Problema inusual con la mascarilla laríngea Supreme™. Caso del trimestre
Rare problem with the insertion of a Supreme™ laryngeal mask airway device. Case of the trimester
Visitas
1199
Sistema Español de Notificación en Seguridad en Anestesia (SENSAR) 1
Sistema Español de Notificación en Seguridad en Anestesia y Reanimación (SENSAR), España
Este artículo ha recibido
Información del artículo
Resumen
Texto completo
Bibliografía
Descargar PDF
Estadísticas
Figuras (1)
Resumen

Paciente a la que se realizó una anestesia general para realizarle una tumorectomía de mama. Tras la inducción se procedió a insertar una mascarilla laríngea Supreme™ para control de la vía aérea. La inserción de la mascarilla, la introducción de una sonda de aspiración a través del tubo de drenaje gástrico y la ventilación mecánica fueron muy dificultosos desde el inicio. Al retirar completamente la mascarilla para subsanar el problema se observó una rotura del tubo de drenaje gástrico con salida de la sonda de aspiración a su través, hacia la zona anterior o laríngea de la mascarilla. Posteriormente se comprobó que esa mascarilla y todas las del mismo lote de fabricación habían sufrido un cambio en su diseño: el manguito era de un material más blando y sin refuerzo en la punta, y el tubo de drenaje gástrico presentaba un defecto de termosellado que facilitó la rotura. Se contactó con el distribuidor local y el fabricante para comunicarles el incidente, retirándose el lote defectuoso del que procedía dicha mascarilla. Asimismo se difundió el incidente a través de SENSAR a otros hospitales, permitiendo alertar rápidamente a otros usuarios de los posibles problemas secundarios al cambio de diseño de la mascarilla laríngea Supreme™.

Palabras clave:
Anestesia
Seguridad del paciente
Mascarilla laríngea
Análisis de fallos de equipamiento
Abstract

A breast tumor was resected under general anesthesia. After induction, the airway was managed with a Supreme™ laryngeal mask airway device. The insertion of the laryngeal mask airway device, the insertion of the orogastric tube through the drain tube, as well as the mechanical ventilation, were very difficult from the beginning. On removing the laryngeal mask airway device to solve the problem, it was observed that the drain tube was broken, and the orogastric tube had passed into the anterior, laryngeal part of the device through the split. It was later found out that the laryngeal mask airway device, as well as the whole manufacturing batch, had suffered a design modification: the cuff was constructed with a softer material without reinforcement in the tip, and the drain tube had a heat-sealing defect that facilitated the break. The incident was reported to the local supplier and the manufacturer, and the defective batch of laryngeal mask airway devices was recalled. The incident was also reported to other hospitals via SENSAR, to warn other users of the potential dangers of the design modification in the Supreme™ laryngeal mask airway.

Keywords:
Anesthesia
Patient safety
Laryngeal masks
Equipment failure analysis

Artículo

Opciones para acceder a los textos completos de la publicación Revista Española de Anestesiología y Reanimación
Suscriptor
Suscriptor de la revista

Si ya tiene sus datos de acceso, clique aquí.

Si olvidó su clave de acceso puede recuperarla clicando aquí y seleccionando la opción "He olvidado mi contraseña".
Suscribirse
Suscribirse a:

Revista Española de Anestesiología y Reanimación

Comprar
Comprar acceso al artículo

Comprando el artículo el PDF del mismo podrá ser descargado

Precio 19,34 €

Comprar ahora
Contactar
Teléfono para suscripciones e incidencias
De lunes a viernes de 9h a 18h (GMT+1) excepto los meses de julio y agosto que será de 9 a 15h
Llamadas desde España
932 415 960
Llamadas desde fuera de España
+34 932 415 960
E-mail
Opciones de artículo
es en pt

¿Es usted profesional sanitario apto para prescribir o dispensar medicamentos?

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos