Compared with the Taxus® stent, the non-polymeric paclitaxel-eluting Amazonia® PAX stent shows no differences in the occurrence of coronary restenosis or clinical events after four months of follow-up. However, the performance of the Amazonia® PAX stent in more complex cases and with longer angiographic follow-up has not been demonstrated.
MethodsThe PAX-B study was a prospective, non-randomised, multicentre study assessing the long-term follow-up of patients treated with the Amazonia® PAX stent. The primary outcome was late in-stent lumen loss.
ResultsThe study included 103 patients with a mean age of 61.3±11.4years; 26.2% were diabetics, 24.3% had acute coronary syndromes, and 71.6% had type B2/C lesions. Multiple stents were implanted in 4.7% of the patients, and angiographic success was achieved in 100% of the cases. During hospitalisation, the periprocedural acute myocardial infarction rate was 3.9%, and one of these events led to target-lesion revascularisation (TLR). At the nine-month angiographic follow-up, the median late in-stent lumen loss was 0.91 [0.50; 1.21] mm. The cumulative rates of major adverse cardiac events at the six-month, nine-month, and 12-month follow-ups were 7.8%, 18.5%, and 21.3%, respectively, mostly due to TLR. There were no deaths or stent thromboses at 12 months.
ConclusionsThe Amazonia® PAX stent showed excellent immediate results and a good safety profile. However, angiographic recurrence rates were relatively high because of the low efficacy of neointimal hyperplasia inhibition.
Seguimento Angiográficoe Clínico Tardio do Novo Stent Farmacológico Não Polimérico Liberador de Paclitaxel para o Tratamento de Lesões Coronárias De Novo: Resultados do Estudo PAX-B
IntroduçãoO stent farmacológico eluidor de paclitaxel, não polimérico, Amazonia® PAX não mostrou diferença na reestenose coronária ou eventos clínicos aos 4 meses de evolução quando comparado com o stent Taxus®. Entretanto, o desempenho do stent Amazonia® PAX em cenários de maior complexidade e com seguimento angiográfico mais longo ainda não foi demonstrado.
MétodosO Estudo PAX-B foi um estudo prospectivo, não randomizado, multicêntrico, que avaliou os resultados tardios de pacientes tratados com o stent Amazonia® PAX. O desfecho primário foi a perda tardia do lúmen intrastent.
ResultadosForam incluídos 103 pacientes com média de idade de 61,3±11,4 anos, 26,2% eram diabéticos, 24,3% apresentaram-se com síndrome coronária aguda e 71,6% tinham lesões tipo B2/C. Implante de múltiplos stents ocorreu em 4,7% dos casos e o sucesso angiográfico foi de 100%. Na fase hospitalar, a taxa de infarto agudo do miocárdio periprocedimento foi de 3,9%, e um desses eventos levou à revascularização da lesão-alvo (RLA). No seguimento angiográfico de 9 meses, a mediana da perda tardia do lúmen intrastent foi de 0,91 [0,50; 1,21] mm. As taxas cumulativas de eventos cardíacos adversos maiores nos seguimentos de 6 meses, 9 meses e 12 meses foram, respectivamente, de 7,8%, 18,5% e 21,3%, principalmente em decorrência de RLA. Não se observou morte ou trombose de stent em 12 meses.
ConclusõesO stent Amazonia® PAX demonstrou excelentes resultados imediatos e alto perfil de segurança. Entretanto, as taxas de recorrência angiográfica foram relativamente altas, em razão da pouca eficácia na inibição da formação de hiperplasia neointimal.
Initially, first-generation drug-eluting stents (DES) with durable polymers were shown to be effective in the treatment of selected coronary lesions due to the significant reduction in neointimal hyperplasia formation, restenosis and, consequently, the need to perform target lesion revascularisation (TLR), when compared to bare-metal stents (BMS).1–3 However, despite the remarkable effectiveness of these devices in different scenarios,4 concerns related to their safety profile and long-term safety 5 have led to the development of new technologies, aiming to optimise results.6–11
Specifically, late stent thrombosis post-DES has been considered a rare and multifactorial phenomenon,12,13 involving delayed strut endothelialisation and healing of the treated segment.14,15 Previous pathological studies demonstrated that the presence of durable polymer debris was associated with inflammation and eosinophil infiltration in the artery wall and is thus a potentially thrombogenic factor.15,16 Therefore, non-polymeric DES were developed as an alternative to earlier generation DES, which used polymers as a drug carrier system.6,17,18
The Amazonia PAX® device (Minvasys SAS – Gennevillieres, France) is a DES that incorporates a low profile platform and an antiproliferative agent, paclitaxel, which is applied directly to the surface of the device without the use of a polymer. Recently published data from the initial study on the device (vs. Taxus® Liberté DES, Boston Scientific – Natick, MA, USA) showed no significant differences in the occurrence of angiographic restenosis at four months or clinical events up to 12 months in a small sample of patients from a single-centre study.19 However, the performance of the Amazonia PAX® stent in more complex scenarios, with more patients and over longer angiographic follow-up has not been determined. The present study aimed to report the nine-month angiographic follow-up and the long-term clinical outcomes of patients treated with the Amazonia PAX® stent in a prospective, multicentre study.
METHODSStudy protocol and populationThe PAX-B study was a prospective, nonrandomised, multicentre, international study that aimed to evaluate the long-term clinical and angiographic outcomes of patients with obstructive coronary artery disease treated with the Amazonia PAX® stent in nine clinical centres, of which seven were located in Europe and two were located in Brazil. The study population included patients aged≥18 years, with stable or unstable angina and/or a stress test positive for ischaemia. The angiographic inclusion criteria were the presence of a de novo coronary lesion with stenosis of≥50% and a length of≤20mm by visual analysis, located in a natural coronary artery, with a reference diameter between 2.5mm and 3.5mm, as well as a favourable anatomy for percutaneous coronary intervention with stent implantation. The main exclusion criteria were acute myocardial infarction within < 72 hours of the index procedure; renal failure with basal serum creatinine≥2mg/ dL; previous percutaneous coronary intervention of the target lesion; previous coronary intervention of the target vessel within nine months of the index procedure; cardiogenic shock; cerebrovascular accident; lesions located in a coronary bifurcation, left main coronary artery or saphenous vein bypass graft; calcification and/or significant tortuosity; total occlusion; left ventricular ejection fraction of < 30%; and hypersensitivity or contraindication to aspirin and/or thienopyridine (clopidogrel).19 According to the protocol, treatment of up to two lesions per patient was permitted, as long as a single lesion was treated per target vessel.
The study was in conducted accordance with the Declaration of Helsinki, was approved by the Ethics Committee of the participating clinical centres, and complied with the mandatory regulations and requirements of the Comissão Nacional de Ética em Pesquisa (Brazilian National Research Ethics Commission, CONEP). Furthermore, all participating patients signed an informed consent prior to study inclusion.
Device descriptionThe design, profile, and characteristics of the Amazonia PAX® stent have been detailed elsewhere.19 In summary, this device incorporates a low-profile metallic cobalt-chromium platform, with a thickness of 73μm, an open-cell modular design, and an antiproliferative agent, paclitaxel, which is applied directly to the abluminal surface of the stent struts by a microdrop spray crystallisation process. This procedure only adds 5μm to strut thickness and aims to ensure the integrity of the drug and a controlled release profile.
ProcedureThe percutaneous procedures were performed according to current guidelines,20 and the final procedure was left to the surgeon’s discretion. Direct stent implantation was allowed. The implantation of an additional stent in the target lesion was allowed if dissection and/or the need for complete coverage of the lesion and the treatment of multiple lesions (maximum of two in distinct target vessels) was permitted according to the protocol. Notably, only patients treated exclusively with Amazonia PAX® stents were considered for the final analysis of the PAX-B study. Amazonia PAX® stents with diameters of 2.5mm, 3mm, and 3.5mm, and extensions of 8mm, 12mm, 16mm, 20mm, and 24mm were available for the study.
Dual antiplatelet therapy consisted of the administration of aspirin (100–325mg/day) and thienopyridine (75mg/day of clopidogrel). In case of failure or intolerance, the administration of 250mg of ticlopidine two times daily was allowed. The administration of clopidogrel was started at least 24 hours before the procedure, and a loading dose of 300mg (or 500mg of ticlopidine) was recommended. After the procedure, aspirin therapy was maintained indefinitely, and thienopyridine therapy was maintained for at least 12 months, as currently recommended.20 Regarding antithrombin therapy during the procedure, unfractionated heparin (UFH) was administered intravenously at a dose of 70–100 U/kg to maintain an activated coagulation time of > 250 seconds (or > 200 seconds in case of glycoprotein IIb/IIIa inhibitor use, which was administered at the surgeon’s discretion).
Regarding additional tests, a 12-lead electrocardiogram was obtained before, immediately after, and 24 hours after the procedure. Laboratory tests for cardiac enzymes (creatine phosphokinase and creatine kinase MB-fraction) were conducted pre-procedure (< 72 hours) and every six to eight hours during the first 24 hours after the procedure or until hospital discharge.
Outcomes and definitionsThe primary study endpoint was late lumen loss in the in-stent segment, as assessed by a quantitative coronary angiography analysis at the nine-month follow-up. Secondary endpoints were target vessel failure, TLR and target-vessel revascularisation (TVR) at the nine-month follow-up, immediate success (of the device, lesion, and procedure), major adverse cardiac events (MACE), and binary restenosis assessed by quantitative coronary angiography at the nine-month follow-up. All deaths were considered to be cardiac unless a noncardiac cause could be clearly established by clinical evaluation or anatomopathological studies. Target vessel failure was defined as the composite endpoint of TVR, myocardial infarction, and cardiac death. MACE were defined by the composite endpoint of death, myocardial infarction, and target-lesion revascularisation (percutaneous or surgical). Stent thrombosis was defined according to the criteria of the Academic Research Consortium.21 Device success was defined as the successful delivery and implantation of the study stent at the lesion site. The lesion success (angiographic) was defined by achieving a stenosis diameter (SD) of < 50% in the target lesion after the procedure. Procedural success was defined as lesion success without the occurrence of MACE during the index hospitalisation.
Angiographic analysisThe acquisition of angiographic images occurred after the intracoronary administration of nitrate (100– 200μg) and included two orthogonal projections (> 30 degrees), which aimed to optimise the visualisation of the target lesion. The qualitative and quantitative angiographic analyses were performed serially pre- and post-procedure and at the nine-month follow-up, and were independently analysed by the Angiographic Analysis Laboratory of the Cardiovascular Research Centre (São Paulo, SP, Brazil). Overall, the qualitative analysis followed previously established definitions, including characteristics of the vessel, lesion morphology, and coronary flow.22 The analyses of quantitative coronary angiography were performed off-line by surgeons who were experienced with the methods and blinded to procedural data, using a computerised quantitative analysis system with semi-automatic detection for luminal borders, which has been validated, is commercially available (QAngio® XA, version 7.2, Medis – Leiden, the Netherlands), and incorporates a tool to perform the segmental analysis. Lesion extension was delimited by the distance between points immediately before and after the target-stenosis considered free of atheromatous processes, i.e., the transition between the stenotic segment and the normal references (5–10mm). The minimal lumen diameter (MLD) and the reference vessel diameter (RVD) were used to calculate SD using the formula: SD (%)=[1−(MLD/RVD)] x 100. Immediate gain was defined as the difference in MLD pre- and post-procedure (MLD post-procedure minus pre-procedural MLD). Late lumen loss was defined as the difference in MLD after the procedure and at the long-term follow-up (MLD post-procedure minus MLD at the nine-month follow-up). Binary restenosis was defined as the presence of SD≥50% in the treated segment at the long-term follow-up and was classified as focal (< 10mm) or diffuse (≥10mm). The restenoses were also classified according to Mehran.23 Quantitative variables were reported in the in-stent and intra-segment segments, which incorporates the in-stent segment and 5mm of the borders.6
Clinical follow-up and databaseClinical follow-up was planned at one month, six months, nine months, and 12 months after the procedure and annually for up to five years. The follow-up consisted of medical consultations or telephone contact, performed in accordance with a pre-defined protocol. At the nine-month follow-up, an angiographic study was mandatory. The electronic data capture system that was used was based on an Oracle platform (Redwood Shores, CA, USA), with restricted access through personal passwords. In 100% of the cases, the data were verified to correct any discrepancies or inconsistent information through remote monitoring. Physical monitoring (in person) was performed in a sample of 15% of the cases, selected at random. For both remote and physical monitoring, the procedures were the responsibility of the study’s data coordinating centre. All adverse events were verified directly with the source documentation and were subsequently adjudicated by an independent clinical events committee consisting of three professionals experienced in clinical and invasive cardiology. For the current analysis, the results of the procedure, the angiographic restudy at nine months, and the clinical follow-up up to 12 months are reported.
