metricas
covid
Buscar en
Enfermedades Infecciosas y Microbiología Clínica (English Edition)
Toda la web
Inicio Enfermedades Infecciosas y Microbiología Clínica (English Edition) Under-utilization of taps in intensive care unit as a cause of reservoirs of non...
Información de la revista
Vol. 36. Núm. 4.
Páginas 214-217 (abril 2018)
Visitas
2174
Vol. 36. Núm. 4.
Páginas 214-217 (abril 2018)
Brief report
Acceso a texto completo
Under-utilization of taps in intensive care unit as a cause of reservoirs of nonfermenting gram-negative bacilli
La infrautilización de grifos en una unidad de cuidados intensivos como causa de reservorios de bacilos gramnegativos no fermentadores
Visitas
2174
Gonzalo de-las-Casas-Cámaraa,
Autor para correspondencia
lalocascam@yahoo.es

Corresponding author.
, María Dolores Martín-Ríosb, Maria Isabel Adillo-Monteroa, María Carmen Muñoz-Egeac, Javier Zapardiel-Ferrerod, Concepción Pérez-Jorge Peremarchd
a Servicio de Medicina Preventiva, Hospital Universitario Infanta Elena, Valdemoro, Madrid, Spain
b Servicio de Medicina Preventiva, Hospital Universitario Rey Juan Carlos, Móstoles, Madrid, Spain
c Servicio de Microbiología y Parasitología, Hospital Universitario Infanta Elena, Valdemoro, Madrid, Spain
d Servicio de Microbiología y Parasitología, IIS-Fundación Jiménez Díaz, Madrid, Spain
Este artículo ha recibido
Información del artículo
Resumen
Texto completo
Bibliografía
Descargar PDF
Estadísticas
Figuras (1)
Tablas (1)
Table 1. Characteristics of patients who presented with infection/colonisation by NFGNB.
Abstract
Introduction

The under-utilisation of taps is associated with the generation of reservoirs of non-fermenting gram-negative bacilli with the ability to disseminate. We describe the detection and approach of the problem in an ICU.

Methods

Observational descriptive study in an ICU with individual cubicles with their own sink. We collected clinical samples from patients and environmental samples from tap aerators and reviewed the unit's hygiene measures.

Results

We detected four cases due to Chryseobacterium indologenes, one to Elizabethkingia meningoseptica and another to Pseudomonas aeruginosa; they were identified both in clinical and the environmental samples. The healthcare professionals reported that almost every hand hygiene opportunity was performed with a hydroalcoholic solution. After considered the daily flushing of water outlets as inefficient, it was decided to remove them.

Conclusions

National recommendations were insufficient for preventing, detecting and controlling tap contamination in units with a high risk of infection. The management of taps in these units needs to be improved.

Keywords:
Surveillance
Non-fermenting gram-negative bacilli
Tap aerator
Intensive care unit
Resumen
Introducción

La infrautilización de grifos se relaciona con reservorios de bacilos gramnegativos no fermentadores en su interior con capacidad de diseminación. Se describe la detección del problema y abordaje en una UCI.

Métodos

Estudio descriptivo en una UCI con boxes individuales con lavabo propio. Se recogieron muestras clínicas de pacientes y ambientales de los aireadores de los grifos. Se revisaron medidas de higiene.

Resultados

Se identificaron 4 casos de Chryseobacterium indologenes, uno de Elizabethkingia meningoseptica y otro por Pseudomonas aeruginosa, tanto en muestras clínicas como ambientales. Los profesionales indicaron utilizar solución hidroalcohólica casi siempre en la higiene de manos. Tras descartar la apertura controlada diaria de grifos por ineficiente, se decidió su retirada.

Conclusiones

Las recomendaciones nacionales resultaron insuficientes para la prevención, detección y control de la contaminación de los grifos de una unidad de alto riesgo de infección. Es necesario mejorar el manejo de los grifos en estas unidades.

