To analyze the results of the Xen45 implant in our hospital center, both in terms of efficacy and safety.
MethodsObservational, retrospective, single-center study of patients undergoing Xen45 surgery at the University Hospital of Jaén between January 1, 2018 and June 31, 2021.
ResultsA total of 73 patients were implanted with the Xen45 device (48 with in combined surgery with phacoemulsification of the crystalline lens and 25 standalone Xen45 surgery). IOP decreased from 19.9 ± 4.9 mmHg at the first visit to a mean of 17.1 ± 5.3 (P = .001) at the sixth month. Drug use required between the first visit and Xen implantation was reduced from 2.1 ± 0.9 to 0.6 (P < .000). A second surgery was necessary in 18 patients (26.6%) in the first 6 months.
ConclusionsGlaucoma surgery by Xen45 implant, alone or combined with cataract surgery, is an effective and safe procedure for IOP lowering.
Analizar los resultados del implante Xen45 en nuestro centro hospitalario, tanto en términos de eficacia como en cuanto a seguridad.
MétodoEstudio observacional, retrospectivo y monocéntrico de los pacientes intervenidos de Xen45 en el Hospital Universitario de Jaén entre el 1 de enero de 2018 y el 31 de junio de 2021.
ResultadosSe implantó el dispositivo Xen45 a un total de 73 pacientes (48 cirugía combinada con facoemulsificación del cristalino y 25 de forma aislada). La PIO disminuyó de 19,9 ± 4,9 mmHg en la primera visita hasta una media de 17,1 ± 5,3 (P = ,001) en el sexto mes. Se redujo de 2,1 ± 0,9 a 0,6 el uso de fármacos necesarios entre la primera visita y el implante Xen (P < ,000). Fue necesaria una segunda cirugía en 18 pacientes (26,6%) en los 6 primeros meses.
ConclusionesLa cirugía de glaucoma mediante implante Xen45, en solitario o combinado con la cirugía de catarata, es un procedimiento eficaz y seguro para la disminución de la PIO.
Artículo
Comprando el artículo el PDF del mismo podrá ser descargado
Precio 19,34 €
Comprar ahora