Statistical analysisCategorical variables are shown as absolute numbers and frequencies. Quantitative variables are shown as mean and standard deviation (±SD) or median and interquartile range [25%; 75%].
RESULTSA total of 103 patients were enrolled between December, 2008 and November, 2009. The baseline characteristics are shown in Table 1. Overall, the mean age was 61.3±11.4years, 26.2% were diabetics, and 41.7% had a previous myocardial infarction. Regarding the clinical presentation, approximately half of the patients had stable angina.
Baseline Clinical Characteristics
Variable | n=103patients |
---|---|
Mean age, years | 61.3±11.4 |
Female gender, n (%) | 24 (23,3) |
Diabetes mellitus, n (%) | 27 (26.2) |
Insulin in use | 5 (4.8) |
Systemic arterial hypertension, n (%) | 74 (71.8) |
Dyslipidemia, n (%) | 64 (62.1) |
Smoking, n (%) | 23 (22.3) |
Previous AMI, n (%) | 43 (41.7) |
Previous PCI, n (%) | 20 (19.4) |
Previous CABG, n (%) | 0 |
Clinical presentation, n (%) | |
Silent ischemia | 29 (28.1) |
Stable angina | 49 (47.6) |
Unstable angina | 25 (24.3) |
AMI=acute myocardial infarction; PCI=percutaneous coronary intervention; CABG=coronary artery bypass graft.
Angiographic data are shown in Table 2. In total, 127 lesions were included (1.24±0.43 lesions per patient) and the anterior descending artery was the target vessel most often treated (39.4%). Additionally, high-complexity lesions (type B2/C) were found in most cases (71.6%). During the procedure, radial access was used in 46.6% of the cases, and a second stent in the target lesion was implanted in 4.7% of the cases (Table 3). At the end of the PCI, both angiographic success and device success were achieved in 100% of the cases.
Angiographic Data
Variable | n=127lesions |
---|---|
Target-vessel, n (%) | |
ADA | 50 (39.4) |
Cx | 44 (34.7) |
RCA | 33 (26) |
Location (segment), n (%) | |
Ostial | 1 (0.8) |
Proximal | 47 (37) |
Medium | 64 (50.4) |
Distal | 15 (11.8) |
Type of lesion (ACC/AHA), n (%) | |
A | 13 (10.2) |
B1 | 23 (18.1) |
B2 | 84 (66.1) |
C | 7 (5.5) |
ACC=American College of Cardiology; AHA=American Heart Association; RCA=right coronary artery; Cx=circumflex artery; ADA=anterior descending artery.
Procedure
Variable | n=103 patients(127 lesions) |
---|---|
Radial access, n (%) | 48/103 (46.6) |
Pre-dilation, n (%) | 54/127 (42.5) |
Nominal balloon extension, mm | 13.9±3.2 |
Nominal balloon diameter, mm | 2.68±0.88 |
Balloon insufflation pressure, atm | 10.6±2.9 |
Implanted study stent, n (%) | 127/127 (100) |
Nominal stent extension, mm | 16.9±3.9 |
Nominal stent diameter, mm | 3.02±0.37 |
Stent release pressure, atm | 13.5±3.5 |
Additional implanted stent, n (%) | 6/127 (4.7) |
Post-dilation, n (%) | 73/127 (57.5) |
Nominal balloon extension, mm | 11.5±2.6 |
Nominal balloon diameter, mm | 3.21±0.41 |
Balloon insufflation pressure, atm | 16.9±3.8 |
Final TIMI flow grade 3, n (%) | 127/127 (100) |
TIMI=thrombolysis in myocardial infarction.
Table 4 shows the pre- and post-procedure results of the quantitative coronary angiography analysis. At the nine-month angiographic follow-up (92.1% of the lesions), the medians of the late intra-segment and in-stent lumen loss were 0.64 [0.32; 0.92] mm and 0.91 [0.50; 1.21] mm, respectively (Table 5). The rate of restenosis was 19.7%, and most recurrent cases were of focal in-stent type (Table 6).
Pre- and Post-procedural Quantitative Angiography
Variable | n=127lesions |
---|---|
Pre-procedure | |
Lesion extension, mm | 12.1 [9.9; 15.8] |
RVD, mm | 2.92 [2.64; 3.21] |
MLD, mm | 0.70 [0.43; 1.00] |
SD, % | 74.8 [64.8; 86.0] |
Post-procedure | |
RVD, mm | 2.95 [2.70; 3.30] |
Intra-segment | |
MLD, mm | 2.46 [2.22; 2.77] |
SD, % | 15 [10.3; 21.1] |
Immediate gain, mm | 1.74 [1.40; 2.06] |
Intrastent | |
MLD, mm | 2.79 [2.53; 3.11] |
SD, % | 5.7 [3.6.; 8.5] |
Immediate gain, mm | 2 [1.72; 2.40] |
Baloon:artery ratio | 1.10 [1.06; 1.16] |
MLD=minimal lumen diameter; RVD=reference vessel diameter; SD=stenosis diameter.
Angiographic follow-up at nine months
Variable | n=117lesions |
---|---|
RVD, mm | 2.81 [2.41; 3.07] |
Intra-segment | |
MLD, mm | 1.84 [1.40; 2.25] |
SD, % | 31.7 [21.5; 44.7] |
Late lumen loss, mm | 0.64 [0.32; 0.92] |
Intrastent | |
MLD, mm | 1.94 [1.41; 2.30] |
SD, % | 27.3 [16.4; 41.8] |
Late lumen loss, mm | 0.91 [0.50; 1.21] |
MLD=minimal lumen diameter; RVD=reference vessel diameter; SD=stenosis diameter.
Binary restenosis
Angiographic follow-up at nine months | n=117lesions |
---|---|
Total, n (%) | 23 (19.7) |
Mehran’s classification,23 n | |
Type 1A | 0 |
Type 1B | 0 |
Type 1C | 18 |
Type 1D | 0 |
Type 2 | 2 |
Type 3 | 0 |
Type 4 | 3 |
Regarding clinical events while in the hospital, four patients (3.9%) had periprocedural myocardial infarctions (all without Q-wave), and one of these events led to TLR. In this case, the patient returned to the haemodynamics laboratory on the same day of the index procedure since he exhibited ischaemic symptoms after the procedure. At the angiographic assessment, a dissection was observed at the border of the previously implanted stent, which was treated with an additional stent implantation, overlapping with the struts of the other stent. Between hospital discharge and 30 days after the procedure, there was one instance of new myocardial infarction (non-Q). Figure 1 shows the cumulative rates of adverse events after hospital discharge up to the 12-month follow-up (100% of patients). In general, rates of target vessel failure (22.3%) and MACE (22.3%) after one year were directly related to the rates of TLR and TVR. There were no deaths from stent thrombosis or any other cause during follow-up.
DISCUSSIONIn this prospective, multicentre, international registry, the non-polymeric paclitaxel-eluting Amazonia PAX® stent showed excellent immediate results (device/angiographic success of 100% and procedural success of 96.1%), as well as sustained safety after one year. However, its clinical effectiveness was lower than that reported for other DES, with and without polymers.1–3,7–11,24–28 Notably, the angiographic findings in the long-term follow-up of the PAX-B study are similar to results reported with other previously tested non-polymeric paclitaxel-releasing stents,6 which also confirms that the Amazonia PAX® stent is less effective than the Taxus® Liberté paclitaxel-releasing stent with durable polymer at suppressing neointimal proliferation.19
In general, first-generation DES, such as Cypher® (Cordis – Miami Lakes, USA) and Taxus®, demonstrated remarkable and sustained clinical effectiveness. 29 However, the durable polymeric components used as drug carriers in these devices were shown to cause an intense local inflammatory response,30,31 which has been associated with recurrence and late thrombotic phenomena. 15,16,32,33 Thus, the second generation of DES introduced more biocompatible polymers and systems with biodegradable polymers. 7,8,10,11 Recent studies suggest that these devices have an improved safety profile compared to the first generation of DES, without compromising their clinical effectiveness.34,35 Nonetheless, the development of DES without polymers (third-generation DES) could, at least in theory, offer additional advantages, such as a) the prevention of adverse effects due to the prolonged or temporary presence of polymeric residues; b) the optimisation of vascular healing; c) the sustained integrity of the stent surface; and d) the maintenance of dual antiplatelet therapy for shorter periods.36 However, the removal of the polymer from the DES may overly compromise its effectiveness, as the polymer is responsible not only for carrying the drug, but also for controlling release kinetics.6 Previous studies have shown that using a non-polymeric stent coating is safe and may lead to a dose-dependent decrease in restenosis rates; however, these studies have also demonstrated that the critical aspect is the elution profile of the antiproliferative drug.6,17,18 In the current analysis, the relatively small sample size (n=103) notwithstanding, impairment in safety was not observed, despite the high prevalence of complex cases and lesions, including an absence of death or stent thrombosis up to the 12-month follow-up. However, the unexpectedly high rates of recurrence indicate the technological challenge of maintaining effectiveness in the absence of a polymer.
In the prospective, randomised, multicentre study DELIVER, which included 1,043 patients, the non-polymeric, paclitaxel-releasing DES Achieve® (Guidant Corp. – Santa Clara, CA, USA) presented a mean late in-stent lumen loss of 0.81±0.60mm at the eight-month follow-up (vs. 0.98±0.57mm in the control group, treated with BMS, P=0.003); however, this difference did not translate into significant decreases in the rates of binary restenosis or TLR. In this case, the device in the active group (Achieve® stent) received a coating of paclitaxel at a dose of 3mg/mm2 of the stent surface area.6 Similar to the first generation Taxus® Liberté DES, which incorporates a durable polymer, the paclitaxel dose is rather low (1μg/mm2 of the stent surface area).3 Nevertheless, preclinical studies have estimated a significant loss of the antiproliferative agent (up to 40% of the total) during stent delivery with the nonpolymeric device when compared to < 10% of drug elution in 30 days with the Taxus® stent.3,6 With the Amazonia PAX® stent, tested in the PAX-A and PAX-B studies, the medians of late in-stent lumen loss were 0.77 [0.47; 1.05] mm (four months) and 0.91 [0.50; 1.21] mm (nine months), respectively. It is noteworthy that the paclitaxel dose (or concentration) was 0.67μg/cm2 of the stent surface area, with 33%–52% of the drug being released within eight hours; 55%–75%, within a week; and 100%, in 45 days.19 Thus, the kinetics of drug release appear to explain, at least in part, the greater efficacy observed with polymeric vs. non-polymeric DES, despite the use of the same drug (in this case, paclitaxel) and a reduced dose.6,19 Greater efficacy has also been observed with stents releasing drugs from the ‘limus’ class. 1,2,7,11,24,25 Late angiographic results with the non-polymeric sirolimus-releasing DES Yukon® (Trans-lumina GmbH, Hechingen, Germany) demonstrated relatively lower efficacy when compared with results of the first generation sirolimus-releasing DES Cypher®. In the prospective, randomised study, Intracoronary Stenting and Angiographic Restenosis (ISAR-TEST), which included 450 patients, the mean late in-stent lumen loss was 0.48±0.61mm with the Yukon® stent (vs. 0.48±0.58mm with the Taxus® stent).24 Similarly, in the ISAR-TEST 3, which compared three sirolimus-eluting stents, including non-polymeric, biodegradable polymer and durable polymer systems (Cypher®), the late in-stent lumen loss was 0.47±0.56mm vs. 0.17±0.45mm vs. 0.23±0.46mm, respectively.37
