from subcultures.

Risk of bias of studies included

The overall quality of the studies included was moderate (Supplementary Table S1). All but two studies26,29 recorded the selection criteria for patients. Only one study reported the simultaneous realization of both tests.24 An adequate gold standard was used as comparator in all studies. Blinding of investigators for the interpretation of the reference test was reflected in only two studies.24,34 Losses that took place during the studies were explained and uninterpretable results were reported in all studies.

Synthesis of results

Overall results for Gram-negative bacteria were reported in 30 studies, and 3163 recorded samples. The correct identification rate (CIR) for these pathogens in the various studies ranged from 0.66 to 1.00 at the species level. There was significant heterogeneity in this meta-analysis (I2: 89.1%; p<0.001) (Fig. 2). The pooled CIR estimated using a random-effects model was 0.92 (95% CI: 0.88–0.95). Publication bias in this meta-analysis was not significant (p=0.02 by Egger's test; Supplementary Fig. S2).

Fig. 2.

The forest plot for the meta-analysis: overall identification of Gram-positive bacteria at the species level (A) and Gram-negative bacteria at the species level (B). CI, confidence interval; W, weight; fixed, fixed-effect model; random, random-effects model; events, number of correct identifications; total, total number of identifications. Gray squares represent the weight of individual studies with the fixed-effect model; horizontal lines through the squares represent 95% confidence intervals; gray diamonds represent the overall estimate and its confidence interval; dotted vertical lines represent the fixed-effect model; and dashed vertical lines represent the random-effects model.

(0.5MB).

The results for correct identification of Enterobacteriaceae were recorded in 28 studies and 2293 samples. The CIRs in the different studies ranged from 0.79 to 1.00 at the species level. Heterogeneity was I2=59.5% (p<0.001; Fig. 2). The pooled CIR estimated with the random-effects model was 0.96 (95% CI: 0.94–0.97). Publication bias in this meta-analysis was not significant (p=0.06 by Egger's test; Supplementary Fig. S2).

Accuracy of identification was also analyzed for each species of Enterobacteriaceae; the results are set out in Table 2. The highest correct identification rates (1.00) were of Enterobacter cloacae, Serratia marcescens, Enterobacter aerogenes, and Citrobacter freundii. In this analysis, heterogeneity in the species of Enterobacteriaceae was not significant, and I2 values ranged from 0% (p>0.70) to 41.3% (p=0.004). The CIRs for other Gram-negative pathogens were: Pseudomonas aeruginosa, 0.98; Acinetobacter baumannii, 0.88; and Haemophilus influenzae 0.81.

Table 2.

Meta-analysis of the accuracy of MALDI-TOF for included Gram-negative and Gram positive pathogens.