Palabras clave:
Vigilancia
Bacilo gramnegativo no fermentador
Aireador de grifo
Unidad de cuidados intensivos
Texto completo
Introduction

Health care-related infections are the second most frequent adverse event during care, after medication errors.1 The latest data from the EPINE study2 show that they affect 7.92% of patients admitted to a hospital, with non-fermenting gram-negative bacilli (NFGNB) being the third most frequent group of microorganisms. They behave like opportunistic pathogens with diverse capacity of dissemination and clinical relevance.3,4

Due to their low requirements and chlorine resistance of some of its types,4 the role that tap aerators can play as reservoirs of NFGNB in units with patients at high risk of infection and that are not monitored in water quality samples for human consumption has been noted.5–7 In addition, the increased number of sinks in these units, accompanied by the diffusion of hydroalcoholic solutions for hand hygiene, favours the underuse of the former with stagnation of water and formation of biofilms8; therefore, measures have been proposed, such as the daily opening of taps that are used infrequently in these units, for their prevention.9

The objective of this work is to describe the detected cases that contributed to the identification of the problem and its approach in an intensive care unit (ICU).

Methods

Descriptive study, whose period is subdivided into 2, one of surveillance of a group of cases of Chryseobacterium indologenes between January and April 2015, whose index case is declared by the intensive care workers to the Preventive Medicine Department for review of special precautions, and another of surveillance of sporadic cases initiated in the same way as a result of the identification of a case of Elizabethkingia meningoseptica in November 2015 and until March 2016, when a case of Pseudomonas aeruginosa was detected. The attack rate was calculated on patients admitted to the ICU during the first period.

The centre studied has an ICU with 8 individual boxes, with its own sink and a surveillance system for microorganisms of epidemiological interest among the Microbiology and Preventive Medicine Departments. The ICU adheres to the national Zero Resistance initiative,10 which means that weekly screening samples of multi-resistant micro-organisms are carried out in the unit, including bronchial aspiration, which was the nosocomial infection/colonisation index location in all cases. Age, sex, week of stay, intrinsic and extrinsic risk factors of infection/colonisation2 and presence of nebulised therapy of patients were collected.

Environmental samples were collected by vortexing the aerators of the taps in sterile physiological serum for 5min, one of the ICU box in which the last 2 cases of the first period arose, and in all the boxes, medication preparation room, living room and rubbish tip, before each of the 2 cases of the second period, 11 locations in total.

All samples were seeded on Trypticase Soy agar+5% sheep blood (TSS), PolyViteX chocolate agar (PVX) and MacConkey agar. The TSS and PVX agar were incubated for 48h in a 37°C oven with 5% CO2, while the MacConkey agar was incubated in an oven at 37°C without CO2. Each plate was checked at 24h of incubation. The identification at the species level of the microorganisms isolated in the study was carried out using the mass spectrometry technique (VITEK MS, Biomerieux) based on the MALDI-TOF. In all the isolates, identification results were obtained with a confidence value of 99.9%, and antibiotic sensitivity studies were carried out based on the Kirby Bauer agar disk diffusion method.

Coinciding with the last case of the first period, controls of water in the consumer's tap were requested from the 8 boxes to the accredited external laboratory with which the hospital performs the water self-monitoring,7 which were repeated at 24h to verify the effectiveness of the corrective action carried out (thermal shock of terminal points and cleaning/disinfection of taps with refill of diffusers of all the boxes), together with sampling of entry point to the centre and points distal to the ICU.

Results

Table 1 summarises the characteristics of the cases and Fig. 1 its chronology, listing of results of environmental samples by locations and control actions.

Table 1.

Characteristics of patients who presented with infection/colonisation by NFGNB.

Age  ≥65 years  ≥65 years  ≥65 years  ≥65 years  50–60 years  40–50 years 
Gender  Male  Male  Male  Male  Male  Female 
Intrinsic risk factors  COPD
DM 
COPD  COPD  –  Neoplasm
DM 
Neoplasm 
Extrinsic risk factors  IMV and nebulised therapy  IMV  IMV and nebulised therapy  IMV and nebulised therapy  IMV  IMV 
Week of hospital stay  Third  Third  Second  Fifth  Fifth  Fourth 
Samples  BAS  BAS  BAS  BAS, blood culture and arterial catheter tip  BAS  BAS and blood culture 
Clinical assessment  Colonisation  Colonisation  Colonisation  Colonisation and infection that did not alter the resolution of the symptoms  Infection that did not alter the resolution of the symptoms  Colonisation and infection that did not alter the resolution of the symptoms 
Isolated  C. indologenes  C. indologenes  C. indologenes  C. indologenes  E. meningoseptica  P. aeruginosa 

BAS: bronchoaspiration; COPD: chronic obstructive pulmonary disease; DM: diabetes mellitus; IMV: invasive mechanical ventilation.