Alternatives tested in non-polymeric DES have included the incorporation of nonpolymeric carrier components and the structural modification of metal surfaces. The VESTAsync system (MIV Therapeutics, Atlanta, GA, USA) has a microporous hydroxyapatite coating that carries the drug sirolimus at a reduced dose (approximately 40% of the dose used in the Cypher® stent). In the initial study in humans (n=15), the late in-stent lumen loss was 0.30±0.25mm at four months and 0.36±0.23mm at nine months. Correspondingly, the obstruction volume percentage with intracoronary ultrasound was 2.8±2.2% at four months and 4±2.2% at nine months. It is important to highlight that the VESTAsync stent had a drug-elution profile similar to that of the Cypher® stent, with total release attained at 90 days.25,28 Additionally, the results of a new nonpolymeric system using active carrier substances released from a ‘bioinductor’ surface (Cre8, CID – Saluggia, Italy) have been recently reported. In this prospective randomised study (n=323), the non-polymeric stent Cre8, which releases a ‘limus’ drug, showed a late in-stent lumen loss at six months of 0.14±0.36mm vs. 0.34±0.40mm for the Taxus® Liberté stent (P < 0.0001).26 Finally, the BioFreedom® stent (Biosensors International – Singapore) incorporates a technology that uses a modified abluminal metal surface and does not use a polymer, resulting in a microporous structure that allows adhesion and the release of the antiproliferative agent Biolimus A9. In the prospective, randomised, BIOFREEDOM study, 182 patients were randomised into three groups and divided into two cohorts according to an angiographic restudy at four months (n=75) or 12 months (n=107). At four months, the median values of late in-stent lumen loss was 0.08 [0.02; 0.14] mm, 0.12 [0.07; 0.25] mm vs. 0.37 [0.14; 0.50] mm for the BioFreedom standard dose, the BioFreedom reduced dose, and the Taxus® Liberté, respectively; at 12 months, the values were 0.17 [0.09; 0.39] mm, 0.22 [0.17; 0.66] mm vs. 0.35 [0.22; 0.57] mm, respectively (P=0.001 for non-inferiority in the comparison of the standard dose BioFreedom group vs. Taxus®).27 Notably, both the Cre8 and the BioFreedom stents showed efficacy that was relatively superior to that of other reported non-polymeric DES.26,27 Possible reasons for the improved efficacy include the use of the ‘limus’ drug family, which has demonstrated superiority in inhibiting neointimal hyperplasia in relation to other classes of drugs, including paclitaxel.38 Moreover, the Cre8 system has drug release kinetics that are similar to the first- and second-generations polymeric DES.26 In contrast, approximately 90% of the drug is released within 48 hours after BioFreedom implantation;27 however, the drug release kinetics do not appear to have an impact on efficacy, as observed in other systems. It is speculated that this effect is probably due to the high lipophilicity of the drug Biolimus A9 (ten times greater than that of sirolimus), thus ensuring optimal drug absorption and a longer-lasting antiproliferative effect. In the case of the Amazonia PAX® stent tested in the present study, a new model, which utilizes a new carrier and release technology for the drug sirolimus based on nanoparticle encapsulation technology (‘LEX’), has been developed (preclinical studies are in progress). This new model should increase the effectiveness in inhibiting neointimal hyperplasia for that system (R. Costa, personal communication, TCT 2011. San Francisco, CA, USA).
CONCLUSIONSIn this prospective, multicentre, international registry, the new drug-eluting, non-polymeric paclita-xel-releasing Amazonia® PAX stent showed excellent immediate results and a high safety profile, including the absence of death or stent thrombosis after one year of follow-up. However, recurrence rates were relatively high because of the low effectiveness of neointimal hyperplasia inhibition.
CONFLICTS OF INTERESTThe authors declare no conflicts of interest.