array:23 [ "pii" => "S0048712020301328" "issn" => "00487120" "doi" => "10.1016/j.rh.2020.11.003" "estado" => "S300" "fechaPublicacion" => "2021-07-01" "aid" => "662" "copyright" => "Sociedad Española de Rehabilitación y Medicina Física" "copyrightAnyo" => "2020" "documento" => "article" "crossmark" => 1 "subdocumento" => "fla" "cita" => "Rehabilitacion. 2021;55:183-9" "abierto" => array:3 [ "ES" => false "ES2" => false "LATM" => false ] "gratuito" => false "lecturas" => array:1 [ "total" => 0 ] "itemSiguiente" => array:18 [ "pii" => "S0048712021000323" "issn" => "00487120" "doi" => "10.1016/j.rh.2021.03.001" "estado" => "S300" "fechaPublicacion" => "2021-07-01" "aid" => "680" "copyright" => "Sociedad Española de Rehabilitación y Medicina Física" "documento" => "article" "crossmark" => 1 "subdocumento" => "fla" "cita" => "Rehabilitacion. 2021;55:190-8" "abierto" => array:3 [ "ES" => false "ES2" => false "LATM" => false ] "gratuito" => false "lecturas" => array:1 [ "total" => 0 ] "es" => array:13 [ "idiomaDefecto" => true "cabecera" => "<span class="elsevierStyleTextfn">ORIGINAL</span>" "titulo" => "Cuidadores principales informales no remunerados: ¿quiénes cuidan a niños, adultos y adultos mayores en situación de discapacidad?" "tienePdf" => "es" "tieneTextoCompleto" => "es" "tieneResumen" => array:2 [ 0 => "es" 1 => "en" ] "paginas" => array:1 [ 0 => array:2 [ "paginaInicial" => "190" "paginaFinal" => "198" ] ] "titulosAlternativos" => array:1 [ "en" => array:1 [ "titulo" => "Informal unpaid primary caregivers: Who takes care of disabled children, adults and seniors?" ] ] "contieneResumen" => array:2 [ "es" => true "en" => true ] "contieneTextoCompleto" => array:1 [ "es" => true ] "contienePdf" => array:1 [ "es" => true ] "resumenGrafico" => array:2 [ "original" => 0 "multimedia" => array:7 [ "identificador" => "fig0010" "etiqueta" => "Figura 2" "tipo" => "MULTIMEDIAFIGURA" "mostrarFloat" => true "mostrarDisplay" => false "figura" => array:1 [ 0 => array:4 [ "imagen" => "gr2.jpeg" "Alto" => 882 "Ancho" => 1508 "Tamanyo" => 72768 ] ] "descripcion" => array:1 [ "es" => "<p id="spar0050" class="elsevierStyleSimplePara elsevierViewall">Gráfico que señala los años de escolaridad de los cuidadores de niños, adultos y adultos mayores.</p>" ] ] ] "autores" => array:1 [ 0 => array:2 [ "autoresLista" => "A. Yelincic, R.A. Cárcamo" "autores" => array:2 [ 0 => array:2 [ "nombre" => "A." "apellidos" => "Yelincic" ] 1 => array:2 [ "nombre" => "R.A." "apellidos" => "Cárcamo" ] ] ] ] ] "idiomaDefecto" => "es" "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S0048712021000323?idApp=UINPBA00004N" "url" => "/00487120/0000005500000003/v2_202202190654/S0048712021000323/v2_202202190654/es/main.assets" ] "itemAnterior" => array:18 [ "pii" => "S0048712020300785" "issn" => "00487120" "doi" => "10.1016/j.rh.2020.06.001" "estado" => "S300" "fechaPublicacion" => "2021-07-01" "aid" => "627" "copyright" => "Sociedad Española de Rehabilitación y Medicina Física" "documento" => "article" "crossmark" => 1 "subdocumento" => "fla" "cita" => "Rehabilitacion. 2021;55:175-82" "abierto" => array:3 [ "ES" => false "ES2" => false "LATM" => false ] "gratuito" => false "lecturas" => array:1 [ "total" => 0 ] "es" => array:13 [ "idiomaDefecto" => true "cabecera" => "<span class="elsevierStyleTextfn">Original</span>" "titulo" => "Eficacia de la magnetoterapia en artrosis erosiva de manos. Un ensayo clínico" "tienePdf" => "es" "tieneTextoCompleto" => "es" "tieneResumen" => array:2 [ 0 => "es" 1 => "en" ] "paginas" => array:1 [ 0 => array:2 [ "paginaInicial" => "175" "paginaFinal" => "182" ] ] "titulosAlternativos" => array:1 [ "en" => array:1 [ "titulo" => "Efficacy of magnetotherapy in hand erosive osteoarthritis. A clinical trial" ] ] "contieneResumen" => array:2 [ "es" => true "en" => true ] "contieneTextoCompleto" => array:1 [ "es" => true ] "contienePdf" => array:1 [ "es" => true ] "resumenGrafico" => array:2 [ "original" => 0 "multimedia" => array:7 [ "identificador" => "fig0005" "etiqueta" => "Figura 1" "tipo" => "MULTIMEDIAFIGURA" "mostrarFloat" => true "mostrarDisplay" => false "figura" => array:1 [ 0 => array:4 [ "imagen" => "gr1.jpeg" "Alto" => 1436 "Ancho" => 1672 "Tamanyo" => 186309 ] ] "descripcion" => array:1 [ "es" => "<p id="spar0050" class="elsevierStyleSimplePara elsevierViewall">Selección, aleatorización y seguimiento de los pacientes en el estudio según el diagrama de flujo CONSORT.</p>" ] ] ] "autores" => array:1 [ 0 => array:2 [ "autoresLista" => "A. Boada-Pladellorens, À. Jaén Manzanera, E. Bertran Garcia, À. Abril Carreras, G. Salvador Alarcón, S. Martínez Pardo" "autores" => array:6 [ 0 => array:2 [ "nombre" => "A." "apellidos" => "Boada-Pladellorens" ] 1 => array:2 [ "nombre" => "À." "apellidos" => "Jaén Manzanera" ] 2 => array:2 [ "nombre" => "E." "apellidos" => "Bertran Garcia" ] 3 => array:2 [ "nombre" => "À." "apellidos" => "Abril Carreras" ] 4 => array:2 [ "nombre" => "G." "apellidos" => "Salvador Alarcón" ] 5 => array:2 [ "nombre" => "S." "apellidos" => "Martínez Pardo" ] ] ] ] ] "idiomaDefecto" => "es" "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S0048712020300785?idApp=UINPBA00004N" "url" => "/00487120/0000005500000003/v2_202202190654/S0048712020300785/v2_202202190654/es/main.assets" ] "es" => array:19 [ "idiomaDefecto" => true "cabecera" => "<span class="elsevierStyleTextfn">Original</span>" "titulo" => "Confiabilidad test-retest de goniometría estándar y G-pro smartphone en el movimiento flexo-extensión del hombro" "tieneTextoCompleto" => true "paginas" => array:1 [ 0 => array:2 [ "paginaInicial" => "183" "paginaFinal" => "189" ] ] "autores" => array:1 [ 0 => array:4 [ "autoresLista" => "A.J. Yarin Achachagua, P. Saravia Saucedo, J. Coveñas Lalupú, D. Esenarro Vargas, V.I. Tafur Anzualdo" "autores" => array:5 [ 0 => array:4 [ "nombre" => "A.J." "apellidos" => "Yarin Achachagua" "email" => array:1 [ 0 => "anwar.yarin@unmsm.edu.pe" ] "referencia" => array:2 [ 0 => array:2 [ "etiqueta" => "<span class="elsevierStyleSup">a</span>" "identificador" => "aff0005" ] 1 => array:2 [ "etiqueta" => "<span class="elsevierStyleSup">*</span>" "identificador" => "cor0005" ] ] ] 1 => array:3 [ "nombre" => "P." "apellidos" => "Saravia Saucedo" "referencia" => array:1 [ 0 => array:2 [ "etiqueta" => "<span class="elsevierStyleSup">b</span>" "identificador" => "aff0010" ] ] ] 2 => array:3 [ "nombre" => "J." "apellidos" => "Coveñas Lalupú" "referencia" => array:1 [ 0 => array:2 [ "etiqueta" => "<span class="elsevierStyleSup">c</span>" "identificador" => "aff0015" ] ] ] 3 => array:3 [ "nombre" => "D." "apellidos" => "Esenarro Vargas" "referencia" => array:1 [ 0 => array:2 [ "etiqueta" => "<span class="elsevierStyleSup">c</span>" "identificador" => "aff0015" ] ] ] 4 => array:3 [ "nombre" => "V.I." "apellidos" => "Tafur Anzualdo" "referencia" => array:1 [ 0 => array:2 [ "etiqueta" => "<span class="elsevierStyleSup">c</span>" "identificador" => "aff0015" ] ] ] ] "afiliaciones" => array:3 [ 0 => array:3 [ "entidad" => "Universidad Nacional Mayor de San Marcos-UNMSM, Lima, Perú" "etiqueta" => "a" "identificador" => "aff0005" ] 1 => array:3 [ "entidad" => "Universidad Privada Norbert Wiener, Lima, Perú" "etiqueta" => "b" "identificador" => "aff0010" ] 2 => array:3 [ "entidad" => "Universidad Nacional Federico Villarreal, Lima, Perú" "etiqueta" => "c" "identificador" => "aff0015" ] ] "correspondencia" => array:1 [ 0 => array:3 [ "identificador" => "cor0005" "etiqueta" => "⁎" "correspondencia" => "Autor para correspondencia." ] ] ] ] "titulosAlternativos" => array:1 [ "en" => array:1 [ "titulo" => "Test-retest reliability of standard goniometry and the G-pro smartphone in shoulder flexion-extension" ] ] "textoCompleto" => "<span class="elsevierStyleSections"><span id="sec0005" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0065">Introducción</span><p id="par0005" class="elsevierStylePara elsevierViewall">La rehabilitación tiene el propósito de mejorar las habilidades de personas que así lo requieran<a class="elsevierStyleCrossRef" href="#bib0110"><span class="elsevierStyleSup">1</span></a>. El presente estudio se encuentra enfocado en la rehabilitación de forma activa, el cual se refiere a que los movimientos son producidos por el mismo organismo mediante su aparato muscular durante los ejercicios de efecto palanca, con los cuales el paciente gradualmente realiza movimientos del miembro superior de una manera completa y correcta. La medición de dichos movimientos durante la fisioterapia facilitaría el proceso de rehabilitación, permitiendo la retroalimentación en-línea del movimiento realizado y una evaluación cuantitativa por parte del especialista<a class="elsevierStyleCrossRefs" href="#bib0115"><span class="elsevierStyleSup">2,3</span></a>. La evaluación precisa del rango articular conforma una tarea esencial en la labor diaria de los fisioterapeutas para poder valorar correctamente la progresión del tratamiento. El goniómetro es un elemento económico, transportable y fácil de usar, constituye el instrumento de medida de referencia en la evaluación de los ángulos articulares, situándose su validez y fiabilidad en la literatura científica como muy buen indicador (entre 0,89-0,98) en función del movimiento considerado<a class="elsevierStyleCrossRef" href="#bib0125"><span class="elsevierStyleSup">4</span></a> El smartphone es actualmente una herramienta que nos permite comunicarnos y ofrece muchas posibilidades gracias a la gran variedad de aplicaciones, software y funcionalidades con las que se le ha dotado en los últimos años. Incluso el Instituto de Informática Sanitaria ha centrado su atención en su uso al analizar más de 43.000 aplicaciones relacionadas con la salud disponibles en 2013<a class="elsevierStyleCrossRef" href="#bib0130"><span class="elsevierStyleSup">5</span></a> en el repositorio de aplicaciones de Google. Es por ello que en esta investigación nos planteamos analizar la fiabilidad y la seguridad de una aplicación G-pro smart en el sistema Android para evaluar sus características y observar su utilidad como goniómetro en la clínica para medir los ángulos articulares<a class="elsevierStyleCrossRefs" href="#bib0135"><span class="elsevierStyleSup">6,7</span></a>.