To evaluate the correlation and agreement in our unit and population of hemoglobin in gasometry versus hematology analyzer, to evaluate errors in transfusion or lack thereof.
Resultsstrong association between Point-of-care (POC) and hematimetry, with P<.001, with a coefficient of determination r2 of 0.56, an intraclass correlation coefficient of 0.63 and a Lin's concordance correlation coefficient of 0.65. For hemoglobins less than 7g/dL, a success rate of 29.41% was obtained.
ConclusionsLow-moderate agreement of POC hemoglobin with standard haemothymetry. High probability of errors in the indication of transfusion based on gasometer hemoglobins, especially in low hemoglobins.
Evaluar la correlación y la concordancia en nuestra unidad y en la población de la cifra de hemoglobina en gasometría versus hematimetría estándar; valorar errores en transfusión o falta de la misma.
ResultadosFuerte asociación entre los resultados del gasómetro (POC) y la hematimetría, con p<0,001, con un coeficiente de determinación r2 de 0,56, un coeficiente de correlación intraclase de 0,63 y un coeficiente de correlación de Lin de 0,65. Valores similares para el hematocrito. Para hemoglobinas menores de 7g/dl se obtiene una tasa de acierto del 29,41%.
ConclusionesConcordancia baja-moderada de la hemoglobina del POC con la hematimetría estándar. Alta probabilidad de errores en la indicación de transfusión en base a hemoglobinas de gasómetro, sobre todo en hemoglobinas bajas.