  Studies (nStrain (nFixed effectRandom-effectsHeterogeneity
      Proportion  95% CI  Proportion  95% CI  p  I2 
Gram-negative
Total Enterobacteriaceae  28  2293  0.96  0.95; 0.97  0.96  0.94; 0.98  <0.001  61% 
K. pneumoniae  28  408  0.98  0.95; 1.00  0.98  0.95; 0.00  0.99  0% 
E. coli  28  1443  0.98  0.97; 0.99  0.98  0.97; 0.00  0.004  41.3% 
E. cloacae  22  116  1.00  0.94; 1.00  0.99  0.92; 0.00  0.21  17.2% 
S. marcescens  21  77  1.00  0.96; 1.00  1.00  0.96; 0.00  0.94  0% 
K. oxytoca  18  68  0.97  0.87; 0.00  0.97  0.87; 0.00  0.77  0% 
P. mirabilis  16  97  0.98  0.91; 0.00  0.98  0.91; 0.00  0.85  0% 
M. morganii  11  31  0.99  0.84; 0.00  0.99  0.84; 0.00  0.92  0% 
E. aerogenes  30  1.00  0.89; 0.00  1.00  0.89; 0.00  0.74  0% 
C. freundii  23  1.00  0.89; 0.00  1.00  0.89; 0.00  0.72  0% 
Other Gram-negative
P. aeruginosa  29  304  0.98  0.92; 0.99  0.95  0.86; 1.00  <0.001  61.2% 
A. baumannii  14  57  0.88  0.73; 0.99  0.76  0.47; 0.97  0.001  56.6% 
H. influenzae  10  23  0.81  0.54; 0.99  0.69  0.28; 0.99  0.046  44.7% 
Gram-positive
Total staphylococci  26  2544  0.69  0.67; 0.71  0.72  0.65; 0.79  <0.001  92.9% 
S.aureus  26  682  0.74  0.70; 0.77  0.83  0.71; 0.82  <0.001  90.4% 
S. epidermidis  20  956  0.69  0.66; 0.72  0.71  0.60; 0.80  <0.001  88.9% 
Other CNS  26  906  0.72  0.68; 0.75  0.78  0.70; 0.86  <0.001  77.1% 
Total streptococci  23  412  0.38  0.32; 0.43  0.50  0.32; 0.68  <0.001  87.7% 
S. pneumoniae  19  212  0.31  0.23; 0.40  0.36  0.14; 0.60  <0.001  79.3% 
S. pyogenes  15  52  0.76  0.58; 0.90  0.83  0.56; 1.00  <0.001  61% 
Other streptococci  14  105  0.09  0.02; 0.18  0.19  0.03; 0.41  <0.001  67.8% 
S. agalactiae  13  43  0.92  0.75; 1.00  0.90  0.71; 1.00  0.27  15.2% 
Total enterococci  24               
E. faecalis  24  252  0.88  0.82; 0.93  0.86  0.74; 0.95  <0.001  67.3% 
E. faecium  22  179  0.84  0.76; 0.91  0.83  0.69; 0.94  <0.001  58.4% 

CNS, coagulase negative staphylococci.

A total of 27 studies, including 3904 samples, reported results for Gram-positive. The CIR in individual studies ranged from 0.17 to 0.98 at the species level. High heterogeneity was found in this meta-analysis (I2: 96.5%; p<0.001) (Fig. 3). The estimated pooled CIR was 0.72 (95% CI: 0.64–0.80) using the random-effects model. The funnel plot and Egger's test (p=0.06) showed that publication bias was not significant (Supplementary Fig. S2). A sensitivity analysis excluding the study 45 with a CIR of 0.17 (95% CI: 0.13–0.22) only reduced the heterogeneity to I2: 93.1 (p<0.001) and the pooled CIR estimated using a random effect model changed to 0.75 (95%CI:0.69–0.88). The rapid identification rates for different Gram-positive bacteria in blood cultures are recorded in Table 2. The rates of correct identification of Staphylococcus aureus, Staphylococcus epidermidis and other coagulase negative staphylococci were over 0.70. The results for streptococci were described in 23 studies and 412 reported samples. The CIRs of these pathogens, especially Streptococcus pneumoniae (0.36) and other alfa-hemolytic streptococci (0.19), were lower than those observed for staphylococci. The CIR of Streptococcus pyogenes was 0.83 and of Streptococcus agalactiae, 0.92. The rates of direct identification of Enterococcus faecalis and Enterococcus faecium were higher than 0.84 for both species. In these analyses, heterogeneity for each species of Gram-positive bacteria was significant (except for S. agalactiae), with I2 values above 60% (p<0.001), (Table 2).

Fig. 3.

The forest plot for the meta-analysis: identification of Enterobacteriaceae at the species level. CI, confidence interval; W, weight; fixed, fixed-effect model; random, random-effects model; events, number of correct identifications; total, total number of identifications. Gray squares represent the weight of individual studies with the fixed-effect model; horizontal lines through the squares represent 95% confidence intervals; gray diamonds represent the overall estimate and its confidence interval; dotted vertical lines represent the fixed-effect model; and dashed vertical lines represent the random-effects model.

(0.67MB).

Sub-group analyses were performed to determine possible variables associated with the poorer rates of identification of Gram-positive bacteria (staphylococci, streptococci). For the sub-analyses, the studies were divided into those that used “in-house” methods for pretreatment of the sample and those that used the commercial Sepsityper method (Table 3). The commercial Sepsityper method gave higher percentages of correct identifications of S. aureus (0.92 vs. 0.58). S. pneumoniae, however, showed superior results with in-house methods, although the identification rate was low with both methods (0.51 vs. 0.17). The results for other staphylococci and streptococci were similar, with no differences between the two methods. In the case of enterococci, the results were slightly higher using the Sepsityper kit (0.92 vs. 0.78). There was lower heterogeneity in these sub-group analyses with the commercial Sepsityper method (I2 values ranged from 0 to 89.4%, but was only significant in one species, S. epidermidis) compared with in-house methods (I2 values ranged from 66.9% to 92.8% and significant in all species).