Fig. 1.

Timeline and listing of results of environmental samples by locations and control actions.

(0.55MB).

During the period of grouped cases of C. indologenes a total of 91 patients were admitted to the ICU with an average age of 59.8 years (SD: 16.3) and a median of days of stay 2 (IQR: 1–4). 54% were men and 46% women. Four cases appeared, with the attack rate being 4.4%.

C. indologenes was isolated in the tap aerator sample from the same box that housed the last 2 cases and coliforms were isolated in consumer water tap samples both in this and in another box in which there were no cases. The other parameters were appropriate. After the corrective action, all the consumer water tap sample verification tests performed were negative for coliforms.

During the period of sporadic cases, a first case was presented with E. meningoseptica, also being identified in the tap sample of its box. In the rest of the locations, 6 more positive for NFGNB and 2 positive for Bacillus spp. were found, not being identified in patient samples. The second case presented P. aeruginosa that was isolated in the tap of its box, with NFGNB being found in 5 more locations that were also not identified in patient samples.

Discussion

The approach to control the group of cases of C. indologenes was aimed at eliminating reservoirs in wet medical devices,4 cleaning/disinfection reinforcement and standard precautions, which has resolved similar situations documented for Chryseobacterium spp.11,12 The intensification of these measures could contribute to the distancing of cases, but recurrence in the same box led to a belief of the taps as a reservoir5,6 and in addition to consumer tap sample controls (thinking of an origin prior to the ICU) it was decided to sample its aerator, resulting in the identification of C. indologenes and the final control after cleaning/disinfection of taps with replacement of diffusers.

The absence of coliforms in the input samples to the centre as well as points distal to the ICU limited its presence to the 2 boxes mentioned, attributing its appearance to an external contamination of the tap during some use of hygiene or cleaning.9

The appearance of a case of E. meningoseptica (formerly Chryseobacterium meningosepticum), whose isolation in clinical samples is documented as being strongly associated with the contamination of taps,5 gave rise to a new surveillance period accompanied by corrective action, verifying that the cleaning and disinfection of ICU taps will be carried out monthly as recommended in units at high risk for control of Legionella.13

The discussion of these findings with professionals revealed that the taps of the boxes were hardly used because almost all hand hygiene opportunities are done with hydroalcoholic solution, which favours the formation of biofilms,8,9 a fact consistent with the negativity of the 2 samples from the medication preparation room, which was identified as the unit's most used tap. Although there may still be hidden reservoirs (including the drains themselves and siphons of the sinks14) and there are limitations to directly relating the contaminations with the infections/colonisations detected, the need to work on these underused taps was made clear.8 After dismissing its daily preventive opening9 for being inefficient (a frequency higher than the recommended weekly opening for the prevention of Legionella15), coinciding with the case of P. aeruginosa, the definitive withdrawal of the taps from the boxes was decided without identifying new cases until the writing of this article.

Despite the low pathogenic capacity of C. indologenes, its persistence was used to initiate the surveillance. Although none of the isolates altered the clinical resolution of the patients, the consequences could have been serious, given the greater potential impact of the other NFGNB as well as the rest of isolates in taps and not in patients (e.g. Stenotrophomonas maltophilia).2,4 The prevention measures and samples of Legionella or quality of water for human consumption recommended in Spain were insufficient for the surveillance and prevention of health care-related infections by NFGNB in a unit with a high risk of infection. The collection of samples for the “Zero Resistance” project implemented in the unit has been useful for the identification of the problem in patients and for proposing a surveillance aimed at its resolution, whilst simultaneously allowing for the effectiveness of the actions performed to be evaluated.

Funding

None declared.

Conflicts of interest

None declared.

Acknowledgements

To Carolina Giráldez-García for her great work coordinating the measurements.