</p></span><span id="sec0010" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0070">Materiales y métodos</span><p id="par0010" class="elsevierStylePara elsevierViewall">Se empleó el goniómetro estándar y un G-pro smartphone con la aplicación para medir de manera pasiva la flexo-extensión de hombro, valorando la movilidad en ambos brazos.</p><p id="par0015" class="elsevierStylePara elsevierViewall">Las mediciones se realizaron en días consecutivos a los mismos sujetos y en distinto orden, teniendo como premisa que el movimiento se realizara sin dolor<a class="elsevierStyleCrossRef" href="#bib0145"><span class="elsevierStyleSup">8</span></a>. Es importante mencionar que, previamente se pidió el consentimiento informado a todos los participantes y se obtuvo la autorización del comité ético del Instituto Superior de Salud.</p><p id="par0020" class="elsevierStylePara elsevierViewall">La población se compuso de estudiantes del segundo ciclo de un Instituto Superior de Salud en Perú, capaces de realizar los movimientos indicados. El muestreo fue intencional y la muestra estuvo compuesta por 30 usuarios, sexo masculino y femenino, entre los 17 y 25 años. Se evaluaron los movimientos de flexión y extensión de hombro según las directrices/recomendaciones marcadas en la literatura científica<a class="elsevierStyleCrossRef" href="#bib0150"><span class="elsevierStyleSup">9</span></a>, siempre respetando el umbral doloroso. El protocolo ejecutado se muestra a continuación:<ul class="elsevierStyleList" id="lis0005"><li class="elsevierStyleListItem" id="lsti0005"><span class="elsevierStyleLabel">•</span><p id="par0025" class="elsevierStylePara elsevierViewall">Necesidad de una segunda persona para estabilización articular, posición y restricción del rango de movimiento<a class="elsevierStyleCrossRef" href="#bib0155"><span class="elsevierStyleSup">10</span></a>.</p></li><li class="elsevierStyleListItem" id="lsti0010"><span class="elsevierStyleLabel">•</span><p id="par0030" class="elsevierStylePara elsevierViewall">Necesidad de un goniómetro estándar y G-pro smart acorde para examinar.</p></li><li class="elsevierStyleListItem" id="lsti0015"><span class="elsevierStyleLabel">•</span><p id="par0035" class="elsevierStylePara elsevierViewall">La medición tiene que ser pasiva y bilateral.</p></li><li class="elsevierStyleListItem" id="lsti0020"><span class="elsevierStyleLabel">•</span><p id="par0040" class="elsevierStylePara elsevierViewall">Se presentan variaciones angulares en distintas razas debido a diferencias en tejido blando y/o a una estructura ósea diferente. La utilización de la extremidad contralateral, si es posible, es la mejor manera de comparar.</p></li><li class="elsevierStyleListItem" id="lsti0025"><span class="elsevierStyleLabel">•</span><p id="par0045" class="elsevierStylePara elsevierViewall">Ciertas patologías o cirugías limitan el rango de movimiento. Por ejemplo, artrosis de codo, de rodilla, inflamación posquirúrgica de ligamentos cruzados, etc.</p></li><li class="elsevierStyleListItem" id="lsti0030"><span class="elsevierStyleLabel">•</span><p id="par0050" class="elsevierStylePara elsevierViewall">La sensación de dolor durante la manipulación articular puede limitar el rango de movimiento; esta limitación puede no estar presente bajo sedación flujo<a class="elsevierStyleCrossRefs" href="#bib0160"><span class="elsevierStyleSup">11,12</span></a><span class="elsevierStyleSup">.</span></p></li></ul></p><span id="sec0015" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0075">Goniometría en miembro superior</span><p id="par0055" class="elsevierStylePara elsevierViewall">Se verificó si el evaluado presenta movimientos de abducción-aducción, flexión-extensión y rotación interna-externa<a class="elsevierStyleCrossRef" href="#bib0170"><span class="elsevierStyleSup">13</span></a>.</p></span><span id="sec0020" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0080">Goniometría en la articulación del hombro en la flexión</span><p id="par0060" class="elsevierStylePara elsevierViewall"><ul class="elsevierStyleList" id="lis0010"><li class="elsevierStyleListItem" id="lsti0035"><span class="elsevierStyleLabel">•</span><p id="par0065" class="elsevierStylePara elsevierViewall">Posición: Paciente en decúbito dorsal; escápula estabilizada contra la camilla; hombro en posición 0; codo en posición 0; antebrazo en pronosupinación 0; muñeca en posición 0.</p></li><li class="elsevierStyleListItem" id="lsti0040"><span class="elsevierStyleLabel">•</span><p id="par0070" class="elsevierStylePara elsevierViewall">Alineación del goniómetro: Goniómetro universal en 0°.</p></li><li class="elsevierStyleListItem" id="lsti0045"><span class="elsevierStyleLabel">•</span><p id="par0075" class="elsevierStylePara elsevierViewall">Eje: Colocado sobre el acromion que corresponde a la proyección del punto central de la cabeza humeral.</p></li><li class="elsevierStyleListItem" id="lsti0050"><span class="elsevierStyleLabel">•</span><p id="par0080" class="elsevierStylePara elsevierViewall">Brazo fijo: Alineado con la línea medio axilar.</p></li><li class="elsevierStyleListItem" id="lsti0055"><span class="elsevierStyleLabel">•</span><p id="par0085" class="elsevierStylePara elsevierViewall">Brazo móvil: Alineado con la línea media longitudinal del húmero tomando como reparo óseo el epicóndilo y superpuesto sobre el brazo.</p></li><li class="elsevierStyleListItem" id="lsti0060"><span class="elsevierStyleLabel">•</span><p id="par0090" class="elsevierStylePara elsevierViewall">Movimiento: Se practica la flexión. El brazo móvil del goniómetro acompaña el movimiento.</p></li><li class="elsevierStyleListItem" id="lsti0065"><span class="elsevierStyleLabel">•</span><p id="par0095" class="elsevierStylePara elsevierViewall">Registro: Se registra el ángulo formado entre la posición 0 y la posición final de flexión.</p></li><li class="elsevierStyleListItem" id="lsti0070"><span class="elsevierStyleLabel">•</span><p id="par0100" class="elsevierStylePara elsevierViewall">Valores normales: Flexión: 0-150°/170° (AO) y 0-180° (AAOS)<a class="elsevierStyleCrossRef" href="#bib0175"><span class="elsevierStyleSup">14</span></a>.</p></li></ul></p></span><span id="sec0025" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0085">Limitación de la aplicación G-pro smart</span><p id="par0105" class="elsevierStylePara elsevierViewall">La alineación sobre la superficie corporal debe realizarse por estimación visual de reparos anatómicos, uno proximal para el brazo fijo, uno distal para el brazo móvil y otro que corresponde al eje de movimiento de la articulación para el eje o axis<a class="elsevierStyleCrossRef" href="#bib0180"><span class="elsevierStyleSup">15</span></a>.</p><p id="par0110" class="elsevierStylePara elsevierViewall">Debido a que el goniómetro debe tomarse con las 2 manos, una para el brazo fijo y la otra para el brazo móvil, el examinador no puede efectuar correctamente la estabilización manual del segmento proximal de la articulación que se evalúa<a class="elsevierStyleCrossRef" href="#bib0185"><span class="elsevierStyleSup">16</span></a>.</p></span><span id="sec0030" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0090">Aplicación G-pro smart</span><p id="par0115" class="elsevierStylePara elsevierViewall">Es la primera aplicación para combinar un goniómetro de lecturas básicas con registros de pacientes. Está validado y es fiable como herramienta basada en investigaciones científicas<a class="elsevierStyleCrossRef" href="#bib0190"><span class="elsevierStyleSup">17</span></a>.</p><span id="sec0035" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0095">Características principales</span><p id="par0120" class="elsevierStylePara elsevierViewall"><ul class="elsevierStyleList" id="lis0015"><li class="elsevierStyleListItem" id="lsti0075"><p id="par0125" class="elsevierStylePara elsevierViewall">Altamente preciso para medir ángulos en las articulaciones grandes, pequeñas articulaciones y también de la columna vertebral.</p></li><li class="elsevierStyleListItem" id="lsti0080"><p id="par0130" class="elsevierStylePara elsevierViewall">Se pueden grabar con perfil distinto para cada paciente para mantener un buen registro para comprobar la mejora secuencial en el rango de movimiento en el seguimiento diferente arriba.</p></li></ul></p></span><span id="sec0040" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0100">Cómo utilizar la app</span><p id="par0135" class="elsevierStylePara elsevierViewall">Los siguientes pasos detallan el uso de la aplicación teniendo en cuenta que estamos midiendo gama de rodilla de movimiento.</p><p id="par0140" class="elsevierStylePara elsevierViewall">Paso 1: colocar el teléfono en la superficie del muslo con «medida contra este borde tocar la superficie».</p><p id="par0145" class="elsevierStylePara elsevierViewall">Paso 2: haga clic en inicio y el brazo de medida del goniómetro aparecerá y será sensible al movimiento.</p><p id="par0150" class="elsevierStylePara elsevierViewall">Paso 3: coloque el goniómetro en la superficie de la tibia de nuevo con”medida contra este borde tocar la superficie».</p><p id="par0155" class="elsevierStylePara elsevierViewall">Paso 4: haga clic en detener; la lectura del goniómetro se podrá ver estando de pie. El goniómetro puede dar lecturas dinámicas también.</p><p id="par0160" class="elsevierStylePara elsevierViewall">Paso 5: haga clic en «guardar lectura» y el panel en el ahorro de los registros se abrirá, siga las instrucciones y rellene el formulario.</p><p id="par0165" class="elsevierStylePara elsevierViewall">Paso 6: los registros pueden ser enviados por correo electrónico o guardados localmente en su teléfono, y luego las lecturas se pueden agregar al perfil del paciente en particular.</p></span></span><span id="sec0045" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0105">Movimiento</span><span id="sec0050" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0110">La amplitud de movimiento</span><p id="par0170" class="elsevierStylePara elsevierViewall">Es el arco de movilidad que ejecuta una articulación o una serie de articulaciones. La postura inicial para medir cualquier amplitud de movimiento, excepto las rotaciones en el plano transversal, es la posición anatómica<a class="elsevierStyleCrossRef" href="#bib0190"><span class="elsevierStyleSup">17</span></a>. Universalmente se emplea el sistema de determinación de 0° a 180°, conocido también como el método del «cero neutro».</p></span><span id="sec0055" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0115">La postura anatómica</span><p id="par0175" class="elsevierStylePara elsevierViewall">Es la posición articular de las extremidades, el punto medio del arco de rotación medial (interna) y lateral (externa) es la de 0° de amplitud para la rotación con el goniómetro estándar<a class="elsevierStyleCrossRef" href="#bib0195"><span class="elsevierStyleSup">18</span></a>.</p><p id="par0180" class="elsevierStylePara elsevierViewall">En Europa, el método fue estandarizado por la Asociación para el Estudio de Osteosíntesis (AO) de Suiza. El método del «cero neutro» se basa en la medición de los movimientos que ocurren en cada uno de los 3 ejes que cortan perpendicularmente los 3 planos del espacio a partir de la posición neutra o posición 0, en la cual todas las articulaciones se encuentran en extensión, salvo el tobillo, donde la posición 0 se verifica en 90° de flexión. A diferencia de la posición anatómica en la que la palma de la mano mira hacia delante, en la posición neutral, el pulgar apunta hacia delante colocando el antebrazo en posición intermedia de pronosupinación. A partir de esta posición, se definen determinadas eminencias óseas palpables que sirven de reparo; para la colocación del goniómetro se debe registrar al arco de movimiento de cada segmento óseo articular que da al inicio de cada movimiento<a class="elsevierStyleCrossRef" href="#bib0165"><span class="elsevierStyleSup">12</span></a>. Los valores normales corresponden al promedio del arco de movimiento que se observa en adultos sanos. Sin embargo, la existencia de importantes variaciones individuales aconseja, además, la comparación con el arco de movimiento del lado contralateral. En resumen, se trata de un método simple y lógico que puede enseñarse y aprenderse fácilmente. Requiere de medios auxiliares muy simples (goniómetro), la terminología empleada no lleva a ninguna confusión, las mediciones siempre se efectúan a partir de la misma posición, los resultados pueden ser transcritos claramente y los puntos de referencia para la colocación del goniómetro están estandarizados. Sin embargo, la mayor dificultad que presenta es la ubicación de los puntos de reparo para la colocación del goniómetro, que debe realizarse por estimación visual<a class="elsevierStyleCrossRef" href="#bib0200"><span class="elsevierStyleSup">19</span></a>.</p></span></span><span id="sec0060" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0120">Confiabilidad y variabilidad intrasujeto</span><span id="sec0065" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0125">Variabilidad intrasujeto</span><p id="par0185" class="elsevierStylePara elsevierViewall">Se administran los ítems, se correlacionan las puntuaciones de los ítems entre sí (debe haber correlaciones altas, especialmente entre ítems que miden una misma dimensión<a class="elsevierStyleCrossRef" href="#bib0160"><span class="elsevierStyleSup">11</span></a>)<a class="elsevierStyleCrossRef" href="#bib0140"><span class="elsevierStyleSup">7</span></a>, y se hacen estimaciones estadísticas para ver si la muestra es representativa.</p><p id="par0190" class="elsevierStylePara elsevierViewall">El investigador correlaciona su medición con el criterio, y este coeficiente es el que se toma como coeficiente de validez<a class="elsevierStyleCrossRef" href="#bib0140"><span class="elsevierStyleSup">7</span></a>. La validez de constructo se suele determinar mediante un procedimiento denominado «análisis de factores» (Harman, 1967)<a class="elsevierStyleCrossRefs" href="#bib0165"><span class="elsevierStyleSup">12,13</span></a>.</p></span><span id="sec0070" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0130">Confiabilidad</span><p id="par0195" class="elsevierStylePara elsevierViewall">Existen diversos procedimientos para calcular la confiabilidad de un instrumento de medición. Todos utilizan fórmulas que producen coeficientes de confiabilidad. Estos coeficientes pueden oscilar entre 0 y 1, donde un coeficiente de 0 significa nula confiabilidad y 1 representa un máximo de confiabilidad (confiabilidad total). Cuanto más se acerque el coeficiente a cero (0), mayor error hay en la medición<a class="elsevierStyleCrossRef" href="#bib0205"><span class="elsevierStyleSup">20</span></a>.</p></span></span></span><span id="sec0075" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0135">Resultados (<a class="elsevierStyleCrossRef" href="#tbl0005">tabla 1</a>)</span><p id="par0200" class="elsevierStylePara elsevierViewall">El resultado de la confiabilidad medido con el alfa Cronbach, del test pretest de la flexión del hombro derecho con el goniómetro universal fue de 0,609. Este valor obtenido indica que la confiabilidad es cuestionable en la medición de la flexión del hombro derecho con el goniómetro universal<a class="elsevierStyleCrossRef" href="#bib0195"><span class="elsevierStyleSup">18</span></a>.</p><elsevierMultimedia ident="tbl0005"></elsevierMultimedia><p id="par0205" class="elsevierStylePara elsevierViewall">En la <a class="elsevierStyleCrossRef" href="#tbl0010">tabla 2</a> evaluada por el autor se muestra la correlación intraclase (ICC); el valor obtenido en el test y retest, fue de 0,425 cuando se aplicó el goniómetro universal en la medición de la flexión del hombro derecho de los estudiantes, siendo esta correlación altamente significativa (p<span class="elsevierStyleHsp" style=""></span><<span class="elsevierStyleHsp" style=""></span>0,01).</p><elsevierMultimedia ident="tbl0010"></elsevierMultimedia><p id="par0210" class="elsevierStylePara elsevierViewall">El valor obtenido de alfa de Cronbach del test y retest en la medición de la flexión del hombro derecho con el gonómietro pro fue de 0,998. Este valor indica que la confiabilidad es excelente en la medición de la flexión del hombro derecho con el goniómetro pro según lo recomendado.</p><p id="par0215" class="elsevierStylePara elsevierViewall">Se encontró una ICC de 0,913 en el test y retest cuando se aplicó el goniómetro pro en la medición de la flexión del hombro izquierdo de los estudiantes, siendo esta correlación altamente significativa (p<span class="elsevierStyleHsp" style=""></span><<span class="elsevierStyleHsp" style=""></span>0,01).</p><p id="par0220" class="elsevierStylePara elsevierViewall">El valor obtenido del alfa de Cronbach en el test y retest de la medición de la extensión del hombro derecho con el G-pro fue de 0,927 por lo que es más confiable que el goniómetro universal en la medición de extensión, con el cual se obtuvo un valor de 0,702 en el test y retest (<a class="elsevierStyleCrossRef" href="#tbl0015">tabla 3</a>).</p><elsevierMultimedia ident="tbl0015"></elsevierMultimedia><p id="par0225" class="elsevierStylePara elsevierViewall">El valor obtenido de ICC indica que este resultado se considera como excelente</p><span id="sec0080" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0140">Correlación entre la medición del goniómetro pro y el goniómetro universal</span><p id="par0230" class="elsevierStylePara elsevierViewall">Para la realización del análisis de correlación se realizó la prueba de la normalidad de las mediciones con el goniómetro pro y el universal, en la flexión del hombro izquierdo.</p><p id="par0235" class="elsevierStylePara elsevierViewall">En la <a class="elsevierStyleCrossRef" href="#tbl0020">tabla 4</a> se muestran los resultados del análisis de la prueba de la normalidad de las mediciones realizadas en el hombro izquierdo con el goniómetro pro y el universal; con excepción del retest con el G-pro, las muestras tienen distribución normal, por lo que se puede asumir la normalidad de los datos y aplicar el coeficiente de correlación de Pearson para el análisis de la asociación de las mediciones.</p><elsevierMultimedia ident="tbl0020"></elsevierMultimedia><p id="par0240" class="elsevierStylePara elsevierViewall">En la <a class="elsevierStyleCrossRef" href="#tbl0025">tabla 5</a> se presentan los resultados del análisis de correlación de Pearson entre el goniómetro pro y el universal en el test y retest de la flexión del hombro izquierdo, donde se aprecia que existe una correlación positiva moderada en el test (0,588) y es altamente significativa entre los 2 instrumentos (p<span class="elsevierStyleHsp" style=""></span><<span class="elsevierStyleHsp" style=""></span>0,01), y una correlación positiva fuerte en el retest (0,473) y es estadísticamente significativa (p<span class="elsevierStyleHsp" style=""></span><<span class="elsevierStyleHsp" style=""></span>0,05).</p><elsevierMultimedia ident="tbl0025"></elsevierMultimedia><p id="par0245" class="elsevierStylePara elsevierViewall">Existe una confiabilidad de excelente (r<span class="elsevierStyleHsp" style=""></span>=<span class="elsevierStyleHsp" style=""></span>0,998) utilizando la aplicación G-pro smartphone, mientras con el goniómetro universal esta confiabilidad es de aceptable a cuestionable (r<span class="elsevierStyleHsp" style=""></span>=<span class="elsevierStyleHsp" style=""></span>0,757 a 0,609) en el movimiento activo de flexo-extensión del hombro.</p><p id="par0250" class="elsevierStylePara elsevierViewall">La confiabilidad medida con el alfa de Cronbach del test y retest de la flexión del hombro derecho con el goniómetro universal fue de 0,609. Este valor obtenido indica que la confiabilidad es cuestionable en la medición de la flexión del hombro derecho con el goniómetro universal; mientras que, el test y retest en la medición de la flexión del hombro derecho con el goniómetro pro fue de 0,998, por lo que se demuestra que el G-pro es más confiable para la medición en flexión de hombro derecho.</p><p id="par0255" class="elsevierStylePara elsevierViewall">La confiabilidad es aceptable en el test y retest de la medición de la flexión del hombro izquierdo con el goniómetro universal, dado que el valor obtenido del alfa Cronbach fue de 0,757.</p><p id="par0260" class="elsevierStylePara elsevierViewall">Asimismo, el valor obtenido del alfa de Cronbach fue de 0,915 en el goniómetro universal; mientras que en el test retest de la medición de la flexión del hombro Izquierdo con el G-pro el valor fue de 0,964. Se demuestra que el G-pro es más confiable para la medición en flexión de hombro izquierdo.</p><p id="par0265" class="elsevierStylePara elsevierViewall">Además, el valor obtenido en el test y retest de la medición de la extensión del hombro izquierdo con el goniómetro universal fue de 0,77. El valor obtenido en el test y retest de la medición del hombro izquierdo con el goniómetro pro fue de 0,896, lo que indica que la confiabilidad es buena en la medición de la extensión del hombro izquierdo con el G-pro<a class="elsevierStyleCrossRefs" href="#bib0200"><span class="elsevierStyleSup">19,21</span></a>.</p><p id="par0270" class="elsevierStylePara elsevierViewall">En el análisis de correlación de Pearson entre el G-pro y el universal en el test y retest de la flexión del hombro derecho se aprecia que existe una correlación positiva débil en el test (0,476) y es altamente significativa entre los 2 instrumentos (p<span class="elsevierStyleHsp" style=""></span><<span class="elsevierStyleHsp" style=""></span>0,01), y una correlación positiva moderada en el retest (0,473) y es estadísticamente significativa (p<span class="elsevierStyleHsp" style=""></span><<span class="elsevierStyleHsp" style=""></span>0,05). Se puede afirmar que en la segunda toma del retest hay una mayor correlación entre los 2 instrumentos.</p></span></span><span id="sec0085" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0145">Discusión de resultados</span><p id="par0275" class="elsevierStylePara elsevierViewall">La prueba de la eficacia entre el goniómetro universal y el goniómetro pro se ajusta como medio de valoración para medir el acondicionamiento para los rangos articulares en el hombro tanto derecho como izquierdo en flexión. Por lo tanto, se recomienda que en futuras investigaciones se tome en consideración esta prueba en abducción, aducción rotación interna y rotación externa.</p><p id="par0280" class="elsevierStylePara elsevierViewall">Se ha encontrado variación en la medición en flexión del hombro (ICC), el valor obtenido en el test entre los 2 instrumentos fue de 0,995 en la medición de la flexión del hombro derecho de los estudiantes, siendo esta correlación altamente significativa (p<span class="elsevierStyleHsp" style=""></span><<span class="elsevierStyleHsp" style=""></span>0,01). Según el método de exploración del aparato locomotor y de la postura<a class="elsevierStyleCrossRef" href="#bib0200"><span class="elsevierStyleSup">19</span></a> se determina que la fiabilidad de nuestros métodos fue sustancial para la flexión total de hombro (ICC<span class="elsevierStyleHsp" style=""></span>=<span class="elsevierStyleHsp" style=""></span>0,82; SEM<span class="elsevierStyleHsp" style=""></span>=<span class="elsevierStyleHsp" style=""></span>12,3°), mientras que todos los otros movimientos demostraron la fiabilidad moderada (rango ICC<span class="elsevierStyleHsp" style=""></span>=<span class="elsevierStyleHsp" style=""></span>0,64 a 0,75), difieren de los obtenidos en dicho estudio. En la medición con el G-pro smartphone se tiene un alfa de Cronbach de 0,896. Ambos estudios presentan que el smartphone ha demostrado que tiene una alta fiabilidad y validez. También se puede medir en los diferentes movimientos como abducción, aducción, rotación interna y externa; nuestra investigación se centró en la flexión y extensión del hombro.</p></span><span id="sec0090" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0150">Conclusiones y recomendaciones</span><p id="par0285" class="elsevierStylePara elsevierViewall"><ul class="elsevierStyleList" id="lis0020"><li class="elsevierStyleListItem" id="lsti0085"><p id="par0290" class="elsevierStylePara elsevierViewall">Se recomienda realizar estudios con muestras con mayor cantidad de estudiantes, con la finalidad de contrastar los resultados obtenidos en el presente estudio. Se sugiere diseñar programas de capacitación sobre el uso de G-pro smartphone para mejorar las herramientas de valoración clínica.</p></li><li class="elsevierStyleListItem" id="lsti0090"><p id="par0295" class="elsevierStylePara elsevierViewall">El acondicionamiento físico demostró beneficios estadísticamente significativos en los valores en la toma de muestra de ambos instrumentos, en la medición del rango articular flexión. Esto es importante, pues no hay que olvidar que la toma de muestra tiene que ser en un buen ambiente, para tomar dichas muestras con los diferentes instrumentos, y debe estar en óptimas condiciones para poder medir los rangos articulares</p></li><li class="elsevierStyleListItem" id="lsti0095"><p id="par0300" class="elsevierStylePara elsevierViewall">Gracias al avance tecnológico podemos mejorar e innovar el resultado de valoración articular para que este sea preciso, y los tratamientos brindados sean adecuados y diferenciados en cada paciente. Esta herramienta es de bajo costo y está al alcance del personal de salud, fácil de utilizar, dado que cuenta con un inclinómetro que trabaja visual-valorativo confiable.</p></li><li class="elsevierStyleListItem" id="lsti0100"><p id="par0305" class="elsevierStylePara elsevierViewall">Se sugiere diseñar programas de acondicionamiento, de cómo se puede utilizar el goniómetro pro en los estudiantes de la carrera de terapia física.</p></li></ul></p></span><span id="sec0095" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0155">Financiación</span><p id="par0310" class="elsevierStylePara elsevierViewall">La presente investigación no ha recibido ayudas específicas provenientes de agencias del sector público, sector comercial o entidades sin ánimo de lucro.</p></span><span id="sec0100" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0160">Conflicto de intereses</span><p id="par0315" class="elsevierStylePara elsevierViewall">Los autores declaran no tener ningún conflicto de intereses.</p></span></span>" "textoCompletoSecciones" => array:1 [ "secciones" => array:13 [ 0 => array:3 [ "identificador" => "xres1667684" "titulo" => "Resumen" "secciones" => array:4 [ 0 => array:2 [ "identificador" => "abst0005" "titulo" => "Antecedentes y objetivos" ] 1 => array:2 [ "identificador" => "abst0010" "titulo" => "Materiales y métodos" ] 2 => array:2 [ "identificador" => "abst0015" "titulo" => "Resultados" ] 3 => array:2 [ "identificador" => "abst0020" "titulo" => "Conclusiones" ] ] ] 1 => array:2 [ "identificador" => "xpalclavsec1481880" "titulo" => "Palabras clave" ] 2 => array:3 [ "identificador" => "xres1667683" "titulo" => "Abstract" "secciones" => array:4 [ 0 => array:2 [ "identificador" => "abst0025" "titulo" => "Background and objective" ] 1 => array:2 [ "identificador" => "abst0030" "titulo" => "Material and methods" ] 2 => array:2 [ "identificador" => "abst0035" "titulo" => "Results" ] 3 => array:2 [ "identificador" => "abst0040" "titulo" => "Conclusions" ] ] ] 3 => array:2 [ "identificador" => "xpalclavsec1481881" "titulo" => "Keywords" ] 4 => array:2 [ "identificador" => "sec0005" "titulo" => "Introducción" ] 5 => array:3 [ "identificador" => "sec0010" "titulo" => "Materiales y métodos" "secciones" => array:6 [ 0 => array:2 [ "identificador" => "sec0015" "titulo" => "Goniometría en miembro superior" ] 1 => array:2 [ "identificador" => "sec0020" "titulo" => "Goniometría en la articulación del hombro en la flexión" ] 2 => array:2 [ "identificador" => "sec0025" "titulo" => "Limitación de la aplicación G-pro smart" ] 3 => array:3 [ "identificador" => "sec0030" "titulo" => "Aplicación G-pro smart" "secciones" => array:2 [ 0 => array:2 [ "identificador" => "sec0035" "titulo" => "Características principales" ] 1 => array:2 [ "identificador" => "sec0040" "titulo" => "Cómo utilizar la app" ] ] ] 4 => array:3 [ "identificador" => "sec0045" "titulo" => "Movimiento" "secciones" => array:2 [ 0 => array:2 [ "identificador" => "sec0050" "titulo" => "La amplitud de movimiento" ] 1 => array:2 [ "identificador" => "sec0055" "titulo" => "La postura anatómica" ] ] ] 5 => array:3 [ "identificador" => "sec0060" "titulo" => "Confiabilidad y variabilidad intrasujeto" "secciones" => array:2 [ 0 => array:2 [ "identificador" => "sec0065" "titulo" => "Variabilidad intrasujeto" ] 1 => array:2 [ "identificador" => "sec0070" "titulo" => "Confiabilidad" ] ] ] ] ] 6 => array:3 [ "identificador" => "sec0075" "titulo" => "Resultados (tabla 1)" "secciones" => array:1 [ 0 => array:2 [ "identificador" => "sec0080" "titulo" => "Correlación entre la medición del goniómetro pro y el goniómetro universal" ] ] ] 7 => array:2 [ "identificador" => "sec0085" "titulo" => "Discusión de resultados" ] 8 => array:2 [ "identificador" => "sec0090" "titulo" => "Conclusiones y recomendaciones" ] 9 => array:2 [ "identificador" => "sec0095" "titulo" => "Financiación" ] 10 => array:2 [ "identificador" => "sec0100" "titulo" => "Conflicto de intereses" ] 11 => array:2 [ "identificador" => "xack587998" "titulo" => "Agradecimientos" ] 12 => array:1 [ "titulo" => "Bibliografía" ] ] ] "pdfFichero" => "main.pdf" "tienePdf" => true "fechaRecibido" => "2019-09-30" "fechaAceptado" => "2020-11-04" "PalabrasClave" => array:2 [ "es" => array:1 [ 0 => array:4 [ "clase" => "keyword" "titulo" => "Palabras clave" "identificador" => "xpalclavsec1481880" "palabras" => array:5 [ 0 => "Goniómetro" 1 => "G-pro" 2 => "Hombro" 3 => "Flexión de hombro" 4 => "Extensión del hombro" ] ] ] "en" => array:1 [ 0 => array:4 [ "clase" => "keyword" "titulo" => "Keywords" "identificador" => "xpalclavsec1481881" "palabras" => array:5 [ 0 => "Goniometer" 1 => "G-pro" 2 => "Shoulder" 3 => "Shoulder flexion" 4 => "Shoulder extension" ] ] ] ] "tieneResumen" => true "resumen" => array:2 [ "es" => array:3 [ "titulo" => "Resumen" "resumen" => "<span id="abst0005" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0010">Antecedentes y objetivos</span><p id="spar0005" class="elsevierStyleSimplePara elsevierViewall">En la actualidad la falta de herramientas para el fisioterapeuta y la calidad que se le da al momento de evaluar no permite hacer un buen diagnóstico. La evaluación precisa del rango articular conforma una tarea esencial en la labor diaria de los fisioterapeutas, para poder valorar correctamente la progresión del tratamiento. En esta investigación nos planteamos analizar la fiabilidad y la seguridad de una aplicación G-Pro smart bajo el sistema Android, para evaluar sus características y observar su utilidad como goniómetro en la clínica para medir los ángulos articulares.</p></span> <span id="abst0010" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0015">Materiales y métodos</span><p id="spar0010" class="elsevierStyleSimplePara elsevierViewall">Se empleó el goniómetro estándar por un tecnólogo médico en terapia física y rehabilitación y la aplicación G-pro smartphone para medir de manera pasiva y bilateral la flexión y rotación de hombro, y la flexión de codo de 30 sujetos que toleraran la bipedestación.</p></span> <span id="abst0015" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0020">Resultados</span><p id="spar0015" class="elsevierStyleSimplePara elsevierViewall">Existe una confiabilidad de buena a excelente (r<span class="elsevierStyleHsp" style=""></span>=<span class="elsevierStyleHsp" style=""></span>0,998 a 0,896; p<span class="elsevierStyleHsp" style=""></span><<span class="elsevierStyleHsp" style=""></span>0,01), utilizando la aplicación G-pro smartphone, mientras que con el goniómetro universal esta confiabilidad es de aceptable a cuestionable (r<span class="elsevierStyleHsp" style=""></span>=<span class="elsevierStyleHsp" style=""></span>0,757 a 0,609) en el movimiento activo de flexo-extensión del hombro.</p></span> <span id="abst0020" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0025">Conclusiones</span><p id="spar0020" class="elsevierStyleSimplePara elsevierViewall">Los resultados obtenidos en este estudio nos muestran que estos dispositivos aplicativos a través de un smartphone suponen un recurso adicional para evaluar los rangos de movilidad articular. Gracias al avance tecnológico podemos mejorar, innovar y objetivar el resultado de valoración articular para que este sea preciso, y por lo tanto los tratamientos brindados sean adecuados y diferenciados para cada paciente. Esta herramienta es de bajo costo y está al alcance del personal de salud, fácil de utilizar, dado que cuenta con un inclinómetro que es confiable tanto visual como valorativo.</p></span>" "secciones" => array:4 [ 0 => array:2 [ "identificador" => "abst0005" "titulo" => "Antecedentes y objetivos" ] 1 => array:2 [ "identificador" => "abst0010" "titulo" => "Materiales y métodos" ] 2 => array:2 [ "identificador" => "abst0015" "titulo" => "Resultados" ] 3 => array:2 [ "identificador" => "abst0020" "titulo" => "Conclusiones" ] ] ] "en" => array:3 [ "titulo" => "Abstract" "resumen" => "<span id="abst0025" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0035">Background and objective</span><p id="spar0025" class="elsevierStyleSimplePara elsevierViewall">Currently, the lack of tools for physiotherapists, and the quality of these tools when physiotherapists assess patients does not allow a good diagnosis. Accurate joint range evaluation is an essential task in the daily work of physiotherapists in order to correctly assess treatment progress. In this study, we examined the reliability and safety of an Android system application called G-Pro to evaluate its characteristics and observe its usefulness as a goniometer to measure joint angles in a clinic.</p></span> <span id="abst0030" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0040">Material and methods</span><p id="spar0030" class="elsevierStyleSimplePara elsevierViewall">The standard goniometer and G-pro smartphone with the application were used to passively and bilaterally measure shoulder flexion and rotation and elbow flexion in 30 participants able to tolerate standing.</p></span> <span id="abst0035" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0045">Results</span><p id="spar0035" class="elsevierStyleSimplePara elsevierViewall">The G-pro smartphone application showed good to excellent reliability (r<span class="elsevierStyleHsp" style=""></span>=<span class="elsevierStyleHsp" style=""></span>0.998 to 0.896, <span class="elsevierStyleItalic">P</span><span class="elsevierStyleHsp" style=""></span><<span class="elsevierStyleHsp" style=""></span>.01), while the universal goniometer showed acceptable to questionable reliability (r<span class="elsevierStyleHsp" style=""></span>=<span class="elsevierStyleHsp" style=""></span>0.757 to 0.609) in active shoulder flexion-extension movement.</p></span> <span id="abst0040" class="elsevierStyleSection elsevierViewall"><span class="elsevierStyleSectionTitle" id="sect0050">Conclusions</span><p id="spar0040" class="elsevierStyleSimplePara elsevierViewall">The results obtained in this study show that these devices represent an additional resource to assess joint mobility ranges. Due to technological progress, we can improve, innovate, and demonstrate the results of joint assessments to ensure their accuracy. This, in turn, ensures that the treatment provided is appropriate and specific to each patient. This tool is low cost and is accessible for health care professionals; moreover, it is easy to use since it has an incliniometer that is reliable both visually and evaluatively.</p></span>" "secciones" => array:4 [ 0 => array:2 [ "identificador" => "abst0025" "titulo" => "Background and objective" ] 1 => array:2 [ "identificador" => "abst0030" "titulo" => "Material and methods" ] 2 => array:2 [ "identificador" => "abst0035" "titulo" => "Results" ] 3 => array:2 [ "identificador" => "abst0040" "titulo" => "Conclusions" ] ] ] ] "multimedia" => array:5 [ 0 => array:8 [ "identificador" => "tbl0005" "etiqueta" => "Tabla 1" "tipo" => "MULTIMEDIATABLA" "mostrarFloat" => true "mostrarDisplay" => false "detalles" => array:1 [ 0 => array:3 [ "identificador" => "at1" "detalle" => "Tabla " "rol" => "short" ] ] "tabla" => array:1 [ "tablatextoimagen" => array:1 [ 0 => array:2 [ "tabla" => array:1 [ 0 => """ <table border="0" frame="\n \t\t\t\t\tvoid\n \t\t\t\t" class=""><thead title="thead"><tr title="table-row"><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Variable \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Media \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Mínimo \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Máximo \t\t\t\t\t\t\n \t\t\t\t\t\t</th></tr></thead><tbody title="tbody"><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Edad \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">20 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">12 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">25 \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Peso \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">57,7 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">42 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">73 \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Talla \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">1,6 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">1,5 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">1,7 \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">índice de masa corporal \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">21,9 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">16 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">24,9 \t\t\t\t\t\t\n \t\t\t\t</td></tr></tbody></table> """ ] "imagenFichero" => array:1 [ 0 => "xTab2839597.png" ] ] ] ] "descripcion" => array:1 [ "es" => "<p id="spar0045" class="elsevierStyleSimplePara elsevierViewall">Estadísticos de la edad, peso, talla e índice de masa corporal de estudiantes del II ciclo de un Instituto Superior de Salud en el año 2016</p>" ] ] 1 => array:8 [ "identificador" => "tbl0010" "etiqueta" => "Tabla 2" "tipo" => "MULTIMEDIATABLA" "mostrarFloat" => true "mostrarDisplay" => false "detalles" => array:1 [ 0 => array:3 [ "identificador" => "at2" "detalle" => "Tabla " "rol" => "short" ] ] "tabla" => array:2 [ "leyenda" => "<p id="spar0055" class="elsevierStyleSimplePara elsevierViewall">Modelo de efectos aleatorios de un factor en el que los efectos de las personas son aleatorios.