Table 3.

Differences between in-house methods and the Sepsityper method among Gram-positive bacteria.

Microorganism  Intervention  Studies (nStrains (nFixed effectRandom effectHeterogeneity
        Proportion  95%CI  Proportion  95%CI  p  I2 
S. aureus  Sepsityper  16  279  0.92  0.88; 0.96  0.92  0.86; 0.97  0.0605  38% 
  In-house  16  389  0.58  0.52; 0.63  0.74  0.54; 0.90  <0.001  92.8% 
S. epidermidis  Sepsityper  12  432  0.63  0.58; 0.68  0.70  0.53; 0.85  <0.001  89.4% 
  In-house  13  512  0.74  0.70; 0.78  0.72  0.59; 0.84  <0.001  88.7% 
Other CNS  Sepsityper  16  494  0.72  0.68; 0.77  0.76  0.68; 0.83  0.006  53.1% 
  In-house  16  410  0.71  0.65; 0.76  0.80  0.63; 0.93  <0.001  85.2% 
S. pneumoniae  Sepsityper  10  34  0.17  0.02: 0.38  0.17  0.02; 0.38  0.590  0% 
  In-house  13  169  0.38  0.28; 0.48  0.51  0.18; 0.83  <0.001  86.5% 
Other streptococci  Sepsityper  43  0.28  0.13; 0.46  0.24  0.03; 0.53  0.031  54.6% 
  In-house  62  0.01  0.00; 0.08  0.16  0.00; 0.48  <0.001  68% 
E. faecalis  Sepsityper  15  99  0.90  0.81; 0.97  0.92  0.80; 0.99  0.083  35.7% 
  In-house  15  149  0.86  0.78; 0.93  0.79  0.56; 0.96  <0.001  78.7% 
E. faecium  Sepsityper  12  79  0.91  0.81; 0.98  0.91  0.81; 0.99  0.400  4.6% 
  In-house  14  98  0.78  0.66; 0.88  0.78  0.55; 0.96  <0.001  66.9% 
Discussion

This meta-analysis synthesizes the available evidence about the validity of the MALDI-TOF system for the accurate identification of bacteria directly from positive blood cultures. An assessment of the results suggests that MALDI-TOF provides highly accurate identification of Gram-negative bacteria at the species level directly from positive blood cultures. For Gram-positive bacteria, overall accuracy is moderate. Individual studies have previously evaluated the different percentages of correct identifications between Gram-negative and Gram-positive bacteria.12,41 This has several potential clinical implications. On the one hand, because the susceptibility patterns of some Gram-negative bacteria (e.g., P. aeruginosa, A. baumannii, some Enterobacteriaceae) are typically different, accurate identification is important and could therefore allow more rapid optimization of targeted therapy. On the other hand, the results found for Gram-positive are somewhat worrisome; the early management of BSI due to S. aureus and coagulase-negative staphylococci can be quite different, so that misidentification of these bacteria and other Gram-positive may have negative consequences for patient care.

It is not yet understood why it is not always possible to identify Gram-positive bacteria directly from positive blood cultures. It has been speculated that the more robust cell wall decreases protein extraction efficacy and that the slow growth of some species can lead to a very small pellet after the extraction. Several studies have shown that, in terms of efficacy (percentage of correct identifications compared to established reference methods), the Sepsityper performs best with Gram-negative bacteria. Nevertheless, even with Gram-positive bacteria, some researchers report successful identifications in around 75% or more of positive blood cultures.48

Various studies have been published comparing the results obtained with the Sepsityper system and different in-house methods before carrying out the MALDI-TOF analysis. However, while some studies have demonstrated that alternative extraction methods may yield results identical to or better than the Sepsityper kit,6,10,41 others have demonstrated the superiority of the Sepsityper extraction method over in-house methods.16,46 To determine the possible advantage of the Sepsityper kit for the correct identification of Gram-positive bacteria, we analyzed separately those studies that had used in-house methods for pretreatment of the sample and those that had used the commercial Sepsityper method. Although the Sepsityper method showed higher identification rates for some species, the in-house methods generally generated similar results to those obtained with Sepsityper.

Overall, our results showed no major differences in the identification of Gram-positive bacteria when the Sepsityper method or in-house methods were used, although the heterogeneity of results arising from studies that used the Sepsityper method was lower than when in-house methods were used. Nevertheless, the rate of correct identifications of S. aureus was clearly higher with the Sepsityper method, and S. pneumoniae showed superior results with the in-house, as against the Sepsityper method. The results for other Staphylococcus spp. and Streptococcus spp. were similar, with no differences between the two methods. For the identification of Enterococcus spp., the results recorded were slightly higher using the Sepsityper, compared to in-house methods. As with previous reports, our results suggest that the decision to implement one or the other depends on the individual laboratory.

In most cases, the methodological quality of the studies included in this review is moderate. The studies have various limitations as well as methodological issues, with problems of both internal and external validity. Some of these limitations were: the selection of samples included in the different studies; lack of blinding; both tests in the included studies were not realized simultaneously; the intervention differed depending on whether the method used was in-house or commercial; likewise, the subsequent interpretation of the data varied as a result of the different cut-offs used; finally the microbiological reference method in the included studies was performed differently, using different techniques and procedures.

To the best of our knowledge, this is the first meta-analysis of studies of diagnostic tests to assess the accuracy of MALDI-TOF for the rapid, direct identification of bacteria from positive blood cultures. Earlier studies have evaluated the MALDI-TOF system for the identification of clinical bacteria; Drancourt49 published a review of MALDI-TOF for the detection of microorganisms in the blood; Morgenthaler et al.48 reviewed only the standard Sepsityper method for rapid identification of microorganisms, including in-house methods, although the review was not systematic. In 2015, Dixon et al.50 carried out a systematic review comparing MALDI-TOF and conventional identification methods for the rapid identification of pathogens in patients with suspected or known BSI, and also collected information on reduced hospital costs, length of stay, and ‘time to appropriate antimicrobial treatment’. Ling et al.,51 in 2014, published a meta-analysis of the accuracy of MALDI-TOF for the identification of clinical pathogenic fungi, but not bacteria.

This meta-analysis has a number of limitations. First, the number and heterogeneity of the studies posed a variety of research questions. A second limitation was potential publication bias; we excluded unpublished papers, gray literature and industry reports. We tried to avoid this bias as much as possible by using several databases and performing searches without language restrictions. In order to avoid possible bias in the application of selection criteria, these were pre-specified a priori. A third limitation is that some studies conducted the test with two or three different interventions (e.g.: Sepsityper and one or two in-house methods). We assumed that each intervention was a new independent study, so that the total number of evaluated strains was higher. A fourth limitation is that some studies presented score cut-off values higher than 2.0 for the MALDI-TOF detection system. We assumed that all results had a cutoff point of ≥2.0 to try and ensure standardization of the results. Another possible limitation was that, although all studies used an adequate gold standard as comparator, these varied across the studies, so that the results would have varied depending on the quality of the gold standard. The results of this study therefore should be interpreted with caution due to these potential limitations. With respect to the high heterogeneity obtained for overall Gram-negative bacteria in the meta-analysis, this may be related to the inclusion of different species where the identification accuracy of the MALDI-TOF system was heterogeneous, since heterogeneity decreased substantially when the accuracy of each specific species of Gram-negative bacteria was analyzed separately, as previously stated. On the other hand, the main reason for high heterogeneity among Gram-positive bacteria could be the inclusion of the two methods (both in-house or commercial) in the same overall meta-analysis, since it was previously confirmed in our separate analyses that heterogeneity for each Gram-positive species was higher with in-house methods. Another possible limitation of our work has been that from the end of it until its possible publication, new articles of interest have been published in the scientific literature, which could partially modify the results of our systematic review.

