Original breve
Disponible online el 22 de abril de 2024
Dispositivos oscilométricos frente a Doppler arterial en la medida del índice tobillo-brazo para el diagnóstico de la enfermedad arterial periférica
Oscillometric devices vs. arterial Doppler in measuring the ankle-arm index for the diagnosis of peripheral arterial disease
Irene Martín-Rubioa,
Autor para correspondencia
irenemartinrubio@yahoo.es

Autor para correspondencia.
, Beatriz Marín-Garcíaa, Elisa Romero-Velascoa, Israel John Thuissard-Vasalloc, Azucena Bautista-Hernándezb, Daniel Abad-Péreza,c
a Servicio de Medicina Interna, Hospital Universitario de Getafe, Getafe, Madrid, España
b Servicio de Medicina Interna, Hospital Universitario de la Princesa, Madrid. España
c Universidad Europea de Madrid, Madrid, España
Información del artículo
Resumen
Texto completo
Bibliografía
Descargar PDF
Estadísticas
Figuras (1)
Tablas (2)
Tabla 1. Características basales de los pacientes incluídos N=187
Tabla 2. Clasificación de los pacientes con cada aparato de medida
Mostrar másMostrar menos
Resumen
Introducción

La enfermedad arterial periférica constituye un marcador de daño vascular que se diagnostica mediante la medida del índice tobillo-brazo (ITB). El objetivo de este estudio es determinar la validez y la concordancia de los dispositivos oscilométricos MESI ABPI-MD y Microlife WatchBP® office-ABI respecto a la prueba de referencia, el Doppler arterial.

Materiales y métodos

Estudio observacional, transversal y descriptivo con pacientes ingresados a quienes se les realizó la medida de ITB con los 3 dispositivos. Se consideran valores normales entre 1-1,4; indeterminados entre 0,91-0,99 y patológicos ≤0,9 y >1,4.

Resultados

Se incluyeron 187 pacientes (54,4% varones) con edad media de 66 años. Los resultados del Doppler fueron inferiores a los de los oscilométricos (mediana [RIC] 1,1 [0,2] vs. 1,2 [0,2], p<0,05), sin diferencias significativas entre los automáticos (p=0,29 para el miembro inferior derecho y p=0,342 para el izquierdo). Ambos dispositivos presentaron elevada especificidad (96,5-99,2%) y baja sensibilidad (29,5-45,4%). La correlación de los resultados fue buena-moderada para MESI y moderada para Microlife. La concordancia entre ambos fue aceptable-moderada.

Conclusión

Los dispositivos oscilométricos automáticos podrían ser de utilidad en pacientes asintomáticos como alternativa al Doppler arterial.

Palabras clave:
Enfermedad arterial periférica
Índice tobillo-brazo
Dispositivos oscilométricos automáticos
Abstract
Introduction

Peripheral arterial disease is a marker of vascular damage that is diagnosed by measuring the ankle-brachial index. The aim of this study was to determine the validity and agreement of the MESI ABPI-MD and Microlife WatchBP® office-ABI oscillometric devices with respect to the gold standard arterial Doppler.

Materials and methods

Observational, cross-sectional, descriptive study of inpatients who underwent ABI measurement with the three devices. Values are considered normal between 1-1.4, indeterminate between 0.91-0.99 and pathological ≤0.9 and >1.4.

Results

A total of 187 patients (54.4% male) with a mean age of 66 years were included. The Doppler results were inferior to those of the oscillometric devices (median [IQR] 1.1 [0.2] vs. 1.2 [0.2], P<.05), with no significant differences between the automated devices (P=.29 for the right lower limb and P=.342 for the left lower limb). Both devices had high specificity (96.5-99.2%) and low sensitivity (29.5-45.4%). The correlation of the results was good-moderate for MESI and moderate for Microlife. The agreement between the two was acceptable-moderate.

Conclusion

Automated oscillometric devices could be useful in asymptomatic patients as an alternative to arterial Doppler.

Keywords:
Peripheral arterial disease
Ankle-brachial index
Automated oscillometric devices

Artículo

Opciones para acceder a los textos completos de la publicación Medicina Clínica
Suscriptor
Suscriptor de la revista

Si ya tiene sus datos de acceso, clique aquí.

Si olvidó su clave de acceso puede recuperarla clicando aquí y seleccionando la opción "He olvidado mi contraseña".
Suscribirse
Suscribirse a:

Medicina Clínica

Comprar
Comprar acceso al artículo

Comprando el artículo el PDF del mismo podrá ser descargado

Precio 19,34 €

Comprar ahora
Contactar
Teléfono para suscripciones e incidencias
De lunes a viernes de 9h a 18h (GMT+1) excepto los meses de julio y agosto que será de 9 a 15h
Llamadas desde España
932 415 960
Llamadas desde fuera de España
+34 932 415 960
E-mail