</p>" "tablatextoimagen" => array:1 [ 0 => array:2 [ "tabla" => array:1 [ 0 => """ <table border="0" frame="\n \t\t\t\t\tvoid\n \t\t\t\t" class=""><thead title="thead"><tr title="table-row"><th class="td-with-role" title="\n \t\t\t\t\ttable-head\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col">Correlación intraclase \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col"> \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " colspan="2" align="center" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Intervalo de confianza 95%</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " colspan="4" align="center" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Prueba F con valor verdadero 0</th></tr><tr title="table-row"><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black"> \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black"> \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Límite inferior \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Límite superior \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Valor \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">gl1 \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">gl2 \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Sig. \t\t\t\t\t\t\n \t\t\t\t\t\t</th></tr></thead><tbody title="tbody"><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Medidas individuales \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,425 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,086 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,676 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">2,475 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">29 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">30 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,008 \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Medidas promedio \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,596 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,159 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,807 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">2,475 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">29 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">30 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,008 \t\t\t\t\t\t\n \t\t\t\t</td></tr></tbody></table> """ ] "imagenFichero" => array:1 [ 0 => "xTab2839600.png" ] ] ] ] "descripcion" => array:1 [ "es" => "<p id="spar0050" class="elsevierStyleSimplePara elsevierViewall">Correlación intraclase de la aplicación del goniómetro universal en la medición de la flexión del hombro derecho de los estudiantes del II ciclo del Instituto Superior de Salud</p>" ] ] 2 => array:8 [ "identificador" => "tbl0015" "etiqueta" => "Tabla 3" "tipo" => "MULTIMEDIATABLA" "mostrarFloat" => true "mostrarDisplay" => false "detalles" => array:1 [ 0 => array:3 [ "identificador" => "at3" "detalle" => "Tabla " "rol" => "short" ] ] "tabla" => array:2 [ "leyenda" => "<p id="spar0065" class="elsevierStyleSimplePara elsevierViewall">Modelo de efectos aleatorios de un factor en el que los efectos de las personas son aleatorios.</p>" "tablatextoimagen" => array:1 [ 0 => array:2 [ "tabla" => array:1 [ 0 => """ <table border="0" frame="\n \t\t\t\t\tvoid\n \t\t\t\t" class=""><thead title="thead"><tr title="table-row"><th class="td-with-role" title="\n \t\t\t\t\ttable-head\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col">Correlación intraclase \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col"> \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " colspan="2" align="center" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Intervalo de confianza 95%</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " colspan="4" align="center" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Prueba F con valor verdadero 0</th></tr><tr title="table-row"><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black"> \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black"> \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Límite inferior \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Límite superior \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Valor \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">gl1 \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">gl2 \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Sig. \t\t\t\t\t\t\n \t\t\t\t\t\t</th></tr></thead><tbody title="tbody"><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Medidas individuales \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,913 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,828 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,958 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">22,088 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">29 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">30 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,000 \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Medidas promedio \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,955 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,906 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,978 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">22,088 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">29 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">30 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,000 \t\t\t\t\t\t\n \t\t\t\t</td></tr></tbody></table> """ ] "imagenFichero" => array:1 [ 0 => "xTab2839596.png" ] ] ] ] "descripcion" => array:1 [ "es" => "<p id="spar0060" class="elsevierStyleSimplePara elsevierViewall">Correlación intraclase de la aplicación del goniómetro pro en la medición de la flexión del hombro izquierdo de estudiantes del II ciclo del Instituto Superior de Salud</p>" ] ] 3 => array:8 [ "identificador" => "tbl0020" "etiqueta" => "Tabla 4" "tipo" => "MULTIMEDIATABLA" "mostrarFloat" => true "mostrarDisplay" => false "detalles" => array:1 [ 0 => array:3 [ "identificador" => "at4" "detalle" => "Tabla " "rol" => "short" ] ] "tabla" => array:2 [ "tablatextoimagen" => array:1 [ 0 => array:2 [ "tabla" => array:1 [ 0 => """ <table border="0" frame="\n \t\t\t\t\tvoid\n \t\t\t\t" class=""><thead title="thead"><tr title="table-row"><th class="td-with-role" title="\n \t\t\t\t\ttable-head\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col">Estadísticos \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col"> \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " colspan="2" align="center" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Goniómetro universal</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " colspan="2" align="center" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Goniómetro pro</th></tr><tr title="table-row"><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black"> \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black"> \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Test \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Retest \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Test \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Retest \t\t\t\t\t\t\n \t\t\t\t\t\t</th></tr></thead><tbody title="tbody"><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">N \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t"> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">30 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">30 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">30 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">30 \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Parámetros normales<a class="elsevierStyleCrossRef" href="#tblfn0005"><span class="elsevierStyleSup">a</span></a><span class="elsevierStyleSup">,</span><a class="elsevierStyleCrossRef" href="#tblfn0010"><span class="elsevierStyleSup">b</span></a> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Media \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">174,57 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">176,40 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">173.47 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">174,53 \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t"> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Desviación típica \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">2,932 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">3,180 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">5,380 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">5,244 \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Diferencias más extremas \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Absoluta \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,159 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,226 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,173 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,269 \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t"> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Positiva \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,140 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0, 129 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,086 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,115 \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t"> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Negativa \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">−0,159 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">−0,226 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">−0,173 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">−0,269 \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Z de Kolmogorov-Smirnov \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,869 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " colspan="2" align="center" valign="\n \t\t\t\t\ttop\n \t\t\t\t">1,237</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,947 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">1,472 \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Sig. asintót. (bilateral) \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0, 436 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " colspan="2" align="center" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,094</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0, 332 \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,026 \t\t\t\t\t\t\n \t\t\t\t</td></tr></tbody></table> """ ] "imagenFichero" => array:1 [ 0 => "xTab2839598.png" ] ] ] "notaPie" => array:2 [ 0 => array:3 [ "identificador" => "tblfn0005" "etiqueta" => "a" "nota" => "<p class="elsevierStyleNotepara" id="npar0005">La distribución de contraste es la normal.</p>" ] 1 => array:3 [ "identificador" => "tblfn0010" "etiqueta" => "b" "nota" => "<p class="elsevierStyleNotepara" id="npar0010">Se han calculado a partir de los datos.</p>" ] ] ] "descripcion" => array:1 [ "es" => "<p id="spar0070" class="elsevierStyleSimplePara elsevierViewall">Prueba de Kolmogorov-Smirnov para una muestra para la medición de la flexión del hombro izquierdo en el test y retest usando el goniómetro universal y el goniómetro pro</p>" ] ] 4 => array:8 [ "identificador" => "tbl0025" "etiqueta" => "Tabla 5" "tipo" => "MULTIMEDIATABLA" "mostrarFloat" => true "mostrarDisplay" => false "detalles" => array:1 [ 0 => array:3 [ "identificador" => "at5" "detalle" => "Tabla " "rol" => "short" ] ] "tabla" => array:1 [ "tablatextoimagen" => array:1 [ 0 => array:2 [ "tabla" => array:1 [ 0 => """ <table border="0" frame="\n \t\t\t\t\tvoid\n \t\t\t\t" class=""><thead title="thead"><tr title="table-row"><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black"> \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black"> \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">R de Pearson \t\t\t\t\t\t\n \t\t\t\t\t\t</th><th class="td" title="\n \t\t\t\t\ttable-head\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t" scope="col" style="border-bottom: 2px solid black">Goniómetro pro \t\t\t\t\t\t\n \t\t\t\t\t\t</th></tr></thead><tbody title="tbody"><tr title="table-row"><td class="td-with-role" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t ; entry_with_role_rowhead " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Goniómetro universal \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Test \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Coeficiente de correlación \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,588 \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t"> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t"> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Sig. (bilateral) \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,001 \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t"> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t"> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">N \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">30 \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t"> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Retest \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Coeficiente de correlación \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,806 \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t"> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t"> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">Sig. (bilateral) \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">0,023 \t\t\t\t\t\t\n \t\t\t\t</td></tr><tr title="table-row"><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t"> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="" valign="\n \t\t\t\t\ttop\n \t\t\t\t"> \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">N \t\t\t\t\t\t\n \t\t\t\t</td><td class="td" title="\n \t\t\t\t\ttable-entry\n \t\t\t\t " align="left" valign="\n \t\t\t\t\ttop\n \t\t\t\t">30 \t\t\t\t\t\t\n \t\t\t\t</td></tr></tbody></table> """ ] "imagenFichero" => array:1 [ 0 => "xTab2839599.png" ] ] ] ] "descripcion" => array:1 [ "es" => "<p id="spar0075" class="elsevierStyleSimplePara elsevierViewall">Análisis de correlación de Pearson entre el goniómetro pro y el universal en el test y retest de la medición de la flexión del hombro izquierdo de estudiantes del II ciclo de un Instituto Superior de Salud del semestre 2016-I</p>" ] ] ] "bibliografia" => array:2 [ "titulo" => "Bibliografía" "seccion" => array:1 [ 0 => array:2 [ "identificador" => "bibs0015" "bibliografiaReferencia" => array:21 [ 0 => array:3 [ "identificador" => "bib0110" "etiqueta" => "1" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Physiotherapy rehabilitation for people with spinal cord injuries" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:1 [ 0 => "L.A. Harvey" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1016/j.jphys.2015.11.004" "Revista" => array:6 [ "tituloSerie" => "J Physiother" "fecha" => "2016" "volumen" => "62" "paginaInicial" => "4" "paginaFinal" => "11" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/26701156" "web" => "Medline" ] ] ] ] ] ] ] ] 1 => array:3 [ "identificador" => "bib0115" "etiqueta" => "2" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Análisis del movimiento de las extremidades superiores aplicado a la rehabilitación física de una persona usando técnicas de visión artificial" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:5 [ 0 => "M. Velarde" 1 => "E. Perugachi" 2 => "D. Romero" 3 => "A. Sappa" 4 => "B. Vintimilla" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Revista" => array:6 [ "tituloSerie" => "Revista Tecnológica ESPOL – RTE" "fecha" => "2015" "volumen" => "28" "paginaInicial" => "1" "paginaFinal" => "7" "itemHostRev" => array:3 [ "pii" => "S0190962218304912" "estado" => "S300" "issn" => "01909622" ] ] ] ] ] ] ] 2 => array:3 [ "identificador" => "bib0120" "etiqueta" => "3" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "A cross-sectional study examining shoulder pain and disability in Division I female swimmers" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:3 [ 0 => "S. Harrington" 1 => "C. Meisel" 2 => "A. Tate" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1123/jsr. 2012-0123" "Revista" => array:6 [ "tituloSerie" => "J Sport Rehabil" "fecha" => "2014" "volumen" => "23" "paginaInicial" => "65" "paginaFinal" => "75" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/23945068" "web" => "Medline" ] ] ] ] ] ] ] ] 3 => array:3 [ "identificador" => "bib0125" "etiqueta" => "4" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Goniometría: una herramienta para la evaluación de las incapacidades laborales" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:1 [ 0 => "C. Taboadeo" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Libro" => array:4 [ "edicion" => "1.<span class="elsevierStyleSup">a</span> ed." "fecha" => "2007" "editorial" => "Asociart ART" "editorialLocalizacion" => "Buenos Aires" ] ] ] ] ] ] 4 => array:3 [ "identificador" => "bib0130" "etiqueta" => "5" "referencia" => array:1 [ 0 => array:1 [ "referenciaCompleta" => "Goniómetro pro gratuito - Aplicaciones Android en Google Play [Internet] [citado 2 May 2016]. Recuperado a partir de:https://play.google.com/store/apps/details?id=com.FiveFufFive.GoniometerProPreviewAndroid&hl=es_419" ] ] ] 5 => array:3 [ "identificador" => "bib0135" "etiqueta" => "6" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "The reliability of a smartphone goniometer application compared with a traditional goniometer for measuring first metatarsophalangeal joint dorsiflexion" "autores" => array:1 [ 0 => array:2 [ "etal" => true "autores" => array:6 [ 0 => "S. Otter" 1 => "A. Brunilda" 2 => "B. Nicola" 3 => "H. Georgie" 4 => "H. Katrina" 5 => "J. Keeley" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1186/s13047-015-0088-3" "Revista" => array:5 [ "tituloSerie" => "J Foot Ankle Res" "fecha" => "2015" "volumen" => "8" "paginaInicial" => "30" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/26207142" "web" => "Medline" ] ] ] ] ] ] ] ] 6 => array:3 [ "identificador" => "bib0140" "etiqueta" => "7" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Reliability of ankle dorsiflexion passive range of motion measurements obtained using a hand-held goniometer and Biodex dynamometer in stroke patients" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:6 [ 0 => "I.-G. Jung" 1 => "J. In-Gui" 2 => "Y. Il-Young" 3 => "K. Soo-Yong" 4 => "L. Dong-Kyu" 5 => "O. Jae-Seop" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1589/jpts.27.1899" "Revista" => array:5 [ "tituloSerie" => "J Phys Therapy Sci" "fecha" => "2015" "volumen" => "27" "paginaInicial" => "1899" "paginaFinal" => "1901" ] ] ] ] ] ] 7 => array:3 [ "identificador" => "bib0145" "etiqueta" => "8" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "The validity and reliability of a new instrumented device for measuring ankle dorsiflexion range of motion" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:6 [ 0 => "J. Calatayud" 1 => "F. Martin" 2 => "P. Gargallo" 3 => "J. García-Redondo" 4 => "J.C. Colado" 5 => "P.J. Marín" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Revista" => array:6 [ "tituloSerie" => "Int J Sports Phys Ther" "fecha" => "2015" "volumen" => "10" "paginaInicial" => "197" "paginaFinal" => "202" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/25883868" "web" => "Medline" ] ] ] ] ] ] ] ] 8 => array:3 [ "identificador" => "bib0150" "etiqueta" => "9" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Validity and reliability of two goniometric mobile apps: Device, application and examiner factors" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:4 [ 0 => "R. Wellmon" 1 => "D. Gulick" 2 => "M. Paterson" 3 => "C. Gulick" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1123/jsr. 2015-0041" "Revista" => array:6 [ "tituloSerie" => "J Sport Rehabil" "fecha" => "2016" "volumen" => "25" "paginaInicial" => "371" "paginaFinal" => "379" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/27632853" "web" => "Medline" ] ] ] ] ] ] ] ] 9 => array:3 [ "identificador" => "bib0155" "etiqueta" => "10" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Confiabilidade da fleximetria e goniometria na avaliação da amplitude de movimento cervical em crianças" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:6 [ 0 => "T. Chaves" 1 => "H. Nagamine" 2 => "J. Belli" 3 => "M. de Hannai" 4 => "D. Bevilaqua-Grossi" 5 => "A. de Oliveira" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1590/s1413-35552008000400006" "Revista" => array:5 [ "tituloSerie" => "Rev Bras Fisioter" "fecha" => "2008" "volumen" => "12" "paginaInicial" => "283" "paginaFinal" => "289" ] ] ] ] ] ] 10 => array:3 [ "identificador" => "bib0160" "etiqueta" => "11" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Knee extension range of motion and self-report physical function in total knee arthroplasty: mediating effects of knee extensor strength" "autores" => array:1 [ 0 => array:2 [ "etal" => true "autores" => array:6 [ 0 => "Y.-H. Pua" 1 => "P. Yong-Hao" 2 => "O. Peck-Hoon" 3 => "C. Hwei-Chi" 4 => "Y. William" 5 => "T. Celia" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1186/1471-2474-14-33" "Revista" => array:5 [ "tituloSerie" => "BMC Musculoskelet Disord" "fecha" => "2013" "volumen" => "14" "paginaInicial" => "33" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/23332039" "web" => "Medline" ] ] ] ] ] ] ] ] 11 => array:3 [ "identificador" => "bib0165" "etiqueta" => "12" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Reliability of the standard goniometry and diagrammatic recording of finger joint angles: A comparative study with healthy subjects and non-professional raters" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:2 [ 0 => "V. Macionis" 1 => "M. Valdas" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1186/1471-2474-14-17" "Revista" => array:6 [ "tituloSerie" => "BMC Musculoskelet Disord" "fecha" => "2013" "volumen" => "14" "paginaInicial" => "17" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/23302419" "web" => "Medline" ] ] "itemHostRev" => array:3 [ "pii" => "S0140673620319073" "estado" => "S300" "issn" => "01406736" ] ] ] ] ] ] ] 12 => array:3 [ "identificador" => "bib0170" "etiqueta" => "13" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Reproducibility of a 3-dimensional gyroscope in measuring shoulder anteflexion and abduction" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:4 [ 0 => "L. Penning" 1 => "N. Guldemond" 2 => "R. de Bie" 3 => "G. Walenkamp" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1186/1471-2474-13-135" "Revista" => array:5 [ "tituloSerie" => "BMC Musculoskelet Disord" "fecha" => "2012" "volumen" => "13" "paginaInicial" => "135" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/22846646" "web" => "Medline" ] ] ] ] ] ] ] ] 13 => array:3 [ "identificador" => "bib0175" "etiqueta" => "14" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Passive range of movement of the shoulder: A standardized method for measurement and assessment of intrarater reliability" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:4 [ 0 => "J. Dougherty" 1 => "D. Justin" 2 => "W. Sarah" 3 => "P.G. Osmotherly" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1016/j.jmpt.2014.11.006" "Revista" => array:6 [ "tituloSerie" => "J Manipulative Physiol Ther" "fecha" => "2015" "volumen" => "38" "paginaInicial" => "218" "paginaFinal" => "224" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/25704219" "web" => "Medline" ] ] ] ] ] ] ] ] 14 => array:3 [ "identificador" => "bib0180" "etiqueta" => "15" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Mobile phone-based joint angle measurement for functional assessment and rehabilitation of proprioception" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:5 [ 0 => "Q. Mourcou" 1 => "A. Fleury" 2 => "B. Diot" 3 => "C. Franco" 4 => "N. Vuillerme" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.1155/2015/328142" "Revista" => array:2 [ "tituloSerie" => "Biomed Res Int." "fecha" => "2015" ] ] ] ] ] ] 15 => array:3 [ "identificador" => "bib0185" "etiqueta" => "16" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Fiabilidad y correlación en la evaluación de la movilidad de rodilla mediante goniómetro e inclinómetro" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:2 [ 0 => "M. Gil Fernández" 1 => "J. Escobar" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Revista" => array:5 [ "tituloSerie" => "Fisioterapia" "fecha" => "2012" "volumen" => "34" "paginaInicial" => "73" "paginaFinal" => "78" ] ] ] ] ] ] 16 => array:3 [ "identificador" => "bib0190" "etiqueta" => "17" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Diagnostic studies for the evaluation of peripheral artery disease" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:4 [ 0 => "H. Banchs Viña" 1 => "H. Claudio" 2 => "M. Mesa" 3 => "A. López-Candales" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Revista" => array:6 [ "tituloSerie" => "Bol Asoc Med P R" "fecha" => "2015" "volumen" => "107" "paginaInicial" => "42" "paginaFinal" => "45" "link" => array:1 [ 0 => array:2 [ "url" => "https://www.ncbi.nlm.nih.gov/pubmed/26742195" "web" => "Medline" ] ] ] ] ] ] ] ] 17 => array:3 [ "identificador" => "bib0195" "etiqueta" => "18" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Métodos de evaluación de la amplitud del movimiento mediante métodos goniométricos" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:1 [ 0 => "J. Mazza" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Revista" => array:3 [ "tituloSerie" => "Academia Research Papers" "fecha" => "2011" "itemHostRev" => array:3 [ "pii" => "S0140673620319103" "estado" => "S300" "issn" => "01406736" ] ] ] ] ] ] ] 18 => array:3 [ "identificador" => "bib0200" "etiqueta" => "19" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:1 [ "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:5 [ 0 => "R.H. Sampieri" 1 => "C.F. Collado" 2 => "P.B. Lucio" 3 => "S.M. Valencia" 4 => "C.P.M. Torres" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Revista" => array:3 [ "tituloSerie" => "Metodología de la investigación." "fecha" => "2014" "paginaInicial" => "600" ] ] ] ] ] ] 19 => array:3 [ "identificador" => "bib0205" "etiqueta" => "20" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Método de exploración del aparato locomotor y de la postura" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:1 [ 0 => "H. Frisch" ] ] ] ] ] "host" => array:1 [ 0 => array:1 [ "Libro" => array:3 [ "fecha" => "2005" "paginaInicial" => "672" "editorial" => "Editorial Paidotribo" ] ] ] ] ] ] 20 => array:3 [ "identificador" => "bib0210" "etiqueta" => "21" "referencia" => array:1 [ 0 => array:2 [ "contribucion" => array:1 [ 0 => array:2 [ "titulo" => "Validation of a research instrument for the design of a self-assessment methodology for the environmental management system" "autores" => array:1 [ 0 => array:2 [ "etal" => false "autores" => array:2 [ 0 => "H.A. Hernández" 1 => "A.E. Pascual" ] ] ] ] ] "host" => array:1 [ 0 => array:2 [ "doi" => "10.22490/21456453.2186" "Revista" => array:5 [ "tituloSerie" => "Revista de Investigación Agraria y Ambiental" "fecha" => "2018" "volumen" => "9" "paginaInicial" => "157" "paginaFinal" => "163" ] ] ] ] ] ] ] ] ] ] "agradecimientos" => array:1 [ 0 => array:4 [ "identificador" => "xack587998" "titulo" => "Agradecimientos" "texto" => "<p id="par0320" class="elsevierStylePara elsevierViewall">Se expresa el más profundo agradecimiento al Instituto Superior de Salud y a todas las personas involucradas para realizar el estudio y probar Confiabilidad test y retest de goniometría estándar y G-pro smartphone en el movimiento flexo-extensión del hombro.</p>" "vista" => "all" ] ] ] "idiomaDefecto" => "es" "url" => "/00487120/0000005500000003/v2_202202190654/S0048712020301328/v2_202202190654/es/main.assets" "Apartado" => array:4 [ "identificador" => "5751" "tipo" => "SECCION" "es" => array:2 [ "titulo" => "Originales" "idiomaDefecto" => true ] "idiomaDefecto" => "es" ] "PDF" => "https://static.elsevier.es/multimedia/00487120/0000005500000003/v2_202202190654/S0048712020301328/v2_202202190654/es/main.pdf?idApp=UINPBA00004N&text.app=https://www.elsevier.es/" "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S0048712020301328?idApp=UINPBA00004N" ]
Información de la revista
Compartir
Descargar PDF
Más opciones de artículo
Original
Confiabilidad test-retest de goniometría estándar y G-pro smartphone en el movimiento flexo-extensión del hombro
Test-retest reliability of standard goniometry and the G-pro smartphone in shoulder flexion-extension