Finally, this meta-analysis highlights the need to perform new diagnostic studies to assess the accuracy of MALDI-TOF for Gram-positive bacteria with standardized methods, using the same comparator to improve the methodology. A faster time-to-result is an advantage for the patient, since, in combination with antibiotic stewardship programs, optimized antibiotic therapy can very often be administered on the basis of a species/genus identification of the underlying microorganisms. Looking to the future, we may suppose that this technology can be used in the detection of bacterial resistance determinants and virulence factors.

In summary, MALDI-TOF MS demonstrated high accuracy for the direct identification of Gram-negative bacteria from blood culture and moderate accuracy for Gram-positive bacteria (low in some species). There were no major differences with respect to the identification of Gram-positive bacteria when either in-house methods or the standard Sepsityper kit were used.

Conflict of interests

All authors confirm that there is no conflict of interest for the publication of this work.

Appendix A
Supplementary data

The following are the supplementary data to this article:

References
[1]
N.K.J. Adhikari, R.A. Fowler, S. Bhagwanjee, G.D. Rubenfeld.
Critical care and the global burden of critical illness in adults.
Lancet, 376 (2010), pp. 1339-1346
[2]
R. Banejee, V. Ozenci, R. Patel.
Individualized approach are needed for optimized blood cultures.
Clin Infect Dis, 63 (2016), pp. 1332-1339
[3]
F.P. Maurer, M. Christner, M. Hentschke, H. Rohde.
Advances in rapid identification and susceptibility testing of bacteria in the clinical microbiology laboratory: implications for patient care and antimicrobial stewardship programs.
Infect Dis Rep, 9 (2017), pp. 6839
[4]
E. Carbonnelle, P. Grohs, H. Jacquier, N. Day, S. Tenza, A. Dewailly.
Robustness of two MALDI-TOF mass spectrometry systems for bacterial identification.
J Microbiol Methods, 89 (2012), pp. 133-136
[5]
M. Christner, H. Rohde, M. Wolters, I. Sobottka, K. Wegscheider, M. Aepfelbacher.
Rapid identification of bacteria from positive blood culture bottles by use of matrix-assisted laser desorption-ionization time of flight mass spectrometry fingerprinting.
J Clin Microbiol, 48 (2010), pp. 1584-1591
[6]
C. Meex, F. Neuville, J. Descy, P. Huynen, M.P. Hayette, P. De Mol, et al.
Direct identification of bacteria from BacT/ALERT anaerobic positive blood cultures by MALDI-TOF MS: MALDI Sepsityper kit versus an in-house saponin method for bacterial extraction.
J Med Microbiol, 61 (2012), pp. 1511-1516
[7]
B. Hazelton, L.C. Thomas, T. Olma, J. Kok, M. O'Sullivan, S.C. Chen, et al.
Rapid and accurate direct antibiotic susceptibility testing of blood culture broths using MALDI Sepsityper combined with the BD Phoenix automated system.
J Med Microbiol, 63 (2014), pp. 1590-1594
[8]
K. Riederer, K. Cruz, S. Shemes, S. Szpunar, J.T. Fishbain.
MALDI-TOF identification of Gram-negative bacteria directly from blood culture bottles containing charcoal: Sepsityper® kits versus centrifugation – filtration method.
Diagn Microbiol Infect Dis, 82 (2015), pp. 105-108
[9]
S. Emonet, H.N. Shah, A. Cherkaoui, J. Schrenzel.
Application and use of various mass spectrometry methods in clinical microbiology.
Clin Microbiol Infect, 16 (2010), pp. 1604-1613
[10]
L. Ferreira, F. Sanchez-Juanes, M. Gonzalez-Avila, D. Cembrero-Fuciños, A. Herrero-Hernández, J.M. González-Buitrago, et al.
Direct identification of urinary tract pathogens from urine samples by matrix-assisted laser desorption ionization-time of flight mass spectrometry.
J Clin Microbiol, 48 (2010), pp. 2110-2115
[11]
P. Gaibani, G. Rossini, S. Ambretti, F. Gelsomino, A.M. Pierro, S. Varani, et al.
Blood culture systems: rapid detection-how and why?.
Int J Antimicrob Agents, 34 (2009), pp. S13-S15
[12]
L. Ferreira, F. Sanchez-Juanes, I. Porras-Guerra, M.I. Garcia-Garcia, J.E. Garcia-Sanchez, J.M. Gonzalez-Buitrago, et al.
Microorganisms direct identification from blood culture by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry.
Clin Microbiol Infect, 17 (2011), pp. 546-551
[13]
G. Prod¿hom, A. Bizzini, C. Durussel, J. Bille, G. Greub.
Matrix-assisted laser desorption ionization-time of flight mass spectrometry for direct bacterial identification from positive blood culture pellets.
J Clin Microbiol, 48 (2010), pp. 1481-1483
[14]
W. Moussaoui, B. Jaulhac, A.M. Hoffmann, B. Ludes, M. Kostrzewa, P. Riegel, et al.
Matrix-assisted laser desorption ionization time-of-flight mass spectrometry identifies 90% of bacteria directly from blood culture vials.
Clin Microbiol Infect, 16 (2010), pp. 1631-1638
[15]
L.G. Stevenson, S.K. Drake, P.R. Murray.
Rapid identification of bacteria in positive blood culture broths by matrix-assisted laser desorption ionization-time of flight mass spectrometry.
J Clin Microbiol, 48 (2010), pp. 444-447
[16]
P.M. Juiz, M. Almela, C. Melción, I. Campo, C. Esteban, C. Pitart, et al.
A comparative study of two different methods of sample preparation for positive blood cultures for the rapid identification of bacteria using MALDI-TOF MS.
Eur J Clin Microbiol Infect Dis, 31 (2012), pp. 1353-1358
[17]
B. La Scola, D. Raoult.
Direct identification of bacteria in positive blood culture bottles by matrix-assisted laser desorption ionisation time-of-flight mass spectrometry.
[18]
D. Moher, A. Liberati, J. Tetzlaff, D.G. Altman, PRISMA Group.
Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement.
Br Med J, 339 (2009), pp. b2535
[19]
P. Whiting, A.W.S. Rutjes, J.B. Reitsma, P.M.M. Bossuyt, J. Kleijnen.
The development of QUADAS: a tool for the quality assessment of studies of diagnostic accuracy included in systematic reviews.
BMC Med Res Methodol, 3 (2003), pp. 1-13
[20]
A. Angeletti, G. Dicuonzo, A. d’Agostino, A. Avola, F. Crea, C. Palazzo, et al.
Turnaround time of positive blood cultures after the introduction of matrix-assisted laser desorption-ionization time-of-flight mass spectrometry.
New Microbiol, 38 (2015), pp. 379-386
[21]
M. Egli, D. Osthoff, J. Goldenberger, S. Halter, J. Schaub, M. Steiger.
Matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF) directly from positive blood culture flasks allows rapid identification of bloodstream infections in immunosuppressed hosts.
Transpl Infect Dis, 17 (2015), pp. 481-487
[22]
V. Chebotar, O.A. Ponomarenko, A.V. Lazareva, O.V.T. Karaseva, V. Tepaev.
MALDI-TOF technique availability for identification of septic agents in pediatric practice.
CTM, 7 (2015), pp. 68-72
[23]
R.M. Martinez, E.R. Bauerle, F.C. Fang, S.M. Butler-Wu.
Evaluation of three rapid diagnostic methods for direct identification of microorganisms in positive blood cultures.
J Clin Microbiol, 7 (2014), pp. 2521-2529
[24]
K.M. Schieffer, K.E. Tan, P.D. Stamper, A. Somogyi, S.B. Andrea, S. Wakefield, et al.
Multicenter evaluation of the Sepsityper extraction kit and MALDI-TOF MS for direct identification of positive blood culture isolates using the BD BACTEC FX and VersaTREK diagnostic blood culture systems.
J Appl Microbiol, 116 (2014), pp. 934-941
[25]
J.D. Haigh, I.M. Grenn, D. Ball, M. Eydmann, M. Millar, M. Wilks.
Rapid identification of bacteria from BioMerieux BacT/Alert blood culture bottles by MALDI-TOF.
Br J Biomed Sci, 70 (2013), pp. 149-155
[26]
B. Rodríguez-Sánchez, C. Sánchez-Carrillo, A. Ruiz, M. Marín, E. Cercenado, Rodríguez-Creixems, et al.
Direct identification of pathogens from positive blood cultures using matrix-assisted laser desorption-ionization time-of-flight mass spectrometry.
Clin Microbiol Infect, 20 (2014), pp. O421-O427
[27]
J. Mestas, S. Felsenstein, J.D. Bard.
Direct identification of bacteria from positive BacT/ALERT blood culture bottles using matrix-assisted laser desorption ionization-time of flight mass spectrometry.
Diagn Microbiol Infect Dis, 80 (2014), pp. 193-196
[28]
M. Tadros, A. Petrich.
Evaluation of MALDI-TOF mass spectrometry and Sepsityper Kit for the direct identification of organisms from sterile body fluids in a Canadian pediatric hospital.
Can J Infect Dis Med Microbiol, 24 (2013), pp. 191-194
[29]
T.J. Gray, L. Thomas, T. Olma, J.R. Iredell, S.C. Chen.
Rapid identification of Gram-negative organisms from blood culture bottles using a modified extraction method and MALDI-TOF mass spectrometry.
Diagn Microbiol Infect Dis, 77 (2013), pp. 110-112
[30]
D. Martiny, F. Debaugnies, D. Gateff, M. Gérard, M. Aoun, C. Martin.
Impact of rapid microbial identification directly from positive blood cultures using matrix-assisted laser desorption/ionization time-of-flight mass spectrometry on patient management.
Clin Microbiol Infect, 19 (2013), pp. E568-E581
[31]
W. Jamal, R. Saleem, V.O. Rotimi.
Rapid identification of pathogens directly from blood culture bottles by Bruker matrix-assisted laser desorption laser ionization-time of flight mass spectrometry versus routine methods.
Diagn Microbiol Infect Dis, 76 (2013), pp. 404-408
[32]
C. Leli, E. Cenci, A. Cardaccia, A. Moretti, F. D’Alò, R. Pagliochini.
Rapid identification of bacterial and fungal pathogens from positive blood cultures by MALDI-TOF MS.
Int J Med Microbiol, 303 (2013), pp. 205-209
[33]
J.H.K. Chen, P.L. Ho, G.S.W. Kwan, K.K.K. She, G.K.H. Siu, V.C.C. Cheng.
Direct bacterial identification in positive blood cultures by use of two commercial matrix-assisted laser desorption ionization–time of flight mass spectrometry systems.
J Clin Microbiol, 6 (2013), pp. 1733-1739
[34]
B. Nonnemann, M. Tvede, T. Bjarnsholt.
Identification of pathogenic microorganisms directly from positive blood vials by matrix-assisted laser desorption/ionization time of flight mass spectrometry.
APMIS, 121 (2013), pp. 871-877
[35]
G.A. March-Rosselló, M.F. Muñoz-Moreno, M.C. García-Loygorri-Jordán de Urriés, M.A. Bratos-Pérez.
A differential centrifugation protocol and validation criterion for enhancing mass spectrometry (MALDI-TOF) results in microbial identification using blood culture growth bottles.
Eur J Clin Microbiol Infect Dis, 32 (2013), pp. 699-704
[36]
O. Clerc, G. Prod’hom, C. Vogne, A. Bizzini, T. Calandra, G. Greub.
Impact of matrix-assisted laser desorption ionization time-of-flight mass spectrometry on the clinical management of patients with Gram-negative bacteremia: a prospective observational study.
Clin Infect Dis, 56 (2013), pp. 1101-1107
[37]
Y. Hoyos-Mallecot, C. Miranda-Casas, J.J. Cabrera-Alvargonzalez, C. Gómez-Camarasa, M.D. Pérez-Ramirez, J.M. Navarro-Marí.
Bacterial identification from blood cultures by a rapid Matrix-Assisted Laser Desorption-Ionisation Time-of-Flight mass spectrometry technique.
Enferm Infecc Microbiol Clin, 31 (2013), pp. 152-155
[38]
P.R.S. Lagacé-Wiens, H.J. Adam, J.A. Karlowsky, K.A. Nichol, P.F. Pang, J. Guenther.
Identification of blood culture isolates directly from positive blood cultures by use of matrix-assisted laser desorption ionization–time of flight mass spectrometry and a commercial extraction system: analysis of performance, cost, and turnaround time.
J Clin Microbiol, 10 (2012), pp. 3324-3328
[39]
R.T. Saffert, S.A. Cunningham, J. Mandrekar, R. Patel.
Comparison of three preparatory methods for detection of bacteremia by MALDI-TOF mass spectrometry.
Diagn Microbiol Infect Dis, 73 (2012), pp. 21-26
[40]
A.L.M. Vlek, M.J.M. Bonten, C.H.E. Boel.
Direct matrix-assisted laser desorption Ionization time-of-flight mass spectrometry improves appropriateness of antibiotic treatment of bacteremia.
[41]
D. Martiny, A. Dediste, O. Vandenberg.
Comparison of an in-house method and the commercial SepsityperTM kit for bacterial identification directly from positive blood culture broths by matrix-assisted laser desorption-ionisation time-of-flight mass spectrometry.
Eur J Clin Microbiol Infect Dis, 31 (2012), pp. 2269-2281
[42]
B.W. Buchan, K.M. Riebe, N.A. Ledeboer.
Comparison of the MALDI Biotyper system using Sepsityper specimen processing to routine microbiological methods for identification of bacteria from positive blood culture bottles.
J Clin Microbiol, 2 (2012), pp. 346-352
[43]
A.J.M. Loonen, A.R. Jansz, J. Stalpers, P.F.G. Wolffs, A.J.C. van den Brule.
An evaluation of three processing methods and the effect of reduced culture times for faster direct identification of pathogens from BacT/ALERT blood cultures by MALDI-TOF MS.
Eur J Clin Microbiol Infect Dis, 31 (2012), pp. 1575-1583
[44]
S. Klein, S. Zimmermann, C. Kohler, A. Mischnik, W. Alle, K.A. Bode.
Integration of matrix-assisted laser desorption/ionization time-of-flight mass spectrometry in blood culture diagnostics: a fast and effective approach.
J Med Microbiol, 61 (2012), pp. 323-331
[45]
D. Fuglsang-Damgaard, C.H. Nielsen, E. Mandrup, K. Fuursted.
The use of Gram stain and matrix-assisted laser desorption ionization time-of-flight mass spectrometry on positive blood culture: synergy between new and old technology.
[46]
S. Schubert, K. Weinert, C. Wagner, B. Gunzl, A. Wieser, T. Maier, et al.
Novel, improved sample preparation for rapid, direct identification from positive blood cultures using matrix-assisted laser desorption/ionization time-of-flight (MALDI-TOF) mass spectrometry.
J Mol Diagn, 13 (2011), pp. 701-706
[47]
J. Kok, L.C. Thomas, T. Olma, S.C.A. Chen, J.R. Iredell.
Identification of bacteria in blood culture broths using matrix-assisted laser desorption- ionization SepsityperTM and Time of Flight Mass Spectrometry.
[48]
N.G. Morgenthaler, M. Kostrzewa.
Rapid identification of pathogens in positive blood culture of patients with sepsis: review and meta-analysis of the performance of the sepsityper kit.
Int J Microbiol, 2015 (2015), pp. 827416
[49]
M. Drancourt.
Detection of microorganisms in blood specimens using matrix-assisted laser desorption ionization time-of-flight mass spectrometry: a review.
Clin Microbiol Infect, 16 (2010), pp. 1620-1625
[50]
P. Dixon, P. Davies, W. Hollingworth, M. Stoddart, A. MacGowan.
A systematic review of matrix-assisted laser desorption/ionisation time-of-flight mass spectrometry compared to routine microbiological methods for the time taken to identify microbial organisms from positive blood cultures.
Eur J Clin Microbiol Infect Dis, 34 (2015), pp. 863-876
[51]
H. Ling, Z. Yuan, J. Shen, Z. Wang, Y. Xu.
Accuracy of matrix-assisted laser desorption ionization-time of flight mass spectrometry for identification of clinical pathogenic fungi: a meta-analysis.
J Clin Microbiol, 52 (2014), pp. 2573-2582
Copyright © 2017. Elsevier España, S.L.U. and Sociedad Española de Enfermedades Infecciosas y Microbiología Clínica