References
[1]
Estrategia de Seguridad del Paciente del Sistema Nacional de Salud. Período 2015-2020.
Ministerio de Sanidad, Servicios Sociales e Igualdad, (2015),
[2]
Vaqué J, Grupo de Trabajo EPINE. Informe Global de España. Estudio de Prevalencia de las Infecciones Nosocomiales en España (EPINE-EPPS 2016) en el contexto del “European Prevalence Survey of Healthcare-Associated Infections and Antimicrobial Use (EPPS)”. Available from: http://hws.vhebron.net/epine/Global/EPINE-EPPS%202016%20Informe%20Global%20de%20Espa%C3%B1a%20Resumen.pdf [accessed 26.12.16].
[3]
F. Fernández-Cuenca, L.E. López-Cortés, J. Rodríguez-Baño.
Contribución del laboratorio de microbiología en la vigilancia y el control de brotes nosocomiales producidos por bacilos gramnegativos no fermentadores.
Enferm Infecc Microbiol Clin, 29 (2011), pp. 40-46
[4]
M. Vaneechoutte, L. Dijkshoorn, A. Nemec, P. Kämpfer, G. Wauters.
Acinetobacter, Chryseobacterium, Moraxella, and other nonfermentative gram-negative rods.
Manual of clinical microbiology, 10th ed., pp. 714-738
[5]
W. Jiun-Ling, C. Mei-Ling, E.L. Yusen, C. Shan-Chwen, C. Yee-Chun.
Association between contaminated faucets and colonization or infection by nonfermenting gram-negative bacteria in intensive care units in Taiwan.
J Clin Microbiol, 47 (2009), pp. 3226-3230
[6]
M.L. Cristina, A.M. Spagnolo, B. Casini, A. Baggiani, P. del Giudice, S. Brusaferro, et al.
The impact of aerators on water contamination by emerging gram-negative opportunists in at-risk hospital departments.
Infect Control Hosp Epidemiol, 35 (2014), pp. 122-129
[7]
Real Decreto 140/2003, de 7 de febrero, por el que se establecen los criterios sanitarios de la calidad del agua de consumo humano.
Boletín Oficial del Estado, 45 (2003), pp. 7228-7245
[8]
Department of Health Report on the review of evidence regarding the contamination of wash-hand basin water taps within augmented care units with Pseudomonas.
Departament of Health, (2012),
[9]
Guidelines for the prevention and control of infection from water systems in healthcare facilities.
Prepared by the Prevention and Control of Infection from Water Systems in Healthcare Facilities Sub-Committee of the Health Protection Surveillance Centre Scientific Advisory Committee, (2015),
[10]
Prevención de la emergencia de bacterias multirresistentes en el paciente crítico. Proyecto Resistencia Zero.
Sociedad Española de Medicina Intensiva, Crítica y Unidades Coronarias (SEMICYUC), (2013),
[11]
T.A.M. Hekker, W. van Overhagen, A.J. Schneider.
Pressure transducers: an overlooked source of sepsis in the intensive care unit.
Intensive Care Med, 16 (1990), pp. 511-512
[12]
S. Maraki, E. Scoulica, A. Manoura, N. Papageorgiou, C. Giannakopoulou, E. Galanakis.
A Chryseobacterium meningosepticum colonization outbreak in a neonatal intensive care unit.
Eur J Clin Microbiol Infect Dis, 28 (2009), pp. 1415-1419
[13]
L.M. Sehulster, R.Y.W. Chinn, M.J. Arduino, J. Carpenter, R. Donlan, D. Ashford, et al.
Guidelines for environmental infection control in health-care facilities. Recommendations from CDC and the Healthcare Infection Control Practices Advisory Committee (HICPAC).
American Society for Healthcare Engineering/American Hospital Association, (2004),
[14]
B. Clarivet, D. Grau, E. Jumas-Bilak, H. Jean-Pierre, A. Pantel, S. Parer, et al.
Persisting transmission of carbapenemase-producing Klebsiella pneumoniae due to an environmental reservoir in a university hospital.
Euro Surveill, 21 (2016),
[15]
Real Decreto 865/2003, de 4 de julio, por el que se establecen los criterios higiénico-sanitarios para la prevención y control de la legionelosis.
Boletín Oficial del Estado, 171 (2003), pp. 28055-28069

Please cite this article as: de-las-Casas-Cámara G, Martín-Ríos MD, Adillo-Montero MI, Muñoz-Egea MC, Zapardiel-Ferrero J, Pérez-Jorge Peremarch C. La infrautilización de grifos en una unidad de cuidados intensivos como causa de reservorios de bacilos gramnegativos no fermentadores. Enferm Infecc Microbiol Clin. 2018;36:214–217.

Copyright © 2017. Elsevier España, S.L.U. and Sociedad Española de Enfermedades Infecciosas y Microbiología Clínica
Descargar PDF
Opciones de artículo
es